K021052 · Global Treasure Industries , Ltd. · FLL · Apr 9, 2002 · General Hospital
Device Facts
Record ID
K021052
Device Name
MODEL GT010706 ELECTRONIC THERMOMETER
Applicant
Global Treasure Industries , Ltd.
Product Code
FLL · General Hospital
Decision Date
Apr 9, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Indicated use-to measure patient temperatures-orally, axillary and rectal. Targeted population-Any patient requiring body temperatures measured. Environment of use-Hospital and home.
Device Story
Electronic digital thermometer for measuring patient body temperature. Input: thermal signal from thermistor sensor. Processing: CMOS signal processing converts sensor input to digital temperature reading. Output: temperature value displayed on LCD; audible buzzer alarm indicates measurement completion. Used in hospital and home settings by patients or clinicians. Device features automatic shut-off and removable battery. Benefits patient by providing quick, accurate temperature monitoring.
Clinical Evidence
Bench testing only. Performance testing confirmed temperature range (90.0-109.9°F), accuracy (+/- 0.2°F), and compliance with ASTM E1112 standards. No clinical data provided.
Technological Characteristics
Thermistor sensor; CMOS signal processing; LCD display; 1.5V button battery power; ABS and thermoplastic elastomer case; stainless steel probe. Accuracy +/- 0.2°F. Compliant with ASTM E1112 and EN ISO 10993 (biocompatibility).
Indications for Use
Indicated for measuring body temperature in any patient requiring such measurement, via oral, axillary, or rectal routes. Range: 90.0°F - 109.9°F.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{1}------------------------------------------------
APR = 9 2002
K021652
FS LIMITED
## GLOBAL TREASURE INDUSTRIES LIMITE
Block 2, 5/F, Room 8, Nan Fung Ind. City, No. 18, Tin Hau Road,
Tuen Mun, New Territories, Hong Kong
Tel: (852)24541493; Fax: 24546187; E-mail: glotr@netvigator.com
### The Non-confidential summary of safety and Effectiveness
The summary of 510k safety and effective new information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR807.92. The assigned 510k number is K _____________ (to be assigned)
Dated of summary prepared: March 12, 2002
| 1. | Submitter's identification<br>Global Treasure Industries Limited<br>Block 2,5F., Room 8, Nan Fung Ind. City, 18 Tin Hau Road, Tuen Mun, N.T., Hong Kong<br>Tel:(852)24541883 FAX:(852)24546187 | | | |
|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------|-------------------------------------|
| 2. | Official contact: | Ben Ma - General Manager | | |
| 3. | Trade name: | GT010706 sec Digital thermometer | | |
| 4. | Common name: | Clinical electronic thermometer | | |
| 5. | Classification name: | 80FLL, Clinical electronic thermometer subsection 880.2910 | | |
| 6. | Intended device: | Clinical electronic thermometer | | |
| 7. | Predicate devices: | Wiltec Industries Ltd-Electronic Thermometer-K961879 | | |
| 8. | Device description: | The intended product is an electronic digital thermometer for measuring patient temperature. | | |
| 9. | Intended use: | | | |
| 9.1 | | Indicated use-to measure patient temperatures-orally, axillary and rectal. | | |
| 9.2 | | Targeted population-Any patient requiring body temperatures measured | | |
| 9.3 | | Environment of use-Hospital and home | | |
| 10. | | Comparison to predicate devices: | | |
| 10.1 | | Side by side Comparison Table | | |
| | | Element of comparison | Subject device Global<br>Treasure Industries<br>GT010706 | Claimed SE device<br>Wiltec:K961879 |
| a. | Use | Indicated for taking temperature | Yes | Yes |
| 10.1 | Side by side Comparison Table (continued) | | | |
| b. | Types of temperature | | | |
| | Oral | Yes | Yes | |
| | Underarm | Yes | Yes | |
| | Rectal | Yes | Yes | |
| | Digital/electronic thermometer | Yes | Yes | |
| c. | Design | | | |
| | LCD display | Yes | Yes | |
| | Temperature increments of 0.1° F | Yes | Yes | |
| | Sensor type- thermistor | Yes | Yes | |
| | Signal processing-CMOS | Yes | Yes | |
| | Power -Button battery 1.5V | Yes | Yes | |
| | On / off button | Yes | Yes | |
