← Product Code [DQY](/productcode/DQY) · K022061

# CVSI PERIPHERAL BALLOON CATHETER; CVSI INFLATER (K022061)

_Cryo Vascular Systems, Inc. · DQY · Sep 20, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K022061

## Device Facts

- **Applicant:** Cryo Vascular Systems, Inc.
- **Product Code:** [DQY](/productcode/DQY.md)
- **Decision Date:** Sep 20, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The CVSi™ Peripheral Balloon Catheter and the CVSi™ Inflator are indicated for percutaneous, transluminal angioplasty of stenotic lesions in superficial femoral and popliteal arteries. The CVSi™ Peripheral Balloon Catheter System is not indicated for use in other vasculature, including coronary arteries, carotid arteries, synthetic vascular grafts, and vein grafts. The CVSi™ Peripheral Balloon Catheter System is not indicated for stent delivery or stent expansion.

## Device Story

CVSi™ Peripheral Balloon Catheter System performs percutaneous transluminal angioplasty (PTA) of stenotic lesions in superficial femoral and popliteal arteries. System components include a balloon catheter, inflator, connecting cable, and rechargeable battery pack. Inflation medium is liquid nitrous oxide supplied in a disposable 12g cylinder. Operated by physicians in clinical settings. Device uses cryo-technology for balloon inflation/dilatation. Output is mechanical dilatation of stenotic arterial lesions. Clinical benefit includes restoration of blood flow in peripheral arteries. Safety and effectiveness supported by IDE human clinical studies and GLP animal studies.

## Clinical Evidence

Evidence includes human clinical experience from an IDE study and additional case studies, plus GLP-compliant animal studies. Testing confirmed device performance against intended use and established comparative safety and effectiveness relative to predicate devices.

## Technological Characteristics

System includes balloon catheter, inflator, connecting cable, and rechargeable battery. Inflation medium: liquid nitrous oxide (12g disposable cylinder). Biocompatibility: ISO 10993-1 compliant for external communicating device in contact with circulating blood (limited contact <24 hours).

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- FOCUS PV Balloon Dilatation Catheter by CardioVascular Dynamics ([K944016](/device/K944016.md))
- OPTA LP PTA Catheter Cordis/A Johnson & Johnson Company ([K971448](/device/K971448.md))
- ACS 20/20 Indeflator by Guidant/Advanced Cardiovascular Systems, Inc. ([K896849](/device/K896849.md))
- Breeze Digital Balloon Inflation Device by NAMIC USA ([K953676](/device/K953676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## SEP 2 0 2002

CVSi™ Peripheral Balloon Catheter System 510(k) Premarket Notification

## 510(k) Summary CryoVascular Systems, Incorporated CVSi™ Peripheral Balloon Catheter System

This 510(k) summary for the CVSi™ Peripheral Balloon Catheter System is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

#### GENERAL INFORMATION

| Manufacturer:      | CryoVascular Systems, Inc.<br>160 Knowles Drive<br>Los Gatos, California 95032<br>(408) 376-3675 (phone)<br>(408) 376-3677 (fax)<br>Est. Reg. No. (to be applied for) |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:    | R. Michael Crompton<br>Vice President, Regulatory / Clinical Affairs<br>and Quality Assurance                                                                         |
| DEVICE DESCRIPTION |                                                                                                                                                                       |

| Classification:      | Class II                                                                                                  |
|----------------------|-----------------------------------------------------------------------------------------------------------|
| Trade Name:          | CVSi™ Peripheral Balloon Catheter System                                                                  |
| Generic/Common Name: | Balloon Dilatation Catheter (21 CFR § 870.1250)<br>Injector and Syringe, Angiographic (21 CFR § 870.1650) |

#### PREDICATE DEVICES

- FOCUS PV Balloon Dilatation Catheter by CardioVascular Dynamics (K944016) (1)
- OPTA LP PTA Catheter Cordis/A Johnson & Johnson Company (K971448) (2)
- ACS 20/20 Indeflator by Guidant/Advanced Cardiovascular Systems, Inc. (K896849) (3)
- Breeze Digital Balloon Inflation Device by NAMIC USA (K953676) (4)

#### INTENDED USE

The CVSi™ Peripheral Balloon Catheter and the CVSi™ Inflator are indicated for percutaneous, transluminal angioplasty of stenotic lesions in superficial femoral and popliteal arteries.

