The EMBOL-X Aortic Filter is indicated for use with the EMBOL-X Aortic Cannula in cardiac surgery procedures to capture and remove particulate emboli from the ascending aorta and heart during and following cross clamp removal in patients aged 60 years and older undergoing first time non-emergent CABG or isolated aortic or mitral valve repair/replacement.
Device Story
The EMBOL-X Aortic Filter is a surgical device used during cardiopulmonary bypass. It consists of a distal heparin-coated mesh filter on a flexible frame, a locking cartridge housing for attachment to an aortic cannula side port, and a proximal plunger mechanism. The device is introduced into the aorta via the cannula; the filter basket captures particulate emboli as blood passes through. It remains in situ for up to 60 minutes. Operated by surgeons in an OR setting, the device aims to reduce embolic debris, potentially improving patient outcomes by preventing post-operative complications like stroke or organ damage. The filter is withdrawn using the plunger mechanism before procedure completion.
Clinical Evidence
The ICEM 2000 trial was a prospective, multi-center, 1:1 randomized controlled study of 1289 patients (645 treatment, 644 control) undergoing CABG or valve procedures. Primary safety endpoint (composite of death, stroke, renal insufficiency, MI, GI complications, limb ischemia) showed equivalence (17.1% treatment vs 18.9% control, p<0.001). Primary effectiveness endpoint (particulate capture) was met in 96.8% of filters (p<0.001).
Technological Characteristics
Device comprises a heparin-coated mesh filter, locking cartridge, and plunger mechanism. Uses conventional medical-grade materials. Single-use, sterile. Designed for integration with the EMBOL-X Aortic Cannula via a side port. Mechanical filtration principle.
Indications for Use
Indicated for patients aged 60+ undergoing first-time, non-emergent CABG or isolated aortic/mitral valve repair/replacement to capture/remove particulate emboli from the ascending aorta and heart during/following cross-clamp removal.
Regulatory Classification
Identification
A cardiopulmonary bypass arterial line blood filter is a device used as part of a gas exchange (oxygenator) system to filter nonbiologic particles and emboli (blood clots or pieces of foreign material flowing in the bloodstream which will obstruct circulation by blocking a vessel) out of the blood. It is used in the arterial return line.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Arterial Line Blood Filter 510(k) Submissions.”
Predicate Devices
Medtronic PercuSurge GuardWire Plus Temporary Occlusion and Aspiration System (K003992)
{1}------------------------------------------------
### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
The following 510(k) Summary of Safety and Effectiveness information is provided in accordance with the requirements of 21 CFR §807.92 and SMDA 1990.
| 510(k) Number: | K022071 | |
|----------------------|-------------------------------------------------------------------------------------|--------------|
| Date Prepared: | February 13 2003 | FEB 1 4 2003 |
| Applicant: | EMBOL•X, Inc. | |
| Address: | 645 Clyde Avenue, Mountain View, CA 94043 | |
| Phone Number: | 650/390-0280 | |
| Fax Number: | 650/390-0282 | |
| Contact Person: | Jean Chang | |
| Trade Name: | EMBOL•X® Aortic Filter | |
| Common Name: | Arterial Line Blood Filter | |
| Classification Name: | Filter, Blood, Cardiopulmonary Bypass, Arterial Line; 21 CFR §870.4260;<br>Class II | |
The EMBOL • X Aortic Filter device consists of three primary components: Device 1.) a distal heparin-coated mesh filter, mounted onto a flexible frame to Description: form a filter basket for particulate emboli capture and retention; 2.) a locking cartridge housing for attachment to the EMBOL•X Aortic Cannula side port, permitting access to the aorta and to ensure correct orientation of the filter during use; and 3.) a proximal syringe-like plunger mechanism to deploy and withdraw the distal basket into and from the aorta, via the cannula, during surgery. The filter is introduced surgically into the aorta via the previously placed cannula, and particulates are captured and removed as blood passes through the filter basket. The filter may remain in situ for up to 60 minutes. The EMBOL.X Aortic Filter utilizes conventional medical grade materials and processes, and is provided packaged, labeled, and sterile, intended for single-use.
{2}------------------------------------------------
- The EMBOL-X Aortic Filter is indicated for use with the EMBOL-X Intended Use: Aortic Cannula in cardiac surgery procedures to capture and remove particulate emboli from the ascending aorta and heart during and following cross clamp removal in patients aged 60 years and older undergoing first time non-emergent CABG or isolated aortic or mitral valve repair/replacement.
