The Matritech NMP22 BladderChek Test is indicated for professional and prescription home use as an aid in monitoring bladder cancer patients, in conjunction with standard diagnostic procedures.
Device Story
NMP22 BladderChek is an in vitro immunochromatographic lateral flow assay for qualitative detection of nuclear matrix protein (NMP22) in voided urine. Device consists of plastic-encased test strip with capture and reporter monoclonal antibodies conjugated to colloidal gold. User adds urine to sample well; if NMP22 antigen is present, it forms immune complex with reporter conjugate. Mixture flows through membrane; antigen-conjugate complexes bind to capture antibody in Test (T) zone, creating visible line. Procedural Control (C) zone confirms device function. Used in professional clinical settings and prescription home use for monitoring bladder cancer recurrence. Results are interpreted visually by clinician or patient. Output aids clinical decision-making alongside standard diagnostic procedures (e.g., cystoscopy). Benefits include non-invasive monitoring for patients with history of bladder cancer.
Clinical Evidence
Prospective clinical trial (n=668) evaluated NMP22 BladderChek as an aid in monitoring bladder cancer recurrence. Sensitivity 45.9% (95% CI: 35.8%, 56.3%); specificity 86.3% (95% CI: 83.2%, 89.0%). PPV 36.6%; NPV 90.3%. Performance varied by tumor stage/grade (higher sensitivity for T2-T3 and Grade 3/4). Non-clinical reproducibility studies (n=1800 reads) showed 99.2% correct read rate. Lay user vs. professional concordance was 96.4%.
Technological Characteristics
Immunochromatographic lateral flow assay; monoclonal antibodies; colloidal gold reporter; plastic-encased test strip. Qualitative visual readout. Analyte: NMP22. Sample: voided urine. No electronic components, software, or external energy source.
Indications for Use
Indicated for professional and prescription home use as an aid in monitoring bladder cancer patients, in conjunction with standard diagnostic procedures.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
{1}------------------------------------------------
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K021231
| Name: | Matritech, Inc.<br>330 Nevada St.<br>Newton, MA 02460 |
|--------------------------|----------------------------------------------------------------------------------------|
| Contact: | Melodie R. Domurad PhD<br>617 928 0820 ext 265<br>mdomurad@matritech.com |
| Date: | April 15, 2002 |
| Catalog Number: | D1200 24-test kit<br>D1201 1-test kit<br>D1250 Control kit |
| Device Name: | NMP22® BladderChek™ Kit<br>NMP22® Control Kit |
| Common: | NMP22 BladderChek |
| Classification: | Class II |
| Manufacturer: | Unotech Diagnostics, Inc.<br>2235 Polvorosa Ave.<br>Suite 220<br>San Leandro, CA 94577 |
| Registration Number: | 1225597 Matritech, Inc.<br>2953160 Unotech Diagnostics, Inc |
| Substantially Equivalent | RTA Stat Test |
# Predicate Device
Description of Device
The NMP22 BladderChek Test is an in vitro diagnostic assay for the qualitative detection of the nuclear matrix protein NMP22 in human urine.
{2}------------------------------------------------
#### Intended Use
The Matritech NMP22 BladderChek Test is indicated for professional and prescription home use as an aid in monitoring bladder cancer patients, in conjunction with standard diagnostic procedures.
#### Summary and Explanation of the Test
The NMP22 BladderChek Test for nuclear matrix protein NMP22 is an immunochromatographic assay utilizing monoclonal antibodies in a lateral flow strip encased in plastic. Two different antibodies are used, one as a capture and one as a reporter antibody. Unprocessed patient urine is added to the sample well of the cartridge and allowed to react with the colloidal gold conjugated reporter antibody. If the antigen is present in urine, it will interact with the reporter conjugate to form an immune complex. The reaction mixture flows through the membrane, which contains zones of immobilized antibodies. In the Test (T) zone, antigen-conjugate complexes are trapped by the capture antibody, forming a visible line if the concentration of antigen in urine is elevated. The procedural Control (C) zone contains an immobilized goat anti-mouse IgG-specific antibody that will capture the conjugated antibody independently of the presence or absence of the antigen, thereby always producing a visible line in the Control window. This procedural control assures the operator that each device is working properly.
