B-F200 BRONCHOSCOPE WITH BSS-F21 ENDOSHEATH SYSTEM
K021344 · Vision-Sciences, Inc. · EOQ · May 24, 2002 · Ear, Nose, Throat
Device Facts
Record ID
K021344
Device Name
B-F200 BRONCHOSCOPE WITH BSS-F21 ENDOSHEATH SYSTEM
Applicant
Vision-Sciences, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
May 24, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The VSI Flexible, Fiberoptic Bronchoscope with EndoSheath® System is used for flexible endoscopic examination of the trachea and other major passages of the lungs, to gather specimens, and/or to find and endoscopically remove foreign objects from the lungs. The EndoSheath® is a protective covering for the bronchoscope.
Device Story
System consists of modified flexible fiberoptic bronchoscope and sterile, disposable EndoSheath®. Sheath provides protective barrier, eliminating need for high-level disinfection of scope between procedures. Modification introduces 'slide-on' installation technique for sheath, replacing vacuum-based method. Suction valve relocated from sheath to scope. Used by clinicians for visualization of trachea and lung passages, specimen collection, and foreign object retrieval. Fiberoptic technology transmits images to eyepiece/monitor for physician guidance. Benefits include reduced scope reprocessing requirements and maintained sterility via microbial barrier.
Clinical Evidence
Bench testing only. Validation included microbial barrier testing, sheath tensile/elongation testing, sheath leak testing, sheathed scope articulation testing, and sheathed scope image quality evaluation. No clinical data provided.
Technological Characteristics
Flexible fiberoptic bronchoscope with disposable protective sheath. Features: 170°/120° down articulation, 90° field of view, 6.0 mm working channel ID. Slide-on installation mechanism. Microbial barrier protection. Design controls per 21 CFR 820.30.
Indications for Use
Indicated for patients requiring flexible endoscopic examination of the trachea and major lung passages, specimen collection, or foreign object removal. Contraindications not specified.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Bronchoscope with Disposable EndoSheath® (K963795)
Modified EndoSheath® for Flexible ENT Scopes (K990354)
Submission Summary (Full Text)
{0}------------------------------------------------
**MAY 24 2002**
K021344
Vision-Sciences, Inc. April 24, 2002
pecial 510(k) Premarket Notification: Modified Flexible Fiberoptic Bronchoscope and EndoSheath® System
# 510(k) Summary
| Trade Name: | Vision-Sciences Flexible Fiberoptic Bronchoscope and EndoSheath®<br>System |
|----------------------|-----------------------------------------------------------------------------------------------------------------|
| Sponsor: | Vision-Sciences, Inc.<br>9 Strathmore Road<br>Natick, MA 01760<br>Registration #1223490 |
| Device Generic Name: | Flexible, fiberoptic bronchoscope with protective sheath system |
| Classification: | According to Section 513 of the Federal Food, Drug, and<br>Cosmetic Act, the device classification is Class II. |
| Predicate Devices: | K963795 – Bronchoscope with Disposable EndoSheath®<br>K990354 – Modified EndoSheath® for Flexible ENT Scopes |
| Manufactured by: | Vision-Sciences, Inc.<br>9 Strathmore Road<br>Natick, MA 01760 |
Product Description: The device system described in this 510(k) consists of a modified sterile, disposable sheath designed to fit the modified VSI flexible fiberoptic bronchoscope. The use of an EndoSheath® eliminates the need for high-level disinfection of the scope following each procedure. The sheath and scope have been modified to accommodate the "slide-on" sheath installation technique.
#### Indications for Use:
The Bronchoscope with EndoSheath® System is used for flexible endoscopic examination of the trachea and other major passages of the lungs, to gather specimens, and/or to find and endoscopically remove foreign objects from the lungs.
### Safety and Performance:
This submission is a Special 510(k): Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), Vision-Sciences has provided certification of compliance to 21 CFR 820.30 Design Control requirements, and a description of the internal Risk Analysis procedure. Validation testing including microbial barrier testing, sheath tensile/elongation testing, sheath leak testing, sheathed scope articulation testing and sheathed scope image quality evaluation is included in Design Validation and Verification planning.
