K022453 · Welch Allyn Protocol, Inc. · DSI · Aug 21, 2002 · Cardiovascular
Device Facts
Record ID
K022453
Device Name
MODIFICATION TO ACUITY CENTRAL MONITORING SYSTEM
Applicant
Welch Allyn Protocol, Inc.
Product Code
DSI · Cardiovascular
Decision Date
Aug 21, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Acuity® system is intended to be used by skilled clinicians for the central monitoring of neonatal, pediatric, and adult patients in health care facilities. The Acuity system consists of the Acuity Central Monitoring Station, Acuity software, FlexNet™ communications network and Welch Allyn vital signs monitors. The vital signs monitors can operate standalone (according to their individual Intended Use Statement) or be networked to the Acuity central monitoring station through the Acuity FlexNet communication network. In addition to the central monitoring of patient data, waveforms, alarms and alerts, the Acuity software can include optional modules to provide extended recording of patient data, arrhythmia monitoring and ST analysis. Full disclosure stores patient data for up to 96 hours. Arrhythmia monitoring module provides real-time monitoring and alarms for specific changes in cardiac rhythms. The clinician is responsible for determining the clinical significance of each detected arrhythmia event or alarm. The arrhythmia module is not intended for use with neonatal patients. ST analysis module provides real-time monitoring and alarms for ST segment deviations, from a reference beat, for patients with suspected heart disease and anomalies. The clinician is responsible for determining the clinical significance of each detected ST segment deviation or alarm. The ST analysis module is not intended for use with neonatal patients. The most likely locations for patients monitored by an Acuity system are step-down units, telemetry departments, general med/surg floors and emergency departments. The Acuity system is available for sale only upon the order of a physician or licensed health care professional.
Device Story
Acuity Central Monitoring Station is a Unix-based workstation for central patient monitoring in clinical settings (step-down, telemetry, med/surg, ED). System integrates patient monitors via FlexNet network (wired/wireless Ethernet). Device receives ECG signals from patient monitors; processes data using Mortara Instrument's arrhythmia and ST analysis software module. System provides real-time monitoring, alarms for cardiac rhythm changes and ST segment deviations, and full disclosure storage (up to 96 hours). Clinicians review outputs on the station to determine clinical significance of events. System aids in patient management by providing centralized oversight of waveforms and vital signs. Benefits include improved arrhythmia detection and ST analysis capabilities compared to previous versions.
Clinical Evidence
Bench testing only. Performance of the Mortara software module was compared against the previous module using playback files. Testing on a fully configured (networked) system showed 100% QRS complex detection, <3% event false positives, and <1% ventricular beat mislabeling. Integration testing confirmed minimal impact on existing system services.
Technological Characteristics
UNIX-based workstation (Sun Microsystems architecture). Connectivity via hardwired or wireless Ethernet (IEEE 802.3/802.11). Software includes Mortara arrhythmia and ST analysis module. System components: Central Monitoring Station, Acuity software, FlexNet network, Welch Allyn vital signs monitors. Quality system certified to ISO 9001 and EN46001.
Indications for Use
Indicated for central monitoring of neonatal, pediatric, and adult patients in healthcare facilities. Arrhythmia and ST analysis modules are indicated for pediatric and adult patients only; not for neonatal use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
# AUG 2 1 2002
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
July 25, 2002
#### Submitter
Welch Allyn Protocol, Inc. 8500 S.W. Creekside Place Beaverton, OR 97008-7107 USA
Telephone: (503) 526-8500 Fax: (503) 526-4200
Contact: Mr. Donald M. Abbey, Vice President, Quality Systems
#### Device Name
Trade Name: Common Name: Cardiac Arrhythmia Monitor Classification Name: Arrhythmia Detector and Alarm (Reference, 21 CFR 870.1025, April 1, 2001).
#### Device Classification
Regulatory Class III
#### Predicate Device
Substantial equivalence is based on the current Acuity system, which consists of the Acuity Central Monitoring Station, Acuity software, FlexNet™ communications network, Welch Allyn vital signs monitors, and the Mortara H-Scribe Holter system. The Acuity Central Monitoring Station and Acuity software were cleared for market by the FDA under 510(k) submissions K913193, K935846 and K972121. The FlexNet 100 communications network was cleared for market as part of K002725. The FlexNet™ communications network has not changed from the original submission. The software in the Acuity® Central Monitoring Station has changed. The Arrhythmia monitoring and ST analysis software module has changed. Welch Allyn Protocol, Inc. has purchased the rights to use an arrhythmia monitoring and ST analysis software module developed by Mortara Instruments, Inc. It contains the same software used in the Mortara H-Scribe Holter System that was cleared for market under K004017. This software replaces the Arrhythmia monitoring and ST analysis software modules currently in the Acuity® Central Monitoring Station.
