AART MALAR IMPLANT

K021419 · Aesthetic and Reconstructive Technologies, Inc. · LZK · Jul 2, 2002 · General, Plastic Surgery

Device Facts

Record IDK021419
Device NameAART MALAR IMPLANT
ApplicantAesthetic and Reconstructive Technologies, Inc.
Product CodeLZK · General, Plastic Surgery
Decision DateJul 2, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the AART Malar Implant is for augmentation or reconstruction of the check areas of the face.

Device Story

AART Malar Implant is a medical-grade silicone elastomer device; crescent-shaped, concave-convex design; available in four styles as right/left mirror image pairs. Used by surgeons for cheek augmentation and reconstruction. Implant provides structural support to facial tissue. Offered sterile or non-sterile; sterilized via Gamma radiation. Device is a passive implant; no electronic or software components.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Medical grade silicone elastomer; crescent-shaped, concave-convex geometry; smooth surface; various dimensions (length, width, height). Sterilization via Gamma radiation per ANSI/AAMI/ISO 11137-1994.

Indications for Use

Indicated for patients requiring augmentation or reconstruction of the cheek areas of the face.

Regulatory Classification

Identification

A chin prosthesis is a silicone rubber solid device intended to be implanted to augment or reconstruct the chin.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL - 2 2002 ## ATTACHMENT 5 ## 510(k) Summary April 29,2002 Contact Information: Aesthetic and Reconstructive Technologies, Inc. (AART) 3545 Airway Drive, Suite 108 Reno, NV 89511 (775) 853-6800 / FAX (775) 853-6805 Winston A. Andrews Proprietary Name: Common Name: Classification Name: AART Malar Implant Silicone Elastomer Malar Implant Implant, Malar Substantial Equivalence: The AART Malar Implant is substantially equivalent in function, design, performance and materials to the Duralastic Anatomical Malar Implants marketed by Allied Biomedical Corporation of Ventura, CA and the Seare Biomedical Malar Implants marketed by Seare Biomedical Corp. of Salt Lake City, Utah. Device Description: The AART Malar Implants are manufactured from a medical grade silicone elastomer that has been molded into various crescent shaped concave convex implants. They are provided in four styles in pairs of right and left mirror images with dimensions varying in length, width, and height. The AART Malar Implants are intended to be used for augmentation and reconstruction of the cheek areas of the face. The surface characteristic of the implants is smooth, The AART Malar Implants will be offered sterile and non-sterile. Intended Use: The intended use for the AART Malar Implant is augmentation and reconstruction of the cheek areas of the face. Predicate Device: The AART Malar Implant is substantially equivalent in material, design, function, and performance to the Duralastic Anatomical Malar Implants marketed by Allied Biomedical Corp. and the Seare Biomedical Malar Implants marketed by Seare Biomedical Corp. All products have identical intended uses and are offered in similar shapes and sizes. Sterilization Cycle: The AART Malar Implant will be sterilized by Gamma radiation. The sterilization cycle will be determined and validated following the ANSUAAMI/ISO 11137-1994 standard "Sterilization of health care products - Requirements for validation and routine control -Radiation sterilization". {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing. The profiles are stacked on top of each other, creating a sense of depth and unity. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL = 2 2002 Ms. Catherine Riple Consultant Aesthetic and Reconstructive Technologies, Inc. 3545 Airway Drive, Suite 108 Reno. NV 89511 Re: K021419 Trade/Device Name: AART Malar Implant Regulation Number: 878.3550 Regulation Name: Chin prosthesis Regulatory Class: II Product Code: LZK Dated: April 29, 2002 Received: May 3, 2002 Dear Ms. Riple: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Page 2 - Ms. Catherine Riple This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Hypt Rhodes Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## AMENDMENT 1 Page of | 510 (k) NUMBER (IF KNOWN) : | K021419 | |-----------------------------|--------------------| | DEVICE NAME: | AART Malar Implant | INDICATIONS FOR USE: The intended use of the AART Malar Implant is for augmentation or reconstruction of the check areas of the face. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.】 . Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109 OR Over-The-Counter-Use (Optional Format 1-2- Styph Ruddy (Division Sign-Off) Division of General, Restorative and Neurological Devices 510(k) Number_ KOZI419
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...