← Product Code [LGD](/productcode/LGD) · K013837

# BIO-RAD PLATELIA TOXO IGM TMB (K013837)

_Bio-Rad · LGD · Sep 30, 2002 · Microbiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K013837

## Device Facts

- **Applicant:** Bio-Rad
- **Product Code:** [LGD](/productcode/LGD.md)
- **Decision Date:** Sep 30, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3780
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K013837 · Sep 30, 2002 | BIO-RAD PLATELIA TOXO IGM TMB | Bio-Rad | Routine clinical laboratory serum samples | Retrospective analysis of 445 serum samples from pregnant women collected during routine laboratory activities to determine expected values for the assay. | Expected values; Routine clinical laboratory samples; Pregnant women |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of routine clinical samples | Pregnant women; Sample Size: 445 | Not applicable for this study | Distribution of serum / cut-off ratio values |

## Indications for Use

The Platelia Toxo IgM TMB kit is an in-vitro diagnostic test kit for the qualitative and quantitative detection of anti-Toxoplasma gondii IgM in human serum or plasma (EDTA, Heparin, Citrate). Note: - Patient testing with the Platelia® Toxo IgM assay must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay. - The Platelia® Toxo IgM assay is presumptive for the detection of anti-Toxoplasma gondii IgM antibodies and presumptive for the diagnosis of acute, recent or reactivated Toxoplasma gondii infection. - The performance of the Platelia® Toxo IgM assay has not been established for neonate testing. - The Platelia® Toxo IgM assay has not been cleared/approved by the FDA for blood/plasma donor screening.

## Device Story

Platelia Toxo IgM TMB is an in vitro diagnostic enzyme immunoassay (EIA) for detecting anti-Toxoplasma gondii IgM antibodies in human serum or plasma. The device uses a double sandwich method in 96-well breakaway microplates. Samples are incubated in wells coated with anti-human IgM antibodies; captured IgM binds to T. gondii antigen-conjugate (monoclonal antibody labeled with horseradish peroxidase). After washing, a TMB substrate/chromogen solution is added, causing color development proportional to the amount of bound conjugate. The reaction is stopped with acid, and optical density is measured via spectrophotometer at 450 nm. Results are determined by comparing sample optical density to a cut-off control. Used in clinical laboratories by trained personnel. Output aids clinicians in diagnosing acute, recent, or reactivated toxoplasmosis when used alongside IgG testing.

## Clinical Evidence

Bench testing only. Performance evaluated using 348 patient samples and 97 specimens from 40 seroconversion panels. Combined results showed 99.53% agreement with a commercial EIA. Excluding equivocal samples, negative agreement was 100% (271/271) and positive agreement was 98.72% (154/156). CDC panel testing (32 positive, 65 negative) showed 100% agreement with both positive and negative specimens. Precision studies (intra-assay and inter-assay) conducted across three sites and with various plasma anticoagulants (EDTA, Heparin, Citrate) demonstrated acceptable reproducibility.

## Technological Characteristics

96-well breakaway microplate EIA. Sensing principle: immunocnzymatic double sandwich. Energy source: spectrophotometer (450 nm). Materials: microplate wells coated with anti-human IgM; T. gondii antigen; monoclonal antibody-HRP conjugate; TMB chromogen solution; synthetic control matrix (Tris-NaCl, BSA, glycerol, colorant).

## Regulatory Identification

Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Toxoplasma gondii from clinical specimens. The identification aids in the diagnosis of toxoplasmosis caused by the parasitic protozoan Toxoplasma gondii and provides epidemiological information on this disease. Congenital toxoplasmosis is characterized by lesions of the central nervous system, which if undetected and untreated may lead to brain defects, blindness, and death of an unborn fetus. The disease is characterized in children by inflammation of the brain and spinal cord.

## Predicate Devices

- Kallestad Diagnostics Platelia® Toxo IgM (K89498)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a logo with the text "O+RAD" in bold, white letters against a black background. The logo has a rounded rectangular shape. The plus sign in "O+RAD" is slightly smaller than the other letters.

