MICROSCAN MICROSTREP PLUS PANEL NEW ANTIMICROBIAL-CLINDAMYCIN
K021184 · Dade Microscan, Inc. · LTT · Jun 20, 2002 · Microbiology
Device Facts
Record ID
K021184
Device Name
MICROSCAN MICROSTREP PLUS PANEL NEW ANTIMICROBIAL-CLINDAMYCIN
Applicant
Dade Microscan, Inc.
Product Code
LTT · Microbiology
Decision Date
Jun 20, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1640
Device Class
Class 2
Intended Use
To determine bacterial antimicrobial agent susceptibility
Device Story
MicroScan MICroSTREP plus Panel is a miniaturized broth dilution susceptibility test. Input: bacterial colonies grown on solid media, inoculated into Mueller-Hinton broth supplemented with 2-5% lysed horse blood and 50 mM HEPES. Device contains various antimicrobial agents diluted in water/buffer. Process: incubation for 20-24 hours at 35°C in non-CO2 incubator. Output: visual observation of lowest antimicrobial concentration inhibiting growth (MIC). Used in clinical laboratories by trained personnel. Results assist healthcare providers in selecting appropriate antimicrobial therapy for streptococcal infections.
Clinical Evidence
Bench testing only. External evaluation conducted using fresh and stock Efficacy isolates and stock Challenge strains. Performance compared against NCCLS frozen Reference Panel. Results showed 96.2% Essential Agreement for Clindamycin. Reproducibility and precision testing demonstrated acceptable results. Quality control testing met acceptance criteria.
Technological Characteristics
Miniaturized broth dilution panel. Reagents: Mueller-Hinton broth, 2-5% lysed horse blood, 50 mM HEPES buffer. Antimicrobial agents: Clindamycin (0.015-2 mcg/ml). Incubation: 20-24 hours at 35°C +/- 1°C in non-CO2 incubator. Manual visual reading of MIC. Standalone diagnostic test.
Indications for Use
Indicated for determining antimicrobial susceptibility of aerobic streptococci, excluding Streptococcus pneumoniae, using the MicroScan MICroSTREP plus Panel with Clindamycin (0.015-2 mcg/ml).
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
K021169 — MICROSTREP PLUS PANEL, AZITHROMYCIN · Dade Microscan, Inc. · Jun 6, 2002
K020939 — MICROSCAN MICROSTREP PLUS PANEL-TETRACYCLINE · Dade Microscan, Inc. · May 17, 2002
K020686 — MICROSCAN MICROSTREP PLUS PANEL NEW ANTIMICROBIAL - TRIMETHOPRIM/SULFAMETHOXAZOLE · Dade Microscan, Inc. · Apr 30, 2002
K021111 — MICROSCAN MICROSTREP PLUS PANEL NEW ANTIMICROBIAL - CEFOTAXIME · Dade Microscan, Inc. · Jun 6, 2002
K062293 — MICROSCAN MICROSTREP PLUS PANEL CHLORAMPHENICOL (1 TO 32 MCG/ML) · Dade Behring, Inc. · Aug 24, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
# JUN 2 0 2002
#### 510(k) Summarv Information:
| Device Manufacturer: | Dade MicroScan Inc. |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Contact name: | Cynthia Van Duker, Regulatory Affairs Manager |
| Fax: | 916-374-3144 |
| Date prepared: | April 12, 2002 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® MICroSTREP plus™ Panel |
| Intended Use: | To determine bacterial susceptibility to Clindamycin |
| Indication for Use | For determining antimicrobic susceptibility with aerobic streptococci, other<br>than Streptococcus pneumoniae |
| Predicate device: | MicroScan® Streptococcus MIC Panel (K963641). |
### 510(k) Summary:
The MicroScan® MICroSTREP plus™ Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 ul Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® MICroSTREP plus™ Panel demonstrated substantially equivalent performance with streptococcal isolates when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000.
