SIMPLICTY SIMPLEX P INFUSION SET FOR USE WITH THE MINIMED PARADIGM INFUSION PUMPS
K021216 · Sterling Medivations, Inc. · FPA · May 3, 2002 · General Hospital
Device Facts
Record ID
K021216
Device Name
SIMPLICTY SIMPLEX P INFUSION SET FOR USE WITH THE MINIMED PARADIGM INFUSION PUMPS
Applicant
Sterling Medivations, Inc.
Product Code
FPA · General Hospital
Decision Date
May 3, 2002
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The intended use of the Simplicity Simplex P Infusion Set is to provide a means to infose or inject medication subcutaneously when attached to the Minibed Paradigm infinion pump or syringc.
Device Story
Simplicity Simplex P Infusion Set is an intravascular administration set for subcutaneous medication delivery. It consists of a 25-gauge PTFE indwelling catheter, a connecting tube, and a Simplicity YP connector. The device is introduced into subcutaneous tissue using a removable 27-gauge AISI 304 stainless steel insertion needle. Once inserted, the needle is removed, and the connector needle (AISI 304 stainless steel) is attached to the catheter to form a leak-proof seal. The proximal end connects to a MiniMed Paradigm pump or syringe via the YP connector. Used by patients with diabetes to facilitate continuous or intermittent medication flow. The device allows the patient to disconnect the tubing from the indwelling catheter. It is a sterile, single-use device.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including compliance with ISO 10555 (sterile, single-use intravascular catheters), ISO 11135 (ethylene oxide sterilization), ISO 10993-1 (biocompatibility), and ISO 9626 (stainless steel needle tubing).
Technological Characteristics
Materials: PTFE (catheter), AISI 304 stainless steel (needles), polyethylene (tubing). Dimensions: 25-gauge catheter, 27-gauge insertion needle. Connectivity: Mechanical interface to MiniMed Paradigm pumps. Sterilization: Ethylene oxide (ISO 11135). Standards: ISO 10555, ISO 11607, ISO 8535, ISO 10993-1, ISO 9626.
Indications for Use
Indicated for people with diabetes requiring subcutaneous infusion or injection of medication via MiniMed Paradigm infusion pump or syringe.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
Sterling Medivations Simplicity Soft YP Infusion Set (K011187)
Sterling Medivations Simplicity Simplex Soft Infusion Set (K020629)
Related Devices
K021530 — SIMPLICITY EASY P SOFT INFUSION SET FOR INFUSION AND/OR INJECTION OF FLUIDS INTO THE BODY BELOW THE SURFACE OF THE SKIN · Sterling Medivations, Inc. · May 21, 2002
K021296 — SIMPLICITY QD EASY P INFUSION SET · Sterling Medivations, Inc. · May 14, 2002
K021487 — SIMPLICITY EASY ACCESS P INFUSION SET · Sterling Medivations, Inc. · May 17, 2002
K011580 — SIMPLICITY SOFT QD SILVER INFUSION SET · Sterling Medivations, Inc. · Jun 5, 2001
K013104 — SIMPLICITY QD SOFT INFUSION SET FOR USE BY PEOPLE WITH DIABETES IN INFUSE INSULIN SUBCUTANEOUSLY FROM A PUMP OR SYRINGE · Sterling Medivations, Inc. · Oct 2, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
# MAY 0 3 2002
K021216
Sterling Medivations, Inc. 66 Neptune Drive Groton, CT 06340 650-814-4083(voice) 770-242-3178(fax)
510(k) SUMMAKY
Date Submitted: April 16, 2001
| Submitter: | Sterling Medivations, Inc. 66 Neptune Drive, Groton, CT 06340<br>Company Phone 650-814-4083, Company Fax 770-242-3178 | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact: | Jocl Douglas, Chief Technology Officer<br>Sterling Medivations, Inc.<br>Applicant Phone 650-814-4083<br>Applicant Fax 770-242-3178 | |
| Trade Name of Device: | Simplicity™ Simplex P Infusion Set for use with<br>the MiniMed Paradigm infusion pumps. | |
| Common Name of Device: | Intravascular administration set. | |
| Classification Name: | Percutaneous intravascular catheter. | |
| Predicate Device | Sterling Medivations, Sterling Medivations | |
Sterling Medivations Sterling Medivations Simplicity Simplex Soft Infusion Set, FDA 510 (k) K 020629 and Sterling Medivations Simplicity Soft YP Infusion Set K01187.
Description of the New Device: Sterling Medivations Inc.'s ("SMI") Simplicity™ Simplex P Infusion Set is designed to provide a nicans to infuse or inject medication subcutaneously when attached to the MiniMed Paradigni infusion pump or syringe.
The Simplicity Simplex P Infusion Set is designed for use by people with diabetes to provide a means to infuse or inject medication subcutaneously when attached to the MiniMcd Paradigm influsion purpor syringe. It is inserted into the subcutaneous tissue of a patient and the Simplicity YP concector is stached to the medication reservoir. This is substantially equivalent to the predicate device and it has the same intendcd use.
The device cousists of four main parts: (1) an infusion catheter made from PTFE, (2) an infusion lub that provides the patient the capability of disconnecting the comnecting tube from the infusion catheren (3) a connecting tube and (4) a Simplicity YP connector consisting of a needle hub and reused causes. (1) a and reservoir connector.
The Simplicity Simplex P Infusion Set is an infusion administration set, connecting to the Paradig m reservoir by means of a Simplicity YP connector consisting of a needle but and reuseble purpose and rescrvoir connector and subcutancously in the patient through an indwelling catheter made of Polytetrafluorthylene (PTFE). The Stcrling Medivations Simplicity Simplex P Infusion Set may je used with any paradigm infusion device that delivers continuous or internattent flow. The connecting tubing is made from a polyethylene tube.
