ZEUS MICROWRIST ROBOTIC SURGICAL SYSTEM AND ACCESSORIES

K021152 · Computer Motion, Inc. · NAY · Sep 24, 2002 · Gastroenterology, Urology

Device Facts

Record IDK021152
Device NameZEUS MICROWRIST ROBOTIC SURGICAL SYSTEM AND ACCESSORIES
ApplicantComputer Motion, Inc.
Product CodeNAY · Gastroenterology, Urology
Decision DateSep 24, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ZEUS® MicroWrist™ Surgical System (ZEUS System) and Accessories are intended to be used to assist a surgeon during procedures such as Laparoscopic Cholecystectomy and Nissen Fundoplication, to hold and position an endoscope, and to control laparoscopic instruments in performance of the surgical tasks of grasping, sharp cutting, blunt dissection, electro-cautery and suturing with knot placement. The ZEUS System is intended to be used by surgeons who are trained in minimally invasive surgery, have successfully completed a ZEUS System training program, and are certified in accordance with their respective hospital's customary practice for ZEUS System use. The ZEUS® MicroWrist™ Surgical System is intended to be used in an operating room environment in which the ZEUS System, the operating surgeon and patient are in the same room.

Device Story

ZEUS MicroWrist Surgical System is a robotic-assisted platform for minimally invasive surgery; consists of surgeon console, three table-mounted arms, and AESOP endoscope positioner. HERMES Control Center provides voice-recognition for ancillary device control. Surgeon operates console in same room as patient; system manipulates endoscopic instruments to perform grasping, cutting, dissection, electro-cautery, and suturing. System provides steady endoscopic view and robotic control of surgical tasks. Benefits include robotic assistance for complex laparoscopic procedures. Surgeons must complete specific training and certification. Clinical trials demonstrated safety and effectiveness compared to standard laparoscopic surgery.

Clinical Evidence

Two prospective, randomized, concurrently controlled clinical trials compared ZEUS System to standard laparoscopic surgery. Study included over 200 patients. Primary endpoints focused on safety and effectiveness for Laparoscopic Cholecystectomy and Nissen Fundoplication. Results demonstrated successful task completion and procedure times; surgeons required a minimum 13-case learning curve to approximate control procedure times.

Technological Characteristics

Robotic-assisted surgical platform; includes surgeon console, three table-mounted arms, and voice-controlled endoscope positioner (AESOP). Complies with EN60601-1, EN60601-1-1, EN60601-1-2, UL 2601-1, and various EN61000-4-x immunity standards. Software-assisted control of surgical instruments. System is intended for operating room use.