The Premarket Notification (510[k]) presents data in support of the new MCroSTREP plus™ Panel with Clindamycin.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed MICroSTREP plus™ Panel by comparing its performance with an NCCLS frozen Reference panel. The MCroSTREP plus™ Panel demonstrated acceptable performance with an overall Essential Agreement of 96.2% for Clindamycin when compared with the frozen Reference panel.
Reproducibility testing demonstrated acceptable reproducibility and precision with Clindamycin.
Quality Control testing demonstrated acceptable results for Clindamycin.
{1}------------------------------------------------
K021184
# JUN 2 0 2002
#### 510(k) Summary Information:
| Device Manufacturer: | Dade MicroScan Inc. |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Contact name: | Cynthia Van Duker, Regulatory Affairs Manager |
| Fax: | 916-374-3144 |
| Date prepared: | April 12, 2002 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® MICroSTREP plus™ Panel |
| Intended Use: | To determine bacterial susceptibility to Clindamycin |
| Indication for Use | For determining antimicrobic susceptibility with aerobic streptococci, other than Streptococcus pneumoniae |
| Predicate device: | MicroScan® Streptococcus MIC Panel (K963641). |
#### 510(k) Summary:
The MicroScan® MICroSTREP plus™ Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including Streptococcus pneumoniae. After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test. Various antimicrobial agents are diluted in water, buffer or minute concentrations of broth to concentrations bridging the range of clinical interest. Panels are rehydrated with 115 ul Mueller-Hinton broth supplemented with 2-5% lysed horse blood (LHB) and buffered with 50 mM HEPES, after inoculation of the broth with a standardized suspension of the organism in saline. After incubation in a non-CO2 incubator for 20-24 hours, the minimum inhibitory concentration (MIC) for the test organism is manually read by observing the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® MICroSTREP plus™ Panel demonstrated substantially equivalent performance with streptococcal isolates when compared with an NCCLS frozen Reference Panel, as defined in the FDA DRAFT document "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", dated March 8, 2000.
The Premarket Notification (510/k]) presents data in support of the new MICroSTREP plus™ with Clindamycin.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed MICroSTREP plus™ Panel by comparing its performance with an NCCLS frozen Reference panel. The MICroSTREP plus™ Panel demonstrated acceptable performance with an overall Essential Agreement of 96.2% for Clindamycin when compared with the frozen Reference panel.
Reproducibility testing demonstrated acceptable reproducibility and precision with Clindamycin.
Quality Control testing demonstrated acceptable results for Clindamycin.
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
## JUN 2 0 2002
Ms. Cynthia Van Duker Regulatory Affairs Manager Dade Behring Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691
Re: k021184
> Trade/Device Name: MicroScan® MICroSTREP plus™ Panel For Clindamycin 0.015-2ug/ml Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Regulatory Class: Class II Product Code: LTT Dated: April 12, 2002 Received: April 15, 2002
Dear Ms. Van Duker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indication for Use Statement
·
| 510(k) No.: | K021184<br>(To be assigned by FDA) |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | MicroScan <sup>®</sup> MICroSTREP plus™ Panel |
| Intended Use | To determine bacterial antimicrobial agent susceptibility |
| Indications for Use: | The MicroScan <sup>®</sup> MICroSTREP plus™ Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of aerobic streptococci, including <i>Streptococcus pneumoniae</i> . After inoculation, panels are incubated for 20 - 24 hours at 35°C +/- 1°C in a non-CO2 incubator, and read visually according to the Package Insert. |
| | This particular submission is for the addition of the antimicrobial Clindamycin at concentrations of 0.015 to 2 mcg/ml to the test panel |
| | The organisms which may be used for Clindamycin susceptibility testing in this panel are: |
| | Streptococci |
| | The MicroScan <sup>®</sup> MICroSTREP plus™ Panel with Clindamycin is not intended for use with: |
| | <i>Streptococcus pneumoniae</i> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF
NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubves(Division Sign-Off)
Division of Clihical Laboratory Devices KO21184 510(k) Number_
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
8
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.