SHARPOINT COATED, FAST ABSORBING POLYGLYCOLIC ACID SYNTHETIC SUTURE, UNDYED
K014021 · Surgical Specialties Corp · GAM · Jan 28, 2002 · General, Plastic Surgery
Device Facts
Record ID
K014021
Device Name
SHARPOINT COATED, FAST ABSORBING POLYGLYCOLIC ACID SYNTHETIC SUTURE, UNDYED
Applicant
Surgical Specialties Corp
Product Code
GAM · General, Plastic Surgery
Decision Date
Jan 28, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Coated FAST ABSORBING PolyGlycolic Acid suture is indicated only for use in superficial general soft tissue approximation of the skin and mucosa where only short-term wound support (7-10 days) is required. Coated FAST ABSORBING PolyGlycolic Acid suture is not intended for use in ligation, ophthalmic, cardiovascular or neurological procedures.
Device Story
Synthetic absorbable braided surgical suture composed of polyglycolic acid homopolymer; coated with polycaprolactone and polyglycolic acid copolymer. Used by surgeons for superficial soft tissue approximation. Provides short-term wound support (7-10 days) before absorption. Noncollagenous and nonantigenic materials. Device is sterile and flexible.
Clinical Evidence
No clinical data. Evidence based on bench testing (USP 24 physical testing for diameter, needle attachment, and tensile strength) and animal implant studies to evaluate biocompatibility per ISO 10993 and tensile strength/mass loss rates.
Technological Characteristics
100% polyglycolic acid braided multifilament suture; coated with polycaprolactone and polyglycolic acid copolymer. Sterile, flexible form factor. Performance testing per USP 24 and ISO 10993.
Indications for Use
Indicated for superficial soft tissue approximation of skin and mucosa requiring short-term (7-10 days) wound support. Contraindicated for ligation, ophthalmic, cardiovascular, or neurological procedures.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Ethicon, Inc. Coated VICRYL RAPIDE (polyglactin 910) braided synthetic absorbable, suture, undyed (K944110)
Submission Summary (Full Text)
{0}------------------------------------------------
## JAN 2 8 2002
## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS SUBSTANTIAL EQUIVALENCY
Surgical Specialties Corporation Submitter: Address: 100 Dennis Drive Reading, PA 19606 610 404 1000, ext. 2231 Telephone: Elizabeth Lazaro Contact Person: Regulatory Affairs Specialist December 3, 2001 Date Prepared: Sharpoint® Coated FAST ABSORBINGTM Name of Device: PolyGlycolic Acid braided, synthetic absorbable suture, undyed. Common / Usual Suture, Absorbable, Synthetic, PolyGlycolic Acid Classification Name: Predicate Device: Ethicon, Inc. Coated VICRYL RAPIDE (polyglactin 910) braided synthetic absorbable, suture, undyed. Coated FAST ABSORBING PolyGlycolic Acid Device Description Suture is a synthetic absorbable braided, sterile, surgical suture composed of a homopolymer of Glycolic Acid. The suture material is coated with a copolymer of Polycaprolactone and PolyGlycolic Acid. The substances contained in the coating and suture are noncollagenous and nonantigenic.
Continue on next page
Coated FAST ABSORBING PolyGlycolic Acid suture, undyed Surgical Specialties Corporation
/
{1}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (continued)
| Intended Use: | Coated FAST ABSORBING PolyGlycolic Acid<br>suture is indicated only for use in superficial general<br>soft tissue approximation of the skin and mucosa<br>where only short-term wound support (7-10 days)<br>is required. |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Coated FAST ABSORBING PolyGlycolic Acid<br>suture is not intended for use in ligation,<br>ophthalmic, cardiovascular or neurological<br>procedures. |
| | The coated FAST ABSORBING PolyGlycolic Acid<br>suture, undyed has the same intended use as the<br>predicate device. |
| Technological Characteristics: | Coated FAST ABSORBING PolyGlycolic Acid<br>suture, undyed is composed of 100% PolyGlycolic<br>Acid, a material equivalent to the material<br>comprising the predicate VICRYL Rapide device.<br>The suture has the same design as the VICRYL<br>Rapide predicate devices, being a sterile, flexible<br>suture available in a coated braided multifilament<br>form. |
| Performance Data: | Physical testing was conducted on the Coated<br>FAST ABSORBING PolyGlycolic Acid suture to<br>USP 24, including <861> Sutures - Diameter,<br><871> Sutures- Needle Attachment, and <881><br>Tensile Strength. Animal testing was performed for<br>conformance to ISO 10993 for biocompatibility and<br>implant studies in animals to demonstrate rates of<br>tensile strength and mass loss. |
Continue on next page
K014021
page 2/3
Coated FAST ABSORBING PolyGlycolic Acid suture, undyed Surgical Specialties Corporation
{2}------------------------------------------------
K014021
page 3/3
9
The FAST ABSORBING PolyGlycolic Acid Synthetic Absorbable surgical suture is similar in intended use, materials, design, and performance characteristics to the Ethicon VICRYL Rapide Synthetic Absorbable Surgical Suture (K944110).
The determination of substantial equivalence for this device was based on a detailed device description, physical testing (U.S.P. 24) and animal testing for conformance with performance standards e.g. ISO 10993-1 Biological Evaluation of Medical Devices and the FDA Guidance Document for Surgical Suture 510(k)s issued on August 10, 2000.
Based on the 510(k) summaries (21 CFR 807) and the information provided herein, we conclude that the new device is substantially equivalent to the Predicate Device under the Federal Food, Drug, and Cosmetic Act.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 8 2002
Ms. Elizabeth Lazaro Regulatory Affairs Specialist Surgical Specialties Corporation 100 Dennis Drive Reading, Pennsylvania 19606
Re: K014021
Trade/Device Name: Sharpoint® Coated FAST ABSORBING™ Polyglycolic Acid Suture Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(glycolide/L-lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: December 05, 2001 Received: December 06, 2001
Dear Ms. Lazaro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Elizabeth Lazaro
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Miriam C. Provost
Of Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Page _1_ of _1__
6
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Sharpoint® Coated FAST ABSORBING™ PolyGlycolic Acid Device Name: Synthetic
Absorbable Surgical Suture
Indications for Use:
Sharpoint® Coated Fast Absorbing™ PolyGlycolic Acid Suture is indicated only for use in superficial soft tissue approximation of the skin and mucosa where only short term wound support (7 – 10 days) is required. Coated Fast Absorbing™ PolyGlycolic Acid Suture is not intended for use in ligation, ophthalmic cardiovascular or neurological procedures.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Muriam C. Provost
(Division Sign-Off) Division of Coneral, Rostorative and Neurological Devices
510(k) Number KO1 402
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.