SCORPIO INSET PATELLAR COMPONENT
K020830 · Howmedica Osteonics Corp. · JWH · Jun 6, 2002 · Orthopedic
Device Facts
| Record ID | K020830 |
| Device Name | SCORPIO INSET PATELLAR COMPONENT |
| Applicant | Howmedica Osteonics Corp. |
| Product Code | JWH · Orthopedic |
| Decision Date | Jun 6, 2002 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3560 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Scorpio® Inset Patellar Component is intended to articulate with any commercially available Scorpio® Total Knee Femoral Component, ,any commercially available Osteonics® Series 7000 Total Knee Femoral component, and any commercially available Osteonics® Omnifit® Total Knee Femoral Component. The subject patellar components are single use devices, intended for cemented applications on the surgically prepared posterior patella as part of primary or revision total knee arthroplasty. This component replaces the patellar articulating surface of the knee joint to simulate the normal function of the knee.
Device Story
Scorpio® Inset Patellar Component is a single-use orthopedic implant; fabricated from UHMWPE; designed for cemented application to surgically prepared posterior patella. Device replaces patellar articulating surface to simulate normal knee function. Used in primary or revision total knee arthroplasty; implanted by orthopedic surgeons in clinical/OR settings. Component features central peg and anterior pocket/undercuts for cement interdigitation. Available in 10mm thickness and four diameters (26-32mm). Provides mechanical articulation interface for compatible femoral components.
Clinical Evidence
Bench testing only. No clinical data presented.
Technological Characteristics
Material: Ultra-high molecular weight polyethylene (UHMWPE). Form factor: Circular button with central peg; 10mm thickness; 26mm, 28mm, 30mm, 32mm diameters. Fixation: Cemented. Application: Primary or revision total knee arthroplasty.
Indications for Use
Indicated for patients with painful, disabling knee joint disease due to degenerative, rheumatoid, or post-traumatic arthritis; post-traumatic loss of knee joint configuration/function; moderate varus, valgus, or flexion deformity with stable ligamentous structures; or revision of unsuccessful knee replacement. Contraindicated in patients with active/latent knee infection; mental/neuromuscular disorders risking instability or fixation failure; compromised bone stock; skeletal immaturity; severe knee instability due to collateral ligament absence; or obesity.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
- Osteonics® Recessed Patellar Component (K863668)
- Duracon® Recessed Patellar Component (K951655)
Reference Devices
- Scorpio® PS Femoral Component (K962152)
- Scorpio® CR Femoral Component (K974556)
- Scorpio TS Femoral Component (K994128)
- Osteonics® Series 7000 Total Knee Femoral component (K912147, K914565, K925372, K941128, K944285)
- Osteonics® Omnifit® Total Knee Femoral Component (K862837, K863668, K884410, K910989)
Related Devices
- K972967 — OSTEONICS SCORPIO TOTAL KNEE UNIVERSAL DOME PATELLAR COMPONENT;OSTEONICS SCORPIO TOTAL KNEE RECESSED PATELLAR COMPONENT · Osteonics Corp. · Jan 16, 1998
- K971652 — P.F.C. SIGMA INSET PATELLA · Johnson & Johnson Professionals, Inc. · Jul 16, 1997
- K960535 — APOLLO TOTAL KNEE SYSTEM SINGLE LUG PATELLA · Intermedics Orthopedics · Apr 23, 1996
- K080029 — MAKO SURGICAL CORP. PATELLOFEMORAL KNEE IMPLANT SYSTEM · Mako Surgical Corp. · May 16, 2008
- K061340 — 23 MM SINGLE-PEG PATELLA COMPONENT · Biomet Manufacturing, Inc. · Sep 25, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
6 2002 Jun
K020830
page 1 of 3
# 510(K) SUMMARY
#### Scorpio® Inset Patellar Component
The Scorpio® Inset Patellar Component is intended to articulate with any commercially available Scorpio® Total Knee Femoral Component (Scorpio® PS Femoral Component premarket notification K962152; Scorpio® CR Femoral Component - premarket notification K974556, and Scorpio TS Femoral Component - premarket notification K994128 ), any commercially available Osteonics® Series 7000 Total Knee Femoral component ( premarket notifications K912147, K914565, K925372, K941128, K944285) and any commercially available Osteonics® Omnifit® Total Knee Femoral Component (premarket notifications K862837, K863668, K884410, and K910989). The subject patellar components are single use devices, intended for cemented applications on the surgically prepared posterior patella as part of primary or revision cemented total knee arthroplasty. This component replaces the patellar articulating surface of the knee joint to simulate the normal function of the knee.