Indications for Use

Indicated for patients undergoing laparoscopic procedures, specifically Laparoscopic Cholecystectomy and Nissen Fundoplication, requiring endoscopic positioning and robotic-assisted surgical instrument control for grasping, cutting, dissection, electro-cautery, and suturing.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ # SEP 2 4 2002 z1152 #### 510(k) Summary ## ZEUS® MicroWrist™ Surgical System and Accessories In accordance with 21 CFR section 807.92 Computer Motion is submitting the following safety and effectiveness summary. #### 1) Submitter Information Computer Motion. Inc. 130-B Cremona Drive Goleta, CA 93117 Contact: David Munjal, Ph.D., R.A.C. Vice President, CA/RA/QA #### 2) Name of Device: | Proprietary Name: | ZEUS® MicroWrist™ Surgical System (ZEUS System) and Accessories | |----------------------|-----------------------------------------------------------------| | Common Names: | ZEUS System, ZEUS MicroWrist, ZEUS MW | | Classification Name: | Laparoscope, General & Plastic Surgery | | Regulation Number: | 876.1500 | | Product Code: | GCJ | | Class: | Class II. | #### 3) Substantial Equivalence: This submission establishes the substantial equivalence of the ZEUS® MicroWrist™ Surgical System and Accessories to two predicate devices: - A standard laparoscope with hand-held minimally invasive surgical instruments 1. - The Intuitive Surgical da Vinci^4 Endoscopic Instrument Control System and 2. Endoscopic Instruments, cleared to market via K990144. #### 4) Description of the Device: The ZEUS® MicroWrist™ Surgical System (ZEUS System) and Accessories, consisting of a surgeon console and three table-mounted arms, serves as a platform for holding, positioning, and manipulating endoscopic instruments in order to perform selected surgical tasks. One arm of the ZEUS System incorporates the AESOP® endoscope positioner, which provides the surgeon with a steady view of the internal operating field. The HERMESTM Control Center, which uses voice-recognition technology to control devices outside the sterile field, is a standard component of the ZEUS® MicroWrist™ Surgical System and Accessories, and has been cleared to operate with many ancillary devices. {2}------------------------------------------------ The ZEUS® MicroWrist™ Surgical System (ZEUS System) and Accessories is designed and tested to the following standards: - International Standard for Medical Electrical Equipment EN60601--1 ● - General Requirements for Safety -- Collateral Standard EN60601-1-1 - Emissions and Immunity Test Measurements EN 60601-1-2 . - Underwriters Laboratory UL 2601-1 ● - EN55011/A1; CISPR 11 Conducted & Radiated Emission ● - EN61000-4-2 ESD Immunity 4kV contact 8kV air ● - EFT Immunity EN61000-4-4 ● - EN61000-4-6. Conducted Immunity Tests ● - EN 61000-4-3 RF Immunity . - EN 61000-4-5 Surge Immunity ● ## 5) Intended Use The ZEUS® MicroWrist™ Surgical System (ZEUS System) and Accessories are intended to be used to assist a surgeon during procedures such as Laparoscopic Cholecystectomy and Nissen Fundoplication, to hold and position an endoscope, and to control laparoscopic instruments in performance of the surgical tasks of grasping, sharp cutting, blunt dissection, electro-cautery and suturing with knot placement. The ZEUS System is intended to be used by surgeons who are trained in minimally invasive surgery, have successfully completed a ZEUS System training program, and are certified in accordance with their respective hospital's customary practice for ZEUS System use. The ZEUS MicroWrist™ Surgical System is intended to be used in an operating room environment in which the ZEUS System, the operating surgeon and patient are in the same room. ### 6) Technological Characteristics in Comparison to the Predicates The difference between a standard laparoscope and the ZEUS System and Accessories is that a standard laparoscope and its associated instruments are hand-held devices not software-assisted. The daVinci predicate is software-assisted, but there are technical differences between this predicate and the ZEUS System and Accessories. However, the ZEUS System and Accessories is substantially equivalent to a standard laparoscope with its associated hand-held minimally invasive surgical instruments, and to the Intuitive Surgical daVinci Endoscope Instrument Control System and Endoscopic Instruments. ## 8) Non-Clinical Tests A hazard analysis and associated validations were completed in response to the new Indication for Use. ## 9) Clinical Trials Two extensive, prospective, randomized, concurrently controlled clinical trials compared the ZEUS® MicroWrist™ Surgical System and Accessories to standard laparoscopic surgery to demonstrate substantial equivalence to the predicate devices cited in terms of safety and effectiveness. {3}------------------------------------------------ ## 10) Conclusion drawn from the Non-Clinical Tests and the Clinical Trials Data included in this submission, including the results from clinical studies with over 200 patients, demonstrate the safety and effectiveness of the ZEUS® MicroWrist™ Surgical System and Accessories for surgical procedures, such as Laparoscopic Cholecystectomy and Nissen fundoplication. Training is demonstrated by the consistent ability and time to complete a task and procedure successfully. For highly selected uncomplicated Laparoscopic Cholecystectomy and Laparoscopic Nissen Fundoplication, during clinical trials, surgeons trained in the ZEUS System and minimally invasive surgery experienced a minimum 13-case learning curve to approximate procedure time with control. {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Dr. David Munjal Vice President Clinical, Regulatory and Quality Assurance Computer Motion, Inc. 130-B Cremona Drive Santa Barbara, California 93117 Re: K021152 Trade/Device Name: ZEUS MicroWrist™ Surgical System and Accessories Regulation Number: 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NAY Dated: July 12, 2002 Received: July 23, 2002 Dear Dr. Munjal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 -- Dr. David Munjal This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark-McMillian Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # INDICATION FOR USE STATEMENT 510(k) Number: K021152 Device Name: ZEUS® MicroWrist™ Surgical System and Accessories The ZEUS® MicroWrist "Surgical System (ZEUS System) and Accessories are intended to be used to assist a surgeon during procedures such as Laparoscopic Cholecystectomy and Nissen Fundoplication, to hold and position an endoscope, and to control laparoscopic instruments in performance of the surgical tasks of grasping, sharp cutting, blunt dissection, electro-cautery and suturing with knot placement. The ZEUS System is intended to be used by surgeons who are trained in minimally invasive surgery, have successfully completed a ZEUS System training program, and are certified in accordance with their respective hospital's customary practice for ZEUS System use. The ZEUS® MicroWrist™ Surgical System is intended to be used in an operating room environment in which the ZEUS System, the operating surgeon and patient are in the same room. ## (PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CRDH, Office of Device Evaluation (ODE) Mark N. Milliken (Division Sign-Off) Di ision of General, Restorative and Neurological Devices 510(k) Number ***_***__ \\$\underline{K02152} Prescription Use (per 21 CFR 801.109) OR Over-the-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...