SINGLE ANALYTE URINE DRUGS OF ABUSE CALIBRATORS AND CONTROLS
Applicant
Lin-Zhi International, Inc.
Product Code
DLJ · Clinical Toxicology
Decision Date
Apr 3, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3200
Device Class
Class 2
Indications for Use
The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine. The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.
Device Story
Device consists of human urine-based liquid calibrators and controls; used for calibration and validation of enzyme immunoassays (EIA) for drugs of abuse (phencyclidine, opiates, cocaine metabolite, amphetamines). Negative calibrator is drug-free human urine; other levels prepared by spiking known drug concentrations. Concentrations confirmed via GC/MS. Used in clinical laboratory settings by laboratory professionals. Output provides reference points for assay calibration and quality control validation; ensures accuracy of patient urine drug screening results. Benefits include standardized, reliable detection of drug analytes in clinical toxicology testing.
Clinical Evidence
No clinical data. Performance characteristics (precision, accuracy, stability) established via bench testing. Nominal concentrations confirmed by GC/MS.
Technological Characteristics
Human urine-based liquid matrix; ready-to-use. Analyte concentrations confirmed by GC/MS. Storage at 2°C to 8°C. No unique technical characteristics; standard chemical calibrator/control formulation.
Indications for Use
Indicated for in vitro diagnostic calibration and validation of enzyme immunoassays detecting phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.
Regulatory Classification
Identification
A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Drugs of Abuse Urine Calibrators and Controls (K983159)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
APR 0 3 2002
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## Introduction
According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.
### Submitter name, Address, and Contact
Lin-Zhi International, Inc. 2391 Zanker Road. Suite 340 San Jose, CA 95131 Phone: (408) 944-0360 (408) 944-0359 Fax:
Contact: Chiu Chin Chang, Ph.D. VP, R&D
## Device Name and Classification
| (a) Classification Name: | Calibrators, Drug Specific;<br>Class II, DLJ (91 Toxicology), 21 CFR 862.3200 |
|--------------------------|---------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Phencyclidine Calibrators,<br>Opiate Calibrators,<br>Cocaine Calibrators, and<br>Amphetamine Calibrators |
| Proprietary Name: | None |
| (b) Classification Name: | Single (Specified) Analyte Controls (Assayed and Unassayed);<br>Class I, LAS (91 Toxicology), 21 CFR 862.3280 |
| Common/Usual Name: | Phencyclidine Controls,<br>Opiate Controls,<br>Cocaine Controls, and<br>Amphetamines Controls |
| Proprietary Name: | None |
#### Legally Marketed Predicate Device(s)
Lin-Zhi International, Inc.' Single Analyte Urine Drugs of Abuse Calibrators and Controls are substantially equivalent to the Drugs of Abuse Urine Calibrators and Controls (Diagnostic Reagents, Inc., now Microgenics Corporation), cleared under premarket notification K983159.
{1}------------------------------------------------
# Device Description
All of the Single Analyte Urine DAU Calibrators and Controls are human urine-based liquid, and ready to use. These Calibrators and Controls do not have any especially unique technical characteristics. When applicable, they contain known concentration of a specific drug analyte.
The Negative DAU Calibrator is a processed, drug-free human urine matrix. The Low, Cutoff, Intermediate, and High Calibrators, as well as the 2 levels of Controls are prepared by spiking known concentrations of drug analyte into the Negative DAU Calibrator matrix. The various concentrations of each drug analyte in their corresponding calibrators and controls are summarized as follows:
| | Phencyclidine EIA | Opiate EIA | Cocaine Metabolite EIA | Amphetamines EIA |
|-------------------------|-------------------|------------|------------------------|-------------------|
| Reference Material | Phencyclidine | Morphine | Benzoylecgonine | d-Methamphetamine |
| Low Calibrator | 12.5 ng/mL | 150 ng/mL | 150 ng/mL | 500 ng/mL |
| Cutoff Calibrator | 25 ng/mL | 300 ng/mL | 300 ng/mL | 1000 ng/mL |
| Intermediate Calibrator | 50 ng/mL | 600 ng/mL | 1000 ng/mL | 1500 ng/mL |
| High Calibrator | 100 ng/mL | 1000 ng/mL | 3000 ng/mL | 2000 ng/mL |
| Control Level 1 | 18 ng/mL | 225 ng/mL | 225 ng/mL | 750 ng/mL |
| Control Level 2 | 32 ng/mL | 375 ng/mL | 375 ng/mL | 1250 ng/mL |
The nominal concentrations of the analyte in the calibrators and controls are determined and confirmed by GC/MS.
