SONOSITE HAND-CARIED UNTRASOUND SYSTEM (POWERSERIES)
Applicant
Sonosite,Inc.
Product Code
IYN · Radiology
Decision Date
May 31, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Diagnostic ultrasound imaging or fluid flow measurement of the human body as follows: Fetal - OB/GYN, Laparoscopic, Intraoperative (abdominal organs and vascular), Abdominal, Small Organ (breast, thyroid, testicle), Trans-vaginal, Trans-rectal, Musculo-skeletal (conventional), Musculo-skeletal (superficial), Neonatal Cephalic, Pediatric, Cardiac (adult), Cardiac (pediatric), Peripheral Vessel.
Device Story
Highly portable, software-controlled ultrasound system; acquires/displays real-time 2D, PW/CW Doppler, Color Power Doppler, and PowerMap Directional Color Power Doppler. Inputs: ultrasound signals from integrated transducers (linear, curved, intracavitary arrays). Processing: piezoelectric transducers convert reflected sound waves to electrical signals; system processes signals for display, measurement, and calculation (anatomical/fetal biometry). Outputs: real-time images, Doppler audio, cine review, measurements. Used in clinical settings by physicians/clinicians for diagnostic assessment and interventional guidance (needle/catheter placement). ECG input used for timing diastolic/systolic function (not for rhythm monitoring). Benefits: portable diagnostic imaging and procedural guidance.
Clinical Evidence
Bench testing only. Device conforms to recognized standards including EN 60601-1, ISO 10993 (biocompatibility), and AIUM/NEMA acoustic output standards. No clinical study data presented.
Technological Characteristics
Portable ultrasound system; piezoelectric transducers (linear, curved, intracavitary arrays); 2.0-10.0 MHz frequency range. Modes: 2D, M-mode, PW/CW Doppler, Color Power Doppler, PowerMap. Connectivity: integrated display, audio output, printing/recording. Software-controlled. Conforms to AIUM/NEMA acoustic output standards (MI max 1.9, ISPTA max 720 mW/cm2). Biocompatible patient-contact materials.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow measurement in adult and pediatric patients, including neonates. Applications include abdominal, fetal, OB/GYN, cardiac, vascular, small organ, musculoskeletal, and interventional guidance (needle/catheter placement). Contraindications: None stated.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
MAY 3 1 2002
K021628
1 of 6
## 510(K) Summary Of Safety And Effectiveness
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with Title 21, Part 807, Section 92.
### Submitter's name, address, telephone number, contact person: 1)
Michael A. Hoffman Director - Regulatory Affairs and Quality Systems SonoSite, Inc.
21919 30th Drive SE Bothell, WA 98021-3904
(425) 951 - 1297
E-mail: michael.hoffman@sonosite.com
Date prepared: April 24, 2002
### Name of the device, including the trade or proprietary name if applicable, 2) the common or usual name, and the classification name, if known:
Common/ Usual Name
Diagnostic Ultrasound System with Accessories
Proprietary Name
SonoSite Hand-Carried Ultrasound System (PowerSeries) (subject to change)
{1}------------------------------------------------
K021628
2076
### Classification Names
| | FR Number | Product Code |
|---------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging<br>System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging<br>System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
### 3) Identification of the predicate or legally marketed device:
SonoSite, Inc. believes that System described in this Submission is substantially equivalent to predicate devices that include the SonoSite Hand-Carried Ultrasound System (K014116) and (K010374), the Advanced Technology Laboratories (ATL) HDI® 5000 Ultrasound System (K961459), and the Dymax Corp.Site-Rite 3 Ultrasound Scanner (K993624). Where applicable, this new Submission references sections of K014116 to signify that the sections in this Submission remain the same as the predicate device.
#### 4) Device Description:
The SonoSite Hand-Carried Ultrasound System (PowerSeries) is a highly portable, software controlled ultrasound system used to acquire and display high-resolution, real-time ultrasound data in 2D and other modes as Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Power Doppler, and PowerMap™ Directional Color Power Doppler, or a combination of these modes with 2D. These modes of operation are dependent on the type of transducer that is attached to the system, with the exception of 2D.
