K021877 · bioMerieux, Inc. · GHH · Aug 7, 2002 · Hematology
Device Facts
Record ID
K021877
Device Name
MDA D-DIMER
Applicant
bioMerieux, Inc.
Product Code
GHH · Hematology
Decision Date
Aug 7, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.7320
Device Class
Class 2
Indications for Use
MDA® D-Dimer is a homogeneous latex particle based immunoassay for the quantitative determination of crosslinked fibrin degradation products containing the D-dimer domain in citrated human plasma. MDA D-Diner is an assay for use in conjunction with a clinical Pre-test Probability Assessment (PTP) model in excluding deep vein thrombosis (DVT) in outpatients suspected of a first episode of DVT. MDA D-Dimer can also be used as and in the assessment and evaluation of patients suspected of pulmonary embolism (PE). The assay is designed for use on the MDA automated coagulation analyzers.
Device Story
MDA D-Dimer is a homogeneous latex particle-based immunoassay for quantitative measurement of cross-linked fibrin degradation products (D-dimer) in citrated human plasma. Used on MDA automated coagulation analyzers in clinical laboratory settings. Principle: latex microparticles coated with specific monoclonal antibodies aggregate in the presence of D-dimer; aggregation rate is photo-optically detected and proportional to D-dimer concentration. Concentration is interpolated from a reference curve. Results aid clinicians in assessing patients suspected of venous thromboembolism (VTE), specifically excluding DVT in low/moderate pre-test probability patients. Benefits include rapid, automated diagnosis, reducing need for invasive procedures or unnecessary treatment.
Clinical Evidence
Prospective cohort study of 556 consecutive outpatients with suspected first DVT episode. Patients classified by pre-test probability (PTP) model. MDA D-Dimer performance (cut-off 0.50 µg FEU/ml) showed 98.2% sensitivity (90-100% CI), 60.4% specificity (56-65% CI), and 99.7% negative predictive value (98-100% CI). Study confirmed utility in excluding DVT when combined with PTP assessment.
Technological Characteristics
Homogeneous latex particle-based immunoassay. Quantitative format. Uses monoclonal antibodies on latex microparticles. Photo-optical detection of aggregation. Designed for use on MDA automated coagulation analyzers. Reference curve range 0-4.0 µg FEU/ml. Sample type: citrated human plasma.
Indications for Use
Indicated for outpatients suspected of a first episode of deep vein thrombosis (DVT) and patients suspected of pulmonary embolism (PE). Used in conjunction with clinical Pre-test Probability Assessment (PTP) models.
{0}------------------------------------------------
# AUG 0 7 2002
### 510(k) SUMMARY
### MDA® D-Dimer
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and the final rule under 21 CFR 807.92 published December 14, 1994.
#### The submitter's name, address, telephone number, a contact person, and the date the summary (۵) (1) was prepared;
Submitter's Name: Submitter's Address: bioMérieux Inc. 100 Rodolphe Street Durbam, North Carolina 27712, USA
Submitter's Telephone: Submitter's Fax: Submitter's Contact: Date 510(k) Summary Prepared: (919) 620-2373 (919) 620-2548 Ron Sanyal June 6, 2002
- The name of the device, including the trade or proprietary name if applicable, the common or (a) (2) usual name, and the classification name, if known;
Trade/Proprietary Name: Common/ Usual Name: Classification Name:
MDA® D-Dimer Fibrin Degradation Product Fibrin Degradation Product
- (a) (3) An identification of the legally marketed device to which the submitter claims substantial equivalence.
Device Equivalent to:
l.
MDA® D-Dimer (K000492)
{1}------------------------------------------------
#### A description of the device(System) (a) (4)
bioMerieux Inc. MDA® D-Dimer is a homogeneous latex particle based immunoassay for the quatitative determination of cross-linked fibrin degradation products containing the D-dimer domain in citrated human plasma.
D-dimer containing fibrin degradation products (FbDP) fragments are released when cross-linked fibrin is degraded by plasmin. Cross-linked fibrin is formed when fibrionogen is cleaved by thrombin to form fibrin monomers, which then spontaneously polymerize and are cross-linked by Factor XIIIa. Thrombin is required to cleave fibringen as well as to activate Factor XIII. Plasmin formation is triggered when a fibrin clot is formed. Plasmin degrades some of the cross-linked fibrin and the resulting level of D-dimer is, therefore, an indirect measure of thrombin generation and subsequent clot formation.
D-dimer is elevated in disseminated intravascular coagulation (DIC), deep vein thrombosis (DVT), pulmonary embolism (PE), sickle cell crisis, pre-eclampsia, some cause of unstable angina, myocardial infarction, some cancers, and following major surgery or trauma.
