← Product Code [BTR](/productcode/BTR) · K021540

# RUSCH EDGAR TUBE (K021540)

_Rusch Intl. · BTR · Aug 1, 2002 · Anesthesiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K021540

## Device Facts

- **Applicant:** Rusch Intl.
- **Product Code:** [BTR](/productcode/BTR.md)
- **Decision Date:** Aug 1, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.5730
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Rüsch Edgar Tube is a sterile, single-use, oral/nasal tracheal tube intended to be used for airway management. The integral instillation system is indicated for drug instillation by the endobronchial route, anesthetic administration and gas monitoring.

## Device Story

Rüsch EDGAR Tube is a sterile, single-use, oral/nasal endotracheal tube; available in cuffed (6.5mm–10.0mm) and uncuffed (2.5mm–6.0mm) versions. Features include radiopaque stripe, centimeter markings for depth, and an integral instillation lumen with a color-coded luer lock connector for drug delivery, anesthetic administration, and gas monitoring. Cuffed versions include a low-pressure/high-volume cuff with a pilot balloon for status indication. Used by clinicians for airway management; provides a conduit for ventilation and direct endobronchial access. Benefits include simplified drug delivery and monitoring without requiring separate access lines.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and technological characteristics compared to legally marketed predicate devices.

## Technological Characteristics

Clear PVC endotracheal tube with radiopaque stripe; cuffed (low-pressure/high-volume) or uncuffed; integral instillation lumen with luer lock connector; sizes 2.5mm–10.0mm. Sterilized using same conditions as predicate Rüsch tubes. Non-body contact components include luer fitting and cap.

## Regulatory Identification

A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.

## Predicate Devices

- Rüsch Safety Tracheal Tubes (Preamendment)
- Rüsch AGT PVC Preformed Nasal Tracheal Tubes ([K931163](/device/K931163.md)/[K931166](/device/K931166.md))
- Rüsch Oral/Nasal (Safety Clear Plus™) Tracheal Tube, Cuffed, Magill/Murphy, Sterile ([K993786](/device/K993786.md))
- Hudson RCI Sheridan ETCO2 Uncuffed Tubes ([K861454](/device/K861454.md))
- Hudson RCI Sheridan LITA Cuffed Tracheal Tubes ([K834463](/device/K834463.md))
- Mallinckrodt Emergency Medical Tube (EMT) (510(k) # Unknown)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K021540

AUG 01 2002

(603) 532-8211 or 6108

510 (k) Summary

- 1. Submitter Name, Address, and Date of Submission:
Rick Lykins Group RA Manager - US Rüsch International Tall Pines Park Jaffrey, NH 03452 Telephone Number: (603) 532-0204 Fax Number: (603) 532-6179

E-Mail: rlykins@tfx.com

Contact: Same as above

- 2. Name of the Device, Common, Proprietary (if known), and Classification:
Classification Name: Tube, Tracheal w/wo Connector

Common Name: Endotracheal or Tracheal Tube

Proprietary Name: Rüsch EDGAR Tube

### 3 . Identification of the legally marketed device to which the submitter claims equivalence:

The Rüsch EDGAR Tube is substantially equivalent in design and materials to:

- Rüsch Safety Tracheal Tubes - Preamendment
- . Rüsch AGT PVC Preformed Nasal Tracheal Tubes - K931163/K931166
- Rüsch Oral/Nasal (Safety Clear Plus™) Tracheal Tube, Cuffed, Magill/Murphy, Sterile - K993786
- The Hudson RCI Sheridan ETCO2 Uncuffed Tubes K861454 ●
- . The Hudson RCI Sheridan LITA Cuffed Tracheal Tubes - K834463
- The Mallinckrodt Emergency Medical Tube (EMT) 510(k) # Unknown .

{1}------------------------------------------------

## Page 2

#### Description of the Device: 4 .

The Rüsch Edgar Tube will be offered in the Magill pattern in two (2) versions - cuffed and uncuffed. Cuffed versions will have a low-pressure/high volume cuff, inflated via a luer activated valve. A pilot balloon will be provided to give a visual indication of cuff status.

