FAMILY OF ALTUS MEDICAL MODIFIED COOLGLIDE AESTHETIC LASERS, MODELS COOLGLIDE I, II AND FUTURE MODELS
K014040 · Altus Medical, Inc. · GEX · Jun 3, 2002 · General, Plastic Surgery
Device Facts
Record ID
K014040
Device Name
FAMILY OF ALTUS MEDICAL MODIFIED COOLGLIDE AESTHETIC LASERS, MODELS COOLGLIDE I, II AND FUTURE MODELS
Applicant
Altus Medical, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 3, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The family of Modified Altus Medical CoolGlide Aesthetic Lasers are intended for use in the medical specialties of general and plastic surgery, dermatology, endoscopic/laproscopic general surgery, gastroenterology, gynecology, otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, pulmonary/thoracic surgery and urology for surgical and aesthetic applications. 1064nm: Dermatology: The Altus Medical Aesthetic Modified CoolGlide laser systems are intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The lasers are also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques. Additionally, the lasers are indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The Modified CoolGlide lasers are also indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment for pseudofolliculitis barbae (PFB). The lasers are also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. The Modified CoolGlide lasers are indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin. The intended use of the integral cooling system in the Altus Modified CoolGlide laser handpiece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesions, and to reduce the potential side effects of laser treatments. Surgical Applications: The lasers are indicated for the incision/excision and cutting, ablation, coagulation/hemostatis of soft tissue in the performance of surgical applications in endosopy/laproscopy, gastroenterology, general surgery, a head and neck/otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, plastic surgery, pulmonary/thoracic surgery, gynecology (e.g. menorrhagia) and urology. For coagulation and hemostasis of vascular and cutaneous lesions in dermatology including, but not limited to, the following general categories: vascular lesions [angiomas, hemangiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions (nevi, lentigines, chloasma, café-au-lait, tattoos (red and green ink); verrucae; skin tags; keratoses; plaques; cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
Device Story
Laser system console with control electronics/software; fiberoptic-coupled handpiece with integrated skin cooling; footswitch/handswitch; remote interlock. Used in clinical settings by physicians. Laser energy targets melanin or vascular structures for coagulation, hemostasis, ablation, or hair reduction. Integrated cooling reduces pain/side effects and allows higher fluences. Output is laser energy delivery to tissue; provider monitors treatment visually. Benefits include precise tissue management and aesthetic improvement.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological identity to predicate devices and established indications for use.
Technological Characteristics
Nd:YAG laser system; 1064nm wavelength; fiberoptic-coupled handpiece; integrated skin cooling system; control console with software/electronics; remote interlock; footswitch/handswitch operation.
Indications for Use
Indicated for patients requiring surgical or aesthetic laser treatment across various specialties including dermatology, general/plastic surgery, ENT, urology, and gynecology. Treats benign vascular/pigmented lesions, unwanted hair (all Fitzpatrick skin types I-VI), hypertrophic/keloid scars, and soft tissue surgical applications. Contraindications not specified.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
K 014040
# JUN 0 3 2002
### Attachment 17 510(k) Summary for the Family of Altus Medical Modified CoolGlide Aesthetic Lasers
#### I. General Information
| Submitter: | Altus Medical, Inc.<br>821 Cowan Road<br>Burlingame, CA 94010 |
|-----------------|---------------------------------------------------------------|
| Contact Person: | Kathy Maynor |
Summary Preparation Date: December 1, 2001
#### II. Names
| Device Names: | Family of Altus Medical Modified CoolGlide Aesthetic<br>Lasers |
|------------------------------|----------------------------------------------------------------|
| Primary Classification Name: | Laser Powered Surgical Instrument (and Accessories) |
#### Predicate Devices III.
- Altus Medical Aesthetic Nd: YAG Laser (K991798, K991234 and K003202); ●
- . VeinLase, manufactured by HGM (K981952);
- . EpiLight PhotoDerm HR, manufactured by ESC Medical Systems (K991935 and K980537);
- . LightSheer Pulsed Diode Array Laser System, manufactured by Star/Coherent Medical (K982940);
- . Lyra and Orion/SL/Lyra Series Laser Systems and SmartScan, manufactured by Laserscope (K990718, K990903, K941841, K933880, K003147 and K003765); and
- . Candela Long Pulse Nd:YAG Laser System, manufactured by Candela (K010104).
