IMMULITE BR-MA, MODEL #LKBR; IMMULITE 2000 BR-MA, MODEL #L2KBR
K013984 · Diagnostic Products Corp. · MOI · Mar 28, 2002 · Immunology
Device Facts
Record ID
K013984
Device Name
IMMULITE BR-MA, MODEL #LKBR; IMMULITE 2000 BR-MA, MODEL #L2KBR
Applicant
Diagnostic Products Corp.
Product Code
MOI · Immunology
Decision Date
Mar 28, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Indications for Use
IMMULITE BR-MA and IMMULITE 2000 BR-MA are clinical use devices intended for in vitro diagnostic use with IMMULITE analyzer - for the quantitative measurement of CA 15-3 antigen in human serum, as an aid in the detection of recurrence in previously treated stage II and stage III breast cancer patients, and in the management of metastatic breast cancer patients by monitoring disease progression or response to treatment. Serial testing for patient CA 15-3 values should be used in conjunction with other clinical methods used for detecting early recurrence in Stage II and stage IIII disease and for monitoring response to treatment in patients with metastatic breast cancer.
Device Story
IMMULITE BR-MA and IMMULITE 2000 BR-MA are automated chemiluminescent enzyme immunoassays for quantitative measurement of CA 15-3 antigen in human serum. Used in clinical laboratories, these devices utilize a two-step sequential immunometric assay format. Patient samples and ligand-labeled anti-CA 15-3 monoclonal antibodies are incubated with anti-ligand-coated solid phase; unbound serum is removed via centrifugal wash. Alkaline phosphatase-labeled anti-CA 15-3 monoclonal antibody is added, followed by another incubation and wash. Addition of phosphate ester of adamantyl dioxetane substrate results in light emission proportional to CA 15-3 concentration, measured by a luminometer. Results assist clinicians in monitoring breast cancer recurrence and treatment response in conjunction with other clinical methods.
Clinical Evidence
Clinical studies evaluated 1400 specimens (IMMULITE BR-MA) and 500 specimens (IMMULITE 2000 BR-MA) from healthy subjects and breast cancer patients. Performance compared to commercial kits showed 87.8%-91% agreement. Longitudinal monitoring of 99 (BR-MA) and 35 (2000 BR-MA) patients followed for recurrence showed 81% and 80% accuracy in reflecting clinical status, respectively. Sensitivity for recurrence detection ranged 72-75%; specificity 76-84%. Monitoring response to treatment showed 66-80% accuracy.
Technological Characteristics
Two-step sequential chemiluminescent enzyme-labeled immunometric assay. Solid phase: anti-ligand-coated. Detection: alkaline phosphatase-labeled monoclonal antibody; phosphate ester of adamantyl dioxetane substrate. Analyzers: IMMULITE and IMMULITE 2000. Calibration range: up to 300 U/mL. Analytical sensitivity: 0.2 U/mL. No significant interference from bilirubin, lipemia, or hemolysis. High dose hook effect: none up to 80,000 U/mL.
Indications for Use
Indicated for quantitative measurement of CA 15-3 antigen in human serum to aid in detecting recurrence in previously treated stage II/III breast cancer patients and managing metastatic breast cancer by monitoring disease progression or treatment response.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
{1}------------------------------------------------
K013984
# 510 (k) Summary of Safety and Effectiveness IMMULITE® and IMMULITE® 2000 BR-MA
This summary of 510(k) safety and effectiveness information is being submitted in Inis sammal y of broke visements of SMDA 1990 and 21 CFR Part 807.92.
