K020364 · Phoenix Diagnostics, Inc. · JGS · Jun 4, 2002 · Clinical Chemistry
Device Facts
Record ID
K020364
Device Name
PHOENIX ISE REAGENTS FOR BECKMAN CX SYSTEMS
Applicant
Phoenix Diagnostics, Inc.
Product Code
JGS · Clinical Chemistry
Decision Date
Jun 4, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1665
Device Class
Class 2
Intended Use
The pHoenix ISE Reagents for Beckman CX Systems are intended for use as ISE Reagents for the determination of Na , K , C1 and C02 in human serum samples on the Beckman CX Systems. The Electrolyte Buffer is diluted with the Wash Solution and introduced into the serum samples for the quantitative determination of Na+, K+, Cl- and C02. The Electrolyte Reference Reagent is to provide reference points for Na+, K+, C1- and C02. The C02 Acid Reagent is to release C02 from serum samples. The C02 Alkaline Buffer is to provide a constant C02 concentration as reference for the C02 electrode. The Wash Concentrate is intended to be diluted with de-ionized water. The working Wash Solution is used to wash the systems sample probe and to dilute reagents on Beckman CX Systems. The Calibrators are used for the calibration of the CX ISE Modules for the quantitative determination of Na+, K+ Cl+and C02 in serum samples.
Device Story
pHoenix ISE Reagents are chemical reagents and calibrators designed as direct replacements for Beckman Coulter reagents used on Beckman CX Clinical Chemistry Systems. The system utilizes Ion Selective Electrode (ISE) technology to measure electrolytes in human serum. Reagents include buffers, reference reagents, acid/alkaline reagents for CO2, and wash concentrates. The device is used in clinical laboratory settings by trained laboratory personnel. The reagents facilitate the quantitative analysis of serum samples; the analyzer processes the electrochemical signals to produce concentration values for Na+, K+, Cl-, and CO2. These results assist clinicians in diagnosing and monitoring electrolyte and acid-base balance in patients. The reagents ensure system calibration and probe maintenance, supporting accurate diagnostic output.
Clinical Evidence
Bench testing only. Precision study (N=80 per analyte) followed NCCLS EP5-T guidelines, measuring 2 levels of serum controls over 20 days; results showed CV% ranging from 0.32% to 5.8%. Method comparison study (N=50 samples) compared pHoenix reagents against Beckman Coulter reagents; linear regression analysis yielded correlation coefficients (r) of 0.999 (Na+), 0.998 (K+), 0.996 (Cl-), and 0.985 (CO2).
Technological Characteristics
Chemical reagents and calibrators for Ion Selective Electrode (ISE) analysis. Includes electrolyte buffers, reference reagents, CO2 acid/alkaline reagents, and wash concentrates. Designed for use on Beckman CX Systems. No electronic or software components; purely chemical/reagent-based system.
Indications for Use
Indicated for the quantitative determination of sodium (Na+), potassium (K+), chloride (Cl-), and carbon dioxide (CO2) in human serum samples using Beckman CX Systems. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A sodium test system is a device intended to measure sodium in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance.
Predicate Devices
Beckman Coulter ISE Reagents for Beckman CX Systems
Related Devices
K052724 — TECO ISE REAGENT SET FOR CX SYSTEM · Teco Diagnostics · Dec 14, 2005
K020148 — pHoenix ISE Reagents FOR Olympus AU Chemistry Systems · Phoenix Diagnostics, Inc. · Feb 4, 2002
K033061 — CALIBRATING MATERIAL, ISE ELECTROLYTE REFERENCE · Diamond Diagnostics, Inc. · Nov 7, 2003
K040977 — GENCHEM ELECTROLYTE BUFFER · Genchem, Inc. · Dec 27, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
JC020364
### 510 (k) Summary
pHoenix ISE Reagents for Beckman CX Systems
Beckman Instruments was the original manufacturer of the Beckman CX Systems. Recently Beckman Instruments acquired Coulter and these Chemistry Systems continued to be referred to as Beckman CX Systems. Beckman Coulter is the name of the new company.
pHoenix Diagnostics, Inc. is submitting a 510 (k) notification for the following: (1) Electrolyte Buffer, (2) Electrolyte Reference Reagent and (3) C02 Acid Reagent, (4) C02 Alkaline Buffer, (5) Wash Concentrate and (6) ISE Calibrator Level 1,2 and 3. These ISE Reagents are intended for use on the ISE Module of the Beckman CX Systems for the quantitative determination of Na+, K+, C1- and C02 in human serum.
