← Product Code [DKZ](/productcode/DKZ) · K020313

# ACON ONE STEP MULTI-DRUG MULTI-LINE SCREEN TEST CARD; ACON ONE STEP MULTI-DRUG MULTI-LINE SCREEN TEST DEVICE (K020313)

_Acon Laboratories Co. · DKZ · May 8, 2002 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K020313

## Device Facts

- **Applicant:** Acon Laboratories Co.
- **Product Code:** [DKZ](/productcode/DKZ.md)
- **Decision Date:** May 8, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3100
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The ACON® One Step Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine samples. The designated cut-off concentrations for these drugs are as follows: Amphetamine at 1,000 ng/ml, Cocaine at 300 ng/ml, Methamphetamine at 1,000 ng/ml, Opiates at 2,000 ng/ml, Marijuana at 50 ng/ml and Phencyclidine at 25 ng/ml. They are intended for healthcare professionals including professionals at the point of care sites.

## Device Story

Rapid chromatographic immunoassay for qualitative, simultaneous detection of 2-6 drugs in human urine; utilizes competitive binding, lateral flow immunochromatographic principle; mouse monoclonal antibodies detect drugs at specific cut-offs (AMP 1000ng/ml, COC 300ng/ml, mAMP 1000ng/ml, OPI 2000ng/ml, THC 50ng/ml, PCP 25ng/ml). Used at point-of-care sites by healthcare professionals; no instrumentation required. Positive result: absence of colored line in test region; negative result: presence of colored line. Procedural control line confirms proper sample volume and membrane wicking. Provides preliminary analytical results; intended to aid clinical decision-making regarding drug screening.

## Clinical Evidence

Clinical evaluation of >1,000 urine specimens compared subject device results against predicate single-drug tests and GC/MS analysis. Results showed high agreement: positive agreement 90-99%, negative agreement 95-99%, and overall agreement 95-99% across all drug analytes. Study included samples with concentrations in the -25% to +25% cut-off range. Physician's office laboratory (POL) study confirmed suitability for point-of-care use.

## Technological Characteristics

Lateral flow immunochromatographic assay; competitive binding principle; mouse monoclonal antibodies; visual qualitative readout; no instrumentation; standalone test card/device format.

## Regulatory Identification

An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.

## Special Controls

*Classification.* Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- ACON AMP One Step Amphetamine Test Strip ([K011673](/device/K011673.md))
- ACON mAMP One Step Methamphetamine Test Strip ([K011672](/device/K011672.md))
- ACON COC One Step Cocaine Test Strip ([K010841](/device/K010841.md))
- ACON OPI One Step Opiates Test Strip ([K011353](/device/K011353.md))
- ACON THC One Step Marijuana Test Strip ([K003557](/device/K003557.md))
- ACON PCP One Step Phencyclidine Test Strip ([K011730](/device/K011730.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

#### 8. SUMMARY OF 510(k) SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

KO20313 The Assigned 510(k) number is

# Submitter:

ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121

Tel.: 858-535-2030 Fax: 858-535-2038

Date: January 28, 2002

Contact Person: Edward Tung, Ph.D.

Product Names:

ACON One Step Multi-Drug Multi-Line Screen Test Card

ACON One Step Multi-Drug Multi-Line Screen Test Device

# Common Name:

Immunochromatographic test for the simultaneously qualitative detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine.

# Device Classification:

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are similar to other FDA-cleared devices for the qualitative and simultaneous detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine specimens. These tests are used to provide a preliminary analytical result only. Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine test systems have been classified as Class II devices with moderate complexity.

# Classification Name:

Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine test system

{2}------------------------------------------------

#### Intended Use:

The ACON® One Step Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine samples. The designated cut-off concentrations for these drugs are as follows: Amphetamine at 1,000 ng/ml, Cocaine at 300 ng/ml, Methamphetamine at 1,000 ng/ml, Opiates at 2,000 ng/ml, Marijuana at 50 ng/ml and Phencyclidine at 25 ng/ml. They are intended for healthcare professionals including professionals at the point of care sites.

#### Description:

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative and simultaneous detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine samples. The test is based on the principle of antigen-antibody immunochemistry. It utilizes mouse monoclonal antibodies to selectively detect elevated levels of Amphetamine, Methamphetamine, Cocaine, Opiates, THC and PCP in urine at the cut-off concentrations of 1,000 ng/ml (AMP), 1,000 ng/ml (mAMP), 300 ng/mL (COC), 2,000 ng/ml (OPI), 50 ng/ml (THC) and 25 ng/ml (PCP). These tests can be performed without the use of an instrument.

A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine at the concentrations below the designated cut-off levels will generate a colored-line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

#### Predicate Devices:

ACON™ of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Drug Test strips were used as the predicate devices for the ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device to compare their performance with clinical urine specimens.

