STRYKER L3 HYDROLERT

K022248 · Stryker Endoscopy · FLN · Aug 6, 2002 · General Hospital

Device Facts

Record IDK022248
Device NameSTRYKER L3 HYDROLERT
ApplicantStryker Endoscopy
Product CodeFLN · General Hospital
Decision DateAug 6, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2420
Device ClassClass 2

Indications for Use

MT Alert Infusion Monitor is intended for use in any healthcare setting where gravityflow infusions are utilized.

Device Story

Passive weighing device for gravity-flow infusion bags; monitors fluid levels to detect near-empty states or track bolus administration. User sets alarm threshold by hanging desired weight. Device mounts to standard poles/rods; operates on alkaline batteries. Provides audio and visual alarms when weight drops below set point. Used in clinical settings by healthcare personnel to manage fluid delivery; assists in monitoring irrigant levels during procedures. Benefits include improved efficiency in fluid management and timely alerts for fluid depletion.

Clinical Evidence

Bench testing only. Validation procedures performed to compare device performance against the predicate system.

Technological Characteristics

Passive weighing device; battery-powered (2 AA batteries); pole-mounted; audio/visual alarm indicators; accommodates infusion bags up to 3 liters; bolus monitoring capability (1 liter mode).

Indications for Use

Indicated for monitoring gravity-flow infusion fluid levels in healthcare settings. Not intended for intravenous infusion monitoring.

Regulatory Classification

Identification

An electronic monitor for gravity flow infusion systems is a device used to monitor the amount of fluid being infused into a patient. The device consists of an electronic transducer and equipment for signal amplification, conditioning, and display.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ iddyp page info AUG 0 6 2002 ## 1.2 510(k) Summary Sierra BioSearch, Inc. 395 Mt. Tom Road Bishop, California 93514 # 510(k) Summary of Safety and Effectiveness # Identification of the Device | Proprietary-Trade Name: | MT Alert Infusion Monitor | |-------------------------|-----------------------------------------------------| | Classification Name: | Monitor, Electric for Gravity Flow Infusion Systems | | Device Class: | II | | Product Code: | FLN | | Common/Generic Name: | Gravity Infusion Monitor | #### Equivalent Predicate Devices The MT Alert Infusion Monitor is substantially equivalent in design and function to the Smith & Nephew Dyonics LeveLert System. ### Indications for Use MT Alert Infusion Monitor is intended for use in any healthcare setting where gravityflow infusions are utilized. ## Description of the Device The MT Alert Infusion Monitor is a passive weighing device that alarms when the infusion bag is near-empty. The alarm point can be set by the user. In addition, MT Alert may also assist in monitoring fluid bolus administration. MT Alert will alarm when a prescribed bolus of fluid has been taken from the infusion bag. MT Alert provides various audio and visual alarms. MT Alert operates from common alkaline batteries and mounts on common poles and rods. ## Safety and Effectiveness in Comparison to Predicate Devices. The validation procedures performed on the system indicate that the new device is as safe and effective as the predicate devices. {1}------------------------------------------------ 4p page 1 of 2 # Substantial Equivalence Chart | Feature | MT Alert™ | LeveLert | |--------------------------------------------------------------------------------------------------------|-------------------------------------------------------|--------------------------| | Used for monitoring the fluid level of an infusion bag. | Yes | Yes | | Automatic alarm when the fluid level has reached a pre-determined weight with audio and visual alarms. | Yes | Yes | | Adjustable near-empty alarm point. | Yes- Easily programmed by hanging the desired weight. | Yes. | | Accommodates most typical infusion bag sizes. | Yes (up to 3 liter size) | Yes (up to 3 liter size) | | Bolus monitoring feature. | Yes (1 liter mode only) | No. | | Passive device, no fluid control functions. | Yes | Yes | | Infusion Container hangar. | Yes | Yes, | | Pole Mounting. | Yes | Yes | | Power | 2- AA batteries, typical 180 day life | Powered externally | | Device Class. | Class II | Class II | Darling Castaneda MB Richard Citrenbaum, MD, President (9) (3 / ス / ス / ス / ス / (Premarket Notification [510(k)] Number) {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized image of an eagle's head and neck, with the head facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle image. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 0 6 2002 Stryker Endoscopy Michael Hilldoerfer 5900 Optical Court San Jose, California 95138 Re: K022248 Trade Name: Stryker L3 Hydroalert Regulation Number: 880.2420 Regulation Name: Electronic Monitor for gravity flow infusion system Regulatory Class: II Product Code: FLN Dated: July 10, 2002 Received: July 12, 2002 Dear Mr. Hilldoerfer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ # Page 2 -- Mr. Michael Hilldoerfer This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. R. Mark N. Millerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ July 10, 2002 510(k) Number if known: ### INDICATION FOR USE: The Stryker L3 Hydrolert is intended to monitor irrigant levels primarily during arthroscopic procedures and to alert operating room personnel of a low irrigant fluid level condition. It will offer both increased efficiency in arthroscopic procedures and universal compatibility with both arthroscopic pump and gravity fluid management systems. It is intended for use with saline solution or any other standard irrigant. It is not intended for use as an intravenous infusion monitor. The alarm is designed to be reusable. ### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) tion (ODE) Mark N. Milkenon 510(k) Number Prescription Use OR Over-the-Counter Use__________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...