The Electrolyte Buffer is intended to provide constant ionic strength for measuring electrolytes in human serum samples. The Electrolyte Buffer is combined with the Wash Solution and introduced into the serum samples for the quantitative determination of Na+, K+, C1- and C02. The Electrolyte Reference Reagent is to provide reference points for Na+, K+, C1- and C02. The C02 Acid Reagent is to release C02 from serum samples, The C02 Alkaline Reagent is to provide a constant C02 concentration as reference for the C02 electrode. The Wash Concentrate is intended to be diluted with de-ionized water. The working Wash Solution is used to wash the systems sample probe and to dilute reagents on Beckman CX Systems. The Calibrators are used for the calibration of the CX ISE Modules.
pHoenix Diagnostics, Inc. is claiming substantial equivalence to predicate devices manufactured by Beckman Coulter, the original manufacturer of these reagents and Beckman CX Systems.
The products encompassed by this 510 (k) submission are Class I and Class II In Vitro Diagnostic Solutions manufactured by pHoenix Diagnostics, Inc., 8 Tech Circle, Natick, MA 01760. These pHoenix ISE Reagents are intended to serve as direct replacements to like named products manufactured by Beckman Coulter.
{1}------------------------------------------------
## 510 (k) Summary cont.
| pHoenix<br>Cat.# | Beckman<br>Cat. # | Description | Models | Class |
|------------------|-------------------|-------------------------------|----------------|-------|
| 47-2095 | 472095 | Electrolyte Buffer | All CX Systems | 1 |
| 45-0214 | 450214 | Electrolyte Reference Reagent | All CX Systems | 2 |
| 44-3320 | 443320 | C02 Alkaline Buffer | All CX Systems | 1 |
| 44-3330 | 443330 | C02 Acid Reagent | All CX Systems | 1 |
| 44-3335 | 443335 | Wash Concentrate | All CX Systems | 1 |
| 46-5908 | 465908 | ISE Calibrator Level 1 | All CX Systems | 2 |
| 46-5909 | 465909 | ISE Calibrator Level 2 | All CX Systems | 2 |
| 46-5910 | 465910 | ISE Calibrator Level 3 | All CX Systems | 2 |
Listed below are pHoenix products and their Beckman Coulter equivalents.
pHoenix uses a similar composition, description and packaging design as that used by Beckman Coulter in its products. pHoenix has shown performance equivalence of its products to Beckman instrument products in the following manner:
- . Through a method comparison where results obtained on Beckman CX Clinical Chemistry Systems calibrated with pHoenix products and compared with results obtained on the same analyzer calibrated with Beckman CX Systems products; and
- . Through a precision study where pHoenix products were installed on Beckman CX Systems Clinical Chemistry Systems and samples were measured one run per day for a 20 day period.
A summary of the results of these studies follows:
Precision data was collected from the analysis of 2 levels of serum controls measured 2 runs per day, 2 times per run for 20 days on Beckman CX Systems for Na , K , Cl and C02 installed with pHoenix reagents. The NCCLS Guideline for precision evaluation, EP5-T, was followed. Typical Results are as follows: Level 2
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|-------|-------|------|------|------|
