THE IZZZY-TENS (WL-2401), THE SLIM-TENS (WL-2402), THE MENS-TENS (WL2403), THE COMPACT TENS (WL-2302)
K020020 · Well-Life Healthcare, Inc. · GZJ · Dec 30, 2002 · Neurology
Device Facts
Record ID
K020020
Device Name
THE IZZZY-TENS (WL-2401), THE SLIM-TENS (WL-2402), THE MENS-TENS (WL2403), THE COMPACT TENS (WL-2302)
Applicant
Well-Life Healthcare, Inc.
Product Code
GZJ · Neurology
Decision Date
Dec 30, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Mini-TENS series including WL-2401, WL-2402, WL-2403, and WL-2302, is a portable, batterypowered, transcutaneous electrical nerve stimulators (TENS device) that are used in the treatment of pain syndromes, that is intended for use on intact skin, and that do not require surgical intervention or violation of the skin surface.
Device Story
Portable, battery-powered TENS device for pain relief; applies electrical current to electrodes attached to patient's intact skin. Stimulus generator produces adjustable output current; transmitted via output port to electrodes. Operates in burst mode with fixed pulse rate, frequency, and timer; amplitude is user-adjustable. Used by patients for symptomatic relief of chronic intractable pain. Device models (WL-2401, WL-2402, WL-2403, WL-2302) vary by stimulation cycle, waveform combinations, and timer settings. Healthcare providers prescribe device; patient applies electrodes and adjusts amplitude for pain management. Benefits include non-invasive, drug-free pain relief.
Clinical Evidence
Bench testing only. Verification and validation tests confirm performance specifications and safety compliance with ANSI/AAMI NS4-1985, EN 60601-1, EN 60601-1-1, and EN 60601-1-2. No clinical data presented.
Technological Characteristics
Portable TENS device; plastic housing; 3V Lithium battery power. Biphasic, rectangular waveform output. Regulated current (RC) delivery. Type BF applied part. Connectivity: standalone. Sterilization: not applicable (electrodes provided). Firmware-controlled stimulation cycles. Dimensions vary by model (e.g., 60x47.3x14.6mm to 56x26x8mm).
Indications for Use
Indicated for symptomatic relief of chronic intractable pain in patients requiring non-surgical, non-invasive transcutaneous electrical nerve stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the word "ANNEX" followed by the Roman numeral "II", all enclosed in a rectangular box. The text is in a bold, sans-serif font and appears to be a title or heading. The black ink of the text contrasts with the white background, making it easily readable.
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _K020020
DEC 3 0 2002
1. Submitter's Identifications: Well-Life Healthcare Inc. Room 6C01, No. 5, Sec. 5, Hsin Yi Rd., Taipei, Taiwan, R.O.C.
Contact: Ms. Grace Chang Sales Manager
Date of Summary Preparation: October, 2001.
- 2. Name of the Device: Well-Life TENS (Transcutaneous Electrical Nerve Stimulation Device), Model Mini-TENS series. including WL-2401, WL-2402, WL-2403, and WL-2302.
- 3. Information of the 510(k) Cleared Device (Predicate Device):
## 4. Device Description:
The Mini-TENS series, including WL-2401, WL-2402, WL-2403, and WL-2302 are transcutaneous electrical nerve stimulator used for pain relief by applying an electrical current to electrodes, which are attached on the patient's skin. The output and waveform is adjustable according to the situation of patient.
Mini-TENS series, models WL-2401, WL-2402, WL-2403, and WL-2302, consist mainly of two parts: the stimulus generator, electrode. The stimulus generator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the patient's skin so as to transmit this stimulus current to the patient for pain relief.
The stimulation mode for Mini-TENS is burst mode with fixed pulse rate, frequency, and fixed timer. Only amplitude is adjustable. This operation way is considered the simplification from the burst mode of comparison clear model, WL-2102A. Every model of Mini-TENS has its individual stimulation operation cycle.
The Mini-TENS series (WL-2401, WL-2402, WL-2403, and WL-2302) are almost identical in function outlook and specification. The main differences between these four models are indicated in the following table.
