The Pre-Implantation Genetic Diagnosis Pipettes are used for the aspiration of blastomeres for pre-implantation genetic diagnosis. These devices are intended for one-time use and will be marketed sterile. "This tool is indicated for embryo or blastomere biopsy, which may be done in order to perform preimplantation genetic diagnosis (PIGD) on the genetic material in the biopsied cell(s). Tests for PGD are currently developed and their performance characteristics are determined by the individual laboratories for their own use. The performance of these tests may vary depending on the particular assay and disease evaluated. Currently these tests have not been cleared or approved by the Food and Drug Administration.
Device Story
Pre-Implantation Genetic Diagnosis Pipettes are sterile, single-use borosilicate glass microtools used for the aspiration of blastomeres. Operated by clinicians in assisted reproduction settings, these pipettes facilitate the biopsy of embryos or blastomeres to obtain genetic material for PIGD. The device functions as a mechanical aspiration tool; it does not perform genetic analysis itself, as PIGD assays are developed and validated by individual laboratories. By enabling the safe removal of cells for diagnostic testing, the device assists in the identification of genetic conditions prior to implantation.
Clinical Evidence
Bench testing only. Mouse Embryo Toxicity testing and Endotoxin testing were performed on the borosilicate glass material; results met all requirements.
Technological Characteristics
Borosilicate glass construction; manual aspiration microtool; sterile, single-use; no electronic components or software.
Indications for Use
Indicated for embryo or blastomere biopsy to perform preimplantation genetic diagnosis (PIGD) on genetic material in biopsied cells.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
Cook Australia Pipettes
Related Devices
K033904 — PRE-IMPLANTATION GENETIC DIAGNOSIS PIPETTES (BIOPSY PIPETTES) · Cook Urological, Inc. · Jan 16, 2004
K133257 — BIOPSY PIPETTES · Research Instruments , Ltd. · May 16, 2014
K012811 — BLASTOMERE BIOPSY MICROPIPET, POLAR BODY MICROPIPET, MODELS 10-MBB,10-MBB-B,10-MPB,10-MPB-B · Humagen Fertility Diagnostics, Inc. · Feb 11, 2002
K022643 — BLASTOMERE BIOPSY PIPETTE, MODELS H-55530 AND H-55540 · Swemed Lab International AB · Nov 5, 2002
K092554 — BLASTOMERE BIOPSY PIPETTES, POLAR BODY BIOPSY PIPETTES · Sunlight Medical, Inc. · Apr 30, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K014160
# FEB 0 8 2002
8
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS I.
Submitted By:
Debbie Schmitt Cook OB/GYN 1100 West Morgan Street Spencer, Indiana 47460 (812) 829-6500 December 14, 20001
#### Device:
Trade Name:
Pre-Implantation Genetic Diagnosis Pipettes (Biopsy Pipettes)
| Proposed Classification Name: | Class II Assisted Reproduction Microtools |
|-------------------------------|-------------------------------------------|
| | 85MQH |
| CFR Reference: | 884.6130 |
#### Predicate Devices:
Cook OB/GYN understands due to the recent reclassification there are no predicate devices. We have used Cook Australia devices as our predicate to illustrate safety and effectiveness.
The Pre-Implantation Genetic Diagnosis Pipettes (Biopsy Pipettes) are substantially equivalent to other pipettes in terms of indications for use, design, construction and material equivalence.
Specifically, these devices are similar to the Pipettes manufactured and distributed in Europe by Cook Australia, 12 Electronics Street, Brisbane Industrial Park, Eight Miles Plains, Queensland, 4113, Australia.
#### Device Description:
The Pre-Implantation Genetic Diagnosis Pipettes (Biopsy Pipettes) are used for the aspiration of blastomeres for Pre-Implantation genetic diagnosis. These devices are intended for one-time use and will be marketed sterile.
These devices are manufactured entirely from borosilicate glass. Mouse Embryo Toxicity testing and Endotoxin testing have been performed on the borosilicate glass. Results show the material meets the requirements of these tests.
#### Substantial Equivalence:
This device will be manufactured according to specified process controls and a Quality Assurance Program. Being similar with respect to indications for use, materials and physical construction to predicate devices, these devices meet the requirements for section 510(k) substantial equivalence.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a representation of the human form. The image is in black and white.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 8 2002
Ms. Debbie Schmitt Regulatory Affairs Manager COOK® Urological 1100 W. Morgan Street SPENCER IN 47460
Re: K014160 Trade/Device Name: Pre-Implantation Genetic Diagnosis Pipettes (Biopsy Pipettes) Regulation Number: 21 CFR §884.6130 Regulation Name: Assisted reproduction microtools Regulatory Class: II Product Code: 85 MQH Dated: December 14, 2001 Received: December 19, 2001
Dear Ms. Schmitt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "COOK" in a bold, sans-serif font. A circled R symbol is present to the right of the letter K, indicating a registered trademark. The text is black against a white background.
Cook Ob/Gyn 1100 W. Morgan Street Spencer, IN 47460 USA Phone: 812-829-6500 Fax: 812-829-1801 www.cookgroup.com
### PREMARKET NOTIFICATION
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K014160
Device Name:
Pre-Implantation Genctic Diagnosis Pipettes (Biopsy Pipettes)
The Pre-Implantation Genetic Diagnosis Pipettes are used for the aspiration of blastomeres for pre-implantation genetic diagnosis. These devices are intended for one-time use and will be marketed sterile.
"This tool is indicated for embryo or blastomere biopsy, which may be done in order to perform preimplantation genetic diagnosis (PIGD) on the genetic material in the biopsied cell(s). Tests for PGD are currently developed and their performance characteristics are determined by the individual laboratories for their own use. The performance of these tests may vary depending on the particular assay and disease evaluated. Currently these tests have not been cleared or approved by the Food and Drug Administration.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Usc
(Per 21 CFR 801.109)
OR
Over-The-Counter Use __
Nancy bradlen
Division of Reproductive, Abdominal,
and Radiological Devices K014160
510(k) Numb
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.