← Product Code [FGE](/productcode/FGE) · K021729

# MODIFICATION TO PROTEGE SELF-EXPANDING NITINOL STENT WITH STARPORT DELIVERY TECHNOLOGY (K021729)

_Sulzer Intratherapeutics, Inc. · FGE · Jun 21, 2002 · Gastroenterology, Urology · SESU_

**Canonical URL:** https://fda.innolitics.com/device/K021729

## Device Facts

- **Applicant:** Sulzer Intratherapeutics, Inc.
- **Product Code:** [FGE](/productcode/FGE.md)
- **Decision Date:** Jun 21, 2002
- **Decision:** SESU
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5010
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Protégé™ Self-Expanding Nitinol Stent with StarPort™ Delivery Technology (Biliary Indication) is intended as a palliative treatment of malignant neoplasms in the biliary tree.

## Device Story

Self-expanding nitinol stent; used for palliative treatment of malignant neoplasms in biliary tree. Stent lengths 20-80 mm (plus new lengths >80 mm) and diameters 6-12 mm. Premounted on 80 or 120 cm over-the-wire delivery catheter for transhepatic access. StarPort delivery system features inner/outer sheath with safety lock. Upon deployment, stent exerts constant outward force to maintain biliary duct patency. Used by clinicians in clinical settings. Not for vascular use.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and performance comparisons.

## Technological Characteristics

Self-expanding nitinol (nickel-titanium alloy) stent; electropolished. Dimensions: 20-80+ mm length, 6-12 mm diameter. Delivery system: 80 or 120 cm over-the-wire catheter with inner/outer sheath and safety lock. Mechanical expansion mechanism.

## Regulatory Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Protégé™ Self-expanding Nitinol Stent with StarPort™ Delivery Technology ([K012066](/device/K012066.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 1

## Premarket Notification (510(k)) Summary

| 510(k) Number:            | K021729                                                                                        |
|---------------------------|------------------------------------------------------------------------------------------------|
| Product Name:             | Protégé™ Self-expanding Nitinol Stent with StarPort™ Delivery Technology (Biliary Indication)  |
| Common Name:              | biliary stent                                                                                  |
| Class:                    | Class II, 21 CFR 876.5010                                                                      |
| Submitter's Name:         | Sulzer IntraTherapeutics Inc.<br>651 Campus Drive<br>St. Paul, MN 55112                        |
| Official Contact:         | Maria E. Brittle<br>Regulatory Affairs Manager<br>Telephone: 651-697-2018<br>Fax: 651-697-4808 |
| Summary Preparation Date: | May 23, 2002                                                                                   |

This summary is provided in compliance with section 513(D)(3)(A) of the Act and summarizes the safety and effectiveness information contained in this premarket notification submission for a modification to the Protége™ Self-expanding Nitinol StarPort™ Delivery Technology.

The Protége™ Self-expanding Nitinol Stent with StarPort™ Delivery Technology is intended as a palliative treatment of malignant neoplasms in the biliary tree.

The Protége™ stent is a self-expanding nitinol (nickel-titanium alloy) stent provided in multiple lengths, (20 - 80 mm) and diameters (6 - 12 mm). The stent is electropolished. The stent is premounted on an 80 or 120 cm working length, 0.035" over-the-wire delivery catheter designed for transhepatic access. Upon deployment, the stent achieves its predetermined diameter and exerts a constant, gentle outward force to establish patency in the biliary ducts. The disposable StarPort Delivery Technology system is comprised of an inner and outer sheath, which are locked together with a safety lock.

The modified Protégé™ Self-expanding Nitinol Stent with StarPort™ Delivery Technology is substantially equivalent to the currently marketed Protégé™ Self-expanding Nitinol Stent with StarPort™ Delivery Technology (K012066) in intended use, materials, technological characteristics and performance. Stent lengths > 80 mm have been added to further expand the product offering.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines representing the wings and a stylized head and body.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 1 2002

Ms. Maria E. Brittle Regulatory Affairs Manager Sulzer IntraTherapeutics, Inc. 651 Campus Drive ST. PAUL MN 55112

Re: K021729

> Trade/Device Name: Protégé™ Self-Expanding Nitinol Stent with StarPort™ Delivery Technology (Biliary Indication) Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: May 23, 2002 Received: May 24, 2002

Dear Ms. Brittle:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{2}------------------------------------------------

## Page 2 – Ms. Maria E. Brittle

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Sincerely yours,

Bernard E. Hart

Bernard E. Statland, M.D., Ph.D. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page_1 of 1

510(k) Number (if known): K021729

Device Name: Protégé™ Self-Expanding Nitinol Stent with StarPort™ Delivery Technology (Biliary Indication)

FDA's Statement of the Indications For Use for device:

The Protégé™ Self-Expanding Nitinol Stent with StarPort™ Delivery Technology (Biliary Indication) is intended as a palliative treatment of malignant neoplasms in the biliary tree.

Prescription Use レ OR (Per 21 CFR 801.109)

Over-The-Counter Use

David L. Segmon

Division Sign-

grass

---

**Source:** [https://fda.innolitics.com/device/K021729](https://fda.innolitics.com/device/K021729)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
