← Product Code [JAQ](/productcode/JAQ) · K020015

# IMPLANT GUIDANCE SYSTEM APPLICATOR SET (K020015)

_Nucletron Corp. · JAQ · Mar 13, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K020015

## Device Facts

- **Applicant:** Nucletron Corp.
- **Product Code:** [JAQ](/productcode/JAQ.md)
- **Decision Date:** Mar 13, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.5700
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

Nucletron Implant Guidance System Applicator Set is intended for interstitial brachytherapy procedures involving the Nucletron remote afterloading equipment: mHDR. The applicator provides a means of delivering the prescribed radiation dose to the treatment area. The applicator is a closed system to prevent the radioactive source from coming in contact with body fluids. The Nucletron Implant Guidance System Applicator Set is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.

## Device Story

Nucletron Implant Guidance System Applicator Set functions as an accessory to mHDR remote afterloading equipment for interstitial brachytherapy. System components include insertion needles, plastic sheaths, closed-end plastic catheters, buttons, and treatment needles. Procedure: insertion needle/sheath placed into target volume; needle removed; catheter inserted into sheath; button placed at skin surface; catheter cut and attached; treatment needle inserted into catheter and locked with metal stopper. Radiographic imaging (CT/MR) confirms applicator location for treatment planning. Treatment needles connect to mHDR equipment via transfer tubes. A check cable run verifies path patency before radioactive source delivery. Applicator acts as a closed system, preventing source contact with body fluids. Used in clinical settings by trained medical professionals. Benefits include precise radiation delivery to target tissue while maintaining a closed, safe path for the radioactive source.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Closed-system applicator set consisting of insertion needles, plastic sheaths, closed-end plastic catheters, and metal locking stoppers. Designed for use with Nucletron mHDR remote afterloading equipment. Provides a physical path for radioactive source delivery. Non-electronic, mechanical accessory.

## Regulatory Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

## Predicate Devices

- Speiser Needle Set ([K953946](/device/K953946.md))
- Flexible Implant Tubes ([K953946](/device/K953946.md))
- Interstitial Needle Set ([K953946](/device/K953946.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Special 510(k) K020015: reply to questions Nucletron Implant Guidance System Applicator Set Februari 2002

Image /page/0/Picture/1 description: The image shows the word "Nucletron" in bold, black font. To the left of the word is a square containing a black circle with two white circles on either side. The black circle is tilted at an angle within the square. The word "Nucletron" is written in a simple, sans-serif font.

# MAR 1 3 2002

Fax

K020015
p. 1 of 2

NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133

+31 318 550485

Department of Health and Human Services Center of Device and Radiological Health Office of Device Evaluation Special 510(k) section

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION

as required by section 807.92(c)

## Submitter of 510(k):

Company name: Registration number: Address:

Phone: Fax: Correspondent: Nucletron Corporation 1121753 7080 Columbia Gateway Drive Columbia, MD 21046-2133 410-312-4100 410-312-4197 Lisa Dimmick Regulatory Affairs Manager

Modified Device Name:

| Trade/Proprietary Name: | Implant Guidance System Applicator Set                          |
|-------------------------|-----------------------------------------------------------------|
| Common/Usual Name:      | Remote Afterloading for interstitial brachytherapy applications |
| Classification Name:    | Remote controlled radionucide applicator system accessory       |
| Classification          | 21 CFR 892.5840, Class II                                       |

# Legally Marketed Device(s)

Our device is substantially equivalent to the legally marketed predicate device cited in the table below:

| Manufacturer   | A Device Device C<br>のではないという<br>No. of the Comment of the<br>A 1440 A Product of Property Top A 2012 2 | 510(k) # = |
|----------------|---------------------------------------------------------------------------------------------------------|------------|
| Nucletron B.V. | Speiser Needle Set                                                                                      | K953946    |
| Nucletron B.V  | Flexible Implant Tubes                                                                                  | K953946    |
| Nucletron B.V  | Interstitial Needle Set                                                                                 | K953946    |

# Description:

The Nucletron Implant Guidance System as described in this submission is designed as an accessory to the Nucletron remote afterloading equipment, mHDR, and is intended for interstitial brachytherapy procedures.

