SOF-PACH™ Reusable Cutaneous Neurostimulation Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. SOF-PACH™ Reusable Electrodes are designed and intended to be used with marketed. Electrical Stimulators i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) of PGF (Pulsed Galvanic Stimulation). These electrodes will include the precaution statement: Federal Law restricts the device to sale by or on the order of a licensed practitioner or therapist.
Device Story
SOF-PACH™ Reusable Neurostimulation Electrodes serve as a conductive interface between patient skin and electrical stimulators (TENS, EMS, IF, PGF). Device consists of a multi-layer flexible structure: top layer of fabric/foam with adhesive; middle layer of carbon/conductive mesh; bottom layer of biocompatible conductive hydrogel. Electrodes attach directly to skin via hydrogel adhesive; no external securing materials required. Connects to stimulation devices via lead wire assembly or male snap assembly. Used in clinical or home settings under practitioner guidance. Output is electrical stimulation delivered to the patient. Benefits include low-profile design for comfort and reduced visibility under clothing. Effectiveness verified by comparing point-to-point impedance levels (125-250 ohms at 1 kHz) against predicate devices.
Clinical Evidence
Bench testing only. Biocompatibility testing (Cytotoxicity, Sensitization, Primary Skin Irritation) performed per ISO 10993-1 and Tripartite Biocompatibility Guidance. Electrical performance validated via point-to-point impedance testing (1 kHz) comparing subject device to predicate devices; results (125-250 ohms) were comparable.
Technological Characteristics
Multi-layer flexible electrode. Materials: Tricot/polyester fabric, polyethylene foam, polypropylene substrate, biocompatible adhesive, carbon/Ag/AgCl conductive layer, conductive hydrogel. Connectivity: Lead wire (21 CFR 898 compliant) or male snap assembly. Dimensions: Low profile. Sterilization: Non-sterile. Energy: Passive conductive interface for TENS/EMS/IF/PGF stimulators.
Indications for Use
Indicated for use as a conductive adhesive interface between patient skin and electrical stimulators (TENS, EMS, IF, or PGF) for patients requiring neurostimulation therapy. Prescription use only.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
K020735
# ເງິນເ 12 2002
## SOF-PACH™ Reusable Neurostimulation Electrodes 510(k) Summary
The following information is submitted in accordance with the requirements of 21CFR807.92
| Submitter's name: | Anne Jen<br>CATHAY INTERNATIONAL LTD.<br>8F-3, NO. 29, CHUNGSHAN N.RD. SEC. 3,<br>TAIPEI, TAIWAN, R. O. C.<br>TEL: 02-2591-7938<br>FAX: 02-2591-7912<br>E-Mail: annepete@seed.net.tw |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Anne Jen |
| Date Prepared: | 17 January 2002 |
| Device Name: | SOF-PACH™ Reusable Neurostimulation Electrodes |
| Comment/Usual Name: | Neurostimulation Electrodes |
| Classification Name: | Electrodes, cutaneous |
| Predicate devices: | K963125 (M & R Manufacturing, Inc.)<br>K932849 (Pepin Mfg., Inc.)<br>K011411 (Pain Management Technologies) |
Device description: SOF-PACH™ Reusable Neurostimulation Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application:
> First layer - Various cloths, tapes, etc. including Tricot/polyester fabric, polyethylene foam or a polypropylene substrate, coated with biocompatible adhesive.
- Second layer Electrically conductive carbon coated with or without Ag/AgCl, and Activated Carbon Mesh material.
- Third layer Biocompatible conductive hydrogel coupling media (CATHAY's Gel) which has passed the required skin sensitivity testing criteria as specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization and Primary Skin Irritation Tests.
{1}------------------------------------------------
K020735
The electrodes are designed for single-patient/multiple application use. They have a low profile construction designed for comfort under clothing and utilize tan-tone fabric for reduced visibility. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin. The electrodes have two different types of contact points that can be used to link the stimulation device to the electrodes. These contact points are compatible with standard marketed Neurostimulation devices.
