← Product Code [KWZ](/productcode/KWZ) · K021661

# RINGLOC CONSTRAINED LINERS (K021661)

_Biomet Orthopedics, Inc. · KWZ · Oct 25, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K021661

## Device Facts

- **Applicant:** Biomet Orthopedics, Inc.
- **Product Code:** [KWZ](/productcode/KWZ.md)
- **Decision Date:** Oct 25, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3310
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Ringloc® Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.

## Device Story

Ringloc® Constrained Acetabular Liner; polyethylene liner with 10° beveled inner edge and locking ring; captures modular femoral head to prevent dislocation. Used in total hip arthroplasty; indicated for high-risk patients (e.g., prior dislocation, neuromuscular disease, instability). Implanted by orthopedic surgeons in clinical/hospital settings. Device provides mechanical constraint to stabilize hip joint; reduces risk of recurrent dislocation in complex revision or primary cases.

## Clinical Evidence

Device evaluated under IDE #G990138. Clinical data referenced in annual progress report submitted to FDA on March 8, 2002.

## Technological Characteristics

Polyethylene acetabular liner with 10° beveled inner edge and locking ring mechanism. Metal/polymer constrained prosthesis. Class II device (21 CFR 888.3310).

## Regulatory Identification

A hip joint metal/polymer constrained cemented or uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device prevents dislocation in more than one anatomic plane and has components that are linked together. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular component made of ultra-high-molecular-weight polyethylene with or without a metal shell, made of alloys, such as cobalt-chromium-molybdenum and titanium alloys. This generic type of device is intended for use with or without bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis.”

## Predicate Devices

- Johnson & Johnson S-ROM® Poly-Dial Constrained Liner ([P960054](/device/P960054.md))
- Osteonics Constrained Acetabular Insert ([P960047](/device/P960047.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 25 2002

BIOMET

KC2K-1
page 1 of 2

OCT 2 5 2002

## SUMMARY OF SAFETY AND EFFECTIVENESS

| Applicant or Sponsor: | Biomet Orthopedics, Inc.<br>P.O. Box 587<br>Warsaw, IN 46581-0587 |
|-----------------------|-------------------------------------------------------------------|
| Contact Person:       | Dalene T. Binkley<br>Telephone: (219) 267-6639                    |
| Proprietary Name:     | Ringloc® Constrained Liner                                        |
| Common Name:          | Constrained Acetabular Insert                                     |

Classification: Prosthesis, hip, constrained, metal/polymer (CFR 888.3310).

Device Classification: Class II

Legally Marketed Device to which Substantially Equivalence is Claimed: Johnson & Johnson S-ROM® Poly-Dial Constrained Liner PMA #P960054 and Osteonics Constrained Acetabular Insert PMA #P960047.

Device Description: The Ringloc® Constrained Acetabular Liners are polyethylene liners with a 10° beveled inner edge that, with a locking ring, captures the modular head.

Indications for Use: The Ringloc® Constrained Liners are indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intra-operative instability, and for whom all other options to constrained acetabular components have been considered.

Summary of Technologies: The Ringloc® Constrained Liner-the materials, design, sizing, and indications are similar or identical to the predicate devices.

Non-Clinical Testing: Mechanical testing, published medical literature, and engineering justifications determined that the Ringloc® Constrained Liner presented no new risks and is, therefore, substantially equivalent to the predicate device.

> MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587

> > .

SHIPPING ADDRESS 56 E. Bell Drive Warsaw, IN 46582

()[FFICE 574.267.6639

FAX 574.267.81.37

E-MAIL biomet@biomet.com

{1}------------------------------------------------

.c.x
1. 196

Clinical Testing: This device was part of the Ringloc® Constrained Liner Clinical Study- IDE #G990138. Please refer to the clinical data in the annual progress report that was submitted to the FDA on March 8, 2002.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a symbol that resembles three stylized human figures or profiles facing to the right, stacked on top of each other.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# OCT 2 5 2002

Ms. Dalene T. Binkley Regulatory Affairs Specialist Biomet, Inc. P.O. Box 587 Warsaw. Indiana 46581-0587

Re: K021661

Trade Name: RingLoc® Constrained Liner Regulation Number: 21 CFR 888.3310 Regulation Name: Hip joint metal/polymer constrained cemented or uncemented prosthesis Regulatory Class: II Product Code: KWZ Dated: August 19, 2002 Received: August 20, 2002

Dear Ms. Binkley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Ms. Dalene T. Binkley

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sincerely yours,

L. Mark McMillan

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510 (k) NUMBER (IF KNOWN):

DEVICE NAME: Ringloc® Constrained Liner

### INDICATIONS FOR USE:

The Ringloc® Constrained Liner is indicated for use as a component of a total hip prosthesis in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability, and for whom all other options to constrained acetabular components have been considered.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ***_***_
(Per 21 CFR 801.109)
OR

(Division Sign-Off)
Division of General, Restorative
and Neurological Devices

Over-The-Counter-Use (Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K021661](https://fda.innolitics.com/device/K021661)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