| Na+ | Total | 80 | 121 | 1.47 | 1.2 | 120 | 124 |
| | Run to Run | 20 | 121 | 0.67 | 0.55 | 120 | 123 |
| K+ | Total | 80 | 1.93 | 0.104 | 5.42 | 1.8 | 2.1 |
| | Run to Run | 20 | 1.93 | 0.040 | 2.00 | 1.85 | 2.0 |
| Cl- | Total | 80 | 70.6 | 1.42 | 2.01 | 68 | 73 |
| | Run to Run | 20 | 70.6 | 0.87 | 1.23 | 69 | 72 |
| C02 | Total | 80 | 16.25 | 0.94 | 5.8 | 14.1 | 17.2 |
| | Run to Run | 20 | 16.25 | 0.57 | 3.57 | 14.7 | 17.2 |
{2}------------------------------------------------
### 510 (k) Summary cont.
| 1<br>T<br>0110<br>پ<br>ﺮ<br>CI<br>L | 1<br>t |
|-------------------------------------|--------|
|-------------------------------------|--------|
| Analyte | | N | Mean | STD | CV% | Min | Max |
|---------|------------|----|-------|-------|------|-----|-----|
| Na+ | Total | 80 | 164.2 | 2.58 | 1.57 | 160 | 169 |
| | Run to Run | 20 | 164.2 | 1.07 | 0.65 | 162 | 166 |
| K+ | Total | 80 | 6.50 | .062 | 0.96 | 6.4 | 6.6 |
| | Run to Run | 20 | 6.50 | .028 | 0.44 | 6.4 | 6.5 |
| Cl- | Total | 80 | 147.5 | 0.824 | 0.56 | 146 | 149 |
| | Run to Run | 20 | 147.5 | .473 | .32 | 147 | 148 |
| C02 | Total | 80 | 34.1 | 1.87 | 5.48 | 30 | 39 |
| | Run to Run | 20 | 34.1 | 1.28 | 3.76 | 32 | 37 |
Correlation with Beckman Coulter Reagents
Correlation data was collected from 50 samples (patient serum samples, control samples and spiked samples) for Na , K , CI and C02 measured on Beckman CX Systems installed with pHoenix reagents (Electrolyte Buffer, Electrolyte Reference Reagent, C02 Alkaline Buffer, C02 Acid Reagent, Wash Cincentrate, and Calibrators) as compared with Beckman Coulter reagents separately. A Linear Regression Analysis was performed using Beckman data as the independent X Variable and pHoenix Diagnostics Data as the Dependent Y Variable in the equation Y = a + bX. Typical results are as follows:
| Analyte | N | Slope | Intercept | Correlation<br>Coefficient | Range |
|---------|----|-------|-----------|----------------------------|-----------|
| Na+ | 50 | 1.027 | 0.60 | 0.999 | 100 - 200 |
| K+ | 50 | 1.04 | 0.348 | 0.998 | 1 - 15 |
| Cl- | 50 | 0.977 | 7.17 | 0.996 | 50 - 200 |
| C02 | 50 | 0.964 | 1,058 | 0.985 | 5 - 40 |
I hope you find this information useful and informative.
Ram Nunna, President
2/1/02
Date
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JUN 0 4 2002
Mr. Ram Nunna President pHoenix Diagnostics, Inc. 8 Tech Circle Natick, MA 01760
Re: k020364
> Trade/Device Name: pHoenix ISE Reagents for Beckman CX Systems 700/900 series Regulation Number: 21 CFR 862.1665; 21 CFR 862.1600; 21 CFR 862.1170; 21 CFR 862.1120; 21 CFR 862.1150 Regulation Name: Sodium test system; Potassium test system; Chloride test system; Blood gases (Poo2, Po2) and blood pH test system; Calibrator Regulatory Class: Class II Product Code: JGS: CEM; CGZ, CHL, JIX Dated: April 2, 2002 Received: April 8, 2002
Dear Mr. Nunna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and ' additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807,97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".
Sincerely yours.
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
### 510(k) Number (if known): K0 Z036
Device Name: pHoenix ISE Reagents for Beckman CX Systems 700/900 series
Indications For Use:
#### Intended Use:
The pHoenix ISE Reagents for Beckman CX Systems are intended for use as ISE Reagents for the determination of Na , K , C1 and C02 in human serum samples on the Beckman CX Systems.
The Electrolyte Buffer is diluted with the Wash Solution and introduced into the serum samples for the quantitative determination of Na+, K+, Cl- and C02.
The Electrolyte Reference Reagent is to provide reference points for Na+, K+, C1- and C02.
The C02 Acid Reagent is to release C02 from serum samples.
The C02 Alkaline Buffer is to provide a constant C02 concentration as reference for the C02 electrode. The Wash Concentrate is intended to be diluted with de-ionized water. The working Wash Solution is used to wash the systems sample probe and to dilute reagents on Beckman CX Systems.
The Calibrators are used for the calibration of the CX ISE Modules for the quantitative determination of Na+, K+ Cl+and C02 in serum samples.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
✓
_
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use · (Optional Format 1-2-96)
Dean
Coopy
(Division Sign-Off)
Division of Clinical Laboratory Devices
16. 26364
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.