- 1. Lead wire assembly 6" wire with .080 inch diameter recessed female socket connected to one side of the wire. The lead wire assembly is in compliance with the requirements of FDA performance standard 21 CFR Part 898
- 2. Male snap assembly Two-part sensor and stud mated together.
Technological Characteristics: SOF-PACH™ Reusable Neurostimulation electrodes are technologically equivalent to the predicate devices. They are physically and technically similar to those currently being marketed for "Neurostimulation" i.e., TENS (Transcutaneous Electrical Nerve Stimulation), MENS (Microcurrent Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) and PGS (Pulsed Galvanic Stimulation).
Safety and Effectiveness: CATHAY considers their SOF-PACH™ Neurostimulation electrodes to be as safe and effective as the M&R Disposable Neurostimulation electrodes, the PMI Neurostimulation electrodes and the PMT Neurostimulation electrodes, which were previously found to be substantially equivalent via 510(k) Premarket Notifications. The safety issues analyzed were skin irritation. The effectiveness of the electrodes was determined by comparing impedance levels in point-to-point testing using a voltmeter.
The first safety issue considered was whether the gel, which is used to adhere the electrode to the skin and which is the only portion of the electrode to maintain skin contact, would cause any skin irritation. CATHAY's Gel has passed the required skin sensitivity testing criteria specified in the Tripartite Biocompatibility Guidance for Medical Devices and ISO 10993-1 requirements for skin contact. These tests include Cytotoxicity, Sensitization and Primary Skin Irritation tests.
Because there are no published performance standards for Neurostimulation electrodes, CATHAY uses impedance levels as the criteria for effectiveness testing. When measuring the electrodes' impedance, driven at 1 K-Hz, using a standard voltmeter, the point-to-point impedance testing for all manufacturers' electrodes (including SOF-PACH™) ranged from 125-250 ohms from the top center portion of the electrode to the underside of the outside edge. Once this testing was complete, it was determined that SOF-PACH™s electrode impedance values were comparable to other marketed electrodes tested.
For the above reason, CATHAY considers its Reusable TENS/NMES electrodes to be as safe and effective as the predicate devices, as well as to multiple other TENS/NMES electrodes currently being marketed, including those manufactured by:
Uni-Patch (K962910), Katecho Incorporated (K000870), MSB Iimited (K980229), Axelgaard (K983741), Selective Med Components, Inc. (K946230), Medtronic, Inc. (K875284), Medical Design & Manufacturing Corp. (K895604), and Labeltape Meditect, Inc. (K900656, K902719 and K894043).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest wings or feathers.
JUL 12 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mrs. Anne Jen Cathay International Ltd. 8F-3, No. 29, Chungshan N.Rd. Sec.3, Taipei, Taiwan, R.O.C.
Re: K020735
Trade/Device Name: SOF-PACH Reusable Neurostimulation Electrodes Models: CM4040, CM505, CM50100, and CM 75 Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: Class II Product Code: GXY Dated: June 11, 2002 Received: June 14, 2002
Dear Mrs. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic
{3}------------------------------------------------
Page 2 - Mrs. Anne Jen
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in vour Section 510(k) Page 2 premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Sincerely yours,
Arto Salin, M. Witte, PhD, M.D.
lia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K020735
# DEVICE NAME: SOF-PACH™ REUSABLE NEUROSTIMULATION ELECTRODES
## INDICATIONS FOR USE:
SOF-PACH™ Reusable Cutaneous Neurostimulation Electrodes are intended for use as the disposable, conductive adhesive interface between the patient's skin and the Electrical Stimulator. SOF-PACH™ Reusable Electrodes are designed and intended to be used with marketed. Electrical Stimulators i.e. TENS (Transcutaneous Electrical Nerve Stimulation), EMS (Electrical Muscular Stimulation), IF (Interferential) of PGF (Pulsed Galvanic Stimulation).
These electrodes will include the precaution statement: Federal Law restricts the device to sale by or on the order of a licensed practitioner or therapist.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription use X (Per 21 CFR 801, 109) OR:
Over-the Counter Use (Optional Format 1-2-96)
Stupt. Qurchu
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K020735
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.