#### Substantial Equivalence to Predicate Device
Both the NMP22 BladderChek Test and BTA Stat are single use, in vitro diagnostic devices for detecting recurrence in patients with a history of bladder cancer. The assay reactions for the two devices are immunochromatographic, utilizing monoclonal antibodies, and produce qualitative results. Both detect analytes from voided urine. Targeted environment for the two tests is prescription home use. The NMP22 BladderChek Test detects the nuclear matrix protein NMP22, and the BTA Stat Test detects Bladder Tumor Antigen.
{3}------------------------------------------------
| | NMP22 BladderChek<br>K021231 | BTA Stat<br>K974845 & K964151 |
|-----------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Intended Use | Aid in the monitoring of bladder<br>cancer patients in conjunction<br>with standard diagnostic<br>procedures | Aid in the management of<br>bladder cancer patients in<br>conjunction with cystoscopy |
| Sample Matrix | Voided urine | Voided urine |
| Procedural Steps | Add 4 drops of urine | Add 5 drops of urine |
| Result Interpretation | +/-, Qualitative | +/-, Qualitative |
| Analyte Measured | NMP22<br>Nuclear Matrix Protein | BTA<br>Bladder Tumor Antigen |
| Targeted Environment | Professional and Prescription<br>Home Use | Laboratory and Prescription<br>Home Use |
| Assay Reaction | Immunochromatographic | Immunochromatographic |
## Non-Clinical Tests
Reproducibility by Laboratory Technicians
Precision studies were conducted to determine the percent of devices correctly read. Specimen panels with NMP22 levels of 0, 5, 15, and 25 U/mL were used. Three laboratory technicians each read ten devices per level in random order, (40 devices/lab tech) for five different days. This produced 150 individual reads per panel level (3 readers x 10 devices per panel x 5 days). The experiment was conducted on three separate lots of devices.
The percent of correctly read NMP22 BladderChek devices was very consistent across lots (n=3), lab techs (n=3), and days (n=5), where urine specimens with NMP22 levels of 0 and 5 U/mL should have been read as negative and urine specimens with NMP22 levels of 15 and 25 U/mL should have been read as positive.
The overall percent of correct reads was 99.2% (1786/1800). Twelve of the 14 incorrect results were at 5 U/mL (438/450 or 97%), which could be expected near the test cut-off.
{4}------------------------------------------------
## Reproducibility by Lay Users Compared to Professional Readers
A precision study with a three-level precision panel of prepared samples was conducted to evaluate lay user reproducibility. The three levels were targeted to the concentrations of 2, 10 and 15 U/mL to represent negative, low positive and positive values. Five lay readers each conducted assays on two blinded and randomized aliquots of the above samples, resulting in a total of ten results per level. In addition, two professional readers each tested five blinded and randomized replicates of the three-level precision panel, producing 10 results for each level. There was 100% concordance among the lay readers. between professional readers, and between the lay and professional readers. All samples at 2 U/mL were read as negative, and all samples at 10 and 15 U/mL were read as positive.
## Performance of Lay Users Compared to Professionals
Field studies to evaluate the ability of lay users compared to professional users of the BladderChek test were conducted at three locations. Each volunteer, aged 50 years or older, some with a history of bladder cancer, tested his or her own urine with one device. At each site professional medical staff re-read the devices used by the lay users, and also conducted their own testing of the volunteers' samples using new devices.
To ensure the presence of some positive results in the study, fourteen samples were spiked with NMP22 to approximately 15 U/mL and fourteen with approximately 25 U/mL. In addition, eight devices were made invalid by not having a working control line.