| Name:<br>Address: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, California 90045 | |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Telephone Number: | (310) 645-8200 | |
| Contact Person | Edward M. Levine, Ph.D. | |
| Date of Preparation: | March 22, 2002 | |
| Catalog Number: | LKBR1, LKBR5 (100, 500 tests)<br>L2KBR2 (200 tests) | |
| Device Name<br>Trade: | IMMULITE® BR-MA and<br>IMMULITE® 2000 BR-MA | |
| Common: | Reagent system for the determination of CA15-3 antigen in<br>serum. | |
| Classification: | MOI, Class II device | |
| Manufacturer of<br>IMMULITE BR-MA: | Euro/DPC Limited<br>Glyn Rhonwy<br>Llanberis, Gwynedd LL55 4EL<br>United Kingdom<br>(Manufactured under a Quality System-<br>ISO9002/ISO13488/EN46002) | |
| Sole U. S. Importer of<br>IMMULITE BR-MA: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, CA 90045-5597 | |
| Manufacturer of<br>IMMULITE 2000 BR-MA: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, CA 90045-5597<br>(Manufactured under a Quality System-<br>ISO9001/21CFR,Part 820/EN46001) | |
1
{2}------------------------------------------------
### Establishment Registration #: Euro/DPC - Not applicable DPC Registration number is 2017183
Substantially Equivalent Predicate Devices:
Baver Immuno 17M CA 15-3TM (K964703) Chiron ACS 180 BR (CA27.29)
Description of Device:
IMMULITE BR-MA and IMMULITE 2000 BR-MA are clinical use devices intended for use with their respective IMMULITE and IMMULITE 2000 Automated Immunoassay Analyzers for the quantitative measurement of CA15-3 antigen in serum.
#### Intended Use of the Device:
IMMULITE BR-MA and IMMULITE 2000 BR-MA are clinical use devices intended for in vitro diagnostic use with IMMULITE analyzer - for the quantitative measurement of CA 15-3 antigen in human serum, as an aid in the detection of recurrence in previously treated stage II and stage III breast cancer patients, and in the management of metastatic breast cancer patients by monitoring disease progression or response to treatment. Serial testing for patient CA 15-3 values should be used in conjunction with other clinical methods used for detecting early recurrence in Stage II and stage IIII disease and for monitoring response to treatment in patients with metastatic breast cancer.
#### Summary and Explanation of the Test:
CA 15-3 is a high molecular weight (300 to 450 kDa) polymorphic epithelial mucin. The heterogeneous breast cancer associated mucins consist of a repeated polypeptide core sequence and an outer shell of carbohydrate. Serum CA 15-3 values increase with clinical stage of breast cancer, the highest values occurring in metastatic disease. Serial determinations of CA 15-3 are most useful as an indicator of response to therapy. The measurement of CA 15-3 antigen is more sensitive and specific than the determination of carcinoembryonic antigen (CEA), having a lower percentage positivity with benign breast lesions, liver cirrhosis and other carcinomas -- 99.9% of serum donated by healthy volunteers contained less than 40 U/mL of CA 15-3 is not elevated during pregnancy. The percentage of raised values found in breast cancer can be as high as 98%, but this depends primarily on the tumor stage of the patient population studied. Elevated levels have also been found in patients with lung cancer (63%) and ovarian cancer (80%).
IMMULITE BR-MA and IMMULITE 2000 BR-MA are two-step sequential chemiluminescent enzyme immunoassays, based on ligand-labeled monoclonal antibody and separation by anti-ligand-coated solid phase.
{3}------------------------------------------------
### Performance Equivalence - Technology Comparison:
IMMULITE BR-MA is a chemiluminescent enzyme immunoassay and Immuno 1 CA 15-3 is a magnetic separation assay. The technology in DPC's IMMULITE BR-MA and IMMULITE 2000 BR-MA is identical to technology used in previously cleared and commercially marketed IMMULITE and IMMULITE 2000 products.
IMMULITE BR-MA is a two-step sequential chemiluminescent enzyme-labeled immunometric assay, based on ligand-labeled monoclonal antibody and separation by anti-ligand-coated solid phase. The patient sample and a ligand-labeled anti-CA15-3 monoclonal antibody are simultaneously introduced into the Test Unit, containing immobilized anti-ligand, and incubated for approximately 30 minutes at 37℃ with intermittent agitation. During this time, CA 15-3 antigen in the sample binds to the ligand-labeled monoclonal antibody, which, in turn, binds to the anti-ligand on the solid phase. Unbound serum is then removed by centrifugal wash. An alkaline phosphataselabeled anti-CA 15-3 monoclonal antibody is introduced, and the Test Unit is incubated for another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Test Unit is incubated for a further 10 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex and thus also the photon output, as measured by the luminometer - is proportional to the concentration of CA 15-3 in the sample.