### Conclusion:
Based on the indications for use, technological characteristics, and comparison to predicate devices, the modified VSI Flexible Fiberoptic Bronchoscope and EndoSheath® System has been shown to be safe and effective for its intended use.
{1}------------------------------------------------
Special 510(k) Premarket Notification.
Design Compariso Table D.1
| Characteristic | Proposed Modified<br>Bronchoscope with<br>Slide-on<br>EndoSheath® System<br>(Current Submission) | Currently Marketed,<br>Unmodified<br>Bronchoscope with<br>EndoSheath® System<br>(K963795) | Currently<br>Marketed Slide-on<br>EndoSheath®<br>System for use with<br>ENT Scopes<br>(K990354) | Currently Marketed<br>VSI Flexible GI<br>Scope/EndoSheath®<br>Systems* | Substantial Equivalence<br>Comparison:<br>Proposed Modified<br>Bronchoscope with Slide-on<br>EndoSheath® System |
|---------------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Down Articulation<br>(sheathed scope) | 170°/120° | 170°/120° | | | Equivalent to unmodified<br>VSI Bronchoscope/sheath<br>system |
| Field of View<br>(apter) | 90° | 90° | | | |
| Depth of Field | +2/-8 | +2/-8 | | | |
| Insertion Tube OD | 3 - 50 mm | 3 - 50 mm | | | |
| Working Channel ID | 6.0 mm | 6.0 mm | | | |
| Working Length | 2.1 mm | 2.1 mm | | | |
| Submersible scope | 550 mm | 550 mm | | | |
| Microbial barrier claim | Yes | Yes | Yes | Yes | Identical to unmodified VSI<br>Bronchoscope/sheath system |
| Listing scope models | VSI Bronchoscope | VSI Bronchoscope | Various | VSI GI Scopes | Identical to VSI ENT<br>Scope/sheath system |
| Sheath installation method | Slide-on | Vacuum | Slide-on | Pressure | Identical to VSI ENT<br>Scope/sheath system |
| Sheath wall thickness | .0020 - .0045" | .010 ± .002" | .0020 - .0045" | No S.E. comparison required - different<br>sheath material | |
Currenty marketed Vision-Sciences G1 Endoscope with EndoSheath System (Sigrem (Sigrem (Sigrem (Sigrem (Sigrem (Sigrem (Sigrem (Signem (Signem (Sig
{2}------------------------------------------------
| Change Description | Where/How Evaluated (V&V Tests) |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Sheath material changed | • Sheath burst/leak test<br>• Sheath tensile/elongation test<br>• Sheath installation/system functional test<br>• Microbial barrier test |
| Sheath working channel<br>construction changed | • Sheath burst/leak test<br>• Sheath tensile/elongation test<br>• Sheath installation/system functional test<br>• Microbial barrier test |
| Suction valve moved from<br>sheath to scope | • Sheath installation/system functional test<br>• Pinch valve cycle test |
Table D.3 Product Changes/Testing Summary
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
MAY 2 4 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vision Sciences. Inc. C/O Delphi Medical Device Consulting 5 Whitcomb Avenue Ayer, MA 01432 Attn: Ms. Pam Papineau
Re: K021344
Trade/Device Name: B-F200 Bronchoscope with BSS-F21 Endosheath System Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope and accessories Regulatory Class: Class II Product Code: EOQ Dated: April 24, 2002 Received: April 29, 2002
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
## Page 2 - Ms. Pam Papineau
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
Special 510(k) Premarket Notification: Device Modification Modified Flexible Fiberoptic Bronchoscope and EndoSheath® System
Page
510(k) Number (if known): DJ2 13 44
Device Name: Flexible Fiberoptic Bronchoscope with EndoSheath® System
Indications for Use:
The VSI Flexible, Fiberoptic Bronchoscope with EndoSheath® System is THE VST rexible, I lochopic Eronomooootic examination of the trachea and other major assages of the lungs, to gather specimens, and/or to find and passagos of the fange, to gign objects from the lungs. The EndoSheath® endoscopioally romove role protective covering for the bronchoscope.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use √
21 CFR 801.109)
(Divisio Division of Ophthalmic Ear, Nose and
510(k) Number K02 1344
Over-the -Counter Use
000008
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.