The System provides measurement capabilities for anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The System may include PW and CW Doppler audio output feature and cine review, image zoom, labeling, biopsy, non-diagnostic ECG capability for the collection of data for M-mode measurements, measurements and calculations, image storage and review, printing, and recording capabilities. Additionally, the system provides ultrasound guidance to assist in the placement of needles and catheters in vascular or other anatomical structures.
The SonoSite Hand-Carried Ultrasound System (PowerSeries) may have curved or linear transducers of the types and frequency listed in the table below. All actions affecting the performance of the transducer are activated from the main system control panel. The transducer is not detachable from the System.
{2}------------------------------------------------
| - محمد الموارث الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموقع الموق | י מי | ্ |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|--------------|
| 2000 | 8 | ﮯ ﮐﮩﺎ ﮐﮧ ﺍﯾﮏ |
| 20 | 1<br>15 | |
| Frequency Range: | 2.0 - 10.0 MHz |
|-------------------|-----------------------------------------------------|
| Transducer Types: | Linear array<br>Curved array<br>Intracavitary array |
The SonoSite Hand-Carried Ultrasound System is designed to comply with the standards listed below.
| EN 60601-1:1997 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CAN/CSA C22.2, No. 601.1:1990 + Amendments |
| UL 2601-1:1997 |
| JIS-T-100X-Series |
| EN 60601-1-1:1993 + Amendments |
| EN 60601-1-2:1993 + Amendments |
| RTCA/D0160D: 1997 |
| ISO 10993 Biocompatibility Series |
| ANSI/AAMI EC53:1995 + Amendments, except for sections 4.4 and 4.5.9 |
| CEI/IEC 61157:1992 |
| Medical Ultrasound Safety, American Institute of Ultrasound in Medicine (AIUM),<br>1994 |
| Standard for Real-Time Display of Thermal and Mechanical Acoustic Output<br>Indices on Diagnostic Ultrasound Equipment, American Institute of Ultrasound in<br>Medicine, 1998 |
| Acoustic Output Measurement and Labeling Standard for Diagnostic Ultrasound<br>Equipment, American Institute of Ultrasound in Medicine, 1993 |
| Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment,<br>NEMA UD2-1998 |
## 5) Intended Use:
The intended clinical applications of the SonoSite Hand-Carried Ultrasound System (PowerSeries), as defined by FDA guidance documents are below. These applications are cumulative, as specific applications are dependent on the transducer attached to the System.
| Fetal - OB/GYN | Musculo-skeletal (conventional) |
|---------------------------------------------------|---------------------------------|
| Laparoscopic | Musculo-skeletal (superficial) |
| Intraoperative (abdominal organs and<br>vascular) | Neonatal Cephalic |
| Abdominal | Pediatric |
| Small Organ (breast, thyroid, testicle) | Cardiac (adult) |
| Trans-vaginal | Cardiac (pediatric) |
| Trans-rectal | Peripheral Vessel |
{3}------------------------------------------------
K021628
4 of 6
Typical examinations performed using the SonoSite Hand-Carried Ultrasound System (PowerSeries) may include the following. Again, the clinical applications themselves, and the modes of operation discussed within a clinical application, are dependent on the transducer attached to the system.
## Abdomen:
This system transmits ultrasound energy into the upper and lower quadrants of the abdomen of an adult or pediatric patient to obtain 2D, CPD, PowerMap DCPD, Tissue Harmonic Imaging, or PW and CW Doppler images, which can be used to assess the presence and extent of some diseases and injuries.
## Small Parts And Superficial Structures:
This system transmits ultrasound energy into the superficial structures of body to obtain 2D, CPD, or PW Doppler images of normal structure and some pathologies of the breast, thyroid, superficial soft tissue, shoulder joints, wrist, ankle, and knee, which can be used to assess the presence and extent of some diseases and iniuries.