## DEVICE DESCRIPTION

The CVSi™ Peripheral Balloon Catheter System consists of a CVSi™ Peripheral Balloon Catheter, CVSi™ Inflator, connecting cable, and a rechargeable battery pack with recharging unit and battery receptacle. The inflation medium (liquid nitrous oxide) is provided in a disposable 12 gr cylinder.

{1}------------------------------------------------

## SUBSTANTIAL EQUIVALENCE

The CVSi Peripheral Balloon Catheter System is substantially equivalent to the predicate devices identified previously. The CVSi Peripheral Balloon Catheter System is substantially equivalent to the predicate devices with regard to intended use, function, materials, and sterilization method.

Testing performed on the CVSi Peripheral Balloon Catheter System demonstrates the device is substantially equivalent to the predicate devices and does not raise new questions regarding safety and effectiveness with respect to balloon dilatation catheters.

#### FUNCTIONAL PERFORMANCE TESTING

Functional testing was conducted on the CVSi Peripheral Balloon Catheter System to ensure that the product will function according to its Instructions for Use. All testing conducted confirmed the acceptability of the CVSi Peripheral Balloon Catheter System to perform as intended.

### BIOCOMPATIBILITY EVALUATION

Biocompatibility testing was conducted on the CVSi Peripheral Balloon Catheter System materials to ensure acceptability when used as directed. The CVSi Peripheral Balloon Catheter System materials passed the necessary biocompatibility tests, in conformance with ISO 10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing for an External Communicating Device in contact with Circulating Blood for a contact duration of A-Limited (< 24 hours).

#### HUMAN CLINICAL EXPERIENCE / ANIMAL STUDIES

Human clinical experience collected in conformance with an IDE study and additional case studies were supplied in the 510(k) Premarket Notification to provide reasonable assurance of the safety and effectiveness of the CVSi Peripheral Balloon Catheter System to predicate devices. Data from animal studies conducted in conformance with GLP regulations and additional animal studies were supplied in the 510(k) Premarket Notification, which together with the human clinical data, provided comparative safety and effectiveness informantion of the new characteristics of the device compared to the predicate devices.

## CONCLUSION

As described in this 510(k) Summary, all testing deemed necessary was conducted on the CVSi Peripheral Balloon Catheter System to ensure that the device is safe and effective for its intended use when used in accordance with its Instructions for Use.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 0 2002

CryoVascular Systems, Inc. c/o R. Michael Crompton Vice President, Regulatory/Clinical Affairs and Quality Assurance 160 Knowles Drive Los Gatos, CA 95032

Re: K022061

Trade Name: CVSi™ Peripheral Balloon Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY and LIT Dated: Juen 20, 2002 Received: June 25, 2002

Dear Mr. Crompton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2 - R. Michael Crompton

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in vour Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Elisa Diaz for

Bram D. Zuckerman, M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Statement of Indications for Use

510(k) Number (if known): K ஜூ२० اے (

Device Name: CVSi™ Peripheral Balloon Catheter System

Indications for Use: The CVSi™ Peripheral Balloon Catheter and the CVSi™ Inflator are indicated for percutaneous, transluminal angioplasty of stenotic lesions in the superficial femoral and popliteal arteries. The CVSi™ Peripheral Balloon Catheter System is not indicated for use in other vasculature, including coronary arteries, carotid arteries, synthetic vascular grafts, and vein grafts. The CVSi™ Peripheral Balloon Catheter System is not indicated for stent delivery or stent expansion.

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division of Cardiovascular & Respiratory Devices
510(k) Number K220061

for BD2 3/20/02

Prescription Use (Per 21 C.F.R. § 801.109

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

**Prescription Use**
(Per 21 CFR 801.109)

OR

---

**Source:** [https://fda.innolitics.com/device/K022061](https://fda.innolitics.com/device/K022061)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
