MODIFICATION TO ISOBAR SPINAL SYSTEM

K020245 · Scien'Tx USA, Inc. · MNH · Feb 1, 2002 · Orthopedic

Device Facts

Record IDK020245
Device NameMODIFICATION TO ISOBAR SPINAL SYSTEM
ApplicantScien'Tx USA, Inc.
Product CodeMNH · Orthopedic
Decision DateFeb 1, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ISOBAR Closed Pedicle Screw is to be used in conjunction with the ISOBAR Spinal System, a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). As a pedicle screw system, the ISOBAR Spinal System is intended for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebrae in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. When used as a posterior, non-cervical, non-pedicle screw fixation system, the ISOBAR Spinal System is intended for hook fixation from T1 to the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by history patient history and radiographic studies), deformities (scoliosis, kyphosis and lordosis), tumor, pseudoarthrosis, trauma (fracture or dislocation) and/or previous failed fusion surgery.

Device Story

ISOBAR Closed Pedicle Screw is a spinal fixation component used with the ISOBAR Spinal System. Device consists of pedicle screws with U-shaped heads, rods, nuts, and crosslink members. Input: surgical placement into spinal segments. Operation: set-screw secures rod into U-slot of screw to fix position. Output: rigid spinal stabilization. Used in clinical settings by surgeons for spinal fusion procedures. Benefits: immobilization of unstable spinal segments to facilitate fusion and treat deformities/trauma.

Clinical Evidence

Bench testing only. Static and fatigue compression testing performed according to ASTM F1717-96 standards to demonstrate mechanical equivalence to the predicate system.

Technological Characteristics

Materials: Titanium alloy (Ti-6Al-4V) conforming to ASTM F136. Design: Pedicle screw with U-shaped head and closed set-screw mechanism. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified. Software: None.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions including degenerative spondylolisthesis, fractures, dislocations, scoliosis, spinal tumors, pseudarthrosis, severe L5-S1 spondylolisthesis (Grade 3-4), degenerative disc disease, kyphosis, and lordosis.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020245 # FEB 012002 ## 510(k) Summary of Safety and Effectiveness | (1) Submitter's name: | Scient'x | |----------------------------------------|---------------------------------------| | Submitter's address: | Guyancourt, France | | Contact telephone number: | (512) 834-6255 | | Contact person: | Joanna Droege | | Date summary prepared: | January 22, 2002 | | (2) Trade or proprietary device name: | ISOBAR Closed Pedicle Screw | | Common or usual name: | Pedicle screw spinal system | | Classification name: | Class II | | (3) Legally marketed predicate device: | ISOBAR U-Line Spinal System (K990118) | #### (4) Subject device description: The ISOBAR Spinal System consists of pedicle screws, rods, nuts and crosslink members. It can be used for single or multiple level fixation. The U-line pedicle screws, rods, nuts and crosslink members were cleared on Single of manufactured from titanium alloy (Ti-6Al-4V) that conforms to ASTM F136. The additional component that is the subject of this Special 510(k) submission is a pedicle screw with a closed set-screw. It has a U shaped head to accept a rod in the same manner as the U-line screw. The difference is the set-screw used to secure the rod into place. As the set-screw is tightened the rod is secured in the U slot of the screw and is fixed in position. #### (5) Subject device intended use: The ISOBAR Closed Pedicle Screw is to be used in conjunction with the ISOBAR Spinal System, a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). As a pedicle screw system, the ISOBAR Spinal System is intended for the treatment of severe spondylolisthesis (Grade 3 and 4) of the LS-S1 vertebrae in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. When used as a posterior, non-cervical, non-pedicle screw fixation system, the ISOBAR Spinal System is intended for hook fixation from T1 to the illum/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by history patient history and radiographic studies), deformities (scoliosis, kyphosis and lordosis), tumor, pseudoarthrosis, trauma (fracture or dislocation) and/or previous failed fusion surgery. #### (6) Performance data: The Food and Drug Administration have established no performance standards applicable to pedicle screw spinal systems. However, static and fatigue compression testing was performed according to ASTM F1717-96. #### Basis for substantial equivalence: (7) The ISOBAR Closed Pedicle Screw is similar in design, materials and indications ISOBAR U-Line Screw (K990118). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a circular design with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 01 2002 Scient'X c/o Ms. Joanna Droege Regulatory/QA Manager Encore Orthopedics 9800 Metric Boulevard Austin, Texas 78758 Re: K020245 Trade Name: ISOBAR Closed Pedicle Screw Trade Name: 100Dr 112 CFR 888.3070, 21 CFR 888.3070, 21 CFR 888.3050 Regulation Name: Pedicle Screw Spinal System, Spondylolisthesis Spinal Fixation Device System, Spinal Interlaminal Fixation Orthosis, Regulatory Class: II Product Code: MNH, MNI, KWP Dated: January 22, 2002 Received: January 24, 2002 Dear Ms. Droege: We have reviewed your Section 510(k) notification of intent to market the device referenced we have have determined the device is substantially equivalent (for the indications for use above and we nave acterings and marketed in interstate commerce prior to May 28, 1976, the stated in the encrosure) to device Amendments, or to devices that have been reclassified in enaculient date of the Modical Doves Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, manact the device, casyone requirements for annual registration, listing of devices, Controlls provisions of the rist heading, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remarket Approval), it may of two code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good 675. A substaintially equirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspections; the rood and Dragon may result in regulatory action. In addition, FDA may publish Contiply with the GMI Yogatareing your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might {2}------------------------------------------------ Page 2 - Ms. Joanna Droege have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Sincerely, yours, Mark N. Melkerson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): _(020245 Device Name: __ ISOBAR Closed Pedicle Screw Indications For Use: # ISOBAR Closed Pedicle Screw ## Indications For Use The ISOBAR Closed Pedicle Screw is to be used in conjunction with the ISOBAR Spinal System, a The IDODIAN Closed Forceded to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with mistatine evidence of neurological impairment, facture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis). KO20245 As a pedicle screw system, the ISOBAR Spinal System is intended for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebrae in skeletally mature patients receiving fusion spondylonsucsis (Orado 9 and 1) or are sttached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. When used as a posterior, non-cervical, non-pedicle screw fixation system, the ISOBAR Spinal System within used as a postellor, non our roal, the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (as defined by back pain of discogenic origin with spondylonsdies), dogenerative and elbistory patient history and radiographic studies), deformities (scoliosis, kyphosis and lordosis), tumor, pseudoarthrosis, trauma (fracture or dislocation) and/or previous failed fusion surgery. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | <span style="text-decoration: underline;">✓</span> | |------------------|----------------------------------------------------| |------------------|----------------------------------------------------| | | OR (per 21 CFR 801.109) (Optional Format 1-2-96) | |--|--------------------------------------------------| |--|--------------------------------------------------| | | Over-The-Counter Use | |--|----------------------| |--|----------------------| (Division Sign-Off) Division of General Restorat and Neurological Devices | 510(k) Number | K020245 | |---------------|---------| |---------------|---------|
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...