The 25 gauge-indwelling catheter is introduced into the subcutancous tissue by a removable 27-gauge insertion necdle formed from a lumen made of AISI 304 stainless stecl. The insertion needle is removed and a connector needle is attached to the indwelling catheter. This connector ne-dle mates with the indwelling catheter forming a seal that permits the infusion of medication without leakage. The connector needle is made from AISI 304 stainless stcel and it is connected to the connecting tubing with a connector housing. The connector tubing proximal end is attached to a Simplicity YP connector consisting of a needle bub and reusable pump adapter and reservoir connection to a MiniMed Paradign pump and reservoir. The connecting tube is solvent bonded to the connector housing and to the aecdle hub
Sterling Mcdivations, Inc. Simplicity Simplex P Infusion Set 510(k) Summary l'age 1 of 2
{1}------------------------------------------------
connector.
Intended Use of the New Device: The intended use of the Simplicity Simplicx P Infusion Set is to provide a means to infose or inject medication subcutaneously when attached to the Minibed Paradigm infinion pump or syringc. The Simplicity Simplex Infision Set is substantially equivalent to the Sterling pulling of Syring. "The Bimplex Set K011187 and Simplicity Simplex Soft Infusion Set, ITTA S10 (k) K 020629.
Comparison of the Technological Features of the New Device and Predicate Device:
The Simplicity Simplex P Infusion Set proposed for conmercial distributor in similar in all signification The onlinessy Business I Thereon Simplicity Soft YP Infusion Set K011187 and Simplicity Simplex Soft Infusion Set, FDA 510 (k) K 020629.
The matcrials and manufacturing processes are substantially equivalcnt, the labeling is substantially equivalent and it has the same intended use as the Sterling Medivations Simplicity Simplicity Simplex P Infusion.
The differences that exist between the new and predicate device are as follows:
- The new device proximal end is a Simplicity YP connector consisting of needle hub and reusable 1. pump adapter and reservoir connector used to attach to a MiniMed Paradigm infusion pu mps and pullip adapter and reservour comments asserts the Sterling Medivations Simpling Simpling Simples of YP Infusion Set K011187 and the Device otherwise similar to the Simplicity Simplicity Simplex Soft . Infusion Set, FDA 510 (k) K 020629.
Performance Data Supporting Substantial Equivalence: To prove substantial equivalence both I Criol mance bata Bappor this Decessions Medivations Simplicity Simplex Infusion Sct FDA 5 . 0(x) K010373 meet the catheter requirements of:
- CDRH 21 C.F.R. Section 880.5440 Intravascular administration set .
- ISO 10555 Sterile, single use intravascular catheters (Part 1: General Requirements) .
- ISO 10555 Stcrile, single use intravascular catheters (Part 5: Peripheral Catheters). .
- ISO 11135:1994 Medical devices -- Validation and routine control of cthylenc oxide sterilization .
- 100 11133.1794 Sterilization of health care products -- Biological indicators -- Part 2: Biological . indicators for ethylene oxide sterilization.
- ISO 9626: 1991 Stainless Steel needle tubing for the manufacture of medical devices. 9
- ISO 11607: 1997 Packaging for terminally sterilized medical devices. .
- ISO 8535: 1991 Sterile single use syringes, with or without needle, for insulin. .
- 100 0355. 1991 bearing of the Limulus Amebocyte Lysate (LAL) Test as an end product . PDA Onlines on validation of the Enemeral drugs, biological products, and medical devices.
- ODE Blue Book Memorandum #K90-1. ●
- ISO 10993-1, Biological evaluation of medical devices Part 1: Evaluation and testing. .
And the design process adhcred to is the Center for Devices and Radiological Health. DESC(GN And the ucsign process bances to is and Comes CTURERS. This Guid: acc CONTROL GOLDANCE FOR MEDICAL DE . CONTRO 9001. This is substantially equivalent to the predicate device.
Signed,
Joel S. Douglas Chicf Technology Officer
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Joel Douglas Chief Technology Officer Sterling Medivations, Incorporated 66 Neptune Drive Groton, Connecticut 06340
MAY 0 3 2002
Re: K021216
Trade/Device Name: Simplicity Simplex P Infusion Set Regulation Number: 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: April 16, 2002 Received: April 17, 2002
#### Dear Mr. Douglas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{3}------------------------------------------------
## Page 2 - Mr. Joel Douglas
You must comply with all the Act's requirements, including, but not limited to: registration r ou indst comply with art 807); labeling (21 CFR Part 801); good manufacturing practice allu listing (21 CFR Part 807), lacelity systems (QS) regulation (21 CFR Part 820); and if requirents as set form in the quality is a control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 110(k) premarket notification. The FDA finding of substantial equivalence of your device to 910(t) promaried nowledge device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 If you desire specific ad 1100 for in vitro diagnostic devices), please contact the and additionally 21 CF (301) 594-4618. Additionally, for questions on the promotion and Office of Confightance wee, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and the Act may be obtained from the Brumber (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{4}------------------------------------------------
K02/2/6
## 510(k) Number (if known):
Device Name: Simplicity Simplex P Infusion Set
## Indications For Use:
The intended use of the Simplicity Simplex P Infusion Set is to provide a means for infusion and/or injection of fluids into the body the surface of the skin when attached to a MiniMed Paradigm pump or syringe.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (PER 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Patrice Accente
action Control, ﺔ ﺍﻟﻤ
Page 7 of 98
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.