#### Device Description
The Acuity system comprises four components: the Acuity Central Monitoring Station, Acuity software, the FlexNet™ communications network, and Welch Allyn vital signs monitors. The Acuity Central Monitoring Station is the heart of the system, using hardware provided by Sun Microsystems, Inc. The Acaity Central Monitation Station, Shippetwork devices on the FlexNet Betweek uring hardwired Ethernet connections (IEEE 802.3-compliant) or wireless Ethernet connections (IEEE 802.11compliant).
The Acuity software operating system is Solaris™, the Unix®-based operating system from Sun Microsystems. Welch Allyn Protocol's application software provides the routines and services required to accomplish the tasks associated with central patient monitoring. These tasks include the user interface, communication with patient monitors, management of alarms and settings, and recording and review of patient data.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a sequence of alphanumeric characters, specifically 'K022453'. The characters are handwritten in a clear, legible style, with each digit and letter distinctly formed. The sequence appears to be a code or identifier, possibly a serial number or product code.
The Archythmia Analysis and ST Analysis options are being replaced with Mortara Instrument's arrhythmia algorithm software module. Mortara Instrument's arrhythmia software will now provide the data analysis for both the Arrhythmia Analysis and the ST Analysis options.
The Acuity application software receives the output from the Mortara software module and records the results in the same manner as the previous version of Acuity software. There are only a few minor changes to the user interface as described in the Acuity Directions For Use.
The Acuity hardware and software operating system are not changed.
### Indications for Use
The Acuity® system is intended to be used by skilled clinicians for the central monitoring of neonatal, pediatric, and adult patients in health care facilities. The Acuity system consists of the Acuity Central Monitoring Station, Acuity software, FlexNet™ communications network and Welch Allyn vital signs monitors. The vital signs monitors can operate standalone (according to their individual Intended Use Statement) or be networked to the Acuity central monitoring station through the Acuity FlexNet communication network.
In addition to the central monitoring of patient data, waveforms, alarms and alerts, the Acuity software can include optional modules to provide extended recording of patient data, arrhythmia monitoring and ST analysis.
- Full disclosure stores patient data for up to 96 hours.
- . Arrhythmia monitoring module provides real-time monitoring and alarms for specific changes in cardiac rhythms. The clinician is responsible for determining the clinical significance of each detected arrhythmia event or alarm. The arrhythmia module is not intended for use with neonatal patients.
- . ST analysis module provides real-time monitoring and alarms for ST segment deviations, from a reference beat, for patients with suspected heart disease and anomalies. The clinician is responsible for determining the clinical significance of each detected ST segment deviation or alarm. The ST analysis module is not intended for use with neonatal patients.
The most likely locations for patients monitored by an Acuity system are step-down units, telemetry departments, general med/surg floors and emergency departments. The Acuity system is available for sale only upon the order of a physician or licensed health care professional.
#### Technological Comparison to the Predicate Device
The Acuity system is comprised of 4 components, the Acuity Central Monitoring Station, Acuity Software, FlexNet communication network, and Welch Allyn vital signs monitors. The Acuity Central Monitoring Station is the heart of the system. It provides the user interface, control, data processing and data storage for the system. The central monitoring station is a UNIX workstation that only supports the Sun Microsystems architecture. Except for the Acuity software, the Acuity system is the same as system(s) see cash ramber ander the sit overnoled 510gle submitsloms.
The arrhythmia module in the Acuity software has changed. It has been replaced by a proprietary software package developed and marketed by Mortara Instruments, Inc. The Mortara software module contains the same arrhythmia algorithms used in their H-Scribe Holter system. The Mortara software module is a selfcontained software package that receives and process ECG signals from the systems, and provides outputs in the form of measurements and event markers. The Acuity Software processes the outputs to create displays, alarms, messages, trends and other system services such as printing.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a sequence of alphanumeric characters. The sequence starts with the letter 'K', followed by the number '0', and then the numbers '22453'. The characters are written in a clear, sans-serif font, and they appear to be handwritten.