SEP 3 0 2002

#### ATTACHMENT I 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K013837

| DATE:                   | September 10, 2002                                                           |
|-------------------------|------------------------------------------------------------------------------|
| APPLICANT:              | Bio-Rad<br>3, Boulevard Raymond Poincaré<br>92430 Marnes-la-Coquette, France |
| PHONE:                  | 33-1-47-95-6138                                                              |
| FAX:                    | 33-1-47-95-6242                                                              |
| OFFICIAL CORRESPONDENT: | Sylvie Confida                                                               |
| PRODUCT TRADE NAME:     | Bio-Rad Platelia® Toxo IgM TMB                                               |
| COMMON NAME:            | Toxoplasma IgM Enzyme Immunoassay (EIA)                                      |
| CLASSIFICATION NAME:    | 21 CFR 866.3780, Enzyme Linked Immunoabsorbent Assay,<br>Toxoplasma Gondii   |
| PREDICATE DEVICE:       | Kallestad Diagnostics Platelia® Toxo IgM (K89498)                            |

#### DEVICE DESCRIPTION

The Platelia® Toxo IgM TMB assay is a qualitative assay which utilizes an immunocnzymatic double sandwich method. Diluted samples and controls are placed into microplate wells coated with antibodies to human IgM. The IgM antibodies present in the samples are captured by the solid phase. Remaining antibodies (including any of the IgG class) and other serum proteins are removed by washing. A solution containing T. gondii antigen and conjugate (monoclonal antibody to T. gondii labeled with horseradish peroxidase) is placed into each well. T. gondii IgM antibodies in the test sample that are captured on the solid phase bind the T. gondii antigen-complex. Excess T. gondii antigen and conjugate are removed by washing. A peroxidase substrate and chromogen solution is added which reacts with the conjugate to initiate a color development reaction is stopped by the addition of an acid. The optical density readings for the test samples are obtained with a spectrophotometer set at a wavelength of 450 nm. The presence of T. gondii IgM antibodies in an individual sample is determined by comparing the optical density reading for the sample to the optical density reading of the cut-off control serum.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in black and white and features the company name in bold, sans-serif letters. The logo is enclosed in a rounded rectangle.

## INTENDED USE

The Platelia® Toxo IgM TMB kit is an in vitro diagnostic test kit allowing the qualitative detection of anti-Toxoplasma gondii in human serum or plasma (EDTA, Heparin, Citrate).

#### Note:

- Patient testing with the Platelia® Toxo IgM assay must be performed in conjunction with an anti-. Toxoplasma gondii IgG antibody assay.
- . The Platelia® Toxo IgM assay is presumptive for the detection of anti-Toxoplasma gondii IgM antibodies and presumptive for the diagnosis of acute, recent, or reactivated Toxoplasma gondii infection.
- . The performance of the Platelia® Toxo IgM assay has not been cleared/approved by the FDA for blood/plasma donor screening.

### TECHNOLOGICAL CHARACTERISTICS

The Platelia® Toxo IgM TMB kit is a modified version of the Platelia® Toxo IgM kit and remains similar in form and function. The following comparison table indicates significant similarities and differences between the kits:

| Characteristics                | Platelia® Toxo IgM (Predicate)                                                                                                               | Platelia® Toxo IgM TMB                                                                                                                                                     |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Format and Test Method         | 96-well microplate EIA,<br>non-breakaway wells                                                                                               | 96-well microplate EIA,<br>breakaway wells                                                                                                                                 |
| Intended Use                   | Assay for the qualitative<br>detection of anti- <i>Toxoplasma</i><br><i>gondii</i> IgM in human serum                                        | Assay for the qualitative<br>detection of anti- <i>Toxoplasma</i><br><i>gondii</i> IgM in human serum or<br>plasma                                                         |
| Positive and Negative Controls | Pooled human serum, negative<br>for <i>T. gondii</i> , is used to<br>manufacture the negative control<br>and to dilute the positive control. | A synthetic matrix consisting of<br>Tris-NaCl buffer, BSA, glycerol,<br>and colorant is used to<br>manufacture the negative control<br>and to dilute the positive control. |
| Chromogen                      | <i>o</i> . phenylenediamine-2 HCL<br>(OPD) tablets                                                                                           | Tetramethylbenzidine (TMB)<br>solution                                                                                                                                     |
| Wavelength                     | Dual wavelength reading at<br>492 nm and 620 nm.                                                                                             | Dual wavelength reading at<br>450 nm and 620 nm.                                                                                                                           |

#### Comparison Table

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black background. The text is bold and sans-serif.