## Pediatric:
This system transmits ultrasound energy into the abdomen, pelvis and superficial structures of pediatric patients to obtain 2D, CPD, PowerMap DCPD, Tissue Harmonic Imaging, or PW and CW Doppler images of the abdominal organs, great vessels, pelvic structures, and pediatric hips, which can be used to assess the presence and extent of some diseases and injuries.
## Cardiac:
This system allows the clinician to perform focused cardiac studies. This system transmits ultrasound energy into the thorax of adult and pediatric patients to obtain 2D, PowerMap DCPD, M-mode, Tissue Harmonic Imaging, or PW and CW Doppler images of the heart, great vessels, and anatomic or pathologic structures. This system can be used to assess overall cardiac performance and size, determine the presence and location of fluid around the heart and lungs, aid in pericardialcentesis and pleuralcentesis procedures, and visualize blood flow through cardiac valves. Also the system can be used to assess the presence and extent of some injuries and diseases. As available, the ECG is used for accurate timing of diastolic and systolic function. The ECG trace is not used to diagnose cardiac rhythms and is not designed for long term cardiac rhythm monitoring.
{4}------------------------------------------------
K621628
5 of 6
### Neonates:
Abdomen, cranium, pelvis and heart in neonates that weigh less than 1500 / kodoment or are less than 32 weeks gestation: This system transmits ultrasound gramo or are toos than, abdomen, pelvis soft tissue or heart of patients to obtain onergy mis and chanian DCPD, M-mode, or PW and CW Doppler images. These images will be used to assess the presence and extent of some diseases or injuries. Some examples of the pathology that ultrasound is used for include: Cranium - hemorrhage, ischemia, shunt placement and dilated ventricles; Organs - Thomormage, Is disease, gallbladder disease; Pelvis - ovarian / lottominal organo - athology and testicular disease; Soft tissue/superficial tissue - hip, lymph nodes and superficial cysts.
## GYN/Infertility:
This system transmits ultrasound energy into the lower abdomen or vagina of a female patient to obtain 2D, CPD, PowerMap DCPD, M-mode, Tissue Harmonic Imaging, or PW Doppler images of the reproductive system, which can be used to assess the presence and extent of disease in the female pelvic organs, monitor ovarian follicle size, and as an aid in chorionic villi sampling (CVS) procedures.
### Obstetrics:
This system transmits ultrasound energy into the abdomen or vagina of a pregnant woman to obtain 2D, M-mode, Tissue Harmonic Imaging, or PW Doppler images of a fetus, which can be used to estimate gestational age, number and weight, and assess the presence and extent of disease and confirm viability. CPD or PowerMap DCPD imaging is intended for high-risk pregnant women. High risk pregnancy indications include, but are not limited to multiple pregnancy, fetal hydrops, placental abnormalities, as well as maternal hypertension, diabetes, and lupus.
## WARNING:
CPD or PowerMap DCPD images can be used as an adjunctive method, not as a screening tool, for the detection of structural anomalies of the fetal heart and as an adjunctive method, not as a screening tool for the diagnosis of Intrauterine Growth Retardation (IUGR).
## Prostate:
This system transmits ultrasound energy through the prostate of an adult patient to obtain 2D, CPD, or PW Doppler images of structures, which can be used to assess the presence and extent of disease or injury. Measurements are available to calculate the volume of the prostate gland.
{5}------------------------------------------------
K0216.28
60fb
## Vascular:
This system transmits ultrasound energy into various parts of the body using 2D, CPD, PowerMap DCPD, Tissue Harmonic Imaging, or PW Doppler to obtain ultrasound images. The most common structures imaged are carotid arteries, untrasound images. The most common causels in the abdomen, and peripheral line decep volns in the anno and adjology purposes. The system provides ultrasound access, and intervontional fully in the placement of needles and catheters in vascular or other anatomical structures.