{1}------------------------------------------------
| Model | Outlook | Time for<br>each<br>operation<br>cycle | Combination of<br>wave form | Wave form information | | | |
|----------------------|---------|----------------------------------------|--------------------------------------------------------------------------|-----------------------|----------------------|------------------------|------------------|
| | | | | Type | Pulse rate<br>(Hz) | Pulse width<br>(μs) | Time<br>duration |
| Mini-TENS<br>WL-2401 | Type A | 60<br>minutes | Single wave<br>form (A) | A | 30 | 250 | 3s/ON,<br>5s/OFF |
| Mini-TENS<br>WL-2402 | Type B | 15<br>minutes | A/1min→<br>B/1min→<br>C/1min, repeat 5<br>times | A | 30~60<br>auto-change | 250 | 6s/ON,<br>1s/OFF |
| | | | | B | 30 | 150~250<br>auto change | 6s/ON,<br>1s/OFF |
| | | | | C | 30~60<br>auto change | 150~250<br>auto change | 6s/ON,<br>1s/OFF |
| Mini-TENS<br>WL-2403 | Type C | 30<br>minutes | Single wave<br>form (A) | A | 25 | 115 | 4s/ON,<br>1s/OFF |
| Mini-TENS<br>WL-2302 | Type C | 15<br>minutes | A/2min→<br>B/2min→<br>C/2min→<br>B/2min→<br>D/2min→<br>B/2min→<br>C/3min | A | 2 | 115 | 2s/ON,<br>1s/OFF |
| | | | | B | 30 | 115 | 2s/ON,<br>1s/OFF |
| | | | | C | 4 | 115 | ON |
| | | | | D | 15 | 115 | 4s/ON,<br>1s/OFF |
The WL-2402 is the most complicate in these three models, so the assessment results for the features of performance and safety of this model could cover that of the other three models.
While comparing with the original 510(k) clear device, WL-2102A. The main difference between burst mode operation of WL-2102A and Mini-TENS are indicated as the following table:
| Comparison<br>features | Clear model | New models | | | |
|---------------------------------------------|-----------------------------|----------------------|----------------------------------------------------------------|----------------------|---------------------------------------------------------------------------------------|
| | WL-2102A | WL-2401 | WL-2402 | WL-2403 | WL-2302 |
| Timer control | 15/30/60 mins<br>adjustable | 60 mins fixed | 15 mins fixed | 30 mins fixed | 15 mins fixed |
| Burst duration<br>time | 0.1s/ON,<br>0.4s/OFF | 3s/ON, 5s/OFF | 6s/ON, 1s/OFF | 4s/ON, 1s/OFF | 2s/ON, 1s/OFF<br>or 4s/ON,<br>1s/OFF or<br>always ON |
| Pulse rate<br>(Hz) | 100Hz fixed | 30Hz fixed | 30~60Hz auto<br>change | 25Hz fixed | 2/4/15/30 Hz<br>auto change |
| Pulse width<br>(μs) | 0~260μs<br>adjustable | 250μs fixed | 150~250μs<br>auto change | 115μs fixed | 115μs fixed |
| Amplitude<br>(mA) | 0~80mA<br>adjustable | 0~50mA<br>adjustable | 0~50mA<br>adjustable | 0~50mA<br>adjustable | 0~50mA<br>adjustable |
| Combination<br>for the type of<br>wave form | Single (A) | Single (A) | 3 wave forms<br>combination:<br>A(1m)→B(1m)→C(1m), 5<br>repeat | Single (A) | 4 wave forms<br>combination:<br>A(2m)→B(2m)→<br>C(2m)→B(2m)→<br>D(2m)→B(2m)→<br>C(3m) |
{2}------------------------------------------------
**ANNEX II**
## 5. Intended Use:
Mini-TENS series including WL-2401, WL-2402, WL-2403, and WL-2302, is a portable, batterypowered, transcutaneous electrical nerve stimulators (TENS device) that are used in the treatment of pain syndromes, that is intended for use on intact skin, and that do not require surgical intervention or violation of the skin surface.
The electrode is included in the packing of Mini-TENS series.
Because that all the performances tested are completely within the limit of standard, and all the labeling is provided according to the requirement of standard, the Well-TENS series fall completely within the scope of clause 1 of ANSI/AAMI NS4-1985.
