K013887 · Spinal Concepts, Inc. · MNH · Feb 12, 2002 · Orthopedic
Device Facts
Record ID
K013887
Device Name
BACFIX SPINAL FIXATION SYSTEM
Applicant
Spinal Concepts, Inc.
Product Code
MNH · Orthopedic
Decision Date
Feb 12, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SCI BacFix® Ti Spinal Fixation System consists of a combination of components which include rods, hooks, locking wedges, screws, transverse connectors, cable-screws, cables and spinous process grommets which are indicated to provide temporary stability of the thoracic, thoracolumbar, or lumbar spine (T1 to S1). When intended for pedicle screw fixation, implants are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar or sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis). Levels of pedicle screw attachment for these indications range to T1 to the sacrum. As a pedicle screw system, the BacFix® Spinal Fixation System is also intended for patients having Grade 3 or Grade 4 spondylolisthesis at LS-S1, when utilizing autologous bone graft, when affixed to the posterior and intended to be removed after solid fusion is established. Levels of pedicle screw fixation for this indication are from L3 to the sacrum. When intended for non-pedicle, posterior screw fixation of the non-cervical spine, the indications are: Idiopathic scoliosis. Neuromuscular scoliosis/kyphoscoliosis with associated paralysis or spasticity. Scoliosis with deficient posterior elements such as that resulting from laminectomy or myelomeningocele. Spinal fractures (acute reduction or late deformity). Degenerative disc disease (back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). Neoplastic disease. Spondylolisthesis. Spinal Stenosis. Failed previous fusion. The cable-screws, cables and spinous process grommets are indicated for: Defect of pars lateralis. Spondylolisthesis. Cables and spinous processes grommets may be used for interspinous wiring if additional stability is needed.
Device Story
BacFix Ti Spinal Fixation System; posterior spinal fixation hardware including rods, screws, hooks, locking wedges, transverse connectors, cables, and spinous process grommets. Used by surgeons in clinical settings to provide temporary stabilization of thoracic, thoracolumbar, or lumbar spine as an adjunct to fusion. Implants provide immobilization of spinal segments; intended for removal after solid fusion. Benefits include stabilization of acute/chronic instabilities or deformities. Mechanical performance validated via ASTM 1717-96 testing.
Clinical Evidence
Bench testing only. Mechanical testing conducted per ASTM 1717-96 to compare construct stiffness of the BacFix system against the Isola Spinal System. No clinical data presented.
Indicated for skeletally mature patients requiring temporary stabilization of thoracic, thoracolumbar, or lumbar spine (T1-S1) as an adjunct to fusion. Conditions: degenerative disc disease, spondylolisthesis (including Grade 3/4 at L5-S1), fracture, dislocation, scoliosis, kyphosis, spinal tumor, failed fusion, spinal stenosis, and pars lateralis defects. Contraindications: not for cervical spine.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
Predicate Devices
Isola Spinal System
Submission Summary (Full Text)
{0}------------------------------------------------
## KO13887 1 of 2
# FEB 1 2 2002
#### BacFix Spinal Fixation System Pedicle Screw Indications 510(k) Premarket Notification
| SUBMITTED BY | Spinal Concepts, Inc.<br>12012 Technology Blvd., Suite 100<br>Austin, TX 78727 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------|
| ESTABLISHMENT<br>REGISTRATION NUMBER | 1649384 |
| CONTACT PERSON | David M. Hooper, Ph.D.<br>Manager, Regulatory and Clinical Affairs<br>Phone: 512-918-2700<br>Fax: 512-918-2784 |
| DATE PREPARED | November 21, 2001 |
| CLASSIFICATION NAME | Pedicle Screw Spinal System<br>Spinal Interlaminal Fixation Orthosis |
| COMMON NAME | Spinal Fixation System |
| PROPRIETARY NAME | BacFix Spinal Fixation System |
### DEVICE DESCRIPTION
DE VICE DESCRIP TION The Bach IT Spinal Fixation System is intended to or also with ASTM F136. The BacFix Spinal System is manufactured from thamall andy, which vehip.comectors and other ancillary components. The FTRaton System moudes secews, 1003, noored for market under K973687, and supplemented through II (calif Bach IX 11 Spillar Fixation System was cleared ink transverse connector), K003351 (addition of additional JTUK Submissions: Troughts (addition of end-to-end and side-by-side connectors).
### INDICATIONS FOR USE
The SCI BacFix® Ti Spinal Fixation System consists of a combination of components which include rods, The SCI Dacrix IT Spulal Fradion System Connectors, cables and spinous process grommets.
hooks, locking wedges, screws, transverse connectors, cables and spinous process g hooks, locking wedges, screws, tability of the thoracic, the racolumbar, or lumbar spine (TI to S1).
