This needle is designed for injection of solutions (such as 2% lidocaine with or without 1:100,000 epinephrine) into the cervix. The application includes local anesthetics such as electro-excision, electro-fulguration, CO2 laser excision, and vaporization, and when required, endocervical curettage and cervical biopsies.
Device Story
Wallach Endocervical Block Needle is a sterile, single-use, disposable 27-gauge needle (3.5" length) with a 21-gauge proximal shaft reinforcement. Device connects via Luer lock to standard piston syringes or intra-vascular administration sets. Used by clinicians in clinical settings to inject local anesthetics (e.g., lidocaine with/without epinephrine) into the cervix prior to gynecological procedures. Facilitates pain management during surgical interventions; enables procedures like biopsies and laser treatments.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
27-gauge stainless steel needle, 3.5" length, with 21-gauge proximal shaft reinforcement. Features a female Luer lock connector. Sterile, single-use, disposable. No electronic components or software.
Indications for Use
Indicated for patients undergoing Ob/Gyn procedures requiring uterine anesthesia, including electro-excision, electro-fulguration, CO2 laser excision/vaporization, endocervical curettage, and cervical biopsies.
Regulatory Classification
Identification
An obstetric anesthesia set is an assembly of antiseptic solution, needles, needle guides, syringes, and other accessories, intended for use with an anesthetic drug. This device is used to administer regional blocks (e.g., paracervical, uterosacral, and pudendal) that may be used during labor, delivery, or both.
Avid Medical 'Avid-Nit Cervical Regional Anesthesia Needle' (K000117)
Terumo Medical 'Terumo 30 Gauge Hypodermic Needle' (K012646)
Nipro Medical Crop 'Nipro Hypodermic Needel' (K013293)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "WALLACH" in a bold, sans-serif font. The letters are black and have a slightly distressed or textured appearance, giving them a vintage or worn look. The word is presented in all capital letters and is the central focus of the image.
JUL 1 2 2002
Surgical Devices, Inc.
235 Edison Road
Orange, CT 06477
(203) 799-2000
Fax: (203) 799-2002
510(k) Premarket Notification Submittal · Wallach Endocervical Block Needle' By Wallach Surgical Devices. Inc. SUMMARY CERTIFCIATION page of SECTION 2 April. 15, 2002 -Page 1 of 5
# 510(k) Summary/Statement Certification
Re: K 02 / 72 4 __
Device Name: "Wallach Endocervical Block Needle"
### Check Only One:
- 1 510(k) Summary. Attached is a summary of safety and effectiveness 1. information upon which an equivalence determination could be based.
### 510(k) Statement, I Certify that, in my capacity as 2. GENERAL MANAGER, of WALLACH SURGICAL DEVICES,
INC.
I will make available all information included in this premarket notification of safety and effectiveness within 30 days of request by any person if the device described in the premarket notification submission is determined to be substantially equivalent. The information I agree to make available will be a duplicate of the premarket notification submission, including any adverse safety and effectiveness information, but excluding all patient identifiers, and trade secret and confidential commercial information, as defined in 21 CFR 20.61.
M. Mels
(Signature)
Michael Malis (Typed or Printed name)
2/15/02
(Dated)
{1}------------------------------------------------
510(k) Premarket Notification Submittal 'Wallach Endocervical Block Needle' By Wallach Surgical Devices, Inc. SUMMARY CERTIFCIATION page of SECTION 2 April. 15, 2002 -Page 2 of 5
## 510(k) Safety and Effectiveness Summary
| Applicant: | Wallach Surgical Devices, Inc,<br>235 Edison Road<br>Orange, CT 06477 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration: | 1219739 |
| Contact: | Michael Malis |
| Phone: | 203-799-2000 |
| Fax: | 203-799-2002 |
| Trade Name: | Wallach Endocervical Block Needle |
| Devices Generic Name: | Endocervical Block Needle |
| Classification Name: | Set, Anesthesia, Parcervical, and Needle, Hypodermic, Single<br>Lumen |
| Classification: | Currently classified as a Class II, under Product<br>Code 85 HEE Regulation Number 884.5100 and Product<br>Code 80 FMI Regulation Number 880.5570 per 21 CFR. |
### Predicate Devices to which we are claiming substantial equivalence:
- Coopersurgical 'Potocky Needle', K910252 1
- Avid Medical 'Avid-Nit Cervical Regional Anesthesia Needle', K000117 2.
- Terumo Medical 'Terumo 30 Gauge Hypodermic Needle', K012646 3.
- Nipro Medical Crop 'Nipro Hypodermic Needel', K013293 4.
### Product Description:
The Wallach Endocervical Block Needle is a Sterile, Disposable, Single Use device. The needle is 27ga., 3 1/2" long. It is used for uterine anesthesia prior to Ob/Gyn procedures. It consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (Luer lock) designed to mate with a male connector (Nozzle) of a piston syringe or an intra-vascular administration set. Over 3 1/4" of its proximal end a 21 gauge metal tube is added to strengthen the shaft.