The Rüsch Edgar Tube consists of a clear endotracheal tube with a radiopaque stripe. The main tube is graduated with multiple centimeter markings to facilitate determination of intubated The instillation lumen is connected to a luer lock length. connector with cap by a tube that is color coded to prevent confusion with the cuff inflation line.

The Rüsch Edgar Tube will be available in a variety of sizes ranging from 2.5mm - 6.0mm uncuffed and 6.5mm - 10.0mm cuffed. Both versions will be available in 0.5mm increments.

#### 5. Intended Use of the Device:

Rüsch Edgar Tube is a sterile, single-use, oral/nasal The tracheal tube intended to be used for airway management. The · integral instillation system is indicated for drug instillation by the endobronchial route, anesthetic administration and gas monitoring.

#### 6. Summary of Technological Characteristics:

The following technological characteristics are the same as or equivalent to the predicate devices listed above:

## Materials:

The Rüsch Edgar Tube is manufactured from the same materials (except for the instillation tube, the luer lock fitting and the cap which are non-body contact components), manufactured by the same processes and sterilized under the same conditions as the Rüsch Safety Tracheal Tubes which were in interstate commerce prior to May 28, 1976, the Rüsch AGT PVC Preformed Nasal Tracheal Tubes, K931163 and K931166 and the Rüsch Oral/Nasal (Safety Clear Plus™) Tracheal Tube, Cuffed, Magill/Murphy, Sterile - K993786.

{2}------------------------------------------------

## Page 3

Rüsch Edgar Tube will differ from the predicate Rüsch The Tracheal tube only in that it will have an integral instillation line, indicated for drug instillation and gas monitoring.

Size Range:

The Rüsch Edgar Tube will be available in a variety of sizes ranging from 2.5mm - 6.0mm uncuffed and 6.5mm - 10.0mm cuffed. Both versions will be available in 0.5mm increments.

The Hudson RCI Sheridan uncuffed tubes are available in sizes ranging from 2.0mm - 5.0mm. The Hudson RCI Sheridan cuffed tubes are available in sizes ranging from 6.0mm - 8.5mm. All versions are available in 0.5mm increments.

The Mallinckrodt uncuffed tubes are available in sizes ranging from 2.0mm - 5.5mm. The Mallinckrodt cuffed tubes are available in sizes ranging from 6.0mm - 8.5mm. All versions are available in 0.5mm increments.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three wavy lines representing snakes intertwined around a staff.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# AUG 01 2002

Mr. Rick Lykins General Manager Rüsch, International Tall Pines Park Jaffrey, New Hampshire 03452

Re: K021540

Trade/Device Name: Rüsch EDGAR Tube Regulation Number: 21 CFR 868.5730 Regulation Name: Endotracheal or Tracheal Tube Regulatory Class: II Product Code: BTR Dated: May 10, 2002 Received: May 10, 2002

Dear Mr. Lykins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced aboye and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Lykins

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html

Sincerely,

Tim A. Ulatowski

Timo Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

## KOZIS 40 510(k) Number (if known):

Rüsch Edgar Tube Device Name:

Indications for Use:

The Rüsch Edgar Tube is a sterile, single-use, oral/nasal tracheal tube intended to be used for airway management. The integral instillation system is indicated for drug instillation by the endobronchial route, anesthetic administration and gas monitoring.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED )

| Concurrence of CDRH, Office of Device Evaluation (ODE) |         |                      |
|--------------------------------------------------------|---------|----------------------|
| (Division Sign-Off)                                    |         |                      |
| Division of Dental, Infection Control,                 |         |                      |
| and General Hospital Devices                           |         |                      |
| 510(k) Number                                          | K021540 |                      |
| Prescription Use                                       | OR      | Over-The-Counter Use |
| (Per 21 CFR 801.109)                                   |         |                      |

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K021540](https://fda.innolitics.com/device/K021540)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
