BACT/ALERT FN CULTURE BOTTLE

K020815 · bioMerieux, Inc. · MDB · Apr 18, 2002 · Microbiology

Device Facts

Record IDK020815
Device NameBACT/ALERT FN CULTURE BOTTLE
ApplicantbioMerieux, Inc.
Product CodeMDB · Microbiology
Decision DateApr 18, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2560
Device ClassClass 1

Indications for Use

BacT/ALERT FN Culture Bottles are used with the BacT/ALERT Microbial Detection Systems in qualitative procedures for enhanced recovery and detection of anaerobic microorganisms from blood and other normally sterile body fluids.

Device Story

BacT/ALERT FN Plastic Culture Bottle is a container for blood or sterile body fluid samples; designed for use with BacT/ALERT Microbial Detection Systems. Device provides nutritional/environmental conditions for anaerobic microorganism growth. Principle of operation: reflectance-based monitoring of CO2 production via a Xylenol Blue indicator in a silicone emulsion sensor. Inoculated bottles are incubated and continuously monitored by the instrument. Output is qualitative detection of microbial presence. Used in clinical laboratory settings by trained personnel. Benefits include enhanced recovery and detection of anaerobic pathogens to aid clinical diagnosis of blood infections.

Clinical Evidence

Bench testing only. Seeded studies performed using 9 microorganisms diluted in human blood, comparing the plastic bottle to the glass predicate. Results demonstrated equivalent recovery of low levels of microorganisms and equivalent detection times.

Technological Characteristics

Plastic culture bottle for anaerobic microbial growth. Detection via reflectance monitoring of CO2-induced color change in a Xylenol Blue/silicone emulsion sensor. Compatible with BacT/ALERT Microbial Detection Systems. Class I device.

Indications for Use

Indicated for adult patients requiring qualitative detection of anaerobic microorganisms in blood and normally sterile body fluids.

Regulatory Classification

Identification

A microbial growth monitor is a device intended for medical purposes that measures the concentration of bacteria suspended in a liquid medium by measuring changes in light scattering properties, optical density, electrical impedance, or by making direct bacterial counts. The device aids in the diagnosis of disease caused by pathogenic microorganisms.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K020815 # 510(k) Summary #### The submitter's name, address, telephone number, a contact person, and the date the (a)(1) summary was prepared; Submitter's Name: bioMérieux, Inc. Submitter's Address: 100 Rodolphe Street. Durham, North Carolina 27712 (919) 620-2373 Submitter's Telephone: Submitter's Contact: Ron Sanyal Ron Sa Date 510(k) Summary Prepared: March 8, 2002 - (a)(2) The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known; Trade or Proprietary Name: BacT/ALERT FN Culture Bottle Common or Usual Name: BacT/ALERT FN Culture Bottle Classification Name: Microbial Growth Monitor - (a)(3) An identification of the legally marketed device to which the submitter claims substantial equivalence; Device Equivalent to: BacT/Alert FN Glass Culture Bottle - A description of the device. (a)(4) Device Description: The BacT/ALERT FN Plastic Culture Bottle was developed for the same intended use as the current BacT/ALERT FN Glass Culture Bottle, to provide suitable nutritional and environmental conditions for organisms commonly encountered in blood infections and normally sterile body fluids. An inoculated bottle is placed into the BacT/ALERT Microbial Detection Instruments where it is incubated and continuously monitored for the presence of microorganisms that will grow in the BacT/ALERT FN Bottle. #### A statement of the intended use of the device. (a)(5) Device Intended Use: BacT/ALERT FN Culture Bottles are used with the BacT/ALERT Microbial Detection Systems in qualitative procedures for enhanced recovery and detection of anaerobic microorganisms from blood and other normally sterile body fluids. {2}------------------------------------------------ #### A summary of the technological characteristics of the new device in comparison to those of (a)(6) the predicate device. The BacT/ALERT FN Plastic Culture Bottle utilizes the same detection technology as the BacT/ALERT FN Glass Culture Bottle. The similarities and/or differences with marketed device are listed in Table (a) (6) 1. | FEATURES | BACT/ALERT FN PLASTIC<br>CULTURE BOTTLE | BACT/ALERT FN GLASS<br>CULTURE BOTTLE<br>(K992432) | |-----------------------------------------|-----------------------------------------|----------------------------------------------------| | Intended Use | Same | Same | | Culture Bottle Material | Plastic | Glass | | Product Code | MDB | MDB | | Technology | Reflectance | Reflectance | | Color change based on<br>CO2 production | YES | YES | | Sensor | Emulsion | Emulsion | | Indicator material | Xylenol Blue in Silicone Emulsion | Xylenol Blue in Silicone Emulsion | | Growth of<br>microorganisms | Same | Same | | Instrument Used | BacT/ALERT Microbial Detection Systems | BacT/ALERT Microbial<br>Detection Systems | | Sample Source | Blood, Body Fluids | Blood, Body Fluids | | Target Population | Adult | Adult | ### TABLE (a)(6).1 {3}------------------------------------------------ #### A brief discussion of the nonclinical tests submitted, referenced, or relied on in the (b)1) premarket notification submission for a determination of substantial equivalency. Testing was performed to establish the performance characteristics of the new device including: Seeded studies were performed on 9 organisms diluted in human blood and inoculated into the BacT/ALERT FN Plastic Culture bottle and the BacT/ALERT FN Glass Culture bottle. #### The conclusions drawn from the nonclinical and clinical tests that demonstrate that the (b)3) device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3). The BacT/ALERT FN Plastic Culture Bottle was substantially equivalent to the BacT/ALERT FN Glass Culture Bottle based on recovery of low levels of the 9 microorganisms included in the study. Detection times were equivalent in both bottles. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with its head turned to the left. The eagle is surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Food and Druq Administration 2098 Gaither Road Rockville MD 20850 Ron Sanyal, M. Pharm., CQE, RAC Manager, Regulatory Affairs Biomerieux, Inc. 100 Rodolphe Street Durham, NC 27712 k020815 Re: Trade/Device Name: BacT/ALERT® FN Culture Bottle Regulation Number: 21 CFR 866.2560 Regulation Name: Microbial Growth Monitor Regulatory Class: Class I Product Code: MDB Dated: March 8, 2002 Received: March 13, 2002 Dear Mr. Sanyal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. APR 1 8 2002 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ ### Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/odrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Page 1 of 1. | | | 510(k) Number (If known):_____________________________________________________________________________________________________________________________________________________ | KOZO815<br>Comments of the consisted to the consisted to the consideration of the control of the contribution of the contribution of the contribution of the contribution of the contribu | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |--|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | | | CALL AND | BacT/ALERT® FN Culture Bottle Device Name: Indications For Usa: BacT/ALERT® FN Culture Bottles are used with the BacT/ALERT Microbial Detection Systems in qualitative procedures for enhanced recovery and detection of anaerobic microorganisms from blood, and other normally sterile body fluids. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concumence of CDRH, Office of Device Evaluation (ODE) | | <div style="text-align:left;"></div> | |-----------------------------------------|--------------------------------------| | (Division Sign-Off) | | | Division of Clinical Laboratory Devices | | | 510(k) Number | KD2 0815 | | Prescription Use | <div> <span style="text-decoration: underline;">✓</span> </div> | OR | Over-The-Counter Use | |----------------------|-----------------------------------------------------------------|----|--------------------------| | (Per 21 CFR 801.109) | | | | | | | | (Optional Format 1-2-96) |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...