## PERFORMANCE SUMMARY

## A. Precision Studies

Inter-assay and intra-assay reproducibility were determined by assaying two T. gondir IgM negative samples and four T. gondii IgM positive samples in triplicate, on three different days at three laboratory sites.

Site 1

|                   |  | Neg 1 |       | Neg 2 |       | Pos 1 |       | Pos 2 |       | Pos 3 |       | Pos 4 |       |
|-------------------|--|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|
|                   |  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |
| N=                |  | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     |
| Mean=             |  | 0.027 | 0.052 | 0.022 | 0.042 | 0.742 | 1.438 | 0.660 | 1.278 | 0.825 | 1.599 | 1.477 | 2.862 |
| Within Run        |  |       |       |       |       |       |       |       |       |       |       |       |       |
| (intra-assay) sd= |  | 0.001 | 0.003 | 0.002 | 0.003 | 0.011 | 0.021 | 0.011 | 0.022 | 0.016 | 0.031 | 0.017 | 0.034 |
| %CV=              |  | 5.2%  | 5.1%  | 7.5%  | 7.6%  | 1.5%  | 1.4%  | 1.7%  | 1.7%  | 1.9%  | 1.9%  | 1.2%  | 1.2%  |
| Total             |  |       |       |       |       |       |       |       |       |       |       |       |       |
| (inter-assay) sd= |  | 0.002 | 0.005 | 0.001 | 0.003 | 0.020 | 0.041 | 0.013 | 0.038 | 0.013 | 0.048 | 0.032 | 0.050 |
| %CV=              |  | 8.1%  | 9.7%  | 6.7%  | 6.3%  | 2.6%  | 2.8%  | 2.0%  | 3.0%  | 1.6%  | 3.0%  | 2.1%  | 1.7%  |

Site 2

|                                 |       | Neg 1 |       | Neg 2 |       | Pos 1 |       | Pos 2 |       | Pos 3 |       | Pos 4 |  |
|---------------------------------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|--|
|                                 | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |  |
| N=                              | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     |  |
| Mean=                           | 0.031 | 0.050 | 0.029 | 0.048 | 0.746 | 1.206 | 0.669 | 1.083 | 0.776 | 1.255 | 1.511 | 2.444 |  |
| Within Run<br>(intra-assay) sd= | 0.007 | 0.011 | 0.007 | 0.011 | 0.025 | 0.040 | 0.015 | 0.024 | 0.009 | 0.015 | 0.051 | 0.082 |  |
| %CV=                            | 22.5% | 23.0% | 22.6% | 23.2% | 3.3%  | 3.3%  | 2.3%  | 2.3%  | 1.2%  | 1.2%  | 3.4%  | 3.4%  |  |
| Total<br>(inter-assay) sd=      | 0.009 | 0.016 | 0.007 | 0.012 | 0.024 | 0.061 | 0.015 | 0.066 | 0.012 | 0.063 | 0.046 | 0.151 |  |
| %CV=                            | 29.9% | 32.8% | 22.3% | 24.4% | 3.2%  | 5.0%  | 2.2%  | 6.1%  | 1.5%  | 5.0%  | 3.0%  | 6.2%  |  |

Site 3

|                   |  | Neg 1 |       | Neg 2 |       | Pos 1 |       | Pos 2 |       | Pos 3 |       | Pos 4 |       |
|-------------------|--|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|-------|
|                   |  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |
| N=                |  | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     | 9     |
| Mean=             |  | 0.032 | 0.057 | 0.024 | 0.042 | 0.769 | 1.393 | 0.705 | 1.278 | 0.684 | 1.227 | 1.346 | 2.426 |
| Within Run        |  |       |       |       |       |       |       |       |       |       |       |       |       |
| (intra-assay) sd= |  | 0.004 | 0.007 | 0.006 | 0.011 | 0.055 | 0.104 | 0.023 | 0.044 | 0.011 | 0.020 | 0.061 | 0.107 |
| %CV=              |  | 11.6% | 12.1% | 24.0% | 25.3% | 7.1%  | 7.5%  | 3.3%  | 3.5%  | 1.6%  | 1.6%  | 4.5%  | 4.4%  |
| Total             |  |       |       |       |       |       |       |       |       |       |       |       |       |
| (inter-assay) sd= |  | 0.014 | 0.022 | 0.010 | 0.016 | 0.073 | 0.203 | 0.077 | 0.212 | 0.119 | 0.150 | 0.098 | 0.150 |
| %CV=              |  | 43.7% | 38.7% | 41.2% | 36.8% | 9.5%  | 14.6% | 11.0% | 16.6% | 17.4% | 12.2% | 7.3%  | 6.2%  |

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is bold and sans-serif.