MDA D-Dimer is a quantitative homogeneous-phase immunoassay using latex microparticles to photo-optically detect binding of specific monoclonal antibody to D-dimer. These latex particles aggregate in the presence of fibrin derivatives containing the D-dimer domain. The rate of latex microparticle aggregation is proportional to the concentration of D-dimer in the sample. D-dimer concentration may be interpolated from a reference curve.
#### (a) (5) A statement of the intended use of the device.
MDA D-Dimer is a homogeneous latex particle based immunoassay for the quantitative determination of cross-linked fibrin degradation products containing the D-dimer domain in citrated human plasma. MDA D-Dimer is an assay for use in conjunction with a clinical Pre-test Probability Assessment (PTP) model in excluding deep vein thrombosis (DVT) in outpatients suspected of a first episode of DVT. MDA D-Dimer can also be used as an aid in the assessment and evaluation of patients suspected of pulmonary embolism (PE). The assay is designed for use on the MDA automated coagulation analyzers.
6
{2}------------------------------------------------
## (a) (6) A summary of the technological characteristics of the new device in comparision to those of the predicate device.
The technological characteristics of the device MDA® D-Dlmer in comparison to those of the 510(k) cleared device MDA® D-Dimer assay (K000492) are given in the table 1 below.
| Category | MDA® D-Dimer assay (K000942) | MDA® D-Dimer |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Medical Device | Yes | Yes |
| Intended Use | MDA® D-Dimer is a homogeneous latex<br>particle based immunoassay for the<br>quantitative determination of cross-linked<br>fibrin degradation products containing the<br>D-dimer domain in citrated human<br>plasma. MDA D-Dimer can be used to<br>aid in the assessment and evaluation of<br>patients suspected of venous<br>thromboembolism (VTE), which is<br>comprised of deep vein thrombosis<br>(DVT) and pulmonary embolism (PE).<br>The assay is designed for use on the<br>MDA automated coagulation analyzers. | MDA® D-Dimer is a homogeneous latex<br>particle based immunoassay for the<br>quantitative determination of cross-linked<br>fibrin degradation products containing the<br>D-dimer domain in citrated human<br>plasma. MDA D-Dimer can be used to aid<br>in the assessment and evaluation of<br>patients suspected of venous<br>thromboembolism (VTE), which is<br>comprised of deep vein thrombosis<br>(DVT) and pulmonary embolism (PE).<br>The assay is designed for use on the MDA<br>automated coagulation analyzers. |
| Regulatory Class | Class II | Class II |
| Product Code | DAP | DAP |
| Classification Panel | Hematology | Hematology |
| C.F.R. Section | 21 CFR 864.7320 | 21 CFR 864.7320 |
| Presentation | Automated Latex Immunoassay | Automated Latex Immunoassay |
| Format | Quantitative | Quantitative |
| Instrument | MDA® automated coagulation analyzers | MDA® automated coagulation analyzers |
| Reagents | Same | Same |
| Principle of the<br>procedure | Same | Same |
| Quality Control | Same | Same |
| Test Procedure | Same | Same |
| Reference Curve<br>Range | 0-4.0 µg FEU/ml | 0-4.0 µg FEU/ml |
Table 1
{3}------------------------------------------------
- (b) (1) A brief discussion of the nonclinical tests submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalency.
Not Applicable
#### A brief discussion of the clinical tests submitted, reference, or relied on in the premarket (b)(2) notification submission for a determination of substantial equivalency.
### Comparison Data:
### Performance Characteristics
#### Specificity
MDA D-Dimer Latex Reagent aggregates in the presence of cross-linked fibrin degradation products D-dimer and D-dimer E.
### Accuracy
Results from MDA D-Dimer reagents obtained on an MDA were compared with a commercially available assay (Fibrinostika® FbDP EIA)for detection of crosslinked and non-crosslinked fibrin degradation product containing the D-dimer. Specimens were tested in duplicate according to NCCLS Approved Guideline EP9-A. " The following results for slope, intercept and correlation were observed for linear least squares regression comparing MDA D-Dimer (y-axis) and the reference method (x-axis):
| Reference Method | n | Slope | Intercept | r |
|------------------------|-----|-------|-----------|------|
| Fibrinostika® FbDP EIA | 175 | 1.005 | 0.293 | 0.91 |
#### Precision
Total precision and within-run precision for the MDA D-Dimer assay were determined in accordance with NCCLS Tentative Guideline EPS-T2.30 Controls were tested in duplicate on an MDA instrument twice daily. Data were collected for 20 days, with a minimum of 40 runs and 80 measurements at each control level. The following precision was observed:
| Sample | Mean<br>(ug FEU/ml) | SD (within-run)<br>(ug FEU/ml) | CV (within-run)<br>(%) | SD(total)<br>(ug FEU/ml) | CV(total)<br>(%) |
|----------------------------------|---------------------|--------------------------------|------------------------|--------------------------|------------------|
| Positive Control | 1.51 | 0.06 | 3.83 | 0.10 | 6.67 |
| MDA Verify 1<br>(Normal Control) | 0.28 | 0.02 | 6.97 | 0.04 | 12.65 |
Image /page/3/Picture/13 description: The image contains a handwritten symbol that resembles a figure eight with a loop at the top. The symbol is drawn with thick, dark lines, giving it a bold appearance. The overall impression is that of a quick, stylized sketch of the number eight.