The indications and contraindications for the use of the Scorpio® Inset Patellar Component in conjunction with a total knee replacement, in keeping with those of other legally marketed Class II patellar components, are as follows:
### Indications:
- . Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
- . Post-traumatic loss of knee joint configuration and function.
- . Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- . Revision of previous unsuccessful knee replacement or other procedure.
{1}------------------------------------------------
K020830
page 2 of 3
### Contraindications:
The contraindications for the subject devices include:
- . Any active or suspected latent infection in or about the knee joint.
- . Any mental or neuromuscular disorder which would create an unacceptable risk of prosthesis instability, prosthesis fixation failure or complications in postoperative care.
- . Bone stock compromised by disease, infection or prior implantation which cannot provide adequate support and/or fixation to the prosthesis.
- Skeletal immaturity.
- Severe instability of the knee joint secondary to the absence of collateral ligament . integrity and function.
- . Obesity. An overweight or obese patient can produce loads on the prosthesis which can lead to failure of the fixation of the device or to failure of the device itself.
## Device Description
The Scorpio® Inset Patellar Component is fabricated from ultra-high molecular weight polyethylene (UHMWPE). The component is available in one thickness (10mm) and four diameters (26mm, 28mm, 30mm, and 32mm). The overall geometry essentially consists of a circular patellar button configuration with a central peg on the anterior surface of the component. The anterior surface also features a pocket and undercuts for cement interdigitation. The subject Scorpio® Inset Patellar Component is mounted to the surface of the surgically prepared patella within a recessed hole when cementing the component into place.
The articular (posterior) surface of this component is identical to the articular surface of the Osteonics® Recessed Patellar Component, found substantially equivalent in premarket
{2}------------------------------------------------
510(k) Summary Scorpio® Inset Patellar Component
Kodo830
page 3 of 3
notification K863668, and cleared for use with Scorpio femoral components in K962152. The anterior (bone fixation) surface is similar to that of the Duracon® Recessed Patellar Component (#K951655). The Duracon® Recessed Patellar Component has been cleared for marketing in two thicknesses (8mm and 10mm), and five diameters (25mm, 27mm, 29mm, 31mm, and 35mm).
Testing was presented to support the claim of substantial equivalence.
For further information contact: Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401 (201) 934-4359 (Telephone) (201) 760-8435 (Fax)
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a stylized eagle-like emblem. The emblem is composed of three curved lines that form the body and wings of the bird, with a simple head and beak on the left side.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
6 2002 JUN
Ms. Margaret F. Crowe Regulatory Affairs Consultant Howmedica Osteonics Corp. 59 Route17 Allendale, New Jersey 07041
Re: K020830
Trade/Device Name: Scorpio® Inset Patellar Component Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: March 13, 2002 Received: March 14, 2002
Dear Ms. Crowe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Margaret F. Crowe
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Melbenzer
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE
(020830
510(k) Number (if known):
page 1 of
Device Name: Scorpio® Inset Patellar Component
Indications for Use:
The Scorpio® Inset Patellar Component is intended to articulate with any commercially available Scorpio® Total Knee Femoral Component, ,any commercially available Osteonics® Series 7000 Total Knee Femoral component, and any commercially available Osteonics® Omnifit® Total Knee Femoral Component. The subject patellar components are single use devices, intended for cemented applications on the surgically prepared posterior patella as part of primary or revision total knee arthroplasty. This component replaces the patellar articulating surface of the knee joint to simulate the normal function of the knee.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR (per 21 CFR 801.109)
Over-the-Counter Use
(Optional Format 1-2-96)
fo Mark N. Mulkerr
Division lign-Off) 1 General, Restorative Divisto: and Neurological Devices KO20830
510(k) Number -