GMRS PRESS FIT STEMS WITH PUREFIX HA

K022403 · Howmedica Osteonics Corp. · LZO · Oct 17, 2002 · Orthopedic

Device Facts

Record IDK022403
Device NameGMRS PRESS FIT STEMS WITH PUREFIX HA
ApplicantHowmedica Osteonics Corp.
Product CodeLZO · Orthopedic
Decision DateOct 17, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Global Modular Replacement System (referred to from this point on as the GMRS) Press Fit Stems with PureFix® HA Coating are intended to be used with components of the proximal femoral segment of the Howmedica Modular Replacement System (referred to from this point on as the MRS) in total hip arthroplasty. These devices are intended for use in total hip arthroplasty indicated as a result of extensive proximal femoral bone loss (from trauma, failed previous arthroplasty, or tumor resection). Adequate bone stock must be present to allow the use of the GMRS Press Fit Stems with PureFix® HA Coating.

Device Story

GMRS Press Fit Stems are modular orthopedic implants for proximal femoral reconstruction. Machined from titanium alloy (Ti-6Al-4V) with CP titanium plasma spray and PureFix® HA coating. Available in straight, straight-fluted, and curved (150mm/200mm) styles. Proximal end features a 2-degree 52-minute male taper for connection to MRS proximal femoral components. Used by orthopedic surgeons in clinical settings for patients with significant bone loss. Provides structural support and fixation to restore hip function. Implanted via press-fit technique; relies on adequate host bone stock for stability.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence established via finite element analysis and mechanical performance testing.

Technological Characteristics

Materials: Titanium alloy (ASTM F-136), CP titanium plasma spray (ASTM F-1580), PureFix® HA coating (ASTM F-1185). Design: Modular femoral stem with male taper (2 deg 52 min). Styles: Straight, straight-fluted, curved (150mm/200mm). Fixation: Press-fit (uncemented).

Indications for Use

Indicated for proximal femoral reconstruction in patients requiring total hip arthroplasty due to extensive proximal femoral bone loss resulting from trauma, failed previous arthroplasty, or tumor resection. Contraindicated in patients with overt infection, rapid disease progression, or inadequate bone stock for press fit fixation.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ GMRS Press Fit Stems with PureFix® HA Confidential 510(k) Premarket Notification OCT 1 7 2002 page 1 of 3 # 510(k) Summary Submission Information Name and Address of Sponsor: Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401 For Information contact: Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401 Device Identification GMRS Press Fit Stems with PureFix® HA Proprietary Name: Modular Stem Common Name: Classification Name and Reference 21 CFR 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Class II Proposed Regulatory Class: Device Product Code: OR(87) LZO #### Intended Use The Global Modular Replacement System (referred to from this point on as the GMRS) Press Fit Stems with PureFix® HA Coating are intended to be used with components of the proximal femoral segment of the Howmedica Modular Replacement System (referred to from this point on as the MRS) in total hip arthroplasty. These devices are intended for use in total hip arthroplasty indicated as a result of extensive proximal femoral bone loss (from trauma, failed previous arthroplasty, or tumor resection). Adequate bone stock must be present to allow the use of the GMRS Press Fit Stems with PureFix® HA Coating. 2 {1}------------------------------------------------ GMRS Press Fit Stems with PureFix® HA Confidential #### Indications Proximal femoral reconstruction secondary to: - Trauma - Failed previous prosthesis . - Tumor resection . #### Contraindications - Overt infection . - Rapid disease progression beyond an acceptable margin . For the use of GMRS Press Fit stems with PureFix® HA Coating, the following additional contraindication should be noted: - . Inadequate bone stock to allow the use of a press fit stem #### Device Description The GMRS Press Fit Stems with PureFix® HA Coating are machined from titanium alloy (Titanium-6Aluminum-4Vanadium) which meets the requirements of ASTM standard F-136. The stems are coated with commercially pure titanium (CP titanium) plasma spray coating that conforms to ASTM specification F-1580. The proximal portion of the stem is also coated with PureFix® HA coating which meets ASTM specification F-1185. The proximal portion of the stem incorporates a male taper with a taper angle of 2 degrees 52 minutes. This male taper mates with the female taper portion of the head/neck section or body section of the proximal femoral section of the Modular Replacement System. The GMRS Press Fit Stems with PureFix® HA Coating are available in four styles: - I Straight Stems: These stems are available in a 125mm length, and diameters from 11mm to 19mm in one millimeter increments. The seat diameter varies from 22 to 36mm, and the seat radius ranges from 8mm to 9.25mm. 510(k) Premarket Notification page 2 of 3 {2}------------------------------------------------ ## GMRS Press Fit Stems with PureFix® HA Confidential - Straight Fluted Stems: These stems are also available in a 125mm length, and 피 diameters from 11mm to 19mm in one millimeter increments. This stem design has four flutes that extend from the seat down 25mm of the length of the stem. These flutes are located 90 degrees apart on the body of the stem. The seat diameter varies from 22mm to 36mm, and the seat radius ranges from 8mm to 9.25mm. 510(k) Premarket N - Curved Stem (150mm): These stems are available in a 150mm length, and diameters I from 11mm to 19mm in one millimeter increments. The seat diameter of this stem ranges from 22mm to 36mm, and the seat radius ranges from 7mm to 10.5mm. The radius of the curvature is 45 inches. - Curved Stem (200mm): These stems are available in a 200mm length, and diameters . from 11mm to 19mm in one millimeter increments. The seat diameter of this stem varies from 22mm to 36mm, and the seat radius ranges from 7mm to 10.5mm. The radius of the curvature is 45 inches. Equivalent products include: - MRS Cemented Stems Howmedica Osteonics 1. - Distal Stem Segments of the Restoration Modular Femoral Stem Howmedica 2. Osteonics Finite element analyses and testing were present to support a claim of substantial equivalence. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and head. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. OCT 1 7 2002 Food and Druq Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 59 Route 17 Allendale, New Jersey 07401 Received: July 23, 2002 Re: K022403 Trade/Device Name: GMRS Press Fit Stems with PureFix® HA Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint/metal/ceramic/polymer/semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Codes: LZO Dated: July 22, 2002 Dear Ms. Crowe: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice {4}------------------------------------------------ ### Page 2 - Ms. Margaret F. Crowe requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Mark A. Milkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ page lofl 510(k) Premarket Notification GMRS Press Fit Stems with PureFix® HA Confidential 510(k) Number (if known): KOL2403 Device Name: GMRS Press Fit Stems with PureFix® HA ## Intended Use The Global Modular Replacement System (referred to from this point on as the GMRS) Press Fit Stems with PureFix® HA Coating are intended to be used with components of the proximal femoral segment of the Howmedica Modular Replacement System in total hip arthroplasty. These devices are intended for use in total hip arthroplasty indicated as a result of extensive proximal femoral bone loss (from trauma, failed previous arthroplasty, or tumor resection). Adequate bone stock must be present to allow the use of the GMRS Press Fit Stems with PureFix® HA Coating. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-the Counter-Use 801.109) Mark N Millman Restorative 2403 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...