There was an overall 96.4% concordance between the results obtained by lay users and professional staff reading the same device, and 95.6% agreement between professional and lay users testing the same urine sample on different devices. All spiked specimens were read correctly by lay and professional readers.
| | Percent Overall Concordance<br>with Lay User Result | | | |
|---------------------------------------------------------------------------------------|-----------------------------------------------------|------------------|------------------|--------------------|
| | Site 1<br>(N=55) | Site 2<br>(N=35) | Site 3<br>(N=47) | Overall<br>(N=137) |
| Professional Re-testing of same urine<br>sample with new NMP22 BladderChek™<br>Device | 96.4%<br>(53/55) | 97.1%<br>(34/35) | 93.6%<br>(44/47) | 95.6%<br>(131/137) |
| Professional Re-Read of Lay User<br>NMP22 BladderChek™ Device | 96.4%<br>(53/55) | 100%<br>(35/35) | 93.6%<br>(44/47) | 96.4%<br>(132/137) |
{5}------------------------------------------------
## Clinical Tests
A prospective clinical trial was performed at 23 sites to determine the utility of NMP22 BladderChek as an aid in monitoring patients with a history of bladder cancer. Voided urine samples were collected from 668 patients prior to surveillance cystoscopy, and clinical staff performed the testing. Physicians conducting the cystoscopies were blinded to the device results.
To determine the sensitivity and specificity, patients were classified as positive or negative for bladder cancer. Patients were considered negative if no tumor was seen endoscopically, or, if a lesion was seen, was pathologically determined to be nonmalignant. There were 291 patients diagnosed as having benign disease(s), and 279 with no evidence of urinary tract disease. Patients were considered positive for bladder cancer if a tumor was seen during cystoscopy, and, if removed was pathologically determined to be malignant. There were 98 recurrences of neoplasms, of which 61 were resected and had stage and grade information available. The remaining 37 patients did not undergo surgical removal, and therefore their tumors could not be staged or graded (Tx, Gx).
{6}------------------------------------------------
Specificity for individuals with no evidence of disease, benign disease by group, history of non-bladder cancer, and active cancers other than of the bladder is presented in the following table.
| | Specificity | (95% CI) |
|------------------------------------------------------------------------|--------------------|----------------|
| Patients with No Evidence of<br>Disease<br>(History of Bladder Cancer) | 83.9%<br>(234/279) | (79.0%, 88.0%) |
| Benign Diseases* | | |
| BPH/prostatitis | 89.6%<br>(120/134) | (83.1%, 94.2%) |
| Cystitis/inflammation/trigonitis/UTI | 85.2%<br>(23/27) | (66.3%, 95.8%) |
| Erythema | 92.5%<br>(49/53) | (81.8%, 97.9%) |
| Hyperplasia/squamous metaplasia/ cysts<br>and polyps | 82.4%<br>(14/17) | (56.6%, 96.2%) |
| Calculi | 100%<br>(5/5) | (47.8%, 100%) |
| Trabeculations | 89.1%<br>(106/119) | (82.0%, 94.1%) |
| Other benign diseases, kidney and<br>genitourinary | 86.4%<br>(51/59) | (75.0%, 94.0%) |
| Cancer History, non-bladder -<br>Inactive Other Cancers^ | 88.8%<br>(71/80) | (79.7%, 94.7%) |
| Cancer History, non-bladder -<br>Active Other Cancers^^ | 71.4%<br>(5/7) | (29.0%, 96.3%) |
*Patients may have more than one benign disease
^ Types of cancers: prostate (n=29), skin (n=18), kidney/renal (n=15), genitourinary, non-bladder, non-prostate (n=5), breast (n=6), GI (n=6), lung/respiratory (n=4), blood (n=1), other (n=3).
