MEDEFIL'S NORMAL SALINE FLUSH SYRINGE IN VARIOUS FILL SIZES - 1 ML FILL IN 6 ML SYRINGE LL; 2 ML FILL IN 6 ML SYRINGE LL
K020999 · Medefil, Inc. · NGT · Jun 20, 2002 · General Hospital
Device Facts
Record ID
K020999
Device Name
MEDEFIL'S NORMAL SALINE FLUSH SYRINGE IN VARIOUS FILL SIZES - 1 ML FILL IN 6 ML SYRINGE LL; 2 ML FILL IN 6 ML SYRINGE LL
Applicant
Medefil, Inc.
Product Code
NGT · General Hospital
Decision Date
Jun 20, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
FOR USE TO MAINTAIN PATENCY OF IN-DWELLING INTRAVENOUS ACCESS DEVICES (IVAD)
Device Story
Single-dose, disposable, sterile, pre-filled syringe containing 0.9% Sodium Chloride Injection, USP. Components include hypodermic barrel, stopper plunger, plunger rod, and tip cap. Available in 6 mL and 12 mL sizes with various fill volumes. Operated manually by clinician; fluid expelled by pressing plunger rod through luer tip into IVAD. Used in clinical settings to maintain IV line patency. Larger barrel diameter reduces infusion pressure, minimizing potential vein damage, particularly in pediatric patients. Manufactured via fully automated aseptic filling process to reduce operator contact and contamination risk.
Clinical Evidence
Bench testing only. Device is pyrogen-free per LAL test method for bacterial endotoxin. No clinical data presented.
Technological Characteristics
Single-dose, sterile, pre-filled hypodermic syringe. Materials: Polypropylene (barrel), silicone (lubricant). Principle: Manual displacement of fluid via plunger. Connectivity: None. Sterilization: Aseptic filling in Class 100 environment. Form factor: 6 mL and 12 mL syringe barrels.
Indications for Use
Indicated for maintaining patency of in-dwelling intravenous access devices (IVAD) in patients requiring intravenous therapy.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Medefil, Inc. Normal Saline I. V. Flush Syringe (K993515)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "MEDEFIL, INC." in a bold, sans-serif font. Below the text is a circular logo featuring a stylized lowercase "m" with a dot above it, resembling an eye. The logo and text are in black and white.
## 510(k) Summary
## Medefil's Normal Saline Flush Syringe
## May 9, 2002
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
- 1. Reason for 510(k):
- 2. Name of Device:
Classification Name Common Name Proprietary Name
Introduction of a modified product
Catheter Intravascular, Small Volume 0.9% Sodium Chloride Injection, USP. Medefil's Normal Saline I. V. Flush Syringe
- 3. Classification:
Name/Class Panel Product Code 21 CFR 880.5200 Class II General Hospital NGT
- 4. Establishment Registration Number: 1423982
- 5. Submitter's Name and Address: Medefil, Inc. 492 Lunt Avenue Schaumburg, Illinois 60193
- 6. Repackaging and sterilization facilities:
Medefil, Inc. 250 Windy Point Drive Glendale Heights, Illinois 60139
000105
- 7. Performance Standards:
No performance standard(s) applicable to this device has been promulgated under Section 514 of the Food, Drug, and Cosmetic Act.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for MEDEFIL, INC. The text "MEDEFIL, INC." is in a bold, sans-serif font. Below the text is a circular logo with a stylized letter "m" inside. The logo is black and white.
- Device Description and Indicated Use: 8.
The Normal Saline Flush Syringe (the DEVICE) is a single dose, disposable, sterile, plastic pre-filled syringe. The DEVICE consists of a hypodermic syringe with a hypodermic barrel, stopper plunger, plunger rod, tip cap and 0.9% Sodium Chloride Iniection, USP. Two different sizes (6 mL, 12 mL) of hypodermic syringes will be utilized in manufacturing. The DEVICE will be marketed in the following dosage forms:
1 mL fill in 6 mL Syringe LL 2 mL fill in 6 mL Syringe LL 2.5 mL fill in 6 mL Syringe LL 3 mL fill in 6 mL Syringe LL 3 mL fill in 12 mL Syringe LL 5 mL fill in 6 mL Syringe LL 5 mL fill in 12 mL Syringe LL 10 mL fill in 12 mL Syringe LL
The barrel of the syringe is the reservoir for the product. It shall be overfilled during repackaging as per Pharmacopoeial Forum, Volume 26, Number 2: page 471. Plunger and plunger rod are the two moving parts of the DEVICE. The solution is delivered by pressing down on the plunger rod, resulting in the expulsion of the fluid from the luer tip.