- Substantial equivalence is derived from a composite of characteristics Predicate from multiple predicate devices. The EMBOL.X Aortic Filter is Devices: substantially equivalent in intended use, clinical application, principle of operation, design and materials, sterility and biocompatibility, and performance to the Medtronic PercuSurge GuardWire Plus Temporary Occlusion and Aspiration System (K003992) and/or the Edwards Lifesciences AF-1025D/AF-1040D Duraflo (heparin treated) Arterial Blood Filter (K820044).
- The EMBOL.X Aortic Filter has similar intended use, design intent, Technological Characteristics: principle of operation, materials, sterility and biocompatibility, accessory requirements, and labeling as that of the predicate devices. Any noted differences between the devices (specific indications for use, method of device delivery, specific physical dimensions and geometry) do not raise new types of safety or effectiveness questions, do not introduce new technological issues, and therefore do not impact the substantial equivalence of the EMBOL•X Aortic Filter.
- The results of biocompatibility, in-vitro (bench), and pre-clinical (animal) Non-Clinical Test tests demonstrate that the EMBOL·X Aortic Filter is sterile, Results: biocompatible, meets established internal performance specifications, and satisfies the requirements of relevant external standards and applicable FDA Guidance.
- Data to support the EMBOL•X Aortic Filter (and associated EMBOL•X Summary of Aortic Cannula) were obtained from the EMBOL.X ICEM 2000 trial. Clinical Studies: The purpose of this prospective, multi-center, 1:1 randomized, controlled equivalency study was to demonstrate the safety and effectiveness of the EMBOL•X Aortic Filter (and the associated EMBOL•X Aortic Cannula) in capturing particulate emboli during first-time non-emergent coronary artery bypass graft (CABG) or aortic/mitral valve repair/replacement procedures utilizing cardiopulmonary bypass. This study was conducted at 22 sites within the United States and Canada, and was comprised of 1289 patients, of which 645 were randomized to the EMBOL•X Aortic Filter, and 644 were randomized to control. An independent Clinical Events Committee (CEC) adjudicated the major clinical endpoints and events.
{3}------------------------------------------------
## Summary of Clinical Studies (continued):
The primary safety measure was a composite endpoint comprised of the following post-operative clinical events, measured from the time of randomization (the operation) through hospital discharge or 30 days. whichever occurred first: Neurologic deficit (mild and severe); Renal insufficiency (with and without dialysis); Gastrointestinal (GI) complications; Perioperative Myocardial Infarction (MI); Limbthreatening peripheral embolism (Limb Ischemia); and Death. The primary effectiveness endpoint was the capture of particulate emboli by the filter (Treatment arm only) in at least 75% of the filtered patients, with particulate debris visually confirmed by light microscopy.
The patient population of this study was limited to patients undergoing first time, non-emergent Coronary Artery Bypass Grafting (CABG), aortic valve replacement or mitral valve repair or replacement only, aged 60 years and older. Of the 1289 patients studied. 927 (71.9%) were male. and 1042 (80.8%) were 65 years of age or older; 65 (5.0%) had a LVEF < 30%, and 388 (30.1%) had a prior MI. None of the demographic or medical history differences between the randomized groups achieved statistical significance (p<0.05).
18 of 22 enrolling centers performed peri-procedural ultrasound imaging (TEE or EPI), resulting in an analysis of a subset (910, or 70.6%) of the total enrolled patients.
The primary safety endpoint was met, with the EMBOL . X Aortic Filter arm composite event rate statistically equivalent to that of standard treatment (17.1% vs. 18.9%, p<0.001). In addition, differences in the stratified events between the randomized groups did not achieve statistical significance (p<0.05). The primary effectiveness endpoint was also met, with particulate capture demonstrated in 96.8% of all filters analyzed (p<0.001).
The following tables summarize the results of the ICEM trial.