When intended for pedicle screw fixation, implants are intended to provide immobilization of when intellection policity setew thation, inferent as an adjunct to fusion in the treatment of the discoses (download so chronic instabilities or deformities of the thoracic, lumbar or sacral spine: degenerative disc disel dice). Inconcernity discogenic back pain with degeneration of the disc confirmed by history and radiographic studies is lambosi spondylolisthesis with objective evidence of neurological impairment, fracture, discusses of the these spondylonsmosts with ogeetive enterious crudarthrosis). Levels of pedicle screw attachment for these indications range to T1 to the sacrum.
As a pedicle screw system, the BacFix® Spinal Fixation System is also intended for patients having Grade 3 or As a pondylolisthesis at LS-S1, when utilizing autologous bone graft, when affixed to the posterior Orade 4 spolidylonistics at 25 OT, when amazing and its established. Levels of pedicle screw fixation for this indication are from L3 to the sacrum.
{1}------------------------------------------------
When intended for non-pedicle, posterior screw fixation of the non-cervical spine, the indications are:
- . Idiopathic scoliosis.
- Neuromuscular scollosis/kyphoscoliosis with associated paralysis or spasticity. .
- Neuromuseular sconosisks phoseenments such as that resulting from laminectomy or . myelomeningocele.
2 cap 2
- Spinal fractures (acute reduction or late deformity). .
- Spillar fractures (double roads.orgin of discogenic origin with degeneration of the disc ● confirmed by history and radiographic studies).
- Neoplastic disease. .
- . Spondylolisthesis.
- Spinal Stenosis. .
- Failed previous fusion. .
The cable-screws, cables and spinous process grommets are indicated for:
- Defect of pars lateralis. ●
- Spondylolisthesis. .
Cables and spinous processes grommets may be used for interspinous wiring if additional stability is needed.
#### PREDICATE DEVICE
PREDICATE DE VICE I he Bacrix Spillal Fixation System is subsumany equivalent is as essent in the pedicle screw fixation.
#### DISCUSSION OF NONCLINICAL TESTS
DISCUSSION OF NONCEINVERD TEST S
Mechanical testing was conducted on the stiffest and most flexible data collected for Mechallical testing was conducted BacFix data were compared to available data ollected for ASTM 1717-90. The experimentally concected that the range of construct stiffiesses of the BacFix system is the 1sola Spinal System. 11 Tras assnof stiffnesses available with the Isola Spinal System.
## COMPARISON TO THE PREDICATE DEVICE
COMPARISON TO THE TREDICATE DEFFICE
The BacFix Ti Spinal Fixation System is substantially equivalent to the Isola Spinal System (DePuy Acromed Inc., Cleveland, OH). Both systems are composed of screws, rods, hooks, transverse comectors Acromed inc., Cleveland, Orr. 'Dour Systems are voting Materials used in both systems are biocompatible and other ancinal y components to the BacFix system is contained within the possible range of and the railer of consuler surmisses of the been in by earling systems are substantially equivalent for all indications.
| Device Name | BacFix Spinal Fixation System | Isola Spinal System |
|-------------------------------|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Indications for use | See above. | Same. |
| Materials | Titanium 6Al-4V | Stainless steel or titanium |
| Components | Screws, rods, hooks, transverse<br>connectors. | Same. |
| Product Labeling | Instructions for use and box labeling<br>including all the necessary warning<br>statements. | Same. |
| Packaging/Sterilization | Non-sterile, single use only. | Same. |
| Biomechanical Test<br>Results | Stiffness range, per ASTM 1717-96,<br>within the range of the Isola Spinal<br>System. | Stiffness range, per ASTM 1717-<br>96, is wider than the BacFix<br>Spinal Fixation System. |
## Table 1: TABLE OF SUBSTANTIAL EQUIVALENCE
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white seal for the U.S. Department of Health & Human Services. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is a stylized image of an eagle with three lines representing its wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 1 2 2002
David M. Hooper, Ph.D. Manager, Regulatory and Clinical Affairs Spinal Concepts Incorporated 12012 Technology Boulevard - Suite 100 Austin, Texas 78727
Re: K013887
Trade Name: BacFix™ Ti Spinal Fixation System Trade Name: 21 CFR 888.3050 and 888.3050 and 888.3070 Regulation Number: 21 CFR 860.3050 and 600.5070
Regulation Name: Spondylolisthesis Spinal Fixation Device System, Pedicle Screw Spinal
Condidentifies Spondylolistics Spinal F System, and Spinal Interlaminal Fixation Orthosis Regulatory Class: II Product Code: MNH, MNI, KWP Dated: November 21, 2001 Received: November 23, 2001
Dear Dr. Hooper:
We have reviewed your Section 510(k) notification of intent to market the device referenced above We have reviewed your Section 310(K) notheaton of the indications for use stated in
and we have determined the device is substantially equivalent (Are 28, 1976, the enactment and we have determined the device is substantial.reprints to May 28, 1976, the enactment date the enclosure) to devices marketed in mich start contracted in accordance with the
