Intended Use

The ONLINE DAT II Methadone II is an in vitro diagnostic test for the qualitative and semi-quantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/ml. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program.

Device Story

ONLINE DAT II Methadone II is an in vitro diagnostic assay for methadone detection in human urine. Operates on automated clinical chemistry analyzers using Kinetic Interaction of Microparticles in Solution (KIMS) technology. Principle: drug in urine competes with drug-polymer conjugate for antibody-bound microparticles; presence of drug inhibits particle aggregation, reducing absorbance change. Output: qualitative (positive/negative) or semi-quantitative methadone concentration. Used in clinical laboratories by technicians/staff to assist in drug testing. Results aid healthcare providers in assessing methadone presence; semi-quantitative results support laboratory quality control programs.

Clinical Evidence

No clinical data provided; substantial equivalence based on analytical performance characteristics of the KIMS assay technology.

Technological Characteristics

In vitro diagnostic immunoassay using KIMS (Kinetic Interaction of Microparticles in Solution) technology. Components include methadone monoclonal antibody-coated microparticles and soluble drug-polymer conjugate. Designed for use on automated clinical chemistry analyzers. No specific materials, energy sources, or software architecture details provided beyond standard automated analyzer compatibility.

Indications for Use

Indicated for qualitative and semi-quantitative detection of methadone in human urine using automated clinical chemistry analyzers at a 300 ng/ml cutoff. Intended for laboratory use.

Regulatory Classification

Identification

A methadone test system is a device intended to measure methadone, an addictive narcotic pain-relieving drug, in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of methadone use or overdose and to determine compliance with regulations in methadone maintenance treatment.

Special Controls

*Classification.* Class II (special controls). A methadone test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ )CO2/505 JUL 16 2002 **510(k) Summary** | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1) Submitter name, address, contact | Roche Diagnostics Corporation<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(317) 521-7637<br>Contact Person: Kerwin Kaufman<br>Date Prepared: May 8, 2002 | | 2) Device name | Proprietary name: ONLINE DAT II Methadone II<br>Common name: Methadone Test System<br>Classification name: Enzyme immunoassay, methadone | | 3) Predicate device | We claim substantial equivalence to the currently marketed Abuscreen OnLine Methadone assay (K983703). | Continued on next page {1}------------------------------------------------ ## 510(k) Summary, Continued The ONLINE DAT II Methadone assay is an in vitro diagnostic test for the 4) Device Description qualitative and semi-quantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/ml. ### Principal of procedure The ONLINE DAT II Methadone II assay is based on the kinetic interaction of microparticles in a solution (KIMS) as measured by changes in light In the absence of sample drug, soluble drug-polymer transmission. conjugates bind to antibody-bound microparticles, causing the formation of particle aggregates. When a urine sample containing the drug in question is present, this drug competes with the conjugate-bound drug derivative for microparticle-bound antibody. Antibody bound to sample drug is no longer available to promote particle aggregation, and subsequent particle lattice formation is inhibited. As the aggregation reaction proceeds in the absence of sample drug, the absorbance increases. Conversely, the presence of sample drug diminishes the increasing absorbance in proportion to the concentration of drug in the sample. Sample drug content is determined relative to the value obtained for a known cutoff concentration of drug. ### Negative Sample drug-polymer conjugate + antibody-bound microparticle = particle aggregates (↑ absorbance) ### Positive Sample sample drug + antibody-bound microparticle = particle aggregation inhibited drug-polymer conjugate + antibody bound microparticle = particle aggregates Continued on next page {2}------------------------------------------------ # 510(k) Summary, Continued ワ . | 5.) Intended<br>Use | The ONLINE DAT II Methadone II is an in vitro diagnostic test for the<br>qualitative and semi-quantitative detection of methadone in human urine on<br>automated clinical chemistry analyzers at a cutoff concentration of 300 ng/ml.<br>Semi-quantitative test results may be obtained that permit laboratories to<br>assess assay performance as part of a quality control program. | |----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 6.) Comparison<br>to the Predicate<br>Device | The Roche ONLINE DAT II Methadone II assay is substantially equivalent to<br>other products in commercial distribution intended for similar use. Most<br>notably, it is substantially equivalent to the currently marketed Roche<br>Abuscreen OnLine Methadone (K983703).<br><br>The Roche ONLINE DAT II Methadone II assay utilizes a modified KIMS<br>technology relative to the currently marketed Abuscreen OnLine Methadone<br>assay. Differences between this application and the cleared assay include:<br>use of a methadone monoclonal antibody attached to<br>microparticles in solution, a soluble drug-polymer conjugate, and use of new calibrators and unassayed controls. | and the comments of the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the consis , {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with flowing lines beneath them. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 2098 Gaither Road # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # # Rockville MD 20850 # UUL 1 6 2002 Mr. Kerwin Kaufman Regulatory Affairs Consultant Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457 Re: k021505 > Trade/Device Name: Roche Diagnostics ONLINE DAT II Methadone II Regulation Number: 21 CFR 862.3620 Regulation Name: Methadone test system Regulatory Class: Class II Product Code: DJR Dated: May 8, 2002 Received: May 9, 2002 Dear Mr. Kaufman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. 。Page 2 26 ・・・ (20) 10月には、 ・ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact.the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsmamain.html". Sincerely yours, Steven Butman 此刻意事 不要有意思了主要感情的意思想 Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use Statement ის შეწყვილების გამოყვარი სახლეობა სამარტი გადავის მწვევა მარტი სამხრეთ მართლები მარტი სამხრეთ მართლები მარტი სამ 彩必赢彩 ాన్స్ స్టేష్ | 510(k) Number (if known): | K021505 | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Roche Diagnostics ONLINE DAT II Methadone II | | Indications for Use: | An in vitro diagnostic test for the qualitative and semi-quantitative detection of methadone in human urine on automated clinical chemistry analyzers at a cutoff concentration of 300 ng/ml. Semi-quantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use (Per 21 CFR 801.109) | <div> <span style="text-decoration: overline;">V</span> OR Over-the-Counter Use ______ </div> | |---------------------------------------|-----------------------------------------------------------------------------------------------------------| | (Optional format 1-2-96) | | Sean Cooper (Division Sign Off) (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number ____________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...