| | Buzzer | Yes | Yes | |
| | Removable battery case | Yes | Yes | |
| | Cleaned with alcohol | Yes | Yes | |
| d. | Material | | | |
| | Case | ABS and Thermoplastic<br>Elastomer | ABS | |
| | Sensor cover | Stainless steel | aluminium | |
| e. | Performance testing | | | |
| | Temperature range | 90.0-109.9° F | 89.6-109.4° F | |
| | Ambient temperature | 60.8-104° F | 60.8-104° F | |
| | Beeps alarum | Yes | Yes | |
| | Response time | About 20 seconds | About 1 minute | |
| | Automatic shut off | Yes | Yes | |
| f. | Accuracy and Performance meets | | | |
| | ASTM E1112 | Yes | Yes | |
| 10.2 | Difference | | | |
{2}------------------------------------------------
# GLOBAL TREASURE INDUSTRIES LIMITED
Block 2, 5/F, Room 8, Nan Fung Ind. City, No. 18, Tin Hau Road, Tuen Mun, New Territories, Hong Kong Tel: (852)24541493; Fax: 24546187; E-mail: glotr@netvigator.com
.
{3}------------------------------------------------
| (a) | There are several difference in between the subject and legally marketed devices: |
|-----|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <ul><li>the subject device has a quicker response time</li><li>the subject device compose of different material: thermoplastic elastomers and stainless steel probe</li><li>the subject device has a slight difference in the working range of measurement</li></ul> |
| (b) | Although there are differences in between the subject device and the legally marketed one, they do not affect the safety, performance and accuracy of the subjected device. All the materials used are non-toxic and in compliance with EN ISO 10993. The performance and accuracy are in compliance with ASTM requirements. The subject device is still considered to be substantially equivalent to the legally marketed one. |
and the comments of the comments of the comments of
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
11 - 1 - 1 - 1 - 1 - 1 -
and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contraction of the contribution of the
.
:
.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, with flowing lines.
## MAY 1 4 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Global Treasure Industries Limited C/O Mr. Ned Devine, Jr Responsible Third Party Official Entela, Incorporated 3033 Madison Avenue, SE Grand Rapids, Michigan 49548
Re: K021052
Trade/Device Name: Model GT010706 Electronic Thermometer Regulation Number: 880.2910 Regulation Name: Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: March 12, 2002 Received: April 1, 2002
Dear Mr. Devine:
This letter corrects our substantially equivalent letter of April 9, 2002 regarding the company name.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [{for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Devine
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours.
Victor
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
### D. Indications for Use Statement
Pursuant to the Notice regarding listing of indications for Use on a separate sheet, the following is per that request.
510(k) Number ________________________________________________________________________________________________________________________________________________________________ (To be assigned)
| Device name: | Clinical electronic thermometer |
|------------------------|---------------------------------------------------------------|
| Indications for use: | |
| Indicated use- | Measure of individual temperature |
| Measurements- | Oral |
| | Axillary -under arm |
| | Rectal |
| Range of measurement- | $90.0^\circ$ - $109.9^\circ$ F ( $32.0^\circ$ - $43^\circ$ C) |
| Accuracy- | +/-0.2° F |
| Targeted population- | Individuals requiring temperature measurements |
| Environment of use- | Hospital and home |
| Disposable / reusable- | Reusable, clean with alcohol |
Concurrence of CDRH, Office of Device Evaluation(ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
PRESCRIPTION USE
OR
OVER-THE-COUNTER USE
(PER 21 CFR 801.109)
. . . . . . . . . .
Patrica Cuscente
(Division Sign-Off) ് wision of Dental, Infection Control, and General Hospital Dev 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.