{4}------------------------------------------------
### Clinical Performance
A multi-center, prospective cohort study was designed to validate the diagnostic utility of the MDA D-Dimer assay to exclude a diagnosis of deep vein thrombosis (DVT).» Consecutive eligible outpatients (n = 556) with a first suspected DVT episode were evaluated at three hospitals. Using a previously validated standardized clinical model » to estimate the probability of DVT, patients were classified as having a high, moderate, or low pre-test- probability -(PTP) of DVT, and had an MDA D-Dimer test performed on presentation.
The D-Dimer assay was performed without knowledge of the PTP assessment results. A clinical cut off value of 0.50 ug FEU/ml previously validated in an accuracy study. A D-Dimer result of >0.50 µg FEU/ml was considered positive, and a result of <0.50 µg FEU/ml was considered negative.
The overall prevalence of DVT in the total population studied was 10.1% (56/556). The sensitivity, specificity and negative predictive value of the MDA D-Dimer assay for a clinical cut off of 0.50 µg FEU/n1 are summarized below with the corresponding 95% confidence interval (CI).
| Patients | N | % Clinical Sensitivity<br>(95% CI) | % Clinical Specificity<br>(95% CI) | % Negative Predictive<br>Value (95% CI) |
|---------------|-----|------------------------------------|------------------------------------|-----------------------------------------|
| Suspected DVT | 556 | 98.2<br>(90-100 CI) | 60.4<br>(56-65 CI) | 99.7<br>(98-100 CI) |
This study was designed as a management clinical trial and patients were grouped according to PTP. Those patients having a negative MDA D-Dimer test result and a low or moderate PTP of DVT underwent no further diagnostic testing and were followed up for 3 months for development of DVT. Patients with a positive MDA D-Dimer and/or high PTP underwent serial compression ultrasound (CUS).
The results of this study suggest that a negative MDA D-Dimer test result in conjunction with Wells clinical Pre-test probability Assessnent (PTP) model excludes clinically significant DVT in out patients with a first suspected episode. In addition, this test provides a rapid, automated diagnostic tool and eliminates the need to expose patients to invasive procedures or unnecessary treatment.
#### The conclusion drawn from the nonclinical and clinical tests that demonstrate that the device is (b) (3) as safe, as effective, and performed as well or better than the legally marketed device identified in (a) (3).
In conclusion, the MDA® D-Dimer has successfully met all aspects of non-clinical and clinical testing and have demonstrated that the device is safe and effective and has performed well and is substantially equivalent to the legally marketed device MDA® D-Dimer assay (K000492),
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is black and appears to be on a white background. The words are all capitalized and evenly spaced.
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Ron Sanyal, M. Pharm, COE, RAC Acting Head of Regulatory Affairs BioMerieux, Inc. 100 Rodolphe Street Durham, North Carolina 27712
Re: k021877
> Trade/Device Name: MDA® D-Dimer Regulation Number: 21 CFR § 864.7320 Regulation Name: Fibrin Degradation Product Regulatory Class: II Product Code: GHH Dated: June 6, 2002 Received: June 7, 2002
Dear Mr. Sanyal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
AUG 0 7 2002
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
.
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page
page 1 of 1
(IF KNOWN): 510 (K) NUMBER DEVICE NAME:
5.0 INDICATIONS FOR USE:
MDA® D-Dimer is a homogeneous latex particle based immunoassay for the quantitative determination of crosslinked fibrin degradation products containing the D-dimer domain in citrated human plasma. MDA D-Diner is an assay for use in conjunction with a clinical Pre-test Probability Assessment (PTP) model in excluding deep vein thrombosis (DVT) in outpatients suspected of a first episode of DVT. MDA D-Dimer can also be used as and in the assessment and evaluation of patients suspected of pulmonary embolism (PE). The assay is designed for use on the MDA automated coagulation analyzers.
Josephine Bautista
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter-Use (Optional Format 1.- 2-96)
4
GAY
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.