^ Patients may have more than one type of prior cancer, and current benign diseases. ^^ Types of cancers: prostate (2/4), 1 kidney (0/1), 1 cervical (0/1), 1 lung/liver (0/1)
Sensitivity for patients with bladder cancer was 45.9%. Patients were considered positive for bladder cancer if a tumor was seen during cystoscopy, and, if removed, was pathologically determined to be malignant.
| Cystoscopy | NMP22 BladderChek Test | | |
|------------|------------------------|----------|-------|
| | Negative | Positive | Total |
| Negative | 492 | 53 | 545 |
| Positive | 78 | 45 | 123 |
| Total | 570 | 98 | 668 |
{7}------------------------------------------------
| | Sensitivity<br>(95% Exact CI) | Specificity<br>(95% Exact CI) | Positive Predictive<br>Value<br>(95% Exact CI) | Negative Predictive<br>Value<br>(95% Exact CI) |
|---------------------------------------|------------------------------------|--------------------------------------|------------------------------------------------|------------------------------------------------|
| NMP22<br>BladderChek vs<br>Cystoscopy | 45.9%<br>(35.8%, 56.3%)<br>(45/98) | 86.3%<br>(83.2%, 89.0%)<br>(492/570) | 36.6%<br>(28.1%, 45.8%)<br>(45/123) | 90.3%<br>(87.5%, 92.6%)<br>(492/545) |
Overall sensitivity and specificity for the clinical trial was as follows.
The NMP22 BladderChek test was more sensitive to later stages and higher grades of cancer, although the majority of malignancies were non-invasive. Sensitivity by stage and grade is detailed in the table below.
| | Sensitivity of NMP22<br>BladderChek Test | (95% CI) |
|------------------------------------------------------------|------------------------------------------|----------------|
| Tumor stage: Ta- Tl | 40.0%<br>(22/55) | (27.0%, 54.1%) |
| Tumor stage: T2- T3 | 80.0%<br>(4/5) | (28.4%, 99.5%) |
| Tumor stage: Tx | 51.4%<br>(19/37) | (34.4%, 68.1%) |
| Tumor grade: Well differentiated<br>(Grade 1) | 30.0%<br>(9/30) | (14.7%, 49.4%) |
| Tumor grade: Moderately differentiated<br>(Grade 2) | 33.3%<br>(4/12) | (9.9%, 65.1%) |
| Tumor grade: Poorly differentiated<br>(Grade 3 or Grade 4) | 66.7%<br>(12/18) | (41.0%, 86.7%) |
| Tumor grade: Gx<br>(Grade unknown) | 52.6%<br>(20/38) | (35.8%, 69.0%) |
The incidence rate for bladder cancer recurrence in this study was 14.7%. The following table demonstrates the Positive (PPV) and Negative (NPV) predictive values of the NMP22 BladderChek test at varying incidence rates.
| Incidence<br>Rate | NMP22 BladderChek Test | |
|-------------------|----------------------------------------|-------|
| | Sensitivity=45.9%<br>Specificity=86.3% | |
| | PPV | NPV |
| 10% | 27.1% | 93.5% |
| 14.7%* | 36.6% | 90.3% |
| 20% | 45.6% | 86.5% |
| 30% | 58.9% | 78.8% |
* Actual incidence rate from NMP22 BladderChek Monitoring Study
Concordance of NMP22 BladderChek and NMP22 Test Kit (microplate)
In comparison testing using voided urine samples, the NMP22 BladderChek Test showed good concordance with the NMP22 Test Kit (microplate).
Voided urine samples were collected from 217 individuals at urology clinics. An aliquot of each sample was applied to a BladderChek Test device. An additional aliquot was stabilized immediately and tested by the Matritech NMP22 (microplate) Test Kit to quantify the NMP22 concentration.