#### Summary of Performance Testing
Module tests demonstrated that there was improved performance with the Mortara software module, in that Mortara software module out performed the current arrhythmia module in the Acuity system. It also proved that integration of the Mortara module would have a minimum impact on the existing Acuity software, such as displays, alarms, trending control and communications.
Test data from the two units noted above were compared and found to be identical with the exception of one test case. The Mortara software module labeled one of the 1600 beats incorrectly.
A standalone (non-networked) Acuity system with a single patient monitor connected was also tested with the same playback files. The data from this system was then compared with the data from that of a fully configured (networked) system. Analysis showed that 100% of the ORS complexes were detected and the arrhythmia events were correctly identified (labeled) with less than 3% event false positives. Less than 1% of the ventricular beats were mislabeled. In one test case, which included 1600 beats, three ventricular beats were mislabeled, resulting in one less false positive event identification. This was observed for one of the monitors from the fully loaded system.
A risk analysis, identifying potential hazards and documenting mitigation of the hazards, has been developed and verified/validated as part of Welch Allyn Protocol's product development procedures. Welch Allyn Protocol's Quality System conforms to 21CFR820 and is certified by TUV Product Service to ISO 9001 and EN46001.
#### Conclusions
As stated above, Welch Allyn Protocol's conclusion is that the Acuity system with Cardiac Arrhythmia and ECG ST Analysis options are safe, effective, comply with the appropriate medical device standards, and offer improved performance compared to Acuity system with Cardiac Arrhythmia and ECG ST Analysis options currently on the market.
This 510(k) Summary of Safety and Effectiveness may be copied and submitted to interested parties as required by 21CFR807.92.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract human profiles facing to the right, with flowing lines suggesting movement or progress.
Public Health Service
AUG 2 1 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Welch Allyn Protocol, Inc. c/o Mr. Donald M. Abbey Vice President, Quality Systems 8500 S.W. Creekside Place Beaverton, OR 97008-7107
Re: K022453
Trade Name: Acuity® Central Monitoring Station Regulation Name: Arrhythmia Detector and Alarm Regulation Number: 21 CFR 870.1025 Regulatory Class: Class III (three) Product Code: DSI Dated: July 25, 2002 Received: July 26, 2002
Dear Mr. Abbey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 -- Mr. Donald M. Abbey
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D. Directo Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
Applicant: Welch Allyn Protocol. Inc. 8500 SW Creekside Place Beaverton, OR 97008-7107 USA
Telephone: (503) 526-8500 Fax: (503) 526-4200
510(k) Number: __
Device Name: Acuity® Central Monitoring Station
## Indications for Use:
The Acuity system is intended to be used by skilled clinicians for the central monitoring of neonatal, pediatric, and adult patients in health care facilities. The Acuity system consists of the Acuity® Central Monitoring Station. Acuity software, FlexNet™ communications network and Welch Allyn vital signs monitors. The vital signs monitors can operate standalone (according to their individual Intended Use Statement) or be networked to the Acuity central monitoring station through the Acuity FlexNet communication network.
In addition to the central monitoring of patient data, waveforms, alarms and alerts, the Acuity software can include optional modules to provide extended recording of patient data (Full Disclosure), arrhythmia monitoring and ST analysis.
- . Full Disclosure stores patient data for up to 96 hours.
- Arrhythmia monitoring module provides real-time monitoring and alarms for specific . changes in cardiac rhythms. The clinician is responsible for determining the clinical significance of each detected arrhythmia event or alarm. The arrhythmia module is not intended for use with neonatal patients.
- ST analysis module provides real-time monitoring and alarms for ST segment deviations, . from a reference beat, for patients with suspected heart disease and anomalies. The clinician is responsible for determining the clinical significance of each detected ST segment deviation or alarm. The ST analysis module is not intended for use with neonatal patients.
The most likely locations for patients monitored by an Acuity system are step-down units, telemetry departments, general med/surg floors and emergency departments. The Acuity system is available for sale only upon the order of a physician or licensed health care professional.
(Please do not write below this line - continue on another page if needed)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Prescription Use<br>(Per 21 CFR 801.109) | OR Over-The Counter |
|------------------------------------------|---------------------|
|------------------------------------------|---------------------|
| (Division Sign-Off) |
|-------------------------------------------------------|
| Division of Cardiovascular<br>and Respiratory Devices |
| 510(k) Number | K022453 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.