IMMULITE 2000 BR-MA is a two-step sequential chemiluminescent enzyme-labeled immunometric assay, based on ligand-labeled monoclonal antibody and separation by anti-ligand-coated solid phase. The patient sample and a ligand-labeled anti-CA15-3 monoclonal antibody are simultaneously introduced into the Reaction Tube, containing immobilized anti-ligand, and incubated for approximately 30 minutes at 37°C with intermittent agitation. During this time, CA 15-3 antigen in the sample binds to the ligand-labeled monoclonal antibody, which, in turn, binds to the anti-ligand on the solid phase. Unbound serum is then removed by centrifugal wash. An alkaline phosphataselabeled anti-CA 15-3 monoclonal antibody is introduced, and the Reaction Tube is incubated for another 30-minute cycle. The unbound enzyme conjugate is removed by a centrifugal wash. Substrate is then added, and the Test Unit is incubated for a further 5 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to vield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex and thus also the photon output, as measured by the luminometer - is proportional to the concentration of CA 15-3 in the sample.
{4}------------------------------------------------
The Bayer Immuno 1 CA 15-3 Assay uses a sandwich immunoassay format. 115D8 Antibody Conjugate (R1) and mIMP® (monoclonal ImmunoMagnetic Particle) Reagent are reacted with patient sample (or calibrator containing CA 15-3 reactive determinants) and incubated on the system at 37°C. The DF3 Enzyme Conjugate (R2) is then added. A second incubation occurs during which the antibody complex is bound. After incubation, the complex is washed and the pNPP (para-nitrophenyl phosphate) substrate is added. The alkaline phosphatase (ALP) in the antibody conjugate reacts with the pNPP to form para-nitrophenoxide and phosphate. Increasing absorbance, due to the formation of paranitrophenoxide, is monitored at 405 nm and 450 nm.
A sample have no CA 15-3 Assay value will have the minimum label bound, while a sample having a high CA 15-3 Assay value will have maximum label bound. Thus, the dose response curve is proportional to the DF3 reactive determinants in the sample.
{5}------------------------------------------------
## Performance Equivalence - Performance Characteristics:
| Characteristics | IMMULITE BR-<br>MA | IMMULITE 2000<br>BR-MA | Immuno 1<br>CA 15-3 |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Calibration upper limit | 300 U/mL | 300 U/mL | 200 U/mL |
| Analytical Sensitivity | 0.2 U/mL | 0.2 U/mL | 0.2 U/mL |
| Precision (within-run) | CV range: 5.5-8.1% | CV range: 3.7-4.9% | CV range: 1.3-3.4% |
| Precision (total) | CV range: 7.4-8.8% | CV range: 4.5-6.5% | CV range: 3.0-4.0% |
| Specificity | No significant cross-<br>reactivity with the<br>following compounds<br>5-Fluorouracil, AFP,<br>CA 125, CA 19-9,<br>CEA, Cisplatin,<br>Cyclophosphamide,<br>Doxorubicin,<br>hydrochloride,<br>Mitomycin C,<br>Vincristine | No significant cross-<br>reactivity with the<br>following compounds<br>5-Fluorouracil, AFP,<br>CA 125, CA 19-9,<br>CEA, Cisplatin,<br>Cyclophosphamide,<br>Doxorubicin,<br>hydrochloride,<br>Mitomycin C,<br>Vincristine | Data not available |
| Linearity | %Observed/Expected<br>range: 95-105% | %Observed/Expected<br>range: 93-104% | Data not available |
| Spiking Recovery | %Observed/Expected<br>range:<br>89-114% | %Observed/Expected<br>range:<br>92-108% | Data not available |
| Effect of Bilirubin<br>(unconjugated) | No significant effect | No significant effect | No clinically significant<br>effect |
| Effect of Lipemia | No significant effect | No significant effect | No clinically significant<br>effect |
| Effect of Hemolysis | No significant effect | No significant effect | No clinically significant<br>effect |
| High Dose Hook Effect | No effect up to 80,000<br>U/mL | No effect up to 80,000<br>U/mL | Data not available |
| Stability | Unopened kits kept<br>at specified<br>conditions are stable<br>for 2 months. Kits<br>are stable for one<br>year when kept in<br>long-term storage | Unopened kits kept at<br>specified conditions<br>are stable for 2<br>months. Kits are<br>stable for one year<br>when kept in long-<br>term storage<br>conditions. | Unopened kits kept at<br>specified conditions are<br>stable through the last<br>day of the month on the<br>product label. Long-<br>term stability data not<br>available. |
The information in the table below provides the performance characteristics of IMMULITE and IMMULITE 2000 BR-MA and Immuno 1 CA 15-3.