#### Technological Characteristics: 6)
This device operates identically to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and waves into the electrical signals that are processed and displayed as 2D or Mmode images. Doppler shift caused by blood flow is displayed, as available, as mode linager as spectrum analysis. The cumulative modes of this device (2D, PW Doppler, CW Doppler, Color Power Doppler, and PowerMap™ Directional Color Power Doppler) are the same as a combination of the predicate devices identified in item 3. Transducer patient contact materials are biocompatible.
This device conforms to the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment (AIUM/NEMA, 1998). A user education program provides additional information so users may moderate the system's acoustic output in accordance with the ALARA (as low as reasonably achievable) principle.
The device's acoustic output limits are:
| ISPTA (d) | 720 mW/cm² (Maximum) |
|-----------------------|-----------------------|
| TIS/TIB/TIC | 0.1 - 4.0 (Range) |
| Mechanical Index (MI) | 1.9 (Maximum) |
| ISPPA (d) | 0 - 700 W/cm2 (Range) |
All applications:
The limits are the same as predicate Track 3 devices.
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
**MAY 31 2002**
SonoSite, Inc. % Mr. Mark Job Responsible Third Party Official TÜV Product Service 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891
Re: K021628
Trade Name: SonoSite Hand-Carried Ultrasound System (PowerSeries) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: 90 IYN and IYO Dated: May 16, 2002 Received: May 17, 2002
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SonoSite Hand-Carried Ultrasound System, as described in your premarket notification:
Transducer Model Number
C15/4-2, 4.0 - 2.0 MHz Curved Array L25/10-5, 10.0-5.0 MHz Linear Array C11/7-4, 7.0 - 4.0 MHz Curved Array ICT/7-4, 7.0 - 4.0 MHz Intracavitary Transducer L38/10-5, 10.0-5.0 MHz Linear Array
{7}------------------------------------------------
## C60/5-2, 5.0 - 2.0 MHz Curved Array · L52/10-5, 10.0 -5.0 MHz Linear Array
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device. vou submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G. (enclosed) of the Center's September 30. 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
{8}------------------------------------------------
If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{9}------------------------------------------------
K021628
System: .
SonoSite Hand-Carried Ultrasound System (PowerSeries)
Intended Use:
Diagnostic ultrasound imaging or fluid flow measurement of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|----------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Pediatric | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-vaginal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Cardiac Pediatric | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 |
| | Other (spec.) | | | | | | | |
N= new Indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other indludes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Tirectional Color Power Doppler, 3-D Integing, Tissue Harmonic Imaging for guidance of biopsy previously cleared through : 10(k) K014116. Included with this 510k are imaging to assist in the placement of needles and calheters in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number
RA
{10}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries) System:
Transducer:
Diagnostic ultrasound imaging or fluid flow measurement of the human Intended Use: body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|----------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| Cardiac | Cardiac Pediatric | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Note 1. Other holded Golor Power Joseph Joseph Joseph for guidance of biopsy previously cleared through 510(k) K014116. Included with this 510k are imaging to assist in the placement of needles and catheters in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
# Indications for Use
Division of Reproductive, Abdominat,
and Radiological Devices K021628
510(k) Number
{11}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries) System:
L25/10-5 10.0-5.0 MHz Linear Array Transducer:
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | | Mode of Operation | | | | |
|--------------------------|----------------------------------------------------|---|---|-------------------|-----|------------------|---------------------|------------------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| (Track I only) | (Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
### Additional Comments:
Note 1: Other Includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy previously cleared through 510(k) K014116. Included with this 510k are imaging to assist in the placement of needles and cathers in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
(Division Sign-Off Division of Reproductive, and Radiological Devices 510(k) Number
Indications for Use
{12}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries)
C11/7-4 7.0 - 4.0 MHz Curved Array Transducer:
System:
Diagnostic ultrasound imaging or fluid flow analysis of the human body Intended Use: as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|---------------------------|----------------------------------------------------|---|---|-----|-----|------------------|----------------------|------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | | | | | | | | |
| | Abdominal | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 | |