{4}------------------------------------------------
| Event | Treatment<br>(N=645) | Control<br>(N=644) | P-Value |
|-------------------------------------|----------------------|--------------------|---------|
| Death | 10 (1.6) | 11 (1.7) | 0.82 |
| Neurologic deficit (Stroke/TIA) | 18 (2.8) | 18 (2.8) | 1.00 |
| Renal insufficiency (RI) | 40 (6.2) | 52 (8.1) | 0.19 |
| RI (w/o dialysis) | 33 (5.1) | 43 (6.7) | 0.23 |
| RI (dialysis) | 7 (1.1) | 9 (1.4) | 0.61 |
| Myocardial infarction (MI) | 67 (10.4) | 64 (9.9) | 0.79 |
| Q Wave MI | 21 (3.3) | 18 (2.8) | 0.63 |
| CK-MB Elevation | 46 (7.1) | 46 (7.1) | 0.99 |
| Gastrointestinal complications (GI) | 5 (0.8) | 5 (0.8) | 1.00 |
| Limb ischemia | 3 (0.5) | 3 (0.5) | 1.00 |
| Any event | 110 (17.1) | 122 (18.9) | 0.38 |
### Table 1. Major Adverse Events [Number (%)]
Numbers are for all randomized patients
Death: Death for any cause
Stroke: Central neurologic deficit persisting for > 24 hours
Transient neurologic deficit (TIA): An ischemic event of the central nervous system that causes a neurologic deficit persisting for < 24 hours
Renal insufficiency: Increase of serum creatinine to > 2.0 mg/dl or a 50% or greater increase over abnormal baseline prior to procedure
Renal insufficiency (dialysis): The new requirement for dialysis
O-Wave MI: New pathological Q-Waves in 2 or more contiguous leads
Non Q-Wave MI: CPK > 5x normal and CK-MB > 5x above the upper limit of normal for the institution, in the absence of new Q-Waves
Gastro-Intestinal Complications: include GI bleeding requiring transfusion; Pancreatitis with abnormal amylase/lipase requiring NG suction therapy; Cholecystitis requiring cholecystectomy or drainage; Mesenteric ischemia requiring exploration
Limb-threatening Peripheral Embolism: Acute onset of diminished pulse, altered pallor (discoloration, either hypo- or hyper-), and pain as evidence of limb-threatening peripheral ischemia
### Table 2.
#### Echocardiographically Evident Endothelial Disruptions Observed through Imaging
| Treatment | Control | P-Value |
|--------------|-------------|---------|
| n/N (%) | n/N (%) | |
| 42/456 (9.2) | 9/454 (2.0) | < 0.001 |
# Table 3.
Composite Endpoint Events Stratified by Echocardiographically Evident Endothelial Disruptions Observed through Imaging
| Arm | Events in Patients With<br>Endothelial Disruption<br>n/N (%) | Events in Patients Without<br>Endothelial Disruption<br>n/N (%) | P-Value |
|-----------|--------------------------------------------------------------|-----------------------------------------------------------------|---------|
| Treatment | 2/42 (4.8) | 74/414 (17.9) | 0.03 |
| Control | 1/9 (11.1) | 81/445 (18.2) | 1.0 |
{5}------------------------------------------------
## Table 4. Particulate Capture Effectiveness
| Attribute | Result |
|------------------------------------------------------------------------------|-------------|
| Number EMBOL•X Filters Deployed | 618 |
| Number (%) Filters Which Captured ≥ 1 Particle | 598 (96.8%) |
| Lower 95% Confidence Bound on Percent of Filters Which Captured ≥ 1 Particle | 95.3% |
Conclusions drawn from Study
The results from the clinical investigation demonstrate that the EMBOL . X Aortic Filter, when used in patients undergoing first time, non-emergent Coronary Artery Bypass Grafting (CABG), aortic valve replacement or mitral valve repair or replacement, does not pose any additional risk to the treated patient population when compared to that of the current standard treatment of no filtration, and the Aortic Filter was effective in capturing particulates. The data demonstrate that the EMBOL·X Aortic Filter is compatible with conventional CPB procedures. It was therefore concluded that the EMBOL. X Aortic Filter is safe and effective when used as indicated in the Instructions for Use.
- Review of the device in vitro pre-clinical studies, combined Summary: with the results of the clinical investigation, provides valid scientific evidence and reasonable assurance that the EMBOL.X Aortic Filter is safe and effective for its intended use. Comparison of the product attributes supports the conclusion that the device is substantially equivalent to the commercially marketed predicate devices.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines above them that resemble a bird's wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 4 2003
EMBOL-X, Inc. c/o Ms. Jean Chang Chief Operating Officer and Executive Vice President 645 Clyde Ave. Mountain View, CA 94043-2208
Re: K022071 Trade Name: EMBOL-X® Aortic Filter Regulation Number: 21 CFR 870.4260 Regulation Name: Filter, Blood, Cardiopulmonary Bypass, Arterial Line Regulatory Class: Class II (two) Product Code: DTM Dated: November 22, 2002 Received: November 25, 2002
Dear Ms. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Ms. Jean Chang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/dsma/dsmamain.html
Sincerely yours,
Kala Tiller
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K022071
> EMBOL•X Aortic Filter Device Name:
Indications For Use: The EMBOL-X Aortic Filter is indicated for use with the EMBOL-X Aortic Cannula in cardiac surgery procedures to capture and remove particulate emboli from the ascending aorta and heart during and following cross clamp removal in patients aged 60 years and older undergoing first time non-emergent CABG or isolated aortic or mitral valve repair/replacement.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
DaCattun
Prescription Use
(Per 21 CFR §801.109) X
OR Over-The-Counter Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.