← Product Code [KGO](/productcode/KGO) · K022442

# POWDER FREE GREEN LATEX SURGICAL GLOVES, STERILE WITH ALOE VERA AND PROTEIN CONTENT LABELING CLAIM (50 MICROGRAM LESS) (K022442)

_Wrp Asia Pacific Sdn. Bhd. · KGO · Sep 5, 2002 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K022442

## Device Facts

- **Applicant:** Wrp Asia Pacific Sdn. Bhd.
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** Sep 5, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Powder Free Green Latex Surgical Gloves, Sterile with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less) are made of natural rubber latex intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient's body, fluids, waste, or environment.

## Device Story

Sterile, powder-free natural rubber latex surgical gloves; intended for use by surgeons and operating room personnel. Device acts as a physical barrier to prevent cross-contamination between healthcare personnel and patients, fluids, or the environment. Features Aloe Vera coating and low protein content (≤ 50 µg/g). Used in clinical/surgical settings. Provides protection for surgical wounds.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including physical property testing, pinhole leak testing per FDA 21 CFR 800.20, protein content analysis, and biocompatibility (skin irritation and dermal sensitization) testing.

## Technological Characteristics

Material: Natural rubber latex. Standards: ASTM D 3577-01aE2 (dimensions, physical properties), ASTM D 6124-01 (powder-free), ASTM D 5712-95 (protein level). Performance: Protein content < 50 µg/g; powder residue ≤ 2 mg/glove. Biocompatibility: Passes primary skin irritation and dermal sensitization tests. Sterile.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

VRP Asia Pacific Sdn Bhd 1 4 7 8 1 7 V

KO22442

# 510(k) SUMMARY

#### 1.0 Submitter:

| Name:      | WRP Asia Pacific Sdn Bhd                             |
|------------|------------------------------------------------------|
| Address:   | Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak |
|            | Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA |
| Phone No.: | +60 3 8706 1486                                      |
| Fax No .:  | +60 3 8706 1485                                      |

22 JUL 2002 Date of Summary Prepared:

### 2.0 Contact Person:

| Name:      | Mr. Yue Wah, CHOW |
|------------|-------------------|
| Phone No.: | +60 3 8706 1486   |
| Fax No.:   | +60 3 8706 1485   |

#### 3.0 Name of the device:

| Trade Name:          | 1. Aloetouch, and<br>2. Multiple or Customer's Trade Name                                                                        |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Device Name:         | Powder Free Green Latex Surgical Gloves, Sterile with Aloe<br>Vera and Protein Content Labeling Claim (50 micrograms or<br>less) |
| Common Name:         | Surgical Gloves                                                                                                                  |
| Classification Name: | Surgeon's Gloves (per 21 CFR 878.4460)                                                                                           |

### 4.0 Identification of The Legally Marketed Device:

Class I Powder Free natural rubber latex Surgeon's gloves, 79KGO, that meets all
the requirements of ASTM standard D 3577 - 01a22 Type 1 and FDA 21 CFR 800.20.

### 5.0 Description of The Device:

The Powder Free Green Latex Surgical Gloves, Sterile with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less) meets all the requirements of ASTM standard D 3577-01a"2 and FDA 21 CFR 800.20.

{1}------------------------------------------------

Image /page/1/Picture/9 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, enclosed in a black oval shape. To the right of the logo is the company name, "WRP Asia Pacific Sdn Bhd," also in a bold, sans-serif font. Below the company name is the number "147817 V".

K02242

#### Intended Use of the Device: 6.0

The Powder Free Green Latex Surgical Gloves, Sterile with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less) are made of natural rubber latex intended to be worn on the hand of healthcare personnel, operating room personnel and similar personnel to prevent contamination between the healthcare or similar personnel and the patient's body, fluids, waste, or environment.

#### 7.0 Summary of The Technological Characteristics of The Device:

The Powder Free Green Latex Surgical Gloves, Sterile with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less) are summarized with the following technological characteristics compared to ASTM or equivalent standards.

| CHARACTERISTICS       | STANDARDS                                | DEVICE PERFORMANCE                      |
|-----------------------|------------------------------------------|-----------------------------------------|
| Dimensions            | ASTM D 3577 - 01aE2                      | Meets                                   |
| Physical Properties   | ASTM D 3577 - 01aE2                      | Meets                                   |
| Freedom from pinholes | ASTM D 3577 - 01aE2<br>FDA 21 CFR 800.20 | Meets                                   |
| Powder-Free           | ASTM D 6124 - 01                         | Meets<br>2 mg/glove maximum             |
| Protein Level         | ASTM D 5712 - 95                         | < 50 µg/g                               |
| Biocompatability      | Primary Skin Irritation in<br>Rabbits    | Passes<br>(Not a primary skin irritant) |
|                       | Dermal Sensitization                     | Passes<br>(Not a contact sensitizer)    |

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains a sequence of handwritten digits and letters. The sequence starts with the letters 'K0', followed by the numbers '22442'. The handwriting style is cursive, with some connections between the characters.

Image /page/2/Picture/1 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, enclosed in a black oval shape. To the right of the oval is the text "WRP Asia Pacific Sdn Bhd" in a similar font. Below the text is the number "147817V".

#### Substantial Equivalent Based on Assessment of Non-Clinical Performance Data 8.0

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

#### 9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

#### 10.0 Conclusion

It can be concluded that the Powder Free Green Latex Surgical Gloves, Sterile with Aloe Vera and Protein Content Labeling Claim (50 micrograms or less) will perform according to the glove performance standards referenced in section 7 above and meet ASTM standards, and FDA requirements for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is a stylized symbol consisting of three abstract human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

5 2002 SEP

Mr. Yue Wah Chow Head of Department, QA/RA WRP Asia Pacific Sdn. Bhd. Lot 1, Jalan 3, Kawasan Persahaan Bander Baru Salak Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA

Re: K022442

Trade/Device Name: Powder Free Green Latex Surgical Gloves, Sterile with Aloe Vera and Protein Content Labeling Claim ( 50 Micrograms or Less ) Regulation Number: 878.4460 Regulation Name: Surgeon's Gloves Regulatory Class: I Product Code: KGO Dated: July 22, 2002 Received: July 25, 2002

Dear Mr. Chow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Chow

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski

Director Division of Anesthesiology. General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letters "WRP" in bold, white font against a black, rounded rectangular background. The letters are large and take up most of the space within the shape. The logo has a simple, clean design and appears to be for an organization or company with the initials "WRP".

WRP Asia Pacific Sdn Bhd

1 4 7 8 1 7 V

## INDICATIONS FOR USE

KO22442

Applicant:

WRP Asia Pacific Sdn Bhd

510(k) Number (if known):

Device Name:

POWDER FREE GREEN LATEX SURGICAL GLOVES, STERILE WITH ALOE VERA AND PROTEIN CONTENT LABELING CLAIM (50 MICROGRAMS OR LESS)

Indications For Use:

The surgeon's glove is a device made of natural rubber latex intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination.

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | OR | Over-The-Counter |
|----------------------|----|------------------|
| (Per 21 CFR 801.109) |    |                  |

  

(Division Sign-Off)

Division of Anesthesiology, General Hospital,

Infection Control, Dental Devices

| 510(k) Number: | K022442 |
|----------------|---------|
|----------------|---------|

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K022442](https://fda.innolitics.com/device/K022442)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