## 6. Comparison to the 510(k) Cleared Device (Predicate Device):
| Features | Clear model | New model | | | | |
|------------------------------------------------------------------------------------|-----------------------------------|-------------------------|-------------------------|-------------------------|-------------------------|--|
| | WL-2102A | WL-2401 | WL-2402 | WL-2403 | WL-2302 | |
| 1. 510(k) Number | K002874 | K020020 | K020020 | K020020 | K020020 | |
| 2. Device Name, Model | Well-TENS,<br>WL-2102A | Mini-TENS,<br>WL-2401 | Mini-TENS,<br>WL-2402 | Mini-TENS,<br>WL-2403 | Mini-TENS,<br>WL-2302 | |
| 3. Manufacturer | Well-Life | Well-Life | Well-Life | Well-Life | Well-Life | |
| 4. Power Source(s)<br>- Method of Line Current<br>Isolation | 1x9V Alkaline<br>Type BF | 1x3V Lithium<br>Type BF | 1x3V Lithium<br>Type BF | 1x3V Lithium<br>Type BF | 1x3V Lithium<br>Type BF | |
| Normal condition<br>Single fault condition | < 0.001mA<br>Not tested | < 0.001mA<br>Not tested | < 0.001mA<br>Not tested | < 0.001mA<br>Not tested | < 0.001mA<br>Not tested | |
| 5. Number of Output Modes | 3 (B/C/M) | 1 (Burst<br>mode) | 1 (Burst<br>mode) | 1 (Burst<br>mode) | 1 (Burst<br>mode) | |
| 6. Number of Output Channels<br>- Synchronous or | 2 | 1 | 1 | 1 | 1 | |
| Alternating?<br>- Method of Channel<br>Isolation | Synchronous<br>Type BF | Synchronous<br>Type BF | Synchronous<br>Type BF | Synchronous<br>Type BF | Synchronous<br>Type BF | |
| 7. Regulated Current or<br>Regulated Voltage? | RC | RC | RC | RC | RC | |
| 8. Software/Firmware/microproc<br>essor Control? | Firmware | Firmware | Firmware | Firmware | Firmware | |
| 9. Automatic Overload Trip? | No | No | No | No | No | |
| 10. Automatic No-Load Trip? | No | No | No | No | No | |
| 11. Automatic Shut Off? | Yes | Yes | Yes | Yes | Yes | |
| 12. Patient Override Control? | No | No | No | No | No | |
| 13. Indicator Display:<br>On/Off Status?<br>Low Battery?<br>Voltage/Current Level? | Yes<br>No<br>No | Yes<br>No<br>No | Yes<br>No<br>No | Yes<br>No<br>No | Yes<br>No<br>No | |
| 14. Timer Range | 15/30/60<br>minutes<br>adjustable | 60 minutes | 15 minutes | 30 minutes | 15 minutes | |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the words "ANNEX II" in a rectangular box. The text is in a bold, sans-serif font. The box is made up of a thin, black line.
