DRG ORTHOHEX SS BONE SCREW

K013786 · Doctor'S Research Group, Inc. · HWC · Dec 17, 2001 · Orthopedic

Device Facts

Record IDK013786
Device NameDRG ORTHOHEX SS BONE SCREW
ApplicantDoctor'S Research Group, Inc.
Product CodeHWC · Orthopedic
Decision DateDec 17, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The DRG OrthoHex SS bone screws are intended for orthopedic trauma and reconstruction.

Device Story

DRG OrthoHex SS bone screws are metallic fixation fasteners used in orthopedic trauma and reconstruction procedures. Device provides mechanical stabilization of bone fragments. Used by orthopedic surgeons in clinical/surgical settings. Benefits include stabilization of fractures or reconstruction sites to facilitate healing.

Clinical Evidence

Bench testing only.

Technological Characteristics

Stainless steel alloy bone screw; metallic bone fixation fastener; Class II; Product Code HWC; Regulation 888.3040.

Indications for Use

Indicated for orthopedic trauma and reconstruction in patients requiring bone fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 7 2001 Doctor's Research Group, Inc. Richard Deslauriers, M.D. President 50 Altair Avenue Plymouth, Connecticut 06782 Re: K013786 Trade Name: DRG Orthohex SS Bone Screw Regulation Number: 888.3040 Regulation Name: Smooth or Threaded Metallic Bone Fixation Fastener Regulatory Class: II Product Code: HWC Dated: November 14, 2001 Received: November 14, 2001 Dear Dr. Deslauriers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Page 2 - Dr. Richard Deslauriers This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Celia M. Witten, Ph.D., M.D. ﮐﮯ Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Doctors Research Group, Inc. 50 Altair Avenue Plymouth, CT 06782 (860) 283-1479 **Statement of indications for use:** 510(k) Number (if Known): k013786 Device Name: DRG OrthoHex Stainless Steel Alloy Bone Screw Indications for use: The DRG OrthoHex SS bone screws are intended for orthopedic trauma and reconstruction. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRG, Office of Device Evaluation (ODE) Signature (Division Sign-Off) Division of General, Restorative and Neurological Devices 86 kni's 510(k) Number _ **Prescription Use** (Per 21 CFR 801.109) OR Over-The-Counter Use
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%