← Product Code [ING](/productcode/ING) · K020933

# GARAVENTA SUPER-TRAC TRE-52 (K020933)

_Garaventa Accessibility · ING · Apr 4, 2002 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K020933

## Device Facts

- **Applicant:** Garaventa Accessibility
- **Product Code:** [ING](/productcode/ING.md)
- **Decision Date:** Apr 4, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 890.3930
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

The Super-Trac TRE-52 is intended to mechanically transport a single individual in a wheelchair up and down stairs in a private or public facility. The Super-Trac must be operated by a trained attendant.

## Device Story

Portable wheelchair lift; attendant-operated; transports wheelchair-bound passengers up/down stairs. Input: passenger in wheelchair loaded onto platform via integrated ramp. Mechanism: two 12V DC motors drive endless belt tracks; hydraulic actuator tilts platform to stair-travel position; auxiliary wheels for landing maneuverability. Output: mechanical movement of wheelchair along stairs. Used in indoor/outdoor public/private settings. Attendant controls device via handle-mounted controls; key switch restricts access; stair-slope indicator ensures safe operating angle. Benefits: provides accessibility in buildings without elevators; portable; stores compactly.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Portable wheelchair lift; 12V DC motor drive; 2x 12V 24AH rechargeable batteries; endless belt track propulsion; hydraulic actuator for platform tilt; integrated ramp; key switch; stair-slope indicator. Load capacity: 130kg (285lbs). Dimensions: 935mm H x 1473mm D x 635mm W. Weight: 54kg total.

## Regulatory Identification

A permanently mounted wheelchair platform lift is a motorized vertical or inclined platform lift device permanently installed in one location that is intended for use in mitigating mobility impairment caused by injury or other disease by providing a guided platform to move a person from one level to another, with or without a wheelchair. A portable wheelchair elevator is a motorized lift device that is not permanently mounted in one location and that is intended for use in mitigating mobility impairment caused by injury or other disease by providing a means to move a person, with or without a wheelchair, from one level to another (e.g., portable platform lifts, attendant-operated stair climbing devices for wheelchairs).

## Special Controls

(i) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate that the safety controls are adequate to prevent a free fall of the platform in the event of a device failure; 
(ii) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate the ability of the device to withstand the rated load with an appropriate factor of safety; 
(iii) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized edition of ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must demonstrate the ability of the enclosures to prevent the user from falling from the device; and 
(iv) Appropriate analysis and nonclinical testing (such as that outlined in the currently FDA-recognized editions of AAMI/ANSI/IEC 60601-1-2, “Medical Electrical Equipment—Part 1-2: General Requirements for Safety—Collateral Standard: Electromagnetic Compatibility—Requirements and Tests,” and ASME A18.1 “Safety Standard for Platform Lifts and Stairway Chair Lifts”) must validate electromagnetic compatibility and electrical safety.

## Predicate Devices

- Stair-Trac ([K981518](/device/K981518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 3. Summary

APR = 4 2002

Portable Wheelchair Lift

K020933
page 1 of 3

Common Device Name:

Trade Name:

Super-Trac TRE-52

Predicate Device:

Stair-Trac K981518

Device Description:

The Super-Trac TRE-52 is a portable wheelchair lift designed to carry a passenger in a wheelchair up and down stairs. It is a mobile, attendant-operated device that will accommodate most child and adult wheelchairs, including power wheelchairs and sports wheelchairs.

The Super-Trac TRE-52 consists of two endless durable belt tracks driven by two 12 volt DC motors. Two 12 volt 24 AH rechargeable batteries are used to power the motors. The device includes a set of auxiliary wheels that assist in maneuverability on landings.

The wheelchair fits onto a platform that is mounted to the drive frame. Loading is by means of an integrated ramp that slides out from below the platform. The wheelchair is secured to the platform using four tie-down straps and a seatbelt that goes around both the passenger and wheelchair. The platform is tilted to the stair-travel position by a hydraulic actuator located on the drive frame.

Super-Trac is equipped with a key switch to restrict use to authorized persons. A stair-slope indicator is provided to determine the stair angle is within the operating range.

The attendant operates Super-Trac TRE-52 from controls located on the operating handle. Super-Trac TRE-52 is designed for use on both indoor and outdoor stairways.

See attached brochure.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Garaventa Stair-Trac, a portable wheelchair lift. The logo is in a bold, shadowed font. The image also contains handwritten text at the top right corner, which appears to be a document number and page number, specifically KONO933 and page 2 of 3.

Image /page/1/Picture/2 description: In the image, two people are on a stair lift. The person in the front is operating the stair lift, and the other person is standing next to the stairs. The stair lift is designed to help people with disabilities to go up and down stairs. There is a sign on the bottom left of the image that indicates that the area is accessible to people with disabilities.

The Garaventa Stair-Trac is an easy to use, portable wheelchair lift designed to attach under most manual wheelchairs. Stair-Trac's powerful motor allows a small attendant to safely transport a larger passenger up and down stair-Trac is suitable for indoor or outdoor use in public buildings and private homes. An affordable and reliable portable access solution.

Image /page/1/Picture/4 description: The image is a logo for Garaventa Accessibility. The logo is black and white and features the word "GARAVENTA" in a bold, sans-serif font. Above the word is a semi-circular shape with lines radiating outward. Below the word "GARAVENTA" is the word "ACCESSIBILITY" in a smaller font. At the bottom of the logo is the website address "www.garaventa.ca".