| | Clear model | New model | WL-2401 | WL-2402 | WL-2403 | WL-2302 |
|--------------------------------------------------------------------------------------|------------------|------------------|------------------|------------------|------------------|---------|
| Features | WL-2102A | | | | | |
| 15. Compliance with Voluntary<br>Standards? | ANSI/AAMI<br>NS4 | ANSI/AAMI<br>NS4 | ANSI/AAMI<br>NS4 | ANSI/AAMI<br>NS4 | ANSI/AAMI<br>NS4 | |
| 16. Compliance with 21 CFR<br>898? (*Becomes<br>mandatory beginning May,<br>9, 2000) | Yes | No | No | No | No | |
| 17. Weight (g) | 135 | 19 | 19 | 19 | 19 | |
| 18. Dimensions (in) [W×H×D] | 92×62×25 | 60×47.3×14.<br>6 | 60×47.3×14.<br>6 | 56×26×8 | 56×26×8 | |
| 19. Housing Materials and<br>Construction | Plastic | Plastic | Plastic | Plastic | Plastic | |
# Table for the comparison of output specifications for mini TENS
| | Clear model | New model | | | |
|--------------------------------------------------------------|----------------------|-------------|-----------------------|-------------|----------------------------|
| Specification | WL-2102A | WL-2401 | WL-2402 | WL-2403 | WL-2302 |
| Waveform<br>(e.g., pulsed monophasic, biphasic) | Biphasic | Biphasic | Biphasic | Biphasic | Biphasic |
| Shape<br>(e.g., rectangular, spike, rectified<br>sinusoidal) | Rectangular | Rectangular | Rectangular | Rectangular | Rectangular |
| Maximum Output Voltage (+/-20%) | | | | | |
| @ 500 Ω | 38V | 32V | 32V | 38.2V | 38.2V |
| @ 2 kΩ | 55V | 42V | 42V | 52.8V | 52.8V |
| @ 10 KΩ | 130V | 80.8V | 80.8V | 84.8V | 84.8V |
| Maximum Output Current (+/-20%) | | | | | |
| @ 500 Ω | +76mA | +64mA | +64mA | +76.2mA | +76.2mA |
| @ 2 kΩ | +55mA | +42mA | +42mA | +52.8mA | +52.8mA |
| @ 10 KΩ | +13mA | +8.08mA | +8.08mA | +8.48mA | +8.48mA |
| Pulse Width (us) | 30-260<br>adjustable | 250 | 150~250<br>changeable | 250 | 250 |
| Frequency (Hz) | 2-150<br>adjustable | 30 | 30-60<br>changeable | 25 | 2, 4, 15, 30<br>changeable |
| Beat Frequency<br>(for interferential modes only) | NA | NA | NA | NA | NA |
| Symmetrical phases<br>(for multiphasic waveform only)? | NA | NA | NA | NA | NA |
| Phase Duration<br>(for multiphasic waveform only) | NA | NA | NA | NA | NA |
| Net Charge (µ C per pulse) @500 Ω | 20.8 | 16.0 | 16.0 | 8.76 | 8.76 |
| Maximum phase Charge (µC) | 22 | 12.5 | 12.5 | 9.2 | 9.2 |
| Maximum Current Density (mA/cm²) | 0.18525 | 0.0333 | 0.0265 | 0.006 | 0.007 |
| Maximum Power Density (W/cm²) | 0.0070395 | 0.0010656 | 0.000848 | 0.0002292 | 0.0002674 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the words "ANNEX II" in bold, sans-serif font. The text is enclosed in a rectangular box. The letters are evenly spaced and of uniform size. The image appears to be a scan or a digital reproduction of a document.
## 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, NS4-1985, as well as E.N 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.
### 8. Conclusions
The Mini-TENS series transcutaneous electrical nerve stimulator, WL-2401, WL-2402, and WL-2403, and WL-2302, have the same intended use and technological characteristics as the cleared device WL-2102A. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
ood and Drug Administration 200 Corporate Boulevard Rockville MD 20850
Mr. Tony C. S. Chang Wincent Consultant Co., Ltd. No. 5. Alley 5. Lane Cheng Hsing, Chung Ching Road, Pei Tun District, Taichung, Taiwan, R.O.C.
Re: K020020
Trade/Device Name: Mini-TENS Series TENS Devices Models WL-2401, WL-2402, WL-2403, WL-2302 Regulation Number: 21 CFR 882.5890 Regulation Name: TENS Device for Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: December 11, 2002 Received: December 12, 2002
Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the devices referenced above and have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your devices are classified (see above) into either class II (Special Controls) or class III (PMA), they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your devices in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{6}------------------------------------------------
Page 2 - Mr. Tony C. S. Chang
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your devices as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Matt-1-Mather-
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "ANNEX 1" written in bold, capital letters. The letters are slightly distorted, and the image quality is low. The text is enclosed in a rectangular box.
Page
510(k) Number (if known) k020020
Device Name:__Mini-TENS / Model: WL-2401, WL-2402, WL-2403, WL-2302
Indications For Use:
The Mini-TENS including a series of TENS models (WL-2401, WL-2402, WL-2403, and WL-2302) is intended for symptomatic relief of chronic intractable pain.
J. Hall of Williams
ivision Sign-Off) Division of General, Restorative and Neurological Devices
10(k) Number K020020
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X OR
Over-The-Counter Use
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.