This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis within the uterus, including: - Abnormal uterine bleeding - Evaluation of abnormal hysterosalpingogram - Evaluation of abnormal ultrasound image - Infertility and pregnancy wastage - Pelvic pain
Device Story
The Visera Hysterovideoscope Olympus HYF Type V is a flexible videoscope used for endoscopic diagnosis within the uterus. It captures visual data via an integrated CCD sensor and transmits the signal to an external Olympus video system center, light source, and display monitor. The device features a separated light guide connector and video connector for connection to ancillary equipment. It is intended for use by clinicians in a clinical setting to visualize the uterine cavity. The output is a real-time video feed displayed on a monitor, which assists the physician in diagnosing conditions such as abnormal bleeding, infertility, or pelvic pain. The device benefits patients by providing direct visualization for diagnostic evaluation.
Clinical Evidence
No clinical data provided; bench testing only. The device was evaluated for safety and efficacy through compliance with voluntary safety standards including IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-18.
Technological Characteristics
Flexible videoscope featuring a CCD sensor. Materials are previously cleared and compliant with ISO 10993-1. Meets IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-18 standards. Connects to external video system center and light source via separated connectors.
Indications for Use
Indicated for endoscopic diagnosis within the uterus in patients presenting with abnormal uterine bleeding, abnormal hysterosalpingogram, abnormal ultrasound images, infertility, pregnancy wastage, or pelvic pain.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{1}------------------------------------------------
Ko22445
# SMDA 510(k) SUMMARY
This summary of 510(k safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR Section 807.92.
## A. Submitter's Name, Address, Phone and Fax Number
| 1. Applicant: | Olympus Optical Co., Ltd.<br>2-3-1 Shinjuku Monolis Nishi-shinjuku<br>Shinjuku-ku, Tokyo, Japan, 163-0914 |
|---------------|-----------------------------------------------------------------------------------------------------------|
|---------------|-----------------------------------------------------------------------------------------------------------|
- 2. Registration No. 8010047
- 3. Initial Importer Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157
- 2429304 4. Registration No.
- 5. Contact Person Laura Storms-Tyler Director, Regulatory Affairs Olympus America Inc. Two Corporate Center Drive Melville, NY 11747-3157 Tel (631)-844-5688 Fax(631)-844-5416
### B. Device Name, Common Name
#### 1. Trade/Proprietary Name and Common Name
Trade Name: VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE V Common Name: Hysteroscope
#### 2. Class, Classification Number and Classification Name
| Classification Number | Classification Name | Class |
|-----------------------|------------------------------|-------|
| 21 CFR 884.1690 | Hysteroscope and accessories | II |
| 21 CFR 876.1500 | Endoscopes and accessories | II |
#### Identification of Legally Marketed Devices Which we Claim Substantial Equivalence 3.
The following listed devices are seem to be as predicate devices in consideration of its characteristic and the following table shows regulatory history.
| Model | Device Description &<br>510(k)#/ Date of Cleared | Manufacturer | Class |
|------------------------------------------------|--------------------------------------------------|----------------------|-------|
| BF-240 | #K963033 | Olympus Optical Co., | II |
| Olympus<br>HYF-P Flexible<br>hysterofiberscope | #K891451 | Olympus Optical Co., | II |
#### D. Summary Description of the Device
{2}------------------------------------------------
#### 1. Summary
This subject device is videoscope used for diagnosis within the uterus.
New type of CCD is lead to be more easier observation and diagnosis. In HYF-V, a light guide connector and a video connector are separated each other. The light guide connector is to connect with a light source and the video connector is with a video system center.
### 2. Design
"HYF-V" has been designed, manufactured and tested in compliance with voluntary safety standards.
lt meets the requirement of IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-2-18.
#### 3. Materials
There are no new patient contacting materials. All of patient contacting materials are cleared by previous 510(k). And all materials have been confirmed with ISO 10993-1.
#### 4. Intended Use of the device
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis within the uterus, including:
- Abnormal uterine bleeding -
- Evaluation of abnormal hysterosalpingogram -
- -Evaluation of abnormal ultrasound image
- Infertility and pregnancy wastage -
- Pelvic pain
#### 5. Technological Characteristics
This endoscope does not have special technological characteristics compared to the existence videoscope such as GI tract bronchial apprication.
#### 6. Reason for not requiring
Compare to the predicate device, this subject device "HYF-V" does not incorporate any significant change for its safety and efficacy to the predicate device.
Observation within uterus has been spread, therefore clinical date is not necessary for its evaluation of safety and efficacy.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of an eagle with three lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 4 2003
Olympus Optical Co., Ltd. % Ms. Laura Storms-Tyler Director, Regulatory Affairs and Quality Assurance Olympus America, Inc. Two Corporate Center Drive MELVILLE NY 11747-3157 Re: K022445
Trade/Device Name: Visera Hysterovideoscope Olympus HYF Type V Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: 85 HIH Dated: November 12, 2002 Received: November 13, 2002
Dear Ms. Storms-Tyler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
**OLYMPUS**
# Indications for Use Statement
510(k) Number(if known): 长0.22445 Device Name: VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE V
#### Indications for Use :
This instrument has been designed to be used with an Olympus video system center, light source, documentation equipment, display monitor, and other ancillary equipment for endoscopic diagnosis within the uterus, including:
- Abnormal uterine bleeding •
- Evaluation of abnormal hysterosalpingogram 1
- Evaluation of abnormal ultrasound image 1
- Infertility and pregnancy wastage -
- Pelvic pain
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use_ / OR Over-The-Counter Use
(Prescription 21 CFR 801.109)
Nancy Brogdon (Optional Format 1-2-96)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number ***_***_ K022445
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.