# Intended Use
The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.
The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.
## Comparison to Predicate Device
LZI's Single Analyte Urine DAU Calibrators and Controls are similar in intended use, matrix, and performance to the DRI's Drugs of Abuse Urine Calibrators and Controls.
Similarities:
- Both are for the calibration and validation of DAU enzyme immunoassays to detect the . same commonly abused drug analytes (phencyclidine, opiates, cocaine metabolite, or amphetamines) in human urine.
{2}------------------------------------------------
- The cutoff concentration for each analyte is the same per recommendations of The . Substance Abuse and Mental Health Services Administration (SAMHSA).
- Both use 2 levels of Controls, and the concentrations are set at equal to or approximately . ± 25% of the cutoff concentration for each analyte according to the SAMHSA guideline.
- The nominal concentrations of the analyte in the calibrators and controls are determined . and confirmed by GC/MS.
- . Both are urine-based liquids.
- . Storage condition is the same, at 2℃ to 8℃.
- Performance characteristics on precision, accuracy and stability are similar. �
| Characteristics | DRI's DAU Urine<br>Calibrators and Controls | LZI's Single Analyte Urine<br>DAU Calibrators and Controls |
|--------------------------------------------------|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| No. of Analytes in each<br>Calibrator or Control | All 4 drugs in each Calibrator or Control | Single drug only in each<br>Calibrator or Control. |
| No. of Calibrators | 3 levels* including the<br>Negative Calibrator. | 5 levels including the Negative<br>Calibrator |
| Nomenclature/Labeling<br>of Calibrators | Negative, Low (= Cutoff),<br>and High | Negative, Low, Cutoff,<br>Intermediate, and High |
| Concentration of<br>Analyte(s) | Phencyclidine Controls: 20<br>and 35 ng/mL<br>Cocaine High Calibrator:<br>1000 ng/mL | Phencyclidine Controls: 18 and<br>32 ng/mL,<br>Cocaine High Calibrator: 3000<br>ng/mL |
Differences:
* Additional calibrators are now available. Currently 5 levels of calibrators (Cal 0, 1, 2, 3, and 4) are available from DRI/Microgenics Corp. (under "Multi-drug Urine Calibrators and Controls" product name).
# Conclusion
The information provided in the premarket notification demonstrates that the LZI's Single Analyte Urine Drugs of Abuse Calibrators and Controls are substantially equivalent to previously approved predicate devices, notably the DRI's DAU Urine Calibrators and Controls, and safe and effective for its intended use.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services U.S.A. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract shapes resembling human profiles or figures, stacked on top of each other.
#### Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Chiu Chin Chang, Ph.D. VP, R&D Lin-Zhi International, Inc. 2391 Zanker Road - Suite 340 San Jose, CA 95131
## APR 0 3 2002
k020769 Re:
Trade/Device Name: Single Analyte Urine Drugs of Abuse Calibrators and Controls Regulation Number: 21 CFR 862.3280; 21 CFR 862.3200 Regulation Name: Clinical toxicology control material; Clinical toxicology calibrator Regulatory Class: Class I; Class II Product Code: LAS; DLJ Dated: February 28, 2002 Received: March 7, 2002
Dear Dr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Premarket Notification
# Indications for Use Statement
# 510(k) Number (if known): K020769
Device Name: Single Analyte Urine Drugs of Abuse Calibrators and Controls
#### Indications for Use:
The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.
The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.
P. Bernhardt for J. Cooper
ision of Clinical Laboratory Devices 510(k) Number LO2 O
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.