The AnnuloFlo® System is intended for use in the repair of the human cardiac valve.
Device Story
AnnuloFlo® System is an annuloplasty ring used during cardiac surgery to reinforce the native mitral valve annulus while preserving the native valve apparatus. The system includes the ring and associated surgical instrumentation for sizing and implantation. The ring features a kidney-shaped titanium stiffener enclosed in a silicone and knitted polyester sewing cuff. It is intended for use by cardiac surgeons in an operating room setting. By reinforcing the annulus, the device supports the repair of damaged mitral valves, potentially avoiding the need for full valve replacement. The materials are non-ferrous and MRI-compatible.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material composition, and intended use comparisons to the predicate device.
Technological Characteristics
Annuloplasty ring consisting of a titanium stiffener ring enclosed in a silicone and knitted polyester fabric sewing cuff. Non-ferrous materials. Kidney-shaped geometry. Sizes: 26, 28, 30, 32, 34, and 36 mm. MRI compatible.
Indications for Use
Indicated for reinforcement of human cardiac mitral valve repair in patients with acquired or congenital disease, or as a replacement for a previously implanted annuloplasty ring, provided visual inspection confirms the valve is repairable.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
Predicate Devices
AnnuloFlex™ Annuloplasty System
Related Devices
K970375 — ANNULOFLO SYSTEM · Carbomedics, Inc. · Jul 25, 1997
K992056 — ANNULOFLEX ANNULOPLASTY SYSTEM, MODEL AF800 · Sulzer Carbomedics, Inc. · Oct 12, 1999
K190506 — NeoForm Annuloplasty Ring · Genesee Biomedical, Inc. · Apr 18, 2019
K023185 — ANNULOFLEX ANNUOPLASTY SYSTEM, MODEL AF 800 · Sulzer Carbomedics, Inc. · Dec 17, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K021051
# APR 2 3 2002
### Summary of Safety and Effectiveness per SMDA 1990 and 21 CFR 807.92 AnnuloFlo® Annuloplasty System
| Submitter: | Sulzer Carbomedics, Inc.<br>1300 East Anderson Lane<br>Austin, Texas 78752<br>U.S.A. | Contact: | Teffany Hankinson<br>Regulatory Affairs Associate<br>Telephone: (512) 435-3202<br>Facsimile: (512) 435-3350 |
|------------|--------------------------------------------------------------------------------------|----------|-------------------------------------------------------------------------------------------------------------|
|------------|--------------------------------------------------------------------------------------|----------|-------------------------------------------------------------------------------------------------------------|
Classification Name: Annuloplasty Ring Date of Summary: March 27, 2002 Proprietary Name: AnnuloFlo® System Common Name: Annuloplasty Ring
Predicate Device: AnnuloFlex™ Annuloplasty System
Statement of Intended Use: The AnnuloFlo® System is intended for use in the repair of the human cardiac valve.
Statement of Indications for Use: The AnnuloFlo® annuloplasty ring is indicated as a reinforcement for repair of the human cardiac mitral valve damaged by acquired or congenital Itimoreement for topen of the a previously implanted annuloplasty ring. The annuloplasty ring dibouse, or as a repidentifical inspection confirms that the valve is repairable and does not require replacement.
Description of Device: The AnnuloFlo® System consists of an annuloplasty ring, and a Doserproasis of instrumentation provided in a tray to properly size the annulus and implant the annuloplasty ring. The annuloplasty ring is designed to reinforce the native annulus while retaining the native valve apparatus.
The AnnuloFlo® annuloplasty ring employs materials with a long and satisfactory history of use in cardiovascular applications. All materials are non-ferrous, have extensive history in implants and do not present a significant risk during Magnetic Resonance Imaging (MRJ). The annulonlasty ring consists of a titanium stiffener ring enclosed in a sewing ring of silicone and knitted polyester fabric.
The mitral AnnuloFlo® annuloplasty ring is kidney shaped. The curved posterior segment of the ring corresponds to the native posterior leaflet. The open segment of the ring corresponds to the anterior leaflet.
Technological Comparison: The sewing cuff material of the AnnuloFlo® annuloplasty ring is similar to the sewing cuff material used in the AnnuloFlex™ annuloplasty ring.
{1}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 2002
Sulzer Carbomedics, Inc. c/o Teffany Hankinson Regulatory Affairs Associate 1300 East Anderson Lane Austin, TX 78752
Re: K021051
Trade/Device Name: AnnuloFlo® System, Mitral AR-700 sizes 26, 28, 30, 32, 34, and 36 mm Regulation Number: 21 CFR 870.3800 Regulation Name: Annuloplasty ring Regulatory Class: Class II Product Code: KRH Dated: March 28, 2002 Received: April 1, 2002
Dear Ms. Hankinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
Page 2 – Ms. Teffany Hankinson
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
K970375 510(K) Number:
Device Name:
. ..
#### AnnuloFlo® System
Indications for Use:
The AnnuloFlo® annuloplasty ring is indicated as a reinforcement for repair of the human cardiac mitral valve damaged by acquired or congenital disease, or as a replacement for a previously implanted annuloplasty ring. The annuloplasty ring should be used only in cases where visual inspection confirms that the valve is repairable and does not require replacement.
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
Division of Cardiovascular & Respiratory Devices
510(k) Number K021051
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.