### Indications for Use:
It is designed for injection of solutions (such as 2% lidocaine with or without 1:100,000 epinephrine) into the cervix. The application includes local anesthetics such as electro-excision, electro-fulguration, CO2 laser excision and vaporization, and when required, endocervical curettage and cervical biopsies.
{2}------------------------------------------------
### Safety and Performance:
Substantial equivalence for this device is based on design, operation, intended use, materials, and performance claims. Testing that was performed on the Wallach Endocervical Block Needle indicates that the devices are substantially equivalent in the performance and design of operation.
Hazard analysis evaluations performed on the Wallach Endocervical Block Needle indicated that there were no new hazards presented with the use of the Wallach Endocervical Block Needle as compared to the predicated devices.
{3}------------------------------------------------
510(k) Premarket Notification Submitted
Sy Wallach Endocervical Findocervical Blook News Coming Comments
SUMARY Currently Burgical Processor Comments of Caracter of
April. 15
# Comparison Chart
| | Wallach Surgical Devices, Inc.<br>'Wallach Endocervical Block Needle' | Coopersurgical<br>'Potocky Needle' | Avid Medical Inc.<br>'Avid-Nit<br>Cervial Regional<br>Anesthesia Needle' | Terumo Medical Corp.<br>'Terumo 30 Ga.<br>Hypodermic Needle' | Nipro Medical Corp.<br>'Nipro Hypodermic<br>Needle' |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------|---------------------------------------------------------|
| Feature: | K<br>(pending)<br>Class II, Product Code HEE<br>Regulation # 884.5100 and<br>Product Code FMI<br>Regulation # 880.5570 | K910252 | K000117 | K 012646 | K013293 |
| Intended Use: | For injection of solutions<br>(such 2% lidocaine with or<br>without 1:100,000<br>epinephine) into the cervix.<br>The application includes local<br>anesthetics such as eletro-<br>excision, elecro-fulguration,<br>CO2 laser excision and<br>vaporization, and when<br>required, endocervical<br>curettage and cervical<br>biopsies | Class II,<br>Product Code: HEE<br>Regulation # 884.5100 | Class II,<br>Product Code: HEE<br>Regulation # 884.5100 | Class II,<br>Product Code: FMI<br>Regulation # 880.5570 | Class II,<br>Product Code: FMI<br>Regulation # 800.5570 |
| | | Equivalent | Equivalent | Equivalent | Equivalent |
| Sterile, Disposable Single Use | Yes | Yes | Yes | Yes | Yes |
| Design: | The needle is 27ga., 3 ½"<br>long. It is used for uterine<br>anesthesia prior to Ob/Gyn<br>procedures. It consists of a<br>metal tube that is sharpened | Equivalent | Equivalent | Similar | Similar |
{4}------------------------------------------------
| 510(k) Premarket Notification Submittal<br>'Wallach Endocervical Block Needle'<br>By Wallach Surgical Devices, Inc.<br>SUMMARY CERTIFCIATION page of<br>SECTION 2<br>April. 15, 2002 -Page 5 of 5 | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|------------|------------|
| end joined to a female<br>connector (Luer lock)<br>designed to mate with a male<br>connector (Nozzle) of a<br>piston syringe or an intra-<br>vascular administration set.<br>Over 3 ¼" of its proximal end<br>a 21 gauge metal tube is<br>added to strengthen the shaft. | Equivalent | Equivalent | Equivalent |
| Material<br>Stainless Steel Tubing<br>Plastic Hub | Equivalent | Equivalent | Equivalent |
# Conclusion:
Based on the indications for use, technological characteristics and companson to currently marketed devices, the Wallach
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Michael Malis General Manager of Operations WALLACH Surgical Devices, Inc. 235 Edison Road ORANGE CT 06477
Re: K021224
JUL 12 2002
Trade/Device Name: Endocervical Block Needle Regulation Number: 21 CFR 884.5100 Regulation Name: Obstetric anesthesia set Regulatory Class: II Product Code: 85 HEE Dated: April 15, 2002 Received: April 17, 2002
Dear Mr. Malis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Wallach Surgical Devices, Inc. Section 7
Page 1 of 1
510(k) Number (if known): _ Koz1224
Device Name: _ _ Wallach Endocervical Block Needle
Indications For Use:
This needle is designed for injection of solutions (such as 2% lidocaine with or without 1:100,000 epinephrine) into the cervix. The application includes local anesthetics such as electro-excision, electro-fulguration, CO2 laser excision, and vaporization, and when required, endocervical curettage and cervical biopsies.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | ✓ | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|---|----|--------------------------------------------------|
|------------------------------------------|---|----|--------------------------------------------------|
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K021224 |
|---------------|---------|
|---------------|---------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.