K021963 · Sciences ET Bio Materiaux (Sbm) · MQV · Jan 28, 2003 · Orthopedic
Device Facts
Record ID
K021963
Device Name
BIOSORB RESORBABLE VOID FILLER
Applicant
Sciences ET Bio Materiaux (Sbm)
Product Code
MQV · Orthopedic
Decision Date
Jan 28, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K021963 · Jan 28, 2003
BIOSORB RESORBABLE VOID FILLER
Sciences ET Bio Materiaux (Sbm)
Published retrospective clinical studies
Retrospective clinical literature was used to support the safety and effectiveness of the BIOSORB beta-tricalcium phosphate ceramic bone void filler.
Retrospective study; Clinical literature; Orthopedic surgery
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Use of ß -tricalcium phosphate in foot and ankle surgery: a report of 20 cases; Retrospective clinical study
Patients undergoing foot and ankle surgery; Sample Size: 20 cases
Not applicable for this study
Safety and effectiveness of beta-tricalcium phosphate
Filling of bone defects with tricalcium phosphate beta in traumatology; Retrospective clinical study
Traumatology patients
Not applicable for this study
Safety and effectiveness of beta-tricalcium phosphate
Beta-tricalcium phosphate ceramic as a bone substitute in orthopaedic surgery; Retrospective clinical study
Orthopedic surgery patients
Not applicable for this study
Safety and effectiveness of beta-tricalcium phosphate
Indications for Use
BIOSORB® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure. BIOSORB® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure rigid stabilization of the defect in all plans.
Device Story
BIOSORB is an osseo-conductive, macroporous, synthetic beta-tricalcium phosphate (βTCP) implant used as a bone void filler. It is provided in various shapes (granules, cylinders, blocks, cubes) to fill bony defects. The device features a multidirectional interconnected porosity structure mimicking human cancellous bone. It is implanted by surgeons in clinical settings to fill gaps caused by trauma or surgery. The device is not load-bearing and requires rigid fixation. Over time, the implant resorbs and is progressively replaced by host bone and soft tissue. It serves as a scaffold for bone ingrowth and a local source of calcium for mineralization during the healing process.
Clinical Evidence
Evidence includes bench testing (ASTM F1088-87), biocompatibility testing (ISO 10993: cytotoxicity, sensitization, genotoxicity, systemic toxicity, irritation, pyrogenicity), and animal studies (rabbit condyle model showing >90% osseous rehabilitation by 10 weeks). Clinical evidence includes a human trial demonstrating stable osseous fusion, absence of inflammatory/septic response, and resorption/replacement by bone. Supporting literature includes three retrospective clinical studies on βTCP ceramic in orthopedic and trauma surgery.
Indicated for patients requiring bone void filling in the skeletal system (extremities, spine, pelvis) due to trauma or surgery. Not for use in defects requiring structural stability; rigid fixation recommended.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
VITOSS scaffold synthetic cancellous bone void filler (K994337)
PRO OSTEON 500R Resorbable Bone void filler (K98017)
Submission Summary (Full Text)
{0}------------------------------------------------
| BIOSORB Resorbable Void Filler | SBM (France) |
|---------------------------------------|--------------|
| 510(k) Premarket Notification K021963 | Confidential |
K021963
### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
JAN 2 8 2003
# 1. GENERAL INFORMATION
| Trade Name | BIOSORB® Resorbable Void Filler |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Resorbable Calcium Salt Bone Void Filler |
| Classification Name | Resorbable Calcium Salt Bone Void Filler |
| Class | Unclassified |
| Product Code | MQV |
| CFR section | 888.3045 |
| Device panel | Orthopedic |
| Special controls | As per 21 CFR 888.3045, the following special controls were established for Resorbable Calcium Salt Bone Void Filler: Draft Guidance for Resorbable Calcium Salt Bone Void Filler. FDA Guidance "Use of international standards Organization's ISO 10993 Biological evaluation of medical devices part I: Evaluation and testing" FDA Guidance "510(k) Sterility review Guidance – revision of 2/12/90 (K90-1)" |
| Performance Standard | ASTM F-1088-87 (reapproved 1992) "Standard specification for beta tricalcium phosphate for surgical implantation". |
| Submitter's name and address | Sciences et Bio Matériaux<br>ZI du Monge<br>F 65100 LOURDES - FRANCE |
| Contact | Denis CLEMENT, General Manager<br>Phone : 33 5 62 42 21 01<br>Fax : 33 5 62 42 21 00<br>e-mail : denis.clement.sr@wanadoo.fr |
October 20, 2002 Summary preparation date:
# 2. PREDICATE DEVICES
| Trade Name | VITOSS scaffold synthetic cancellous bone void<br>filler® K 994337<br>PRO OSTEON 500R Resorbable Bone void filler<br>K98017 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Common Name | Resorbable Calcium Salt Bone Void Filler |
| Classification Name | Resorbable Calcium Salt Bone Void Filler |
| Class | unclassified |
| Product Code | MQV |
| CFR section | 888.3045 |
| Device panel | Orthopedic |
{1}------------------------------------------------
| a manager and consistence and consistence and consistence and consistence and the consistence of the consistence of the contribution of the first of the first of the first of | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| BIOSORB Resorbable Void Filler | SBM (France) |
| 510(k) Premarket Notification K021963 | Confidential |
# 3. DEVICE DESCRIPTION
BIOSORB® is an osseo-conductive macroporous implant made of synthetic beta tri Calcium Phosphate (ßTCP) indicated for Bone Void Filling.