The Nucletron Implant Guidance System Applicator Set consists of an insertion needle with a plastic sheath, which is inserted through the skin surface into the target

{1}------------------------------------------------

## Special 510(k) K020015: reply to questions Nucletron Implant Guidance System Applicator Set Februari 2002

volume. The insertion needle is then removed and a closed end plastic catheter is placed into the sheath. A marker on the side of the insertion obturator insures the proper depth of catheter placement. Once the catheter is in place the sheath is removed and a button is placed over the plastic catheter and slid into contact with the skin surface. The catheter is then cut level with the button and attached. A closed end treatment needle is then inserted into the catheter, once in position a metal needle stopper is slid over the needle to the button surface and locked. Radiographic images, planar films or transverse slices, i.e. CT. MR is obtained to determine the precise location of the applicator within the body. This information is then used for brachytherapy treatment planning purposes. When the treatment planning is completed the treatment needle is then attached to the Nucletron remote afterloading equipment (treatment head), mHDR, by the Nucletron transfer tubes. The transfer tubes lock onto the open end of the treatment needles and the remote afterloading equipment (treatment head) prior to treatment. When the applicator is attached, a check cable run is performed to ensure that the applicator is properly attached and that there are no obstructions, which will interrupt treatment. After the check cable run, the radioactive source will step through the applicator to deliver the prescribed dose of radiation. When the treatment session is complete, the treatment needles are detached from the transfer tube and remote afterioading equipment. The treatment needles are then removed from the patient. When the course of treatment is completed the catheters and buttons are removed from the patient.

The applicator is a closed system to prevent the radioactive source from coming in contact with body fluids. The applicator does not control the treatment unit; it strictly provides a treatment path for the radioactive source. The Nucletron remote afterloading system and the clinical staff verify that the applicator is properly attached prior to treatment.

#### Intended use:

Nucletron Implant Guidance System Applicator Set is intended for interstitial brachytherapy procedures involving the Nucletron remote afterloading equipment: mHDR. The applicator provides a means of delivering the prescribed radiation dose to the treatment area. The applicator is a closed system to prevent the radioactive source from coming in contact with body fluids.

### Summary of technological considerations:

The Nucletron Implant Guidance System Applicator Set is substantially equivalent to the cleared predicate devices, Speiser Needle Set (#K953946), Flexible Implant Tubes (#K953946) and Interstitial Needle Set (#K953946).

Leslie E. Fe

Name: U. Lutz Title: Business Seament Manager HDR Brachytherapy Nucletron B.V. Veenendaal, The Netherlands

11.02.02
Date

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human figures in profile, with a triple-bar design above them.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 3 2002

Ms. Lisa C. Dimmick Regulatory Affairs Manager Nucletron Corporation 7080 Columbia Gateway Drive COLUMBIA MD 21046-2133

Re: K020015

Trade/Device Name: Implant Guidance System Applicator Set Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: 90 JAQ Dated: February 11, 2002

Received: February 19, 2002

Dear Ms. Dimmick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nouticate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific ac ne of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx 876.2xxx, 3xxx, 4xxx, 5xxx 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx 892.2xxx, 3xxx, 4xxx, 5xxx Other

(301) 594-4591 (301) 594-4616 (301) 594-4616 (301) 594-4654 (301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding Office of Cotopremarket notification" (21 CFR Part 807.97). Other general information on by relected to premiered the Act may be obtained from the Division of Small Manufacturers, Jour responsiblement assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Special 510(k) Nucletron Implant Guidance System Applicator Set December 10, 2001

Image /page/4/Picture/1 description: The image shows the word "Nucletron" in bold, black font. To the left of the word is a square containing a black circle with two white circles on either side. The word "Nucletron" is written in a sans-serif font and is slightly larger than the square graphic.

#### NUCLETRON B.V.

Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 +31 318 550485 Fax

Department of Health and Human Services Center of Device and Radiological Health Office of Device Evaluation Special 510(k) section

Kozoo15

# Statement of intended use

Device Name:

Implant Guidance System Applicator Set

#### Intended Use:

Nucletron Implant Guidance System Applicator Set is intended for interstitial brachytherapy procedures involving the Nucletron remote afterloading equipment: mHDR. The applicator provides a means of delivering the prescribed radiation dose to the treatment area. The applicator is a closed system to prevent the radioactive source from coming in contact with body fluids.

#### Prescription use:

The Nucletron Implant Guidance System Applicator Set is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.

Leslie-& R

10.12.01
Date

Name: U. Lutz Title: Business Segment Manager HDR Brachytherapy Nucletron B.V. Veenendaal, The Netherlands

Nancy Brandon

---

**Source:** [https://fda.innolitics.com/device/K020015](https://fda.innolitics.com/device/K020015)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
