K013971 · Becton, Dickinson & CO · JKA · Feb 1, 2002 · Clinical Chemistry
Device Facts
Record ID
K013971
Device Name
BD VACUTAINER SAFETY COAGULATION TUBE
Applicant
Becton, Dickinson & CO
Product Code
JKA · Clinical Chemistry
Decision Date
Feb 1, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Indications for Use
The BD Vacutainer™ Safety Coagulation tube is an evacuated blood collection tube that provides a means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anticoagulated specimen that may be used for clinical laboratory coagulation assays. The benefits of a safety plastic coagulation tube with Hemogard Safety Closure Assembly are: - reduced risk of specimen tube breakage . - . reduced exposure to blood by laboratory personnel and to minimize blood splatter during stopper removal These benefits lead to increased safety of laboratory personnel and reduced necessity of repeat specimen collection.
Device Story
BD Vacutainer™ Safety Coagulation tube is a sterile, plastic, evacuated blood collection tube containing 0.109M or 0.129M sodium citrate. Device functions as a closed system for blood collection, transport, and processing. Following collection, specimen is centrifuged to isolate plasma for coagulation parameter analysis. Used in clinical laboratory settings by laboratory personnel. Plastic construction and Hemogard Safety Closure Assembly reduce tube breakage, minimize blood splatter during stopper removal, and decrease exposure risk to personnel. Benefits include enhanced laboratory safety and reduced need for repeat specimen collection due to breakage.
Clinical Evidence
Clinical evaluations compared the plastic BD Vacutainer™ Safety Coagulation tube to the glass VACUTAINER™ Brand Sodium Citrate Tube. Study assessed performance across patient populations including normal donors and those on Warfarin or Heparin therapy. Results demonstrated clinically equivalent coagulation assay results between the plastic and glass tubes.
Technological Characteristics
Sterile, evacuated blood collection tube; plastic construction; Hemogard Safety Closure Assembly; 0.109M or 0.129M sodium citrate additive; closed system design.
Indications for Use
Indicated for the collection, transport, and processing of blood for clinical laboratory coagulation assays in patients requiring coagulation parameter testing, including those undergoing anticoagulation therapy (e.g., Warfarin, Heparin).
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
{0}------------------------------------------------
# I. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
## 510(k) Summary Of Safety and Effectiveness
- I. General Information
This Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 § 807.92
## Establishment:
| • Address: | Becton Dickinson VACUTAINER Systems<br>1 Becton Drive<br>Franklin Lakes, NJ 07417-1885 |
|------------------------|----------------------------------------------------------------------------------------------------------|
| • Registration Number: | 2243072 |
| • Contact Person: | Keith M. Smith<br>Director, Regulatory Affairs<br>Telephone No.:(201) 847-5837<br>Fax No. (201) 847-7040 |
| • Date of Summary: | September 28, 2001 |
## Device
- Trade Name: BD Vacutainer™ Safety Coagulation tube . Classification Name: Tubes, Vials, Systems, Serum Separators, Blood Collection
- . Classification: Class II
- Performance Standards: None Established under 514 of the Food, Drug and Cosmetic Act
{1}------------------------------------------------
#### II. Safety and Effectiveness Information Supporting the Substantial Equivalence Determination
Substantial Equivalence Declaration:
The term "Substantial Equivalence" as used in this 510(k) Premarket Notification is limited to the definition of Substantial Equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR & 807, Subpart E, under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of, substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
- . Device Description:
The BD Vacutainer™ Safety Coagulation tubes are sterile, plastic, evacuated blood collection tubes. The tubes contain 0.109M or 0.129M Sodium Citrate as an anticoagulant intended to prevent whole blood from clotting prior to analysis. The specimen is centrifuged and the plasma portion is analyzed for coagulation parameters to detect clotting time disorders and to monitor patients undergoing anticoagulation therapy. The benefits of a plastic tube decrease the occurrence of accidental breakage, increases the safety of laboratory personnel and reduces the necessity of repeat specimens.
. Intended Use:
> The BD Vacutainer™ Safety Coagulation tube is an evacuated blood collection tube that provides a means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anticoagulated specimen that may be used for clinical laboratory coagulation assays. The benefits of a safety plastic coagulation tube with Hemogard Safety Closure Assembly are:
- reduced risk of specimen tube breakage .
- . reduced exposure to blood by laboratory personnel and to minimize blood splatter during stopper removal
These benefits lead to increased safety of laboratory personnel and reduced necessity of repeat specimen collection.
- Synopsis of Test Methods and Results .
Clinical evaluations were performed to determine the safety and efficacy of the BD Vacutainer™ Safety Coagulation tube. The BD Vacutainer™ Safety Coagulation tube (plastic) was compared to the currently marketed VACUTAINER™ Brand Sodium Citrate Tube (glass). The results of the
{2}------------------------------------------------
clinical evaluation demonstrated that the BD Vacutainer™ Safety clinical evaluation demonstition illy equivalent results when compared to Coagulation tube provides crimidale) equalizare Tube for Normal, Warfarin, Heparin and other patient donors.
- Substantial Equivalence .
Based on comparison of the device features, materials, and intended use, the BDP Based on comparison on the device havenes, macemes, which substantially equivalent to the commercially available predicate device.
The predicate device, K number, and clearance date are identified below:
| Manufacturer | Predicate Device | K-Number | Clearance Date |
|--------------------------------------------|---------------------------------------------|----------|----------------------------------------------------------------------------------------------------------------------------------|
| Becton Dickinson<br>VACUTAINER™<br>Systems | VACUTAINER™<br>Brand Sodium<br>Citrate Tube | N/A | Pre-Amendment Device<br>and, therefore, exempt<br>from premarket<br>notification requirements<br>according to the MDA of<br>1976 |
لغد S.D
Keith M. Smith Director, Regulatory Affairs
11/30/01
Date
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird in flight, composed of three curved, overlapping shapes.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB 01 2002
Mr. Keith M. Smith Associate Director, Regulatory Affairs BD Pharmaceutical Systems Becton Dickinson and Company 1 Becton Drive Mail Code 440 Franklin Lakes, New Jersey 07417-1880
Re: k013971
> Trade/Device Name: BD Vacutainer™ Safety Coagulation Tube Regulation Number: 21 CFR § 862.1675 Regulation Name: Tubes, Vials, Systems, Serum Separators, Blood Collection Regulatory Class: II Product Code: GIM Dated: November 28, 2001 Received: December 3, 2001
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
### Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# B. INDICATIONS FOR USE
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
BD Vacutainer™ Safety Coagulation tube Device Name:
Indications for Use:
The BD Vacutainer™ Safety Coagulation tube is a plastic evacuated blood collection tube that provides a means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anticoagulated specimen that may be used for clinical laboratory coagulation assays. The benefits of a safety plastic coagulation tube with Hemogard Safety Closure Assembly are:
- reduced risk of specimen tube breakage .
- reduced exposure to blood by laboratory personnel and to minimize blood . splatter during stopper removal
These benefits lead to increased safety of laboratory personnel and reduced necessity of repeat specimen collection.
Aurhini Boutta
rsion Sign Sign of Clinical Laboratory Devices
(Please do not Write below this line-continue on another 510(k) Number
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Over-the-Counter Use Or
(Per 21 CFR § 801.109)
(Optional format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.