BIOSORB® implants are available in various shapes and sizes in order to fill various bone defect. Shapes are basic such as granules, cylinders, blocks and cubes.
BIOSORB® presents a multidirectional interconnected porosity structure, similar to that of the human cancellous bone. BIOSORB® implant slowly resorbs during the remodeling and bone defect repair process and is progressively replaced with bone and soft tissues. The progressive resorption of BIOSORB® resorbable void filler is intended to prevent premature resorption.
# 4. INTENDED USE
BIOSORB® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure.
BIOSORB® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure rigid stabilization of the defect in all plans.
# 5. SAFETY AND EFFECTIVENESS TESTING SUMMARY
# 5.1 Biocompatibility
BIOSORB® conforms to the recognized consensus standard specification, ASTM F 1088-87 (reapproved 1992) for surgically implantable beta tricalcium phosphate. FDA has recognized the use of this consensus standard as verification of material characteristics and biocompatibility for surgical application (Recognition List Number 001, effective date: 19 feb 1998).
Moreover, the biocompatibility of ßTCP implants is well documented. As a biomaterial ß TCP has consistently proven to be non toxic, non allergenic, biocompatible and elicits no inflammation. No adverse svstem effects have been reported.
A wide variety of tests was performed on вюѕояв ®
According to ISO 10993 « Biologic evaluation of medical devices » and to the type of medical device (long-term implantable medical device. bone/tissue contact) the following biologic effects have been investigated: Cytotoxicity, Sensitization, Genotoxicity and mutagenicity, Systemic toxicity, Irritation (intradermic injection), Systemic tolerance, Pyrogenicity Testing performed on BIOSORB ® shows an excellent biocompatibility with no significant adverse observations of any kind.
{2}------------------------------------------------
| And Concession Compress of Children Comments of Children Comments of Children Comments of Children<br>BIOSORB Resorbable Void Filler | SBM (France) |
|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| 510(k) Premarket Notification K021963 | Confidential<br>- Real Property of Children Comments of Children Comments of Children Comments of Children |
### 5.2. Osseous rehabilitation
Histo-morphometric analyses were performed to estimate the percentage of osseous rehabilitation and the ingrowth depth within the pores of BIOSORB cvlinders implanted in rabbit condyles up to four months. New bone formation is clearly evident by 4 weeks and a quasi complete (>90%) rehabilitation was observed by 10 weeks.
#### 5.3. In vivo TCP dissolution
The evolution of the local calcium content around irradiated beta tricalcium phosphate ceramic implants has been evaluated in vivo (rabbit). Results indicate that part of the calcium dissolved from the implant is involved in a local mineralization process. The calcium of resorbable TCP implants represents a store which is probably involved in both local mineralization process during bone healing and circulating calcium pool, like physiological bone mineral.
### 5.4 Clinical data
A human clinical trial has been performed to investigate the safety and effectiveness of BIOSORB® bone void filler and more especially the achievement of a stable osseous fusion in consolidation of bone defects. the absence of inflammatory or septic response and the resorption of the BIOSORB implant and its replacement by osseous tissues.
Clinical and radiological data demonstrate the biocompatibility, the osseo integration and the resorption of BIOSORB ® bone void filler.
Several retrospective clinical studies have been published, supporting the use of BIOSORB beta-tricalcium phosphate ceramic (45% porosity -- 250-400 um pore-size).
