MODIFICATION TO EPI-STAR SURGICAL LASER SYSTEM, M0DEL DS-60

K013864 · Nidek, Inc. · GEX · Feb 8, 2002 · General, Plastic Surgery

Device Facts

Record IDK013864
Device NameMODIFICATION TO EPI-STAR SURGICAL LASER SYSTEM, M0DEL DS-60
ApplicantNidek, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateFeb 8, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Epi-Star Surgical Laser System is intended to be used for Plastic Surgery and Dermatology, with the intended uses for the treatment of vascular and pigmented lesions and for the removal of hair in skin types I - VI.

Device Story

Epi-Star Surgical Laser System is an 800 nm continuous wave laser for plastic surgery and dermatology. Device delivers laser energy to skin to treat vascular/pigmented lesions and remove hair. System features self-limiting parameters for spot size (2-5mm), output power (5-60W), dwell time, density, and fluence (up to 400 J/cm²). Model A utilizes a scanner handpiece with a 2-TE chip cooling window; Model B utilizes a scanner with a cold air system. Operated by clinicians in clinical settings. System provides cooling to protect skin during high-fluence procedures. Output parameters are adjusted based on skin type. Benefits include effective lesion treatment and hair removal with integrated thermal management.

Clinical Evidence

No human clinical data provided. Evidence consists of bench testing based on product specifications and hazard analysis, and a preclinical animal study to evaluate the safety of the device at higher fluence rates. Results indicated the system performs as intended.

Technological Characteristics

800 nm continuous wave laser; 5-60 Watts output power; 5-700 ms pulse duration; 620-650 nm aiming beam. Spot sizes: 2, 3, 4, 5 mm. Cooling: Model A (2-TE chip window), Model B (cold air system). Self-limiting fluence up to 400 J/cm². Class II device.

Indications for Use

Indicated for patients of skin types I-VI requiring treatment of vascular and pigmented lesions or hair removal. Intended for use in plastic surgery and dermatology settings.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 18: This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. # 18.1 SUBMITTER INFORMATION | a. | Company Name: | Nidek, Inc. | |----|--------------------------------------|-----------------------------------------------| | | b. Company Address: | 47651 Westinghouse Drive<br>Fremont, CA 94539 | | c. | Company Phone:<br>Company Facsimile: | (510) 226-5700<br>(510) 226-5750 | | | d. Contact Person: | Hiro Matsuzaki<br>Quality Assurance Manager | | | e. Date Summary Prepared: | January 4, 2002 | ## DEVICE IDENTIFICATION 18.2. | a. | Trade/Proprietary Name: | Epi-Star Surgical Laser System, Models A<br>and B | |----|-------------------------|-----------------------------------------------------| | b. | Classification Name: | Surgical Laser Instrument<br>21 CFR 878.4810 78 GEX | ## IDENTIFICATION OF PREDICATE DEVICES 18.3 | Company | Device | 510(k) No. | Date Cleared | |-------------|-----------------------------------|------------|--------------| | Nidek, Inc. | Epi-Star Surgical Laser<br>System | K990119 | 07/27/2000 | ## DEVICE DESCRIPTION 18.4 The Epi-Star Surgical Laser System is an 800 nm continuous wave device coupled with cooling accessory. The Epi-Star laser is self-limiting in the size of 1/3 {1}------------------------------------------------ K013864 2/3 Nidek Epi-Star Surgical Laser System 510(k) Amendment, #K013864, January 4, 2002 > spots (2, 3, 4, and 5mm), output power, dwell time, density and fluence rates. The Epi-Star Surgical Laser can achieve fluence rates of up to 400 J/cm² when coupled with the cooling accessories. The treatment parameters for the classifications of various skin types is provided. The Epi-Star laser can be coupled with either a scanner handpiece that has a cooling window (Model A) or with a scanner and cold air system (Model B). #### 18.5 SUBSTANTIAL EQUIVALENCE The Epi-Star Surgical Laser System with the proposed modifications is substantially equivalent to the current Epi-Star Surgical Laser System that is currently in commercial distribution by Nidek, Inc. The fundamental technical characteristics of the Epi-Star Surgical Laser System are the same as those of the predicate device and the indications for use has not changed with the modifications of the device. #### 18.6 INDICATIONS FOR USE The Epi-Star Surgical Laser System is intended to be used for Plastic Surgery and Dermatology, with the intended uses for the treatment of vascular and pigmented lesions and for the removal of hair in skin types I - VI. #### 18.7 TECHNOLOGICAL CHARACTERISTICS The modifications to the Epi-Star Laser System allow for higher fluence rates of up to 400 J/cm2 for the treatment of vascular and pigmented lesions and the removal of hair. The increased fluence rates are dependent on the skin types to be treated and are self-limiting by the laser system. The output power of the system is from 5 - 60 Watts, the aiming beam is 620 - 650 nm and the pulse duration is 5 - 700 ms. Spot sizes of 2, 3, 4, and 5mm can be accomplished with the Epi-Star. Model A includes a scanner with a 2-TE chip cooling window and Model B includes a scanner attached to a cold air system. The current Epi-Star Laser {2}------------------------------------------------ K013867 Nidek Epi-Star Surgical Laser System 510(k) Amendment, #K013864, January 4, 2002 > System allows for fluence rates of up to 50 J/cm². The current Epi-Star includes a scanner with a 1-TE chip cooling window. ## PERFORMANCE DATA 18.8 Performance testing was conducted on the Epi-Star Surgical Laser System. System and component testing was completed based on product specifications and hazard effects determined from the risk analysis. A preclinical animal study was conducted to determine the safety of the Epi-Star Surgical Laser System with the higher fluence rates. The Epi-Star Surgical Laser System was found to perform as intended. ## CONCLUSION 18.9 This notification contains all information required by 21 CFR 807.87. The Epi-Star Surgical Laser System with the higher fluence rates was found to perform as intended during validation testing and in preclinical animal studies. The Epi-Star Surgical Laser System is substantially equivalent to the current Epi-Star Surgical Laser System in commercial distribution. The Epi-Star Surgical Laser System is intended for the treatment of vascular and pigmented lesions and for the removal of hair. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract eagle design with three stylized wing segments. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 0 8 2002 Nidek, Inc. c/o Ms. Carol White Patterson Consulting Group, Inc. 21521 Hummingbird Street Trabuco Canyon, California 92679 Re: K013864 Trade/Device Name: Epi-Star Surgical Laser System, Models A and B and Zimmer Elektromedizin Cryo 5 Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: November 19, 2001 Received: November 21, 2001 Dear Ms. White: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Carol White This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Muriam C. Provost for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Nidek Epi-Star Surgical Laser System 510(k) Amendment, #K013864, January 4, 2002 # INDICATION FOR USE K013864 510(k) Number: Epi-Star Surgical Laser System, Models A and B Device Name: The Epi-Star Surgical Laser System is intended to be used for Indications for Use: Plastic Surgery and Dermatology, with the intended uses for the treatment of vascular and pigmented lesions and for the removal of hair in skin types I - VI. Miriam C. Provost Division of General, Restorative and Neurological Devices 510(k) Number_KO 3864 (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use : ... ..
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...