{6}------------------------------------------------
### Performance Equivalence - Method Comparison (Breast Cancer Patients and Healthy Subjects):
#### IMMULITE BR-MA
The IMMULITE BR-MA assay was compared to Kit A, a commercially available immunoassay that also measured CA 15-3, at two clinical sites in the northeastern and southern United States. A total of 1400 specimens from apparently healthy female subjects and pregnant women, and female patients with breast cancer, malignant and nonmalignant diseases, were evaluated at the two clinical sites. Using the 95th percentile of the normal reference ranges of the two assays (38 U/mL for IMMULITE BR-MA and 34.8 U/mL for Kit A), a qualitative comparison of the IMMULITE BR-MA and Kit A yielded the following results.
| IMMULITE BR-MA | | Positive<br>Agreement | Negative<br>Agreement |
|----------------|----------------------|-----------------------|-----------------------|
| Kit A | > 38 U/mL ≤ 38 U/mL | | |
| > 34.8 U/mL | 334 18 | 94.9% | 85.2% |
| ≤ 34.8 U/mL | 153 895 | | |
Agreement: 87.8%
95% Confidence Limits for positive agreement and negative agreement, respectively: 92.0% - 96.9% and 83.3% - 87.5%.
The CA 15-3 measurements of 1307 specimens that fell within the working range of both assays (300 U/mL for IMMULITE BR-MA and 200 U/mL for Kit A) were compared in a linear regression:
IMMULITE BR-MA = 1.08 (Kit A) + 8.34 r = 0.79
95% Confidence Intervals: Slope: 1.04, 1.13 Intercept: 6.53, 10.2
Means:
39.4 U/mL (IMMULITE BR-MA) 28.7 U/mL (Kit A)
In the clinical study conducted in the southern United States, measurements of CA 15-3 by IMMULITE BR-MA were compared to measurements by Kit B, a commercially available chemiluminescence assay for CA 27-29 on 181 specimens using the respective normal range upper limits of the two assays as the cutoff (38 U/mL for IMMULITE BR-MA and 38.6 U/mL for Kit B).
| | IMMULITE BR-MA | | Positive<br>Agreement | Negative<br>Agreement |
|-------------|----------------|-----------|-----------------------|-----------------------|
| Kit B | > 38 U/mL | ≤ 38 U/mL | | |
| > 38.6 U/mL | 63 | 4 | 94.0% | 87.7% |
| ≤ 38.6 U/mL | 14 | 100 | | |
90.1% Agreement:
95% Confidence Limit for positive agreement: 85.4% - 98.4% 95% Confidence Limit for negative agreement: 80.3% - 93.1%
{7}------------------------------------------------
A subset of the specimens (n=157) within the working ranges of both assays were compared in a linear regression:
IMMULITE BR-MA = 0.86 (Kit B) + 10.4 U/mL r = 0.95 95% Confidence Intervals: 0.82, 0.91 Slope: Slope: Intercept: 7.61. 13.2
Means:
49.1 U/mL (IMMULITE BR-MA) 44.8 U/mL (Kit B)
#### IMMULITE 2000 BR-MA
The IMMULITE 2000 BR-MA assay was compared to Kit A, a commercially available immunoassay that also measured CA 15-3, at one clinical site in the northeastern United States. A total of 500 specimens from apparently healthy female subjects and pregnant women, and female patients with breast cancer, malignant and nonmalignant dieseases, were evaluated at this clinical site. Using the 95th percentile of the normal reference were cranated at this entired. Cor IMMULITE 2000 BR-MA and 34.8 U/mL for Kit A), a qualitative comparison of the IMMULITE BR-MA and Kit A yielded the following results.
| IMMULITE 2000 BR-MA | | | Positive<br>Agreement | Negative<br>Agreement |
|---------------------|-----------|-----------|-----------------------|-----------------------|
| Kit A | > 38 U/mL | ≤ 38 U/mL | | |
| > 34.8 U/mL | 128 | 8 | 94.1% | 84.1% |
| ≤ 34.8 U/mL | 58 | 306 | | |
Agreement: 86.8%
95% Confidence Limits for positive agreement and negative agreement, respectively: 88.7% - 97.4% and 80.3% - 87.8%.