{8}------------------------------------------------
Overall concordance was 91.2% (198/217, CI 86.7%, 94.7%), with a positive concordance of 87.0% (20/23, CI 66.4%, 97.2%) and negative concordance of 91.8% (178/194, CI 86.9%, 95.2%).
| NMP22 Microplate | NMP22 BladderChek Test | | |
|----------------------|------------------------|----------|-------|
| Assay | Negative | Positive | Total |
| Negative (≤10 U/mL) | 178 | 16 | 194 |
| Positive (> 10 U/mL) | 3 | 20 | 23 |
| Total | 181 | 36 | 217 |
#### Conclusion
The NMP22 BladderChek Test and its predicate device, the BTA Stat are technologically similar. The assay reactions (immunochromatography utilizing monoclonal antibodies), sample matrix (voided urine), procedural steps (addition of drops of urine to device), result interpretation (qualitative +/-), patient population (history of bladder cancer), and indication (aid in monitoring), are all similar.
Reproducibility of results with NMP22 BladderChek were excellent between laboratory technicians across lots, days and specimen panel levels. Overall percent of correct reads was 99.2%. A precision study using a prepared specimen panel of negative (2 U/mL). low positive (10 U/mL) and positive (15 U/mL) samples demonstrated 100% concordance between professional staff (n=2) and lay users (n=5). In addition, an investigation to evaluate performance of lay users compared to professional readers showed 96.4% overall concordance when professionals re-read tests performed by lay volunteers using their own urine, and 95.6% when the professionals re-tested the lay users' urine samples. The BTA Stat package insert states that reproducibility studies were performed to determine day-to-day, reader, lot-to-lot, and laboratory-tolaboratory variability. It reports that these studies showed nearly total agreement with the exception of samples near the limit of detection, thus non-clinical performance of the two devices is comparable.
In the prospective clinical trial performed with NMP22 BladderChek sensitivity was 45.9% (CI 35.8%, 56.3%) and specificity was 86.3% (CI 83.2%, 89.0%), with a positive predictive value of 36.6% (CI 28.1%, 45.8%) and a negative predictive value of 90.3% (C1 87.5%, 92.6%). The incidence rate of cancer in the study population was 14.7%. Per its package insert, BTA Stat demonstrated 67% (CI 60%, 73%) sensitivity and 70% (CI 61%, 79%) specificity. Because the performance for this device was determined using retrospective samples selected from a frozen bank, incidence rate could not be determined, so true positive and negative predictive values could not be calculated.
{9}------------------------------------------------
Using a range of theoretical incidence rates of 10%, 20% and 30%, the NMP22 BladderChek Test demonstrated positive predictive values of 27.1% to 58.9%, compared to 19.8% to 48.8% for BTA Stat. In addition, the overall accuracy (defined as test positive for patients with cancer and test negative for patients without cancer divided by the total number of patients) of NMP22 BladderChek was 80.3% compared to 67.9% for BTA Stat. This was due in part to the excellent specificity of BladderChek, resulting in fewer false positive results.
The non-clinical and clinical studies presented in this submission demonstrate that the clinical utility and performance of NMP22 BladderChek is equivalent or better than that of the predicate device, BTA Stat.
{10}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right, with three parallel lines above them. The seal is black and white.
JUL 3 0 2002
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Melodie R. Domurad. Ph.D. Vice President. Clinical and Regulatory Affairs Matritech, Inc. 330 Nevada Street Newton, Massachusetts 02460
Re: k021231 Trade/Device Name: NMP22® BladderChekTM Test Regulation Number: 21 CFR § 866.6010 Regulation Name: Tumor Marker Test Systems - Monitoring-Bladder Regulatory Class: II Product Code: MMW Dated: June 27, 2002 Received: June 28, 2002
Dear Dr. Domurad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{11}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
## INDICATION FOR USE STATEMENT
510(k) Number: K 021231
长021231
Device Name: NMP22® BladderChek™ Test
Indication For Use:
The NMP22 BladderChek test is indicated for use as an aid in monitoring bladder cancer patients, by professional and prescription home use, in conjunction with standard diagnostic procedures.
JP Reeve for S. Altaie
#### (Division Sign Division of Clinical Laboratory Devices
510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The Counter Use
2.1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.