In addition, inter-assay and intra-assay reproducibility with plasma samples were determined by assaying three additional samples (one T. gondii IgM negative sample and two T. gondii IgM positive samples) in triplicate, on three different days, at one laboratory site.

| Serum                            | Neg 1 |       | Pos 1 |       | Pos 2 |       |
|----------------------------------|-------|-------|-------|-------|-------|-------|
|                                  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |
| N =                              | 9     | 9     | 9     | 9     | 9     | 9     |
| Mean =                           | 0.022 | 0.04  | 0.710 | 1.26  | 1.083 | 1.92  |
| Within-Run<br>(intra-assay) sd = | 0.002 | 0.003 | 0.019 | 0.034 | 0.029 | 0.052 |
| %CV =                            | 8.7%  | 8.8%  | 2.7%  | 2.7%  | 2.7%  | 2.7%  |
| Total<br>(inter-assay) sd =      | 0.003 | 0.006 | 0.035 | 0.077 | 0.051 | 0.074 |
| %CV =                            | 14.5% | 14.9% | 4.9%  | 6.1%  | 4.8%  | 3.9%  |
| EDTA                             | Neg 1 |       | Pos 1 |       | Pos 2 |       |
|                                  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |
| N =                              | 9     | 9     | 9     | 9     | 9     | 9     |
| Mean =                           | 0.024 | 0.04  | 0.761 | 1.35  | 1.063 | 1.88  |
| Within-Run<br>(intra-assay) sd = | 0.002 | 0.004 | 0.019 | 0.034 | 0.019 | 0.033 |
| %CV =                            | 8.3%  | 8.3%  | 2.5%  | 2.6%  | 1.7%  | 1.8%  |
| Total<br>(inter-assay) sd =      | 0.002 | 0.004 | 0.036 | 0.030 | 0.056 | 0.076 |
| %CV =                            | 8.5%  | 9.4%  | 4.7%  | 2.2%  | 5.2%  | 4.0%  |
| Citrate                          | Neg 1 |       | Pos 1 |       | Pos 2 |       |
|                                  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |
| N =                              | 9     | 9     | 9     | 9     | 9     | 9     |
| Mean =                           | 0.022 | 0.04  | 0.674 | 1.19  | 1.075 | 1.91  |
| Within-Run<br>(intra-assay) sd = | 0.001 | 0.002 | 0.017 | 0.029 | 0.020 | 0.036 |
| %CV =                            | 5.0%  | 5.0%  | 2.5%  | 2.5%  | 1.9%  | 1.9%  |
| Total<br>(inter-assay) sd =      | 0.002 | 0.004 | 0.031 | 0.038 | 0.043 | 0.091 |
| %CV =                            | 9.9%  | 10.6% | 4.5%  | 3.2%  | 4.0%  | 4.8%  |
| Heparin                          | Neg 1 |       | Pos 1 |       | Pos 2 |       |
|                                  | OD    | S/CO  | OD    | S/CO  | OD    | S/CO  |
| N =                              | 9     | 9     | 9     | 9     | 9     | 9     |
| Mean =                           | 0.022 | 0.04  | 0.719 | 1.27  | 1.034 | 1.83  |
| Within-Run<br>(intra-assay) sd = | 0.001 | 0.002 | 0.012 | 0.020 | 0.022 | 0.039 |
| %CV =                            | 4.4%  | 4.4%  | 1.6%  | 1.6%  | 2.1%  | 2.1%  |
| Total<br>(inter-assay) sd =      | 0.002 | 0.003 | 0.024 | 0.031 | 0.059 | 0.115 |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BIO-RAD. The logo is white text on a black rounded rectangle. The text is in all caps and bolded.

