MICROJET CRONO H

K013855 · Cane S.R.I. · FRN · May 9, 2002 · General Hospital

Device Facts

Record IDK013855
Device NameMICROJET CRONO H
ApplicantCane S.R.I.
Product CodeFRN · General Hospital
Decision DateMay 9, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The portable Crono infusion device has been designed only for use in subcutaneous and intravenous infusion of prescribed liquid medicines.

Device Story

Microjet Crono H is a compact, portable ambulatory infusion pump designed for subcutaneous and intravenous delivery of prescribed liquid medications. Device utilizes a proprietary syringe system to reduce overall size compared to conventional pumps. Mechanism features a direct-drive system acting on the rubber syringe piston, providing up to 3 times higher thrust force than traditional pumps to overcome catheter occlusions. Infusion control system enables automatic resumption of delivery following occlusion clearance. Infusion is highly fractionated (22 ml per impulse) to improve drug absorption. Device includes an LCD displaying delivery time and battery status. Operated by patients in everyday life settings to facilitate continuous therapy, such as iron chelation. Benefits include improved patient compliance due to reduced size and enhanced reliability in drug delivery.

Clinical Evidence

Bench testing, EMC testing, and user testing were performed. No clinical trial data provided. Testing confirmed the device is as safe and effective as the predicate.

Technological Characteristics

Ambulatory infusion pump; 3V Lithium battery power; 77 x 47 x 29 mm dimensions; 115g weight; 10/20 ml capacity. Features direct-drive piston mechanism for high-thrust infusion. Includes LCD for status monitoring. Software-controlled infusion delivery and occlusion management.

Indications for Use

Indicated for subcutaneous and intravenous infusion of prescribed liquid medicines in patients requiring ambulatory infusion therapy.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K013855 ### EXHIBIT 2 CANÈ S.r.l. Via Pavia, 105/I 10090 Rivoli-Cascine Vica (Torino) Italy Tel .: ++39-011-957.48.72 Fax ++39-011-959.88.80 Contact: Mario Cané, President January 30, 2002 ## 510(k) Summary of Safety and Effectiveness - 1. Identification of the Device: Proprietary-Trade Name: Microjet Crono H Classification Name: 80 FRN Common/Usual Name: Ambulatory Infusion Pump - 2. Equivalent legally marketed devices This product is similar in function to the Cadd-Legacy 1 Ambulatory Infusion Pump , Sims Deltec, Inc., K982838 - 3. Indications for Use (intended use) The portable Crono infusion device has been designed only for use in subcutaneous and intravenous infusion of prescribed liquid medicines. - 4. Description of the Device: Canè s.r.l., a company that specializes in the production of ambulatory pumps, has now produced a new generation of compact pumps: Crono H, a perfect combination of high technology and innovative design. The infusion pumps which use normal commercial syringes are inevitably cumbersome and thus difficult to use in everyday life. This provokes in the patients a refuse of the therapy. A special syringe, an integral part of the Crono H allows an efficient reduction of the pump size. In this way, the patients can freely carry out the therapy at any time of the day, even during their everyday life. Patients who have been following a iron chelation therapy for a long time, often experience difficulties in absorbing the drug subcutaneously; this may lead to a catheter occlusion with consequent interruption of the infusion. Crono H has a particular mechanism which pushes directly the rubber syringe piston: so it is possible to reach a thrust force up to 3 times higher with respect to conventional pumps. In case of catheter occlusion, an innovative infusion control system makes it possible to proceed with the infusion automatically and, after the occlusion is eliminated, to complete it. For a better absorption of the drug, Crono H makes the infusion up to 3 times more fractionated with respect to traditional pumps (22 ml per impulse, using a 10/20 ml syringe). Crono H is fitted with a liquid crystal display which shows the time it takes to complete the delivery and battery charge status. - 5. Safety and Effectiveness, comparison to predicate device. The results of bench, EMC, and user testing indicates that the new device is as safe and effective as the predicate device. {2}------------------------------------------------ # 6. Substantial Equivalence Chart | Characteristic | Cadd-Legacy 1<br>Ambulatory Infusion<br>Pump , Sims Deltec, Inc.,<br>K982838 | Microjet CronoH | |----------------|---------------------------------------------------------------------------------------------|-------------------------------------------| | Intended Use: | Intravenous<br>Intra-arterial<br>Subcutaneous<br>Intraperitoneal<br>Epidural<br>Intrathecal | Subcutaneous and<br>intravenous only | | Power Source | 2 AA alkaline batteries, AC<br>Adapter | Lithium battery (3V) of the<br>123 A type | | Size | 4.4 x 3.8 x 1.6 x in (112x<br>95x41 mm) | 3" x 1.85" x 1.14" (77 x 47<br>x 29 mm) | | Weight | 13.8 oz (392 grams) | 4.0 oz (115 g) (battery<br>included). | | Capacity | 10 ml | 10 or 20 ml | | Warranty: | 1 year | 2 years | #### 7. Conclusion After analyzing both bench and user testing data, it is the conclusion of CANÈ S.r.1. that the Microjet Crono is as safe and effective as the predicate device, has few technological differences, and has no new indications for use, thus rendering it substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 0 9 2002 Cane S. R. I C/O Mr. Daniel Kamm Kamm & Associates P.O. Box 7007 Deerfield, Illinois 60015 Re: K013855 Trade/Device Name: MicroJet Crono H Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: February 14, 2002 Received: February 15, 2002 Dear Mr. Kamm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other {4}------------------------------------------------ ### Page 2 - Mr. Kamm requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594 advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Susa Punoy Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {5}------------------------------------------------ # j) Indications for Use 510(k) Number K013855 Device Name: Microjet CronoH ambulatory infusion pump Indications for Use: The Microjet Crowo ambulatory infusion pump device has been designed for use in subcutaneous and intravenous infusion of prescribed liquid medicines. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR P (Per 21 CFR 801.109) Over the Counter Use Pulace Licente (Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number_4615
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...