PHOENIX ELETROLYTE CALIBRATION SET FOR THE ROCHE COBAS ISE MODULE

K020129 · Phoenix Diagnostics, Inc. · JIX · Feb 1, 2002 · Clinical Chemistry

Device Facts

Record IDK020129
Device NamePHOENIX ELETROLYTE CALIBRATION SET FOR THE ROCHE COBAS ISE MODULE
ApplicantPhoenix Diagnostics, Inc.
Product CodeJIX · Clinical Chemistry
Decision DateFeb 1, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The products encompassed by this request are intended for invitro diagnostic use and are intended for use in calibrating Nat, K+, Cli electrodes in Roche Cobas ISE module. Roche is the original equipment manufacturer for the analyzer and predicate reagents. phoenix electrolyte calibration set is intended for use in playe of like named products manufactured by Roche Diagnostics.

Device Story

pHoenix Electrolyte Calibration Set consists of 2-level liquid reagents; serves as direct replacement for Roche Diagnostics calibrators. Used in clinical laboratory settings on Roche Cobas ISE module; operated by laboratory technicians. Device provides known concentrations of Na+, K+, and Cl- to calibrate ion-selective electrodes. System measures patient serum samples; calibrator ensures accuracy of analyzer output. Healthcare providers use resulting electrolyte measurements for clinical decision-making regarding patient fluid and electrolyte status.

Clinical Evidence

Bench testing only. Precision study (NCCLS EP5-T) performed over 20 days (N=80 per level) on Roche Cobas ISE module; total CV% ranged 0.5%-5.0%. Method comparison study (N=50) compared patient serum results using pHoenix calibrators vs. Roche reagents; linear regression showed slopes 0.98-1.01 and correlation coefficients 0.98-0.99.

Technological Characteristics

In vitro diagnostic calibration reagents (2 levels). Ion-selective electrode (ISE) calibration principle. Liquid form factor. Composition similar to Roche OEM reagents. No specific materials or software algorithms described.

Indications for Use

Indicated for in vitro diagnostic use to calibrate Na+, K+, and Cl- electrodes on the Roche Cobas ISE module.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K020129 ## 510 (k) Summary pHoenix Electrolyte Calibration Set for the Roche® Cobas™ ISE Module The products encompassed by this 510 (k) submission are Class II (75 JIX) In Vitro Diagnostic Solutions manufactured by pHoenix Diagnostics, Inc., 8 Tech Circle, Natick, MA 01760. The Calibration Set consists of 2 levels and is intended for use in calibrating Na*, K* and CI on the Roche® Cobas™ ISE Module. Roche Diagnostics is the original equipment manufacturer (OEM) of the system. The pHoenix products stated are currently cleared under docket K902675. Information herein will support pHoenix's position to extend the intended use of these products to the Roche® Cobas™ ISE Module. The Roche® Cobas™ ISE Module measures Na*, K*, and CI by using Ion Selective Electrode Technology. The pHoenix Calibrator Reagents are intended to serve as direct replacements to like named products manufactured by Roche Diagnostics. pHoenix Product 17-200 and 17-201 are equivalent to Roche Product 44224. pHoenix uses a similar composition, description and packaging design as that used by Roche Diagnostics in its products. pHoenix has shown performance equivalence of its products to Roche Diagnostic products in the following manner: - Through a method comparison where results obtained on Roche® Cobas™ ISE . Module Chemistry Systems calibrated with pHoenix products and compared with results obtained on the same analyzer calibrated with Roche products; and - Through a precision study where pHoenix products were installed on Roche® . Cobas 114 ISE Module and samples were measured over multiple runs. A summary of the results of these studies follows: Precision data was collected from the analysis of 2 levels of serum controls measured 2 runs per day, 2 times per run for 20 days on a on Roche® Cobas™ ISE Module calibrated with pHoenix calibrator reagents. The NCCLS Guideline for precision evaluation, EP5-T, was followed. Typical results are as follows: | Analyte | | N | Mean | STD | CV% | Min | Max | |---------|------------|----|-------|-------|------|------|------| | Na+ | Total | 80 | 120.0 | 1.50 | 1.20 | 117 | 122 | | | Run to Run | 20 | 120.0 | 0.80 | 0.60 | 118 | 122 | | K+ | Total | 80 | 2.0 | 0.110 | 5.0 | 1.85 | 2.15 | | | Run to Run | 20 | 2.0 | 0.040 | 1.5 | 1.90 | 2.10 | | Cl- | Total | 80 | 100.0 | 0.60 | 0.60 | 97 | 102 | | | Run to Run | 20 | 100.0 | 0.25 | 0.30 | 98 | 101 | Level 2 {1}------------------------------------------------ ## 510 (k) Summary Cont. Level 4 | Analyte | | N | Mean | STD | CV% | Min | Max | |---------|------------|----|------|------|------|-----|-----| | Na+ | Total | 80 | 165 | 1.20 | 0.70 | 162 | 167 | | | Run to Run | 20 | 165 | 0.70 | 0.40 | 163 | 167 | | K+ | Total | 80 | 6.50 | 0.07 | 1.0 | 6.4 | 6.6 | | | Run to Run | 20 | 6.50 | 0.04 | 0.6 | 6.4 | 6.5 | | Cl- | Total | 80 | 146 | 0.80 | 0.50 | 144 | 148 | | | Run to Run | 20 | 146 | 0.60 | 0.40 | 125 | 148 | Accuracy by Correlation with Roche Standard Reagents Correlation data was collected from patient serum samples and control samples for Na", K* and CI measured on a Roche® Cobas™ ISE Module calibrated with pHoenix reagents as compared with Roche reagents separately. A Linear Regression Analysis was performed using pHoenix data as the independent X Variable and Roche Data as the Dependent Y Variable in the equation Y = a + bX. Typical results are as follows: | Analyte | N | Slope | Intercept | Correlation<br>Coefficient | Range | |---------|----|-------|-----------|----------------------------|-----------| | Na+ | 50 | 1.01 | 1.1 | 0.990 | 120 - 180 | | K+ | 50 | 0.980 | 1.1 | 0.990 | 2.5 - 7.5 | | Cl- | 50 | 0.990 | 1.1 | 0.980 | 78 - 120 | I hope you find this information useful and informative. Ram Nunna, President 1/11/02 Data Date {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 # FEB 01 2002 Mr. Ram Nunna President pHoenix Diagnostic Inc. 8 Tech Circle Natick, MA 01760 Re: k020129 > Trade/Device Name: pHoenix Electrolyte Calibration Set for the Roche® Cobas™ ISE Module Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: November 2, 2001 Received: January 14, 2002 Dear Mr. Nunna: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on.our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/odrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE STATEMENT ### 510(k) Number (if known): K020129 Device Name: pHoenix Electrolyte Calibration Set for the Roche® Cobas™ ISE Module Indications For Use: ### Intended Use: The products encompassed by this request are intended for invitro diagnostic use and are intended for use in calibrating Nat, K+, Cli electrodes in Roche Cobas ISE module. Roche is the original equipment manufacturer for the analyzer and predicate reagents. phoenix electrolyte calibration set is intended for use in playe of like named products manufactured by Roche Diagnostics. (Division Sign-Off) Division of Clinical Laboratory Devices | 510(k) Number | K020129 | |---------------|---------| |---------------|---------| ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...