- . Use of ß -tricalcium phosphate in foot and ankle surgery: a report of 20 cases in Foot and Ankle Surgery- Volume 7 (4)- 217 - December 2001.
- Filling of bone defects with tricalcium phosphate beta in traumatology. In Ann Chir . 2000 Dec:125(10):972-81
- Beta-tricalcium phosphate ceramic as a bone substitute in orthopaedic surgery in Int . Orthop 2002;26(2):109-15
| Criteria | BIOSORB® | VITOSS® | PRO OSTEON® 500 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|----------------------------------------|
| Indication | Synthetic bone void filler | | |
| Intended use | To fill bony void or gaps of the skeletal system (i.e. the extremities, spine and pelvis) resulting from surgery or trauma that are not intrinsic to the stability of the bony structure. | | |
| Labeling | Same intended use, contra-indications, warnings, precautions and adverse invents as predicate | | |
| Chemical | Calcium salt | | |
| Mineral phase | Beta Tri Calcium<br>Phosphate Ca3(PO4)2 | Beta Tri Calcium<br>Phosphate Ca3(PO4)2 | Calcium Carbonate<br>Calcium Phosphate |
| Performance Std. | ASTM F1088-87 | ASTM F1088-87 | |
| Biocompatibility | Established | | |
| Physical structure | Trabecular structure similar to cancellous bone | | |
| Porosity | Interconnected porosity | | |
# 6. SUBSTANTIAL EQUIVALENCE
{3}------------------------------------------------
| BIOSORB Resorbable Void Filler<br>510(k) Premarket Notification K021963 | SBM (France)<br>Confidential |
|-------------------------------------------------------------------------|------------------------------|
|-------------------------------------------------------------------------|------------------------------|
| Criteria | BIOSORB® | VITOSS® | PRO OSTEON® 500 |
|------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| | 40-50%<br>& 60-80% | 88-92 % | 55% |
| Pore size | 250-400 μm | 1-1000 μm | 435 μm<br>[280-779 μm] |
| Mechanical<br>strength | Does not impart mechanical strength to surgical site | | |
| Resorption | | | |
| | 35% at 6 months and<br>72% at 12 months in<br>human | Significantly by 3<br>months in human<br>(76% at 6 weeks,<br>86% at 12 weeks in<br>Canine) | About 6 months in<br>human<br>20% at 6 weeks, 45%<br>at 12 weeks in<br>Canine) |
| Osseous | | | |
| Rehabilitation | New bone formation<br>clearly evident by 4<br>weeks and quasi<br>complete (>90%) by<br>10 weeks (Rabbit) | New bone formation<br>clearly evident by 6<br>weeks - complete by<br>12 weeks (Canine) | |
| Presentation | Various shapes - Various sizes | | |
| Sterility | Sterilized by Gamma radiation<br>Single use only | | |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human figure with three heads, representing health, services, and people. The figure is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Denis Clement General Manager Sciences et Bio Matériaux (SBM) Z.I. du Monge 65100 Lordes France
JAN 28 2003
Re: K021963
Tade Name: BIOSORB® Resorbable Bone Void Filler Regulatory Class: Unclassified Product Code: MQV Dated: October 28, 2002 Received: October 30, 2002
Dear Mr. Clement:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Denis Clement
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark A. Milleson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| BIOSORB Resorbable Void Filler<br>510(k) Premarket Notification K021963 | SBM (France)<br>Confidential |
|-------------------------------------------------------------------------|------------------------------|
|-------------------------------------------------------------------------|------------------------------|
#### 510(k) Number: K021963
#### Device Name : BIOSORB® RESORBABLE BONE VOID FILLER
#### Indications for Use:
BIOSORB® Resorbable Void Filler is a resorbable calcium salt bone void filler intended to fill bony voids or gaps of the skeletal system (i.e. the extremities, spine and pelvis,) caused by trauma or surgery, that are not intrinsic to the stability of the bony structure.
BIOSORB® Resorbable Void Filler does not possess sufficient mechanical strength to support reduction of a defect prior to soft and hard tissue ingrowth. Rigid fixation techniques are recommended as needed to assure stabilization of the defect in all plans.
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
h Mark N. Milken
ral. Restorative
510(k) Number.
Prescription Use (PER 21 CFR 801.109) or
Over-the-Counter Use
(optional Format 1-2-96)
000002
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.