The CA 15-3 measurements of 447 specimens that fell within the working range of both assays (300 U/mL for IMMULITE 2000 BR-MA and 200 U/mL for Kit A) were compared in a linear regression:
IMMULITE 2000 BR-MA = 1.04 (Kit A) + 7.66 r = 0.85
95% Confidence Intervals: Slope: 0.98. 1.10 Intercept: 5.37, 9.95 Means: 37.2 U/mL (IMMULITE 2000 BR-MA) 28.4 U/mL (Kit A)
{8}------------------------------------------------
In the same study, IMMULITE 2000 BR-MA was compared to IMMULITE BR-MA in a linear regression for 466 specimens that fell within the working range of both assays:
IMMULITE 2000 BR-MA = 1.02 (IMMULITE BR-MA) + 0.90 r = 0.99
95% Confidence Intervals: Slope: 1.01, 1.04 Intercept: 0.04, 1.76
Means:
42.8 U/mL (IMMULITE BR-MA) 44.7 U/mL (IMMULITE 2000 BR-MA)
#### Clinical Performance:
In two clinical studies conducted in the northeastern and southern United States, serial samples from 99 breast cancer patients followed for recurrence and 80 patients followed for response to treatment were tested by IMMULITE BR-MA and compared with the clinical history of these patients.
An analysis of these cases considered IMMULITE /IMMULITE 2000 BR-MA as consistent, inconsistent, or equivocal with clinical status if over half, less than half, or exactly half of the specimens tested for a case agreed with the clinical status.
IMMULITE BR-MA measurements accurately reflected the changes in the clinical status in 80 of the 99 (81%) patients followed for recurrence. Of the remainder, 13 (13%) did not parallel the clinical status, and 6 of the 99 (6%) were equivocal as to whether CA 15-3 reflected the clinical status.
For all 294 specimens from patients followed for recurrence, the IMMULITE BR-MA measurements were compared with the clinical status of the breast cancer patients:
| Clinical<br>Status | IMMULITE<br>>38 U/mL | <=38 U/mL |
|------------------------------------|----------------------|-----------|
| Active | 41 | 16 |
| NED* | 39 | 198 |
| Agreement 81% | | |
| Sensitivity (95% CI): 72% (59-83%) | | |
| Specificity (95% CI): 84% (79-88%) | | |
| PPV* (95% CI): 51% (40-63%) | | |
| NPV* (95% CI): 93% (88-96%) | | |
For patients followed for response to treatment, IMMULITE BR-MA measurements accurately reflected the changes in the clinical status in 53 of the 80 (66%) patients. Of
{9}------------------------------------------------
the remainder, 19 (24%) did not parallel the clinical status, and 8 (10%) were equivocal as to whether CA 15-3 reflected the clinical status.
For all 282 specimens from patients followed for response to treatment, the IMMULITE BR-MA measurements were compared with the clinical status of the breast cancer patients:
| Clinical<br>Status | >38 U/mL | <=38 U/mL |
|------------------------------------|----------|-----------|
| Active | 191 | 68 |
| NED* | 10 | 13 |
| Agreement: 72% | | |
| Sensitivity (95% CI): 74% (68-79%) | | |
| Specificity (95% CI): 57% (35-77%) | | |
| PPV* (95% CI): 95% (91-98%) | | |
| NPV* (95% CI): 16% (9-26%) | | |
IMMULITE 2000 BR-MA achieved similar results in one clinical study conducted in the northeastern United States. Serial samples from 35 breast cancer patients followed for recurrence and 15 patients followed for response to treatment were tested by IMMULITE 2000 BR-MA and compared with the clinical history of these patients.
IMMULITE 2000 BR-MA measurements accurately reflected the changes in the clinical status in 28 of the 35 (80%) patients followed for recurrence. Of the remainder, 3 (9%) did not parallel the clinical status, and 4 (11%) were equivocal as to whether CA 15-3 reflected the clinical status.