510(k) Number for these predicate devices are:

| ACON AMP One Step Amphetamine Test Strip      | K011673 |
|-----------------------------------------------|---------|
| ACON mAMP One Step Methamphetamine Test Strip | K011672 |
| ACON COC One Step Cocaine Test Strip          | K010841 |
| ACON OPI One Step Opiates Test Strip          | K011353 |
| ACON THC One Step Marijuana Test Strip        | K003557 |
| ACON PCP One Step Phencyclidine Test Strip    | K011730 |

{3}------------------------------------------------

# Comparison to a Predicate Device:

A comparison of the features of the ACON™ One Step Multi-Line Screen Test Card and Test Device versus the ACON™ Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Tests is shown below:

- Both tests are assays intended for the qualitative detection of Amphetamine, Cocaine, . Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine samples.
- Both tests are intended as a screening method that provides a preliminary analytical test result. .
- Both tests are immunochromatographic, lateral flow assays for the rapid detection of . Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine with a visual, qualitative end result, while ACON Multi-Drug Multi-Line Test detects 2 to 6 of the above drugs simultaneously.
- Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody ● interactions to indicate a positive or negative result.
- Both tests have the same cut-off for each drug test. .

# Safety and Effectiveness Data:

# Accuracy

A clinical evaluation was conducted using clinical urine specimens. This evaluation compared the test results between ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device versus previously FDA-cleared Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single tests; as well as against data obtained from the customary GC/MS analysis. Over 1,000 clinical specimens were employed including approximately 10% of the samples with drug concentrations in the -25% to +25% cut-off range. The comparisons of data obtained from this study yielded the following results:

{4}------------------------------------------------

# ACON One Step Multi-Drug Multi-Line Screen Test Card

vs.

| ACON Test Card | Positive Agreement           | Negative Agreement           | Overall Agreement            |
|----------------|------------------------------|------------------------------|------------------------------|
| AMP            | 129/129=>99%<br>(97% - 99%)* | 172/172=>99%<br>(98% - 99%)* | 301/301=>99%<br>(98% - 99%)* |
| COC            | 112/112=>99%<br>(97% - 99%)* | 186/187=99%<br>(97% - 99%)*  | 298/299=99%<br>(98% - 99%)*  |
| mAMP           | 121/122=99%<br>(96% - 99%)*  | 174/174=>99%<br>(98% - 99%)* | 295 / 296=99%<br>(98%-99%)   |
| OPI            | 131/133=98%<br>(95% - 99%)*  | 164/164=>99%<br>(98% - 99%)  | 295/297 =>99%<br>(97%-99%)   |
| THC            | 124/124=>99%<br>(97% - 99%)* | 175/176=99%<br>(97% - 99%)*  | 299/300=99%<br>(98%-99%)*    |
| PCP            | 71/72=99%<br>(93% - 99%)*    | 160/160=>99%<br>(98%-99%)*   | 231/232=99%<br>(97% - 99%)*  |

Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Tests:

* 95% Confidence Interval

#### ACON One Step Multi-Drug Multi-Line Screen Test Device vs.

Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Test:

| ACON Test Device | Positive Agreement            | Negative Agreement         | Overall Agreement             |
|------------------|-------------------------------|----------------------------|-------------------------------|
| AMP              | 129/129=>99%<br>(97%-99%)*    | 172/172=>99%<br>(98%-99%)  | 301/301=>99%<br>(98%-99%)*    |
| COC              | 112/112=>99%<br>(97%-99%)*    | 186/187=>99%<br>(97%-99%)* | 298/299=>99%<br>(98%-99%)*    |
| mAMP             | 121/121=>99%<br>(97% - 99%)*  | 175/175=>99%<br>(98%-99%)* | 296 / 296 =>99%<br>(98%-99%)* |
| OPI              | 132/133=>99%<br>(96% - 99%)*  | 164/164=>99%<br>(98%-99%)  | 296/297=>99%<br>(98%-99%)*    |
| THC              | 124/124 =>99%<br>(97% - 99%)* | 175/176=>99%<br>(98%-99%)* | 299 / 300=>99%<br>(98%-99%)*  |
| PCP              | 71/72 = 99%<br>(95% - 99%)*   | 160/160=>99%<br>(98%-99%)* | 231/232 = 99%<br>(97% - 99%)* |

* 95% Confidence Interval

{5}------------------------------------------------

Clinical study results of ACON One Step Multi-Line Screen Test Card and Test Device are compared to GC/MS analysis data:

The GC/MS cut-off levels for each of the six drugs tested are as following:

| Amphetamine     | 1,000 ng/ml |
|-----------------|-------------|
| Cocaine         | 300 ng/ml   |
| Methamphetamine | 1,000 ng/ml |
| Opiate          | 2,000 ng/ml |
| Marijuana       | 50 ng/ml    |
| Phencyclidine   | 25 ng/ml    |

Samples with drug concentration above the cut-off level were considered presumptive positive and concentrations below the cut-off are considered negative.