#### IV. Product Description
Family of Altus Medical Modified CoolGlide Aesthetic Lasers are comprised of the following main components:
- · a laser system console (including software and control electronics);
- · a control and display panel;
- · a permanently attached fiberoptic-coupled handpiece;
- · a skin cooling device integrated into the handpiece;
- · a footswitch (or finger-operated exposure switch (handswitch) option integrated into the handpiece)
- · a remote interlock connector (disables laser when treatment room door is opened).
{1}------------------------------------------------
#### V. Indications for Use
The family of Modified Altus Medical CoolGlide Aesthetic Lasers are intended for use in the medical specialties of general and plastic surgery, dermatology, endoscopic/laproscopic pulmonary/thoracic surgery and urology for surgical and aesthetic applications.
## 1064nm:
# Dermatology:
The Altus Medical Aesthetic CoolGlide laser systems are intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The lasers are also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos(significant reduction in the intensity of black and/or blue/black tattoos) and plaques. Additionally, the lasers are indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The CoolGlide lasers are also indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment for pseudofolliculitis barbae (PFB).
The lasers are also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
The CoolGlide lasers are indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
The intended use of the integral cooling system in the Altus handpiece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesions, and to reduce the potential side effects of laser treatments.
# Surgical Applications ::
The lasers are indicated for the incision/excision and cutting, ablation. coagulation/hemostatis of soft tissue in the performance of surgical applications in endosopy/laproscopy, gastroenterology, general surgery, head and neck/otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, plastic surgery, pulmonary/thoracic surgery, gynecology (e.g. menorrhagia) and urology.
{2}------------------------------------------------
### Indications for use (532nm):
For coagulation and hemostasis of vascular and cutaneous lesions in dermatology including, but not limited to, the following general categories: vascular lesions [angiomas, hemangiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions (nevi, lentigines, chloasma, café-au-lait, tattoos (red and green ink); verrucae; skin tags; keratoses; plaques; cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
#### Rationale for Substantial Equivalence VI.
The family of Altus Medical Modified CoolGlide Aesthetic Lasers share the same general indications for use, and therefore is substantially equivalent to the currently marketed Altus Medical Aesthetic Nd:YAG Laser (K991798, K991234 and K003202), the VeinLase, manufactured by HGM (K981952), the EpiLight PhotoDerm HR, manufactured by ESC Medical Systems (K991935 and K980537), the LightSheer Pulsed Diode Array Laser System, manufactured by Star/Coherent Medical (K982940), the Lyra and Orion/SL/Lyra Series Laser Systems and SmartScan. manufactured by Laserscope (K990718, K990903, K941841, K933880, K003147 and K003765), and the Candela Long Pulse Nd: YAG Laser System manufactured by Candela (K010104).
#### VII. Safety and Effectiveness Information
The new indications for use in dermatology, endoscopic/laproscopic general surgery, gastroenterology, general surgery, gynecology, otorhinolaryngology, neurosurgery, oculoplastics, orthopedics, plastic surgery, pulmonary.thoracic surgery, and urology are based upon the indications for use for predicate laser systems.
Technologically, the modified CoolGlide family of aesthetic lasers is identical to the previous predicate CoolGlide family (K003202). Therefore the risks and benefits for the modified CoolGlide laser family are comparable to the predicate devices.
We therefore believe that there are no questions of safety or effectiveness raised by the introduction of this device.
