← Product Code [NHI](/productcode/NHI) · K022241

# VERTIS PNT CONTROL UNIT, MODEL CU 100; VERTIS PNT SAFEGUIDE-CERVICAL, MODEL SG 102-XXX (K022241)

_Vertis Neuroscience, Inc. · NHI · Sep 11, 2002 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K022241

## Device Facts

- **Applicant:** Vertis Neuroscience, Inc.
- **Product Code:** [NHI](/productcode/NHI.md)
- **Decision Date:** Sep 11, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K022241 · Sep 11, 2002 | VERTIS PNT CONTROL UNIT, MODEL CU 100; VERTIS PNT SAFEGUIDE-CERVICAL, MODEL SG 102-XXX | Vertis Neuroscience, Inc. | Post-market surveillance data; Published human clinical trial literature | The sponsor utilized existing post-market performance data and published clinical literature to support the safety and effectiveness of the new cervical electrode kit and cable modifications. | post-market data; clinical literature; cervical pain management |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Post-market data for the Vertis PNT System; Post-market surveillance | Patients treated with the Vertis PNT System | Not applicable for this study | Safety and effectiveness |
| Published human clinical trial data; Literature review of clinical trials | Patients with cervical pain | Not applicable for this study | Cervical pain management |

## Indications for Use

Percutaneous Neuromodulation Therapy (PNT) is indicated for the symptomatic relief and management of chronic or intractable pain and/or as an adjunctive treatment in the management of post-surgical pain and post-trauma pain. The PNT Control Unit is to be used with PNT Lumbar Safeguides for low back pain or PNT Cervical Safeguides for neck and/or upper back pain.

## Device Story

System delivers percutaneous electrical stimulation for pain management. Components: software-driven 5-channel AC-powered control unit; sterile needle electrodes (Safeguides); patient cable. Used in clinical settings by physicians (anesthesiologists, PM&R) or ordered by physicians for use by physical therapists. Control unit generates electrical stimulus delivered via needle electrodes to target nerves. Output affects clinical decision-making by providing symptomatic pain relief. Benefits include management of chronic, intractable, post-surgical, or post-trauma pain.

## Clinical Evidence

Bench testing included electrode performance and dimensional data. Clinical evidence included human MRI/CT imaging data, published human clinical trial data for cervical pain management, and post-market data for the existing PNT system. Results demonstrated acceptable performance for cervical stimulation.

## Technological Characteristics

5-channel AC-powered nerve stimulator. Includes sterile needle electrodes (Safeguides) and patient cable. Software-driven control unit. Class II device.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Vertis Percutaneous Neuromodulation Therapy (PNT) Stimulation System ([K011702](/device/K011702.md))
- EPIX Tens Device System ([K970203](/device/K970203.md))
- EPIX Tens Device System ([K951903](/device/K951903.md))
- SMP-Plus ([K982410](/device/K982410.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K022241

# SEP 1 1 2002

1:2

## Appendix 13 510(k) Summary of Safety and Effectiveness

### Vertis Neuroscience, Inc. Percutaneous Neuromodulation Therapy (PNT) ™ Control Unit and Accessories

#### General Information

| Classification | Class II                                                                                     |
|----------------|----------------------------------------------------------------------------------------------|
| Trade Name     | Percutaneous Neuromodulation Therapy<br>(PNT) Nerve Stimulation System                       |
| Submitter      | Vertis Neuroscience, Inc.<br>2101 Fourth Avenue, Suite 200<br>Seattle, Washington, USA 98121 |
| Contact        | Lori Glastetter<br>Vice President, Regulatory Affairs/<br>Quality Assurance                  |

#### Nature of this 510(k)

The Vertis Percutaneous Neuromodulation Therapy (PNT) System (nerve stimulator and accessories) was cleared for marketing under 510(k) Notification K011702. This submission was filed to request clearance to market a new cervical electrode kit and cable and modify the existing PNT labeling to accommodate these new accessories.

