The Bionx Duet™ Suture Anchor is intended for use to reattach soft tissue to bone in orthopaedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules, to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.
Device Story
Duet Suture Anchor is a threaded, bioabsorbable screw-in device; preloaded on a disposable inserter with two non-absorbable braided polyester #2 sutures. Used by surgeons in OR for arthroscopic or open soft tissue-to-bone reattachment. Device provides mechanical fixation during healing; copolymer retains 90% strength for 20 weeks; resorbs over several years. Surgeon inserts anchor into bone; sutures are used to secure soft tissue. Benefits include stable fixation during healing phase; gradual strength loss matches tissue repair progression. Material is inert, non-collagenous, non-pyrogenic.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
Bioabsorbable screw-in anchor; 6.0mm diameter, 16.5mm length. Material: Self-Reinforced (96/4D) poly-L/D-lactide copolymer. Preloaded with two non-absorbable braided polyester #2 sutures. Sterile, single-use disposable inserter. Mechanical properties: retains 90% strength through 20 weeks; complete resorption over several years.
Indications for Use
Indicated for reattachment of soft tissue (ligaments, tendons, joint capsules) to bone in arthroscopic or open orthopaedic procedures. Contraindicated in: non-listed procedures; conditions compromising fixation (osteopenic/comminuted bone, soft tissue pathology); conditions retarding healing (poor blood supply, infection); active infection; inability/unwillingness to restrict activity; foreign body sensitivity; allergy to implant/suture materials.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K020056 page 193
JUN 2 6 2002
# 510(k) Summary Bionx Implants Inc. Duet Suture Anchor™
# Submitter's Name, Address, Telephone Number, and Contact Person
Bionx Implants, Inc. 1777 Sentry Parkway West Gwynedd Hall, Suite 400 Blue Bell, PA 19422
Gerard S. Carlozzi Contacts: President and Chief Executive Officer Phone: (215) 643-5000 (215) 653-0984 Facsimile:
> Bionx Implants Ltd. Tuija Annala Director, Quality and Regulatory Affairs P.O.Box 3 FIN-33721 Tampere Finland Phone: 358-3-316 5679 Facsimile: 358-3-316 5629
#### Date prepared: December 18, 2001
#### Name of the device:
| A. | Trade or Proprietary Name: | Duet™ Suture Anchor |
|----|----------------------------|--------------------------------------------------------|
| B. | Common Name: | Bioabsorbable suture anchor |
| C. | Classification Name: | Biodegradable soft tissue fixation<br>fastener (87MAI) |
| D. | Device Product Code: | MAI |
| E. | Regulatory Classification: | Class II |
#### Predicate Device:
The Duet™ Suture Anchor is intended for use in the reattachment of soft tissue to bone by arthroscopy or direct visualization. This indication is substantial equivalent with Bionx Implants Inc. SmartAnchor (K972783), Bankart Tack (K973849),
{1}------------------------------------------------
Contour Labral Nail (K992567, K003970), BioCuff™ (K001378) and BioCuff™ C (K013057), Arthrex Inc. Bio-Absorbable Corkscrew (K990361) Bio-Absorbable Corkscrew Suture Anchor (K990987), Bio-Absorbable CorkScrew Suture Anchor (K003227) and TissueTak (K990340), Innovasive Devices, Inc. BioROC EZ (K973381), RC Multi-Suture Anchor (K992458) and Sutureless Anchor (K984490), Instrument Makar, Inc,Biologically Quiet Biosphere (K963396) and Biologically Ouiet Mini-Screw Suture Anchor (K960555), Linvatec Corporation Bio-Anchor (K964805, K963369, K983186) and BioRC Anchor (K990110), Mitek Surgical Products Panalok (K950272), Panalok RC (K970896) and Absorbable Toggle Anchor (K964013), Acufex Inc. Suretac 6.0, 8,0 (K911837) and DePuy Inc. De Puy Dupont Phantom Suture Anchor (K964921).