For all 121 specimens from patients followed for recurrence, the IMMULITE 2000 BR-MA measurements were compared with the clinical status of the breast cancer patients:
| Clinical<br>Status | IMMULITE 2000<br>>38 U/mL | <=38 U/mL |
|------------------------------------|---------------------------|-----------|
| Active | 27 | 9 |
| NED* | 20 | 65 |
| Agreement: 76% | | |
| Sensitivity (95% CI): 75% (58-88%) | | |
| Specificity (95% CI): 76% (66-85%) | | |
| PPV* (95% CI): 57% (42-72%) | | |
| NPV* (95% CI): 88% (78-94%) | | |
For patients followed for response to treatment, IMMULITE 2000 BR-MA measurements accurately reflected the changes in the clinical status in 12 of the 15 (80%)
{10}------------------------------------------------
patients. Of the remainder, 1 (7%) did not parallel the clinical status, and 2 (13%) were equivocal as to whether CA 15-3 reflected the clinical status.
For all 67 specimens from patients followed for response to treatment, the IMMULITE 2000 BR-MA measurements were compared with the clinical status of the breast cancer patients:
| Clinical<br>Status | >38 U/mL | <=38 U/mL | |
|------------------------------------|---------------|-----------|--|
| Active | 55 | 10 | |
| NED* | 1 | 1 | |
| Agreement: 84% | | | |
| Sensitivity (95% CI): 85% (74-92%) | | | |
| Specificity (95% CI): 50% (1-99%) | | | |
| PPV* (95% CI): | 98% (91-100%) | | |
| NPV* (95% CI): | 9% (0-41%) | | |
| * NED - No Evidence of Disease | | | |
PPV - Positive Predictive Value
NPV - Negative Predictive Value
Representative profiles of patients followed for recurrent breast cancer and for response to treatment are shown in the following figures.
Monitoring of a stage II (in May 1988) breast cancer patient with IMMULITE BR-MA (CA 15-3). Longitudinal changes in IMMULITE BR-MA values correlate with changes in disease status.
Image /page/9/Figure/7 description: The figure is a line graph showing the levels of IMMULITE BR-MA (U/mL) over time. The x-axis represents the date, and the y-axis represents the IMMULITE BR-MA levels. The graph shows that the levels were low from 7/27/90 to 7/24/92, and then increased significantly between 7/24/92 and 11/28/94, reaching a level of approximately 1100 U/mL.
| Date | U/mL | Disease Status |
|----------|------|------------------------------------------------------------------------------|
| 7/27/90 | 31.8 | CT scan chest - negative, unchanged since 1988. |
| 1/25/91 | 30.2 | Ct scan chest - no evidence of metastasis. |
| 7/24/92 | 37.7 | Total body scan-increased tracer activity, highly suspicious for metastasis. |
| 5/10/94 | 478 | CT abd, pelvis - Multiple fine nodular radiodensities in lungs. |
| 11/28/94 | 1092 | CT chest-small bilateral pleural effusions. |
{11}------------------------------------------------
Monitoring of a stage IV (in December 1989) breast cancer patient with IMMULITE BR-MA (CA 15-3). Longitudinal changes in IMMULITE BR-MA values correlate with changes in disease status and treatment.
Image /page/10/Figure/1 description: The figure shows a line graph of IMMULITE BR-MA/(U/mL) over time. The x-axis shows the date, and the y-axis shows the IMMULITE BR-MA/(U/mL). The graph shows that the IMMULITE BR-MA/(U/mL) starts at around 55 on 4/1/92, decreases to around 30 on 6/3/92, and then increases to around 190 on 12/10/92.
| Date | U/mL | Disease Status |
|----------|------|------------------------------------------------------------------------------------------|
| 4/1/92 | 55.2 | Radiation therapy 4/91 to 10/91. |
| 5/6/92 | 33.1 | CT showed multiple metastasis, decreasing in number and size. |
| 6/3/92 | 28.0 | Chemo-VAT (Virablastin, Adriamycin, Thiotepa). |
| 7/8/92 | 29.6 | Chest X-Ray pos; Abdominal, pelvis areas no change, stable. |
| 10/23/92 | 111 | Bone scan showed bone metastasis. CT scan showed marked<br>increase in liver metastasis. |
| 12/10/92 | 185 | Terminal breast cancer. |
Monitoring of a stage II (in May 1988) breast cancer patient with IMMULITE 2000 BR-MA (CA 15-3). Longitudinal changes in IMMULITE 2000 BR-MA values correlate with changes in disease status
Image /page/10/Figure/4 description: This image is a line graph showing the levels of IMMULITE 2000 BR-MA (U/mL) over time. The x-axis represents the date, and the y-axis represents the levels of IMMULITE 2000 BR-MA (U/mL). The graph shows that the levels of IMMULITE 2000 BR-MA (U/mL) were relatively low between 7/27/90 and 7/24/92, but then increased significantly between 5/10/94 and 11/28/94, reaching a level of approximately 1400 U/mL.