The DEVICE will be filled using aseptic technique utilizing pre-sterilized components in a Class 100 environment. The sterile components of the DEVICE (hypodermic syringe, tip cap and 0.9% Sodium Chloride Injection, USP) will be purchased from qualified vendors.
The DEVICE is used to maintain patency of in-dwelling intravenous access devices (IVAD). The solution is delivered by pressing down on the plunger rod, resulting in the expulsion of the fluid from the luer tip.
- 9. Packaging:
The DEVICE provided is individually packaged in a plastic pouch (dust cover). There are six individually packaged DEVICES in a "strip". Twenty such strips (120 individually packed syringes) or 30 individually packed devices will be packed in a card board box.
The DEVICE shall be pyrogen free per the LAL test method for bacterial endotoxin.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for MEDEFIL, INC. The text "MEDEFIL, INC." is written in a bold, sans-serif font. Below the text is a circular logo with a stylized lowercase "m" inside. There is a small dot above the "m" inside the circle.
#### 10. Substantial Equivalence:
The Normal Saline Flush Syringe DEVICE is substantially equivalent to the legally marketed predicate device listed below:
## Medefil, Inc. Normal Saline I. V. Flush Syringe (K993515)
The Medefil's Normal Saline Flush Syringe Device is similar to the legally marketed predicate device in that:
- both have the same intended use: to maintain patency of indwelling IVAD a)
- b) both have the same concentration, 0.9% sodium chloride, pre-filled solution.
- c) both function under the same principle of operation including independent moving parts like stopper and plunger rods.
- d) both nozzle tips are compatible with female luer fitting.
- e) both fluid container materials are biocompatible and chemically compatible with 0.9% Sodium Chloride solution.
- both have a sterile, pvrogen free 0.9% sodium chloride fluid and fluid path. f)
- the fluid container material is same for both devices polypropylene g)
- both devices use silicone as a lubricant. h)
- i) the fluid is expelled out of the syringe barrel of both the devices by pressing on the plunger for hypodermic syringe
- both the devices are filled using aseptic filling process. i)
The DEVICE is different from the predicate device, without adverse affect to safety or efficacy, in that:
- a) the dead space (distance between plunger tip and end of the luer lock nozzle tip, when plunger is fully depressed) of the DEVICE is more than that of Medefil's Normal Saline I. V. Flush Syringe.
- b) the DEVICE is available in two different syringe barrel size 6 mL and 12 mL where as predicate device is also available in 3 mL size.
- c) the diameter of the DEVICE syringe barrel is more than that of the predicate device. This is due to the fact that diameter of the barrel is inversely proportional to the amount
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for MEDEFIL, INC. The text "MEDEFIL, INC." is in a bold, sans-serif font. Below the text is a circular logo with a stylized lowercase "m" inside. The logo is black and white.
of "infusion pressure" generated when the plunger is pushed for expulsion of fluid. The larger diameter of the barrel is advantageous so that damage is not done to the patient veins especially pediatric patients.
- d) the DEVICE will be manufactured using fully automates syringe filling machine whereas the predicate device is manufactured using semi - automated syringe filling machine. But both the machines work on the same principle. Semi - automated syringe filling process requires lot of involvement of the operators during the filling and sealing process but a fully automated process decreases operator contact and opportunity for product contamination.
## END OF SUMMARY
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure.
### · Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 2 0 2002
Mr. Pradeep Aggarwal President Medefil, Incorporated 492 Lunt Avenue Schaumburg, Illinois 60193
Re: K020999
Trade/Device Name: Normal Saline IV Flush Syringe Regulation Number: 880.5200 Regulation Name: Device, Flush, Vascular Access Regulatory Class: II Product Code: NGT Dated: May 28, 2002 Received: May 31, 2002
Dear Mr. Aggarwal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
## Page 2 - Mr. Aggarwal
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulstowski
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Medefil, Inc. March, 2002 Sterile Normal Saline I. V. Flush Syringe Indications For Use
Page 1 :
## 510(k) Number (if known): __ Pending__________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________ Normal Saline I. V. Flush Syringe
Indications For Use:
## FOR USE TO MAINTAIN PATENCY OF IN-DWELLING INTRAVENOUS ACCESS DEVICES (IVAD)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
le, Cll. Wtard
(Division Sign-Off) Division of Dental, Infection Control D209999 and General Hospital Devices 510(k) Number -
(Optional Format 3-10-98)
000019
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.