## Intended Use:
The Bionx Duet™ Suture Anchor is a bioabsorbale screw-in suture anchor that is preloaded on a disposable inserter device with two non-absorbable, braided, polyester #2 sutures. The Duet™ Suture Anchor is manufactured from Self-Reinforced (96/4D) PLA Copolymer that retains 90% strength through 20 weeks and completely resorbs over a period of several years in vivo. The unique Self-Reinforced Copolymer provides high initial mechanical strength required for insertion and through the healing phase (20 weeks) with complete absorption occurring over several years. The unique absorption profile of the Copolymer allows the anchor to gradually loose strength as the collagen fibers of the repair form and gain strength. The Copolymer is inert, non-collagenous and non-pyrogenic through the absorption process. The Duet™ Suture Anchor is indicated for use in arthroscopic or open surgical procedures to reattach soft tissue to the bone.
The Duet™ Suture Anchor is contraindicated in the following conditions:
- Surgical procedures other than those listed. 1.
- Conditions that may compromise Anchor fixation (osteopenic, comminuted 2. bone, pathologic conditions in the soft tissues to be attached, etc.)
- Conditions that may retard healing (poor blood supply, past or potential 3. infection, etc).
- Active infection. 4.
- Conditions that may limit the patients ability or willingness to restrict റ് activities or follow directions during the healing period.
- Foreign body sensitivity to materials. 6.
- 7. Patients with suspected or known allergy to the implant or suture materials.
# Device Description:
Duet™ Suture Anchor is threaded, bioabsorbable suture anchor, provided as preloaded with two non-absorbable, braided, polyester #2 sutures, one of them is green and another one is white. Diameter of the anchor is 6.0mm and length 16.5mm. It is assembled with sutures with needles into single use, sterile, disposable driver and packaged into blister with a punch. The Duet™ Suture Anchor is made of poly-L/D-
{2}------------------------------------------------
page 3 of 3
lactide copolymer. The poly-L/D-lactide material equals the raw material used in previously cleared Bionx Implants Inc. SmartNail™ (K993074), Meniscus Arrow™ (K993453), SmartScrew ACL™ (K993073), SmartScrew (K012001), The Wedge™ (K000616), Contour Labral Nail (K003970), BioCuff (K001378) and BioCuff C (K013057).
## Substantial Equivalence:
All of these devices have the same intended use and principles of operation and very similar technological characteristics. Furthermore, the minor technological differences between the Bionx Implants Inc. Duet™ Suture Anchor and the predicate devices do not raise any new issues of safety or effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 2 6 2002
Ms. Tuija Annala Director Quality and Regulatory Affairs Bionx Implants Ltd. P.O. Box 3. FIN-33721 Tampere, Finland
Re: K020056
Trade/Device Name: Duet™ Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: April 5, 2002 Received: April 8, 2002
Dear Ms. Annala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{4}------------------------------------------------
### Page 2 - Ms. Tuija Annala
(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N-Mullene
for Celia M. Witten, Ph.D., M.D.
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE
510(K) Number (if known):
Device Name:
Duet™ Suture Anchor
#### Indications for Use:
The Bionx Duet™ Suture Anchor is intended for use to reattach soft tissue to bone in orthopaedic surgical procedures. The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules, to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.
The Bionx Duet™ Suture Anchor is contraindicated in 1) Surgical procedures other than those listed, 2) Conditions that may compromise Anchor fixation (osteopenic, comminuted bone, pathologic conditions in the soft tissues to be attached, etc., 3)Conditions that may retard healing (poor blood supply, past or potential infection, etc), 4) Active infection, 5) Conditions that may limit the patients ability or willingness to restrict activities or follow directions during the healing period, 6)Foreign body sensitivity to materials, 7) Patients with suspected or known allergy with implant or suture materials.
(Please do not write below this line - continue on another page is needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
No OR Over-The-Counter Use
Mark A. Millman
vision Sign-Division of Genat 91. Restorative and Neurological Devices
510(k) Number.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.