← Product Code [MOS](/productcode/MOS) · K014086

# CTL BODY / SPINE ARRAY COIL (K014086)

_Usa Instruments, Inc. · MOS · Feb 5, 2002 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K014086

## Device Facts

- **Applicant:** Usa Instruments, Inc.
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Feb 5, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The CTL Body / Spine Array Coil is a receive-only phased array RF coil, used for obtaining diagnostic images of the spine region (cervical, thoracic and lumbar anatomy) in Magnetic Resonance Imaging Systems. The indications for use are the same as for standard MR Imaging. The CTL Body / Spine Coil is designed for use with Array the Rhapsody™1.0T MRI scanner manufactured by Siemens Medical Systems. The Siemens Rhapsody™ 1.0Tesla system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images vield information that can be useful in the determination of a diagnosis.

## Device Story

CTL Body/Spine Array Coil is a multi-element, receive-only phased array RF coil; designed for use with Siemens Rhapsody 1.0T MRI scanner. Device conforms to spine contours; includes removable anterior section for cervical imaging and three bridges with coil elements for thoracic/lumbar imaging. Operates by receiving RF signals from hydrogen nuclei; does not transmit RF power. Used in clinical MRI settings by trained technicians/physicians. Decoupling achieved via RF chokes and switching diodes to isolate elements during RF transmission. Non-conductive housing prevents RF burns. Output is raw signal data processed by the MRI system into diagnostic images; interpreted by physicians to aid diagnosis.

## Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness established through comparison of design, materials, and functional characteristics (decoupling, RF absorption, resonant loop prevention) to predicate devices.

## Technological Characteristics

Receive-only phased array RF coil. Materials: Polyurethane and polycarbonate plastic housing. Sensing: Multi-element coil array. Decoupling: RF chokes with switching diodes. Connectivity: Dedicated interface for Siemens Rhapsody 1.0T MRI scanner. Form factor: Contoured spine housing with removable anterior section and three element bridges.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Premier 7000 C/T/L Spine Coil ([K980157](/device/K980157.md))
- Magna 5000 CTL Spine Coil ([K994345](/device/K994345.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

## FEB 0 5 2002

## SUMMARY OF SAFETY AND EFFECTIVENESS

Kol4086

- 1. Device Name :
3. Classification :

Magnetic Resonance Imaging Accessory

CTL Body / Spine Array Coil

2. Proprietary Name :

Class II

4. Establishment Registration #:

5. Manufacture Facility Location:

- 6. Performance Standard:
7. Intended Use:

- 1529041 USA Instruments, Inc 1515 Danner Drive
Aurora, Ohio 44202, USA Telephone: 330-562-1000; Fax: 330-562-1422.

No applicable performance standards have been issued under Section 514 of the Food, Drug and Cosmetic Act.

The CTL Body / Spine Array Coil is a receive-only phased array RF coil, used for obtaining diagnostic images of the spine region (cervical, thoracic and lumbar anatomy) in Magnetic Resonance Imaging Systems. The indications for use are the same as for standard MR Imaging. The CTL Body / Spine Coil is designed for use with Array the Rhapsody™1.0T MRI scanner manufactured by Siemens Medical Systems.

- 8. Device Description: The CTL Body Spine Array Coil is a multi-element phased array receive-only coil. The coil is shaped to conform to the contours of the spine and has a removable anterior section for imaging the anterior cervical region. The coil also has three bridges consisting of coil elements for enhanced imaging of the thoracic and lumbar spine. The elements and associated circuitry are enclosed in a housing made of plastic materials, which are fire rated and have high impact and tensile strength.

{2}------------------------------------------------

## 9. Safety and Effectiveness

.....

| CTL Body / Spine Array Coil Product<br>Features                                                                                                                                                                        | Comparison to predicate device or other 510(k)<br>cleared product                                                                                                                               |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: Imaging of the spine.                                                                                                                                                                                    | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Indications for Use: Identical to routine<br>MRI imaging                                                                                                                                                               | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Coil Material:<br>Polyurethane Plastic<br>Polycarbonate Plastic                                                                                                                                                        | -Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345)                                                                                                    |
| Coil Design: Receive-only phased array<br>design                                                                                                                                                                       | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Decoupling: RF Chokes with Switching<br>Diodes                                                                                                                                                                         | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Prevention of RF Burns: Does not<br>transmit RF Power, Decoupling isolates<br>the coil elements from RF fields during<br>RF transmission, Coil elements and<br>circuitry are enclosed in a non-<br>conductive housing. | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Radio Frequency Absorption: Coil is a<br>receive only coil and does not transmit<br>RF power                                                                                                                           | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |
| Formation of Resonant Loops:<br>Decoupling isolates coil elements from<br>RF fields during RF transmission. Length<br>of cable and stiffness<br>does not permit looping                                                | -Similar to the Premier 7000 C/T/L Spine Coil<br>manufacture by USA Instruments, Inc. (K980157)<br>-Similar to the Magna 5000 CTL Spine Coil<br>manufactured by USA Instruments, Inc. (K994345) |

{3}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three overlapping human profiles, which are meant to represent the department's mission of protecting the health of all Americans.

## Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 0 5 2002

Mr. Rony Thomas Vice President USA Instruments, Inc. 1515 Danner Drive AURORA OH 44202

Re: K014086

Trade/Device Name: CTL Body/Spine Array Coil MRI Specialty Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: December 7, 2001 Received: December 11, 2001

Dear Mr. Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known):____________________________________________________________________________________________________________________________________________________

Device Name: CTL Body / Spine Array Coil

Indications for Use: The CTL Body / Spine Array Coil is designed to provide Magnetic Resonance Images of the spine anatomy. The CTL Body / Spine Array Coil is designed for use with the Siemens Rhapsody™ 1.0Tesla scanner.

> Anatomic Regions: Spine Nuclei Excited: Hydrogen

The indications for use are the same as for standard imaging:

The Siemens Rhapsody™ 1.0Tesla system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images vield information that can be useful in the determination of a diagnosis.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

|  | Concurrence of CDRH, Office of Device Evaluation (ODE)        |         |
|--|---------------------------------------------------------------|---------|
|  | <img alt="signature" src="signature.png"/>                    |         |
|  | (Division Sign-Off)                                           |         |
|  | Division of Reproductive, Abdominal, and Radiological Devices |         |
|  | 510(k) Number                                                 | K014086 |

| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use<br>(Optional Format 1-2-96) |
|------------------------------------------|----|--------------------------------------------------|
|------------------------------------------|----|--------------------------------------------------|

---

**Source:** [https://fda.innolitics.com/device/K014086](https://fda.innolitics.com/device/K014086)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).