#### B. Comparison Studies

Performance of the Platelia Toxo IgM TMB assay was evaluated against another commercially available enzyme immunoassay (EIA) by testing 348 patient samples and 97 specimens from 40 seroconversion panels. The combined results of the patient and seroconversion panel testing demonstrate an agreement of 99.53%.

|     | 1<br>= 445 | TMB |                                                                                                                                                                                             |      |
|-----|------------|-----|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
|     |            | Neg | Equivocal<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Pos  |
| lgM | Neg        | 27  |                                                                                                                                                                                             |      |
| EIA | Equivocal  |     |                                                                                                                                                                                             |      |
|     | Pos        |     |                                                                                                                                                                                             | ાર્ય |

Platelia Toxo IgM TMB vs IgM EIA Correlation Table (combined results)

Excluding equivocal samples / specimens, the combined results of comparative prospective and retrospective testing demonstrated the following:

Negative agreement : 100.00% (271/271). The 95% confidence intervaled is 99.82 - 100.00%. Positive agreement : 98.72% (154/156). The 95% confidence intervale is 96.61 - 100.00%. Overall agreement : 99.53% (425/427). The 95% confidence interval-6 is 98.76% - 100.00%.

#### C. Expected Values

A total of 445 fresh and frozen serum samples obtained from pregnant women during routine laboratory activities in the area of Paris, France were tested with the Platelia® Toxo IgM TMB assay. The distribution of serum / cut-off ratio values is shown in the following chart.

Image /page/4/Figure/9 description: The image is a bar chart titled "Platelia Toxo IgM TMB Expected Values". The x-axis is labeled "Ratio Serum / Cut-Off", and the y-axis is labeled "Frequency". The bar chart shows the frequency of different ratio serum/cut-off values. The bar chart shows that the most frequent ratio serum/cut-off value is less than 0.1, with a frequency of 214, and greater than or equal to 1.0, with a frequency of 154.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Bio-Rad. The logo is white text on a black rounded rectangle. The text reads "BIO-RAD" with a plus sign between "BIO" and "RAD".

### D. CDC Test Panel

The following information is from a serum panel obtained from the CDC and tested by Bio-Rad Laboratories. The results are presented as a means to convey further information on the performance of this assay with a masked, characterized serum panel. This does not imply an endorsement of the assay by the CDC.

The panel consists of 32 positive and 65 negative samples. The Platelia® Toxo IgM TMB assay demonstrated 100% agreement with the positive specimens and 100% agreement with the negative specimens.

Please Note: There should be no other statistical calculation or inferences drawn from the panel results.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are drawn with thick, curved lines. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular fashion around the image.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 3 0 2002

Mr. David Bhend Regulatory Affairs Associate Bio-Rad Laboratories Diagnostics Group 6565 18511 Avenue NE Redmond, WA 98052

Re: k013837

> Trade/Device Name: Platelia® Toxo IgM TMB Regulation Number: 21 CFR 866.3780 Regulation Name: Toxoplasma gondii serological reagents Regulatory Class: Class II Product Code: LGD Dated: July 19, 2002 Received: July 23, 2002

Dear Mr. Bhend:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and : additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

#### ATTACHMENT G INDICATIONS FOR USE STATEMENT

## 510(k) Number (if known): K013837

Device Name: Platelia® Toxo IgM TMB

## Indications for Use:

The Platelia Toxo IgM TMB kit is an in-vitro diagnostic test kit for the qualitative and quantitative detection of anti-Toxoplasma gondii IgM in human serum or plasma (EDTA, Heparin, Citrate).

Note:

- . Patient testing with the Platelia® Toxo IgM assay must be performed in conjunction with an anti-Toxoplasma gondii IgG antibody assay.
- The Platelia® Toxo IgM assay is presumptive for the detection of anti-● Toxoplasma gondii IgM antibodies and presumptive for the diagnosis of acute, recent or reactivated Toxoplasma gondii infection.
- The performance of the Platelia® Toxo IgM assay has not been established for ● neonate testing.
- . The Platelia® Toxo IgM assay has not been cleared/approved by the FDA for blood/plasma donor screening.

Woody Deubler

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K013837

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Professional Use: ___________

OR

Prescription Use: (Per 21 CFR 801.109

Over-The-Counter Use: (Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K013837](https://fda.innolitics.com/device/K013837)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