### VIII. Conclusion
،
The family of Altus Medical Modified CoolGlide Aesthetic Lasers were found to be substantially equivalent to the currently marketed Altus Medical Aesthetic Nd: YAG Laser (K991798, K991234 and K003202), the VeinLase, manufactured by HGM (K981952), the EpiLight PhotoDerm HR, manufactured by ESC Medical Systems (K991935 and K980537), the LightSheer Pulsed Diode Array Laser System, manufactured by Star/Coherent Medical (K982940), the Lyra and Orion/SL/Lyra Series Laser Systems and SmartScan, manufactured by Laserscope (K990718, K990903, K941841, K933880, K003147 and K003765) and the Candela Long Pulse Nd:YAG Laser System (K010104). The family of Altus Medical Modified CoolGlide Aesthetic Lasers share similar indications for use, design features, and similar functional features as, and thus are substantially equivalent to, the currently marketed predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
JUN 0 3 2002
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Kathy Maynor Vice President of Regulatory/Quality Altus Medical, Inc. 821 Cowan Road Burlingame, CA 94010
Re: K014040
Trade/Device Name: Family of Altus Medical Modified CoolGlide Aesthetic Lasers Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and
plastic surgery and in dermatology
Regulatory Class: II Product Code: GEX Dated: March 6, 2002 Received: March 8, 2002
Dear Ms. Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include reguirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 – Ms. Kathy Maynor
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
A Hupka Pluralis
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Attachment 2 Indications For Use Statement as Requested by FDA
510(k) Number (if Known): K014040
## Device Name: Family of Altus Medical Modified CoolGlide Aesthetic Lasers
Indications For Use:
The family of Modified Altus Medical CoolGlide Aesthetic Lasers are intended for use in the medical specialties of general and plastic surgery, dermatology, endoscopic/laproscopic general surgery, gastroenterology, gynecology, otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, pulmonary/thoracic surgery and urology for surgical and aesthetic applications.
### 1064nm:
### Dermatology:
The Altus Medical Aesthetic Modified CoolGlide laser systems are intended for the coagulation and hemostasis of benign vascular lesions such as, but not limited to, port wine stains, hemangiomas, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. The lasers are also intended for the treatment of benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), café au lait macules, seborrheic keratoses, nevi, chloasma, verrucae, skin tags, keratoses, tattoos (significant reduction in the intensity of black and/or blue/black tattoos) and plaques.
Additionally, the lasers are indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative and Neurological Devices
| Prescription Use (Per 21 CFR 801.109) | <table><tr><td>510(k) Number</td><td>K014040</td></tr><tr><td></td><td>OR Over-The-Counter Use</td></tr></table> | 510(k) Number | K014040 | | OR Over-The-Counter Use |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------|---------|--|-------------------------|
| 510(k) Number | K014040 | | | | |
| | OR Over-The-Counter Use | | | | |
(Optional Format 1-2-96)
{6}------------------------------------------------
The Modified CoolGlide lasers are also indicated for the removal of unwanted hair, for the stable long term, or permanent, hair reduction through selective targeting of melanin in hair follicles, and for the treatment for pseudofolliculitis barbae (PFB).
ઠાપ્ર 040
Dase 2 st
The lasers are also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
The Modified CoolGlide lasers are indicated for use on all skin types (Fitzpatrick I-VI), including tanned skin.
The intended use of the integral cooling system in the Altus Modified CoolGlide laser handpiece is to provide cooling of the skin prior to laser treatment, for the reduction of pain during laser treatment, to allow for the use of higher fluences for laser treatments such as hair removal and vascular lesions, and to reduce the potential side effects of laser treatments.
### Surgical Applications ::
The lasers are indicated for the incision/excision and cutting, ablation, coagulation/hemostatis of soft tissue in the performance of surgical applications in endosopy/laproscopy. gastroenterology, general surgery, a head and neck/otorhinolaryngology (ENT), neurosurgery, oculoplastics, orthopedics, plastic surgery, pulmonary/thoracic surgery, gynecology (e.g. menorrhagia) and urology.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div align="center">(Division Sign-Off)</div> |
|--|-----------------------------------------------|
|--|-----------------------------------------------|
Division of General, Restorative and Neurological Devices
| 510(k) Number | K014040 |
|---------------------------------------|---------|
| Prescription Use (Per 21 CFR 801.109) | X |
OR Over-The-Counter Use ________
(Optional Format 1-2-96
{7}------------------------------------------------
For coagulation and hemostasis of vascular and cutaneous lesions in dermatology including, but not limited to, the following general categories: vascular lesions [angiomas, hemangiomas (port wine), telangiectasia (facial or extremities telangiectasias, venous anomalies, leg veins)]; benign pigmented lesions (nevi, lentigines, chloasma, café-au-lait, tattoos (red and green ink); verrucae; skin tags; keratoses; plaques; cutaneous lesion treatment (hemostasis, color lightening, blanching, flattening, reduction of lesion size).
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
OR
Prescription Use (Per 21 CFR 801.109)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Hyt. Plivolu
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K014840
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.