of the Medical Device Amendments, or to accorded to the fore, market the of the Medical Device Amendinents, or to Govern in Act (Act). You may, therefore, market the provisions of the Pederal Food, Drug, and Cosmons of the Act. The general controls provisions of the device, subject to the general controls provisions of the rices, good manufacturing practice, Act melade requirentions against misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III (Premarket If your device is classified (set additional controls. Existing major regulations affecting your Approval), it may be subject to such additions. Title II, Parts 800 to 895. A substantially
device can be found in the Code of Federal Regulations. Title 21, Parts 800 to device can be found in the Code of Techante with the current Good Manufacturing Practices: General equirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General
requirement, as set forth in the Quality System Regulations the Food and Drug requirement, as set form in the Quanty Bystem rogencie (QS) inspections, the Food and Drug
regulation (21 CFR Part 820) and that, through periodic with the GMP regulation regulation (21 C.FK Part 820) and may and the entrely with the GMP regulation Administration (FDA) will verify suctifals) - caublish further announcements concerners concerning
may result in regulatory action. In addition, FDA may promorses notificati may result in regulatory action. In addition, I Dr. I may posse to your premarket notification your device in the Ederal Kegister. Trass note: the response sections 542 of the Act
submission does not affect any obligation you might have under sections of the Federal submission does not affect any obligation you might have around of other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Dr. David Hooper
This letter will allow you to begin marketing your device as described in your 510(k) premarket This letter will allow you to begin narketing your device to a legally marketed notification. The FDA finding of substantial equivalence of your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your devices), please contact the Office of Compliance at (301)
additionally 809.10 for in vitro diagnostic devices), please contact the Off additionally 809.10 for in vitro diagnosite the promotion and advertising of your device, please
594-4659. Additionally, for questions on the promotion and advertising of yo 594-4659. Additionally, for questions on the promotion and mease note the regulation entitled, the regulation entitled, and contact the Office of Compliance at (301) 594-407. Aller general information
"Misbranding by reference to premarket notification of Small Manufacturers "Misbranding by reference to premation (21 or the Division of Small Manufacturers
on your responsibilities under the Act may be obtained from the Division of Small Manufactur on your responsibilities under the Acc inay oc out in at 2011 at 12:00
Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address
1992 "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Muriam C. Provost
Col Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K013887
1α2
## INIDICATIONS FOR USE STATEMENT
Page 1 of 2
K013887 510(k) Number (if known):
Device Name:
# Spinal Concepts, Inc. BacFix Ti Spinal Fixation System
Indications for Use:
The SCI BacFix® Ti Spinal Fixation System consists of a combination of components which include rods, hooks, locking wedges, screws, transverse connectors, cable-screws, cables and spinous process grommets which are indicated to provide temporary stability of the thoracic, thoracolumbar, or lumbar spine (T1 to S1).
When intended for pedicle screw fixation, implants are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar or sacral spine: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor and failed previous fusion (pseudarthrosis). Levels of pedicle screw attachment for these indications range to T1 to the sacrum.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: (Per 21 CFR 801.109) OR
Over-The-Counter: (Optional Format 1-2-96)
Muriam C. Provost
Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K013857
{5}------------------------------------------------
# INIDICATIONS FOR USE STATEMENT
K013887
Page 2 of 2
As a pedicle screw system, the BacFix® Spinal Fixation System is also intended for patients having Grade 3 or Grade 4 spondylolisthesis at LS-S1, when utilizing autologous bone graft, when affixed to the posterior when deling and intended to be removed after solid fusion is established. Levels of pedicle screw fixation for this indication are from L3 to the sacrum.
When intended for non-pedicle, posterior screw fixation of the non-cervical spine, the indications are:
- . Idiopathic scoliosis.
- Neuromuscular scoliosis/kyphoscoliosis with associated . paralysis or spasticity.
- Scoliosis with deficient posterior elements such as that . resulting from laminectomy or myelomeningocele.
- Spinal fractures (acute reduction or late deformity). ●
- Degenerative disc disease (back pain of discogenic origin with . degeneration of the disc confirmed by history and radiographic studies).
- Neoplastic disease. ●
- Spondylolisthesis. ●
- Spinal Stenosis. ●
- Failed previous fusion. .
The cable-screws, cables and spinous process grommets are indicated for:
- Defect of pars lateralis. .
- Spondylolisthesis. .
Cables and spinous processes grommets may be used for interspinous wiring if additional stability is needed.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
X Prescription Use: (Per 21 CFR 801.109)
OR
Over-The-Counter: (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.