#### ACON One Step Multi-Drug Multi-Line Screen Test Card vs. GC/MS Analysis

| ACON Test Card | Positive Agreement            | Negative Agreement              | Overall Agreement               |
|----------------|-------------------------------|---------------------------------|---------------------------------|
| AMP            | 129/136=95%<br>(90% - 98%)*   | 164/165=99%<br>(97% - 99%)*     | 293/301 = 97%<br>(95% - 99%)*   |
| COC            | 112/118 = 95%<br>(89% - 98%)* | 180/181 = 99%<br>(97% - 99%)*   | 292/299 = 98%<br>(95% - 99%)*   |
| mAMP           | 121/134 = 90%<br>(84% - 94%)* | 162/162 = >99%<br>(98% - 100%)* | 283 / 296 = 96%<br>(93% - 98%)* |
| OPI            | 130/131 = 99%<br>(96% - 99%)* | 164/166 = 99%<br>(96% - 99%)*   | 294/297 = 99%<br>(97% - 99%)*   |
| THC            | 116/122 = 95%<br>(90% - 98%)* | 169/178 = 95%<br>(91% - 98%)*   | 285 / 300 = 95%<br>(92% - 97%)  |
| PCP            | 70/78 = 90%<br>(81% - 95%)*   | 153/154 = 99%<br>(96% - 99%)*   | 223/232 = 96%<br>(93% - 98%)*   |

* 95% Confidence Interval

{6}------------------------------------------------

| ACON Test Device | Positive Agreement            | Negative Agreement            | Overall Agreement               |
|------------------|-------------------------------|-------------------------------|---------------------------------|
| AMP              | 128/136=94%<br>(89% - 97%)    | 164/165=99%<br>(97% - 99%)    | 292/301 = 97%<br>(94% - 99%)*   |
| COC              | 112/118 = 95%<br>(89% - 98%)* | 180/181 = 99%<br>(97% - 99%)* | 292/299 = 98%<br>(95% - 99%)*   |
| mAMP             | 120/134 = 90%<br>(83% - 94%)* | 162/162 =>99%<br>(98% - 100%) | 282 / 296 = 95%<br>(92% - 97%)* |
| OPI              | 130/131 = 99%<br>(96% - 100%) | 164/166 = 99%<br>(96% - 99%)* | 294/297 = 99%<br>(97% - 99%)*   |
| THC              | 116/122 = 95%<br>(90% - 98%)* | 170/178 = 96%<br>(91% - 98%)* | 286 / 300 = 95%<br>(92% - 97%)* |
| PCP              | 70/78 = 90%<br>(81% - 95%)*   | 152/154 = 99%<br>(96% - 99%)* | 222/232 = 96%<br>(92% - 98%)*   |

#### ACON One Step Multi-Drug Multi-Line Screen Test Device vs. GC/MS Analysis

* 95% Confidence Interval

# Conclusion:

Clinical study results demonstrate the substantial equivalency between the ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device and the Amphetamine, Methamphetamine, Cocaine, Opiates, THC and PCP single tests, which has already being cleared by FDA and marketed in the United States. It is also demonstrated that these tests are safe and effective in detecting Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine at the following cut-off concentrations: Amphetamine 1,000 ng/ml, Cocaine 300 ng/ml, Methamphetamine 1,000 ng/ml, Opiates 2,000 ng/ml, THC 50 ng/ml and PCP 25 ng/ml. The physician's office laboratory POL study demonstrated that these tests are also suitable for use by professional at point-of-care site.

{7}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a snake winding around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Re:

MAY 0 8 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Edward Tung, Ph.D. Director of Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121

> k020313 Trade/Device Names: ACON® One Step Multi-Line Screen Test Card ACON® One Step Multi-Drug Multi-Line Screen Test Device Regulation Number: 21 CFR 862.3100; 21 CFR 862.3250; 21 CFR 862.3870; 21 CFR 862.3650; 21 CFR 862.3610 Regulation Name: Amphetamine test system; Cocaine and cocaine metabolite test System; Cannabinoid test system; Opiate test system; Methamphetamine test system Regulatory Class: Class II; Class II; Class II; Class II, Class II Product Code: DKZ; DIO; LDJ; DJG; LCM; LAF Dated: April 8, 2002 Received: April 12, 2002

Dear Dr. Tung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{8}------------------------------------------------

Page 2 - .

This letter will allow you to begin marketing your device as described in your 510(k) premarket I nis lotton . The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (Joase contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

#### INDICATIONS FOR USE 10.

510(k) Number:

K020313

ACON® One Step Multi-Drug Multi-Line Screen Test Card Device Name:

ACON® One Step Multi-Drug Multi-Line Screen Test Device

Indications for Use:

The ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine. The designated cut-off concentrations for these drugs are as follows: Amphetamine at 1,000 ng/ml, Cocaine at 300 ng/ml, Methamphetamine at 1,000 ng/ml, Opiates at 2,000 ng/ml, Marijuana at 50 ng/ml and Phencyclidine at 25 ng/ml. They are intended for healthcare professionals including professionals at the point of care sites.

Lean Cooper

(Division Sign-Off)
Division of Clinical L
510(k) Number. K020313

(Please do not write below this point) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Or

Over-The-Counter Use

---

**Source:** [https://fda.innolitics.com/device/K020313](https://fda.innolitics.com/device/K020313)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).