| | Intra-operative (Abdominal<br>organs and vascular) | | | | | | B+M | Note 1 | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | P | | B+M; B+PWD;<br>B+CWD | Note 1 | |
| | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | | |
| | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | P | P | P | P | | B+M; B+PWD:<br>B+CWD | Note 1 | |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Note 1. Other Thouse, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy previously cleared through 510(k) K014116. Induded with this 510k are imaging to assist in the placement of needles and catheters in vascular of other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive,
and Radiological Devices
510(k) Number _
{13}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries) System: ICT/7-4 7.0-4.0 MHz Intracavitary Transducer Transducer: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|---------------------------|----------------------------------------------------|---|-------------------|-----|-----|------------------|---------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | | | | | | | |
| | Intra-operative (Abdominal<br>organs and vascular) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-vaginal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Color Power Doppler, 3-D Imaging, Tissue Harmonic Imaging for guidance of biopsy previously cleared through Ook) 1 ower Bopper, o Dimaging to assist in the placement of needles and catherers in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off)
and Radiological Device 510(k) Number
Indications for Use
{14}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries) System: Transducer: L38/10-5 10.0-5.0 MHz Linear Array
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | | |
|--------------------------|---------------------------|----------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| | General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Abdominal | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Intra-operative (Neuro.) | | | | | | | |
| | | Laparoscopic | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Small Organ (breast,<br>thyroid, testicles.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Adult Cephalic | | | | | | | |
| | | Trans-rectal | | | | | | | |
| | | Trans-vaginal | | | | | | | |
| | | Trans-urethral | | | | | | | |
| | | Trans-esoph. (non-Card.) | | | | | | | |
| | | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Musculo-skel. (Superfic.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Intra-luminal | | | | | | | |
| | | Other (spec.) | | | | | | | |
| Cardiac | | Cardiac Adult | | | | | | | |
| | | Cardiac Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Trans-esophageal (card.) | | | | | | | |
| | | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 |
| | | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Noter friddes Ooks T The Dopper John Citraging, and Imaging for guidance of biopsy proviously clared through 510(k) K014116. Included with this 510k are imaging to assist in the placement of needles and cathers in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy Brogdon
(Division Sign-Off) Division of Reproductive, and Radiological Devices 510(k) Number _
Indications for Use
{15}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries) System: Transducer: C60/5-2 5.0-2.0 MHz Curved Array Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body
as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|----------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | | | B+M; B+PWD | Note 1 |
| | Cardiac Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Color Power Doppler, combined B and Color Power Doppler, combined B and PowerMap™ Directional Now Your Timelator Over Pirson Tissue Harmonic Imaging, and imaging for guidance of biopsy previously cleared through 510(k) K014116. Included with this 510k are imaging to assist in the placement of needles and catherers in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy C. Brogdon
(Division Sign-Off)
and Radiological Devices 510(k) Number
Section 4.3, page 12
Indications for Use
{16}------------------------------------------------
SonoSite Hand-Carried Ultrasound System (PowerSeries) System:
Transducer:
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|----------------------------------------------------|-------------------|---|-----|-----|------------------|---------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Abdominal | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Abdominal<br>organs and vascular) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Small Organ (breast,<br>thyroid, testicles.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Neonatal Cephalic | P | P | P | | | B+M; B+PWD | Note 1 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Musculo-skel. (Superfic.) | P | P | P | | | B+M; B+PWD | Note 1 |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | P | P | P | | | B+M; B+PWD | Note 1 |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | | B+M; B+PWD | Note 1 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
Note 1: Other includes Color Power Dopler, combined B and Color Power Doppler, combined B and Coverty of the until Note 1. Only 1. Ower Roudes Ook 1 Ower Bopping, and imaging for guidance of biopsy proviously cleared through Occar rower opplied on this 510k are imaging to assist in the placement of needles and catherers in vascular or other anatomical structures.
Prescription Use (Per 21 CFR 801.109)
Nancy C hogdon
(Division Sian-O Division of Reproductiv and Radiological Device 510(k) Number
Section 4.3, page 13
Indications for Use
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.