← Product Code [JDS](/productcode/JDS) · K014189

# ODI TALON INTRAMEDULLARY HIP NAIL (K014189)

_Orthopedic Designs, Inc. · JDS · Mar 21, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K014189

## Device Facts

- **Applicant:** Orthopedic Designs, Inc.
- **Product Code:** [JDS](/productcode/JDS.md)
- **Decision Date:** Mar 21, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ODi Talon™ Intramedullary Hip Nail will be used on indications that are common with presently marketed intramedullary hip nail systems. The primary indications are for fixation/stabilization of stable and unstable fractures of the proximal femur including intertrochanteric fractures, pertrochanteric fractures, high subtrochanteric fractures, and combinations of these fractures. The device is intended to stabilize fragments of the fracture until bony union can occur.

## Device Story

ODi Talon™ Intramedullary Hip Nail provides fixation and stabilization for proximal femur fractures. System components include an intramedullary nail, lag screw with deployable tangs for femoral head purchase, slotted sleeve, sleeve lock, compression screw, and end cap. Device is implanted by surgeons in an OR setting. Lag screw tangs provide rotational stability and purchase; compression screw allows axial fracture compression. Polyethylene patch in compression screw prevents rotation. End cap protects nail threads for potential future explantation. Cortical screws provide distal cross-locking. Device stabilizes fracture fragments until bony union occurs.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Implant grade stainless steel construction. System includes intramedullary nail, lag screw with deployable tangs, slotted sleeve, sleeve lock, compression screw with polyethylene patch, and end cap. Mechanical fixation via intramedullary placement and distal cortical screws. Pre-assembled configuration for surgical efficiency.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Talon™ Compression Hip Screw system ([K984331](/device/K984331.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 2 1 2002

14189

# SECTION 12: 510(k) SUMMARY

Image /page/0/Picture/2 description: The image shows the logo for Orthopedic Designs, Inc. The logo consists of a stylized "i" inside a circle, followed by the letters "ODI" in a bold, sans-serif font. Below the letters, the words "ORTHOPEDIC DESIGNS, INC." are written in a smaller, sans-serif font.

# Premarket Notification

510(k) Summary of Safety and Effectiveness Information

For Release Upon Request Only

Date of Preparation:

December 20, 2001

### Regulatory Authority:

Safe Medical Devices Act of 1990, 21 CFR 807.92

## Company Name / Contact:

| Company: | Orthopedic Designs, Inc. (ODi) |
|----------|--------------------------------|
|          | 6965 1st Ave. North            |
|          | St. Petersburg, FL 33710       |

#### John Sodeika Contact: (727) 343-0338

Establishment Registration Number: 1064129

| Classification Name:       | Rod, Fixation, Intramedullary and<br>Accessories, Metallic |
|----------------------------|------------------------------------------------------------|
| Classification Reference:  | 21 CFR § 888.3020                                          |
| Common Used Name:          | Femoral Nail                                               |
| Device Product Code:       | HSB                                                        |
| Classification Panel:      | 87- Orthopedic Devices                                     |
| Trade Proprietary Name:    | ODi Talon™ Intramedullary Hip Nail                         |
| Proposed Regulatory Class: | Class II                                                   |

{1}------------------------------------------------

## Device Description:

The ODi Talon™ Intramedullary Hip Nail is used for fixation and stabilization of fractures of the proximal femur until bony union can occur. The system consists of the following parts:

- An intramedullary nail with portals that allow passage of distal cortical screws . and a proximal lag screw assembly. The nail will be provided in a pre-assembled condition with the sleeve lock and end cap already attached to save time in surgery.
- The Talon™ Lag Screw is completely compatible with ODi's Talon™ ● Compression Hip Screw system and has been previously approved by the FDA in K984331. The Talon™ lag screw has deployable tangs to increase the purchase of the lag screw within the femoral neck/head. These tangs may also be retracted for removal of the lag screw if and when it is necessary. The distal end of the lag screw is keyed with a "double-d" shape, cannulated and internally threaded. This keyed shaft provides rotational stability for better lag screw purchase. The screw is internally threaded to allow the use of the compression screw to compress the fracture fragments.
- A slotted sleeve which passes through the intramedullary nail. The sleeve is . keyed to the lag screw assembly to prevent its rotation while allowing axial translation of the lag screw.
- A sleeve lock which passes through the proximal end inner bore of the ● intramedullary nail. The sleeve lock has 2 positions within the intramedullary nail - "locked" and "unlocked". In the "locked" position, the legs on the sleeve lock mate with the slots in the sleeve thereby preventing rotation and axial translation of the sleeve, but allowing axial translation of the lag screw assembly. The sleeve lock is provided for surgery pre-assembled in the "unlocked" position within the intramedullary nail.
- A compression screw which shoulders against the slotted sleeve and engages . the internal threads in the distal end of the lag screw assembly providing for axial compression of a proximal hip fracture. A polyethylene patch is embedded in the threaded portion of the compression screw to prevent rotation of the screw inside the lag screw after the desired amount of compression is reached.. The compression screw is used to compress the fracture site by drawing the nail and lag screw portions together. The compression screw has been previously approved by the FDA in K984331.
- An end cap with both internal and external threads and a keying slot. The . external threads engage the internal threads in the proximal end of the intramedullary nail and protect them from bony ingrowth for the possible future attachment of nail removal instrumentation for explantation. The internal threads in the end cap mate with the nail installation instrumentation. The end cap is provided for surgery pre-assembled in the intramedullary nail.
- Cortical screws are provided to cross-lock the distal end of the nail to the . femoral shaft to help prevent axial translation of the nail. The cortical screws have been previously approved by the FDA in K984331.

The nail will be provided in a pre-assembled condition with the sleeve lock and end cap already attached to save time in surgery.

ODi will manufacture the implants from implant grade stainless steels.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# MAR 2 1 2002

Mr. John Sodeika Vice President of Engineering Operations Orthopedic Designs, Inc. 6965 1st Avenue N. St. Petersburg, Florida 33710

Re: K014189

Trade/Device Name: Odi Talon™ Intramedullary Hip Nail Regulation Number: 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: JDS Dated: December 20, 2001 Received: December 21, 2001

Dear Mr. Sodeika:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. John Sodeika

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely, yours,

R. Mark N. Mckune

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# SECTION 5: DEVICE INDICATIONS FOR USE

Page 1 of

510(k) Number:

ODi Talon™ Intramedullary Hip Nail Device Name:

Indications For Use:

The ODi Talon™ Intramedullary Hip Nail will be used on indications that are common with presently marketed intramedullary hip nail systems. The primary indications are for fixation/stabilization of stable and unstable fractures of the proximal femur including intertrochanteric fractures, pertrochanteric fractures, high subtrochanteric fractures, and combinations of these fractures. The device is intended to stabilize fragments of the fracture until bony union can occur.

## Contra-Indications:

The ODi Talon™ Intramedullary Hip Nail is not intended for use in patients with the following conditions:

- 1. Active local Infection.
- 2. Metal sensitivity or allergic reaction to foreign bodies.
- 3. Loss of bone stock or insufficient bone quality to support the device.
- 4. Obliterated medullary canal.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CFR 801.109)

Mark A. Millerson

Division Si n-Off) Division of General, Restorative and Neurological Devices

OR Over-The-Counter

(Optional Format 1-2-96)

510(k) Number _

5-1

---

**Source:** [https://fda.innolitics.com/device/K014189](https://fda.innolitics.com/device/K014189)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