| Date | U/mL | Disease Status |
|----------|------|------------------------------------------------------------------------------|
| 7/27/90 | 33.1 | CT scan chest - negative, unchanged since 1988. |
| 1/25/91 | 36.5 | Ct scan chest - no evidence of metastasis. |
| 7/24/92 | 38.9 | Total body scan-increased tracer activity, highly suspicious for metastasis. |
| 5/10/94 | 457 | CT abd, pelvis - Multiple fine nodular radiodensities in lungs. |
| 11/28/94 | 1426 | CT chest-small bilateral pleural effusions. |
{12}------------------------------------------------
Monitoring of a stage IV (in December 1989) breast cancer patient with IMMULITE 2000 BR-MA (CA 15-3). Longitudinal changes in IMMULITE 2000 BR-MA values correlate with changes in disease status and treatment.
Image /page/11/Figure/1 description: This figure shows a plot of IMMULITE 2000 BR-MA (U/mL) vs. Date. The x-axis shows the date, and the y-axis shows the IMMULITE 2000 BR-MA (U/mL). The plot shows that the IMMULITE 2000 BR-MA (U/mL) decreases from 4/1/92 to 6/3/92, then increases from 7/8/92 to 12/10/92. The values range from approximately 30 to 190 U/mL.
| Date | U/mL | Disease Status |
|----------|------|------------------------------------------------------------------------------------------|
| 4/1/92 | 58.1 | Radiation therapy 4/91 to 10/91. |
| 5/6/92 | 35.8 | CT showed multiple metastasis, decreasing in number and size. |
| 6/3/92 | 30.3 | Chemo-VAT (Virablastin, Adriamycin, Thiotepa). |
| 7/8/92 | 32.7 | Chest X-Ray pos; Abdominal, pelvis areas no change, stable. |
| 10/23/92 | 115 | Bone scan showed bone metastasis. CT scan showed marked<br>increase in liver metastasis. |
| 12/10/92 | 189 | Terminal breast cancer. |
### Conclusion:
The conclusions drawn from the clinical and nonclinical studies demonstrate that the device is safe, effective, and is substantially equivalent to the current legally marketed device.
{13}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/12/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads facing to the right. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
MAR 2 8 2002
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation 5700 West 96th Street Los Angeles. California 90045-5597
Re: k013984
> Trade/Device Name: IMMULITE® BR-MA and IMMULITE® 2000 BR-MA Regulation Number: 21 CFR § 866.6010 Regulation Name: Tumor Associated Antigen Immunological Test System Regulatory Class: II Product Code: MOI Dated: March 4, 2002 Received: March 5, 2002
### Dear Dr. Levine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{14}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
## 510(k) Number (if known): K013984 Device Name: IMMULITE® BR-MA IMMULITE® 2000 BR-MA
# Indications For Use:
#### IMMULITE BR-MA
For in vitro diagnostic use with IMMULITE analyzer – for the quantitative measurement of CA 15-3 antigen in human serum, as an aid in the detection of recurrence in previously treated stage II and stage III breast cancer patients, and in the management of metastatic breast cancer patients by monitoring disease progression or response to treatment. Serial testing for patient CA 15-3 values should be used in conjunction with other clinical methods used for detecting early recurrence in Stage II and Stage III disease and for monitoring response to treatment in patients with metastatic breast cancer.
#### IMMULITE 2000 BR-MA
For in vitro diagnostic use with IMMULITE 2000 analyzer – for the quantitative measurement of CA 15-3 antigen in human serum, as an aid in the detection of recurrence in previously treated stage II and stage III breast cancer patients, and in the management of metastatic breast cancer patients by monitoring disease progression or response to treatment. Serial testing for patient CA 15-3 values should be used in conjunction with other clinical methods used for detecting early recurrence in Stage II and Stage III disease and for monitoring response to treatment in patients with metastatic breast cancer.
ed
Division Sign-Off Division of Clinical Laboratory Devices
510(k) Number Ko 13984
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
\/
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.