STAT PROFILE CRITICAL CARE XPRESS ANALYZER

K020297 · Nova Biomedical Corp. · CHL · Jun 10, 2002 · Clinical Chemistry

Device Facts

Record IDK020297
Device NameSTAT PROFILE CRITICAL CARE XPRESS ANALYZER
ApplicantNova Biomedical Corp.
Product CodeCHL · Clinical Chemistry
Decision DateJun 10, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1120
Device ClassClass 2

Intended Use

The Stat Profile Critical Care Xpress Analyzer is intended for in vitro diagnostic use by health care professionals for the quantitative determination of pH, pCO2, pO2, SO2%, Hct, Hb, Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac and Creat in heparinized whole blood, and Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac and Creat in serum and plasma.

Device Story

Stat Profile Critical Care Xpress Analyzer; point-of-care or clinical laboratory diagnostic device. Inputs: heparinized whole blood, serum, or plasma samples. Operation: electrochemical and biosensor-based measurement of pH, pCO2, pO2, SO2%, Hct, Hb, Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac, and Creat. Output: quantitative analyte concentrations displayed to clinician. Usage: healthcare professionals in clinical settings. Clinical impact: aids diagnosis/treatment of acid-base disturbances, electrolyte imbalances, renal conditions, diabetes, and metabolic disorders; facilitates monitoring of critically ill patients.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and performance validation of the analytical measurement system for the specified analytes.

Technological Characteristics

Electrochemical and biosensor-based analytical system. Measures pH, pCO2, pO2, SO2%, Hct, Hb, Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac, and Creat. Designed for in vitro diagnostic use. Benchtop form factor.

Indications for Use

Indicated for health care professionals to perform quantitative in vitro diagnostic analysis of blood gases, electrolytes, metabolites, and hematology parameters in heparinized whole blood, serum, and plasma for patients requiring assessment of acid-base status, electrolyte balance, renal function, and metabolic state.

Regulatory Classification

Identification

A blood gases (PCO2 , PO2 ) and blood pH test system is a device intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. Measurements of blood gases (PCO2 , PO2 ) and blood pH are used in the diagnosis and treatment of life-threatening acid-base disturbances.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of three human profiles facing right, with stylized lines representing hair or feathers above them. The overall design is simple and monochromatic. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 2098 Gaither Road Rockville MD 20850 ## JUN 1 0 2002 Mr. Paul W. MacDonald Director of Quality Assurance/Regulatory Affairs Nova Biomedical 200 Prospect Street Waltham. MA 02454-9141 k020297 Re: Trade/Device Name: Stat Profile Critical Care Xpress Analyzer Regulation Number: 21 CFR 862.1120 Regulation Name: Blood gases(P 002, Po2) and blood pH test system Regulatory Class: Class II Product Code: CHL Dated: April 12, 2002 Received: April 15, 2002 Dear Mr. MacDonald: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html". Sincerely yours. Steven Dutman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number: Device Name: Stat Profile Critical Care Xpress Analyzer Indications for Use: Intended Use The Stat Profile Critical Care Xpress Analyzer is intended for in vitro diagnostic use by health care professionals for the quantitative determination of pH, pCO2, pO2, SO2%, Hct, Hb, Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac and Creat in heparinized whole blood, and Na+, K+, Cl-, Ca++, Mg++, Glu, BUN, Lac and Creat in serum and plasma. #### Clinical Utility The following list includes the clinical utility information for each of the analytes measured on the CCX Analyzer. Blood Gases (PCO2, PO2 and pH): Whole blood measurement of blood gases is used inthe diagnosis, and treatment of life-threatening acid-base disturbances in critically ill patients with numerous metabolic and pulmonary diseases. Oxygen Saturation: Used to assess the oxygenation of hemoglobin and the adequacy of tissue oxygenation in the evaluation of pulmonary function. Also used in the diagnosis and treatment of cyanosis. Hematocrit: Whole blood measurement of hematocrit is used to estimate that red blood cells are present in sufficient quantity to carry oxygen and carbon dioxide. Hemoglobin: Oxygen is carried from the lungs throughout the body by hemoglobin presentin red blood cells. Measurement of hemoglobin provides the clinician with information regarding the evaluation of chronic and acute anemias and also with information pertaining to the potential oxygen transport capability of the hemoglobin. Sodium: Measurements are used in the diagnosis and treatment of aldosteronism, diabetes insipidus, adrenal hypertension, Addison's disease, dehydration, or diseases involving electrolyte imbalance. Potassium: Measurement of potassium is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions characterized by low or high potassium levels. Chloride: Measurement of chloride is used in the diagnosis and treatment of electrolyte {3}------------------------------------------------ and metabolic disorders such as cystic fibrosis and diabetic acidosis. Ionized Calcium: Used in the diagnosis and treatment of hypertension, renal disease, and vitamin D related disorders. Also useful in the diagnosis and treatment of patients with increased total protein and/or albumin levels, as in dehydration. Ionized Magnesium: Measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low levels of magnesium) and hypermagnesemia (abnormally high levels of magnesium). Creatinine: Measurement of creatinine is used in the diagnosis and treatment of certain renal conditions and is used for monitoring adequacy of dialysis, for example, peritoneal dialysis and peritoneal equilibration testing. Glucose: Measurement of glucose is used in the diagnosis and treatment of carbohvdrate metabolism disturbances including diabetes mellitus. neonatal hypoglycemia. and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. Lactate: Measurement of lactic acid (lactate) in whole blood, serum, and plasma is used to evaluate the acid-base status of patients suspected of having lactic acidosis. Urea Nitrogen: Measurement of urea nitrogen is used in the diagnosis and treatment of certain renal and metabolic diseases. Ref. 1.Tietz, N.W. ed. 1986. Textbook of Clinical Chemistry. W. B. Saunders Co. ### (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IFNEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Optional Format 3-10-98) Dean Cooper (Division Sign-Off) Division of Clinical Laboratory Devices 510(k) Number K020297 /b
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...