7(4)

{2}------------------------------------------------

## Specifications

Load Capacity ... ... 130kg 285lbs Power Capacity . . . . . . . . . 650 stairs Maximum Speed

Up Stairs Down Stairs - 12.5m (41.0ft)/mir 12 VDC 12. ........ 12 VDC Motor Batteries . . . . . . . . . 2 x 6V - 20AH Max Charge Rate .... ............. 4 amp Optimum Stair Angle ....

# Dimensions

Height 935mm 36.8ir Depth 1473mm 58.0ir Width 635mm 25 Oir

# Weight

| Drive Frame    | 39kg/86lb  |
|----------------|------------|
| Support Handle | 15kg/33lb  |
| Total Weight   | 54kg/119lb |

# Clearances

Wheelchair loading/unloading requires a minimum straight line clearance of approximately 3m/10f The diagram below shows typical clearances required on turnback stairways Dimensions vary with stairway width and operator expenence. Consult Garaventa or your local representative for oparance versication

Image /page/2/Figure/9 description: The image shows a technical drawing with dimensions. The top part of the drawing indicates a length of 2250mm or 88.6in and a height of 1300mm or 51.18in. The bottom part of the drawing shows two sections, each with a length of 915mm or 36.0in.

## Garaventa Accessibility

PO Brx 1769 B The WA 98231-1760

505 134A Street Surrey, BC Canada V3W 7B3 -604) 594-0422 To Free -- 800-663-6556 - 6041 594-9915 ww.garaventa.ca

FEV (1 23 /

Image /page/2/Picture/15 description: In the image, a person is standing next to a wheelchair. The person appears to be interacting with the wheelchair, possibly assisting someone or adjusting it. The scene is set outdoors, with steps and a platform visible in the background, suggesting a public space or building entrance.

Attach the Stair-Trac to the wheelchair, recline the passenger and move them to the stairway.

Image /page/2/Picture/17 description: In the image, a person is pushing another person in a wheelchair up a set of stairs. The person in the wheelchair is facing forward, and the person pushing the wheelchair is behind them. The stairs are made of concrete, and there is a building in the background. The image is in black and white.

Ascend or descend stairs with the press of a button. The passenger sits in a comfortable, upright position.

Image /page/2/Picture/19 description: In the image, a person is loading an object into the back of a car. The car's trunk is open, and the person is standing behind the car. The person is wearing a hat and a jacket. The object being loaded appears to be a piece of equipment with wheels. The background is blurry, but it appears to be a street or parking lot.

The Stair-Trac is easy to load into vehicles using the optional loading ramp, and fits easily into most car trunks.

cd0933 page 3 of 3

### Features

- · Attaches to most standard wheelchairs with push handles (adult/child)
- Powered to carry over 250 lbs. up approximately 30 floors
- · Use on virtually any indoor or outdoor stairway
- · Stable and self-supporting
- · Durable tracks grip stairs securely
- · Detachable support handle for compact storage and easy transport
- · Quick-release safety belt
- · Key switch restricts use to authorized persons
- · Optional auxiliary seat for non-wheelchair passengers
- · Rechargeable batteries and charger

## Benefits

- · Economical compared to stairway lifts and elevators
- Can be used on various stairways within a building
- · Can be used when electrical pov fails or is not available
- Easily stores in a secure area when not in use
- · Requires minimal maintenance
- · Available immediately

## What our customers say

"The Garaventa Stair-Trac has allower our District to meet its needs at a fraction of the cost of installing an elevator in our building."

- Batavia City Schools, Batavia, N\

"It has proven to be dependable, safe, and functional for our school district." - Hillcrest Elementary, American Falls, II

## Coming soon...The Super-Trac

The portable wheelchair lift for custom and power wheelchairs. Available April, 2002

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem that resembles a bird or a symbol of human services, with flowing lines and curves.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR - 4 2002

Mr. Norm Cooper Director of Sales & Marketing Garaventa Accessibility P.O. Box 1769 Blaine, Washington 98231-1769

Re: K020933

Trade/Device Name: Garaventa Super-Trac TRE-52 Portable Wheelchair Lift Regulation Number: 890.3930 Regulation Name: Wheelchair Elevator Regulatory Class: Class II Product Code: ING Dated: March 20, 2002 Received: March 22, 2002

Dear Mr. Cooper:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may , arores or visions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Norm Cooper

This letter will allow you to begin marketing your device as described in your Section 510(k) rms lotter will and will would be finding of substantial equivalence of your device to a legally premaired notification - The sification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of Compliance at (30) 57 contact the Office of Compliance at (301) 594-4639. Also, please note the your actived, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

L. Mark N. Millhouser

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Statement of Indications For Use

K020933 510(k) Number:

Device Name:

Super-Trac TRE-52

Portable Wheelchair Lift

#### Indications for Use:

The Super-Trac TRE-52 is intended to mechanically transport a single individual in a wheelchair up and down stairs in a private or public facility. The Super-Trac must be operated by a trained attendant.

(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

4144444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444444

Concurrence of CDRH, Office of Device Evaluation (ODE)

for Mark n Milken

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number.

Prescription Use
(Per 21 C.F.R Section 801.109)

OR

Over-the-Counter Use __

---

**Source:** [https://fda.innolitics.com/device/K020933](https://fda.innolitics.com/device/K020933)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