#### Device Description

The Vertis PNT System is designed for delivering percutaneous electrical stimulation (termed: Percutaneous Neuromodulation Therapy -PNT ). The Vertis PNT System is intended to be used in pain management by a physician (e.g., anesthesiologists or physical medicine and rehabilitation physicians) or on the order of a physician (e.g., by a physical therapist) in a clinic environment. The device system includes 3 major components:

- the Vertis PNT Control Unit a software-driven, ● five channel, AC powered nerve stimulator which generates the electrical stimulus;
- the sterile Safeguides which are sterile, needle electrodes; ●
- the Patient Cable - which interconnects the PNT Control Unit to the electrodes

{1}------------------------------------------------

K022241
205-2

#### Indications for use

Percutaneous Neuromodulation Therapy (PNT) is indicated for the symptomatic relief and management of chronic or intractable pain and/or as an adjunctive treatment in the management of post-surgical pain and post-trauma pain.

The PNT Control Unit is to be used with PNT Lumbar Safeguides for low back pain or PNT Cervical Safeguides for neck and/or upper back pain.

#### Substantially Equivalent Devices

| Manufacturer                           | Substantially Equivalent devices                                                                                                                    | 510(k)             |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| Vertis Neuroscience,<br>Inc.           | Vertis Percutaneous Neuromodulation<br>Therapy (PNT) Stimulation System (Vertis<br>PNT Control Unit and Accessories)<br>Model CU 100 and SG 101-xxx | K011702            |
| Empi, Inc.<br>St. Paul, MN             | EPIX Tens Device System<br>Model EPIX VT<br>Model EPIX XL                                                                                           | K970203<br>K951903 |
| Rehabilicare, Inc.<br>New Brighton, MN | SMP-PlusTM<br>Model 4930                                                                                                                            | K982410            |

#### Safety and Effectiveness - Testing

Extensive data were provided that evaluated the use of PNT cervical electrodes and cervical percutaneous electrical stimulation. These data included the following. Data demonstrated acceptable results for the device and therapy.

- electrode performance and dimensional (bench) data ●
- human magnetic resonance images (MRI)/ computerized tomography . (CT) imaging data
- published human clinical trial data for cervical pain management ●
- . post-market data for the Vertis PNT System

#### Summary of Substantial Equivalence

Based on the information provided in this Notification, we believe the described modification to the legally marketed predicate Vertis PNTTM Control Unit and Accessories has been shown to be substantially equivalent to devices in commercial distribution prior to May 28, 1976.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.

SEP 1 1 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Lori J. Glastetter Vice President, Regulatory Affairs/Quality Assurance Vertis Neuroscience, Inc. 2101 Fourth Avenue, Suite 200 Seattle, Washington 98121

Re: K022241

Trade/Device Name: Vertis Percutaneous Neuromodulation Therapy (PNT) Nerve Stimulation System (Vertis PNT Control Unit, Vertis PNT Lumbar Safeguides and Vertis PNT Cervical Safeguides) Regulation Number: 21 CFR 882.5890 and 21 CFR 882.1350 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief and Needle Electrode Regulatory Class: Class II Product Code: NHI and GXZ Dated: July 10, 2002 Received: July 11, 2002

Dear Ms. Glastetter:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Lori Glastetter

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page__________________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): K 0222 4 |

Device Name:__________________________________________________________________________________________________________________________________________________________________

Indications For Use:

Percutaneous Neuromodulation Therapy (PNT) is indicated for the symptomatic relief and management of chronic or intractable pain and/or as an adjunctive treatment in the management of post-surgical pain and post-trauma pain a

The PNT Control Unit is to be used with PNT Lumbar Safeguides for low back pain or PNT Cervical Safeguides for neck and/or upper back pair.

### (Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

Mark N. Milliken

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number -

---

**Source:** [https://fda.innolitics.com/device